MPD-45 series - Total Power International

Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MPD-45
45W Dual Output Medical Type
series
Features :
Universal AC input / Full range
Low leakage current 0.3mA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
EN60601-1
IEC60601-1
SPECIFICATION
MODEL
OUTPUT
MPD-45A
CH1
5V
CH2
CH1
12V
5V
24V
RATED CURRENT
3.2A
2A
3.2A
1.2A
CURRENT RANGE
0.4 ~ 5A
0.2 ~ 2.5A
0.4 ~ 5A
0.2 ~ 1.8A
RATED POWER
40W
OUTPUT POWER (max.)
52W with 18CFM min. Forced air convection
LOAD REGULATION
7.0%
4.0%
7.0%
2.0%
1.0%
2.0%
4.0%
3.0%
4.0%
3.0%
50ms/230VAC
90 ~ 264VAC
FREQUENCY RANGE
47 ~ 440Hz
EFFICIENCY(Typ.)
76%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.2A/115VAC
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
SAFETY &
ISOLATION RESISTANCE
EMC
EMI CONDUCTION & RADIATION
(Note 4)
HARMONIC CURRENT
EMS IMMUNITY
MTBF
DIMENSION
PACKING
NOTE
4.0%
VOLTAGE RANGE
OVER VOLTAGE
150mVp-p
1.0%
HOLD UP TIME (Typ.)
PROTECTION
60mVp-p
CH1:4.5 ~ 5.5V
800ms, 20ms/230VAC
OVERLOAD
OTHERS
120mVp-p
SETUP, RISE TIME
LEAKAGE CURRENT
CH2
44.8W
RIPPLE & NOISE (max.) Note.2 60mVp-p
CH1:4.5 ~ 5.5V
VOLTAGE ADJ. RANGE
VOLTAGE TOLERANCE Note.3
LINE REGULATION
INPUT
MPD-45B
OUTPUT NUMBER
DC VOLTAGE
800ms, 20ms/115VAC at full load
16ms/115VAC at full load
127 ~ 370VDC
78%
0.7A/230VAC
COLD START 15A/115VAC
<0.3mA / 264VAC
30A/230VAC
53 ~ 75W rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
5.75 ~ 6.75V on CH1
Protection type : Hiccup mode, recovers automatically after fault condition is removed
-10 ~ +60 (Refer to output load derating curve)
20 ~ 90% RH non-condensing
-20 ~ +85
, 10 ~ 95% RH
0.04%/ (0 ~ 50 ) on +5V output
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
UL2601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
I/P-O/P, I/P-FG:100M Ohms/500VDC
Compliance to EN55011 (CISPR11) Class B
Compliance to EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN60601-1-2, medical level, criteria A
291.3Khrs min. MIL-HDBK-217F (25
127*76*28mm (L*W*H)
0.2Kg; 72pcs/17.4Kg/1.35CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. Mounting holes M1 and M2 should be grounded for EMI purposes.
Authorized Distributor: Total Power International, Inc. Toll Free: 877-646-0900(US) PH# (978) 453-7272 www.total-power.com
MPD-45
45W Dual Output Medical Type
Mechanical Specification
series
Unit:mm
127
115.8
5.6
5.6
FG
M2
M1
76
64.8
CN2
FS1
1
2
3
4
5
6
AC FUSE
T2A/250V
1
2
CN1
FS2
AC FUSE
T2A/250V
SVR1
43.
2 max.
28
8
DC Output Connector (CN2) : Molex 5273-06 or equivalent
Pin No. Assignment
+V
1
+5V
2,3
COM
4,5
6
NC
AC Input Connector (CN1) : Molex 5277-02 or equivalent
Assignment
AC/N
AC/L
Pin No.
1
2
Mating Housing
Terminal
Molex 5195
or equivalent
Molex 5194
or equivalent
Mating Housing
Terminal
Molex 5195
or equivalent
Molex 5194
or equivalent
Block Diagram
fosc : 45KHz
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
POWER
SWITCHING
FG
O.L.P.
RECTIFIERS
&
FILTER
+V
RECTIFIERS
&
FILTER
+5V
COM
DETECTION
CIRCUIT
PWM
CONTROL
O.V.P.
Static Characteristics
Derating Curve
100
100
90
80
80
70
LOAD(%)
LOAD (%)
60
40
20
-10
0
10
20
30
40
AMBIENT TEMPERATURE (
50
)
60 (HORIZONTAL)
60
50
40
90
100
110
135
155
INPUT VOLTAGE (V) 60Hz
230 264