PMMK130 series

1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
130 Watts Output Power
Single Outputs
Universal 90-264VAC Input
5VDC to 48VDC Outputs
5656VDC Input to Output Isolation
Active Power Factor Correction
3” x 5” x 1.4”
Model Number
Output V
Output I max.
Ripple & Noise
Efficiency(min/nom) Capacitive Load
SINGLE OUTPUT
PMMK130S-5 (U)(-A)
PMMK130S-12 (U)(-A)
PMMK130S-15 (U)(-A)
PMMK130S-24 (U)(-A)
PMMK130S-28 (U)(-A)
PMMK130S-36 (U)(-A)
PMMK130S-48 (U)(-A)
5 VDC
12 VDC
15 VDC
24 VDC
28 VDC
36 VDC
48 VDC
22 Amps
10.8 Amps
8.7 Amps
5.4 Amps
4.6 Amps
3.6 Amps
2.7 Amps
100mV pk-pk
120mV pk-pk
120mV pk-pk
120mV pk-pk
120mV pk-pk
120mV pk-pk
120mV pk-pk
75% / 80%
80% / 82%
80% / 85%
82% / 85%
82% / 85%
82% / 85%
82% / 85%
10,000uF
26,800uF
22,000uF
28,000uF
16,800uF
1,950uF
7,800uF
Note:
1. The PMMK130 Series is offered in an optional U-Channel Chassis as well as with optional 3.96mm pitch 5 and 10 Pin Input and Output connectors (open frame construction and screw terminals are standard respectively). Simply add a “U” at the end of the Model Number to recieve the
U-Channel configuration or a “-A” to receive the pins, or both if both are desired. For example: The Model Number “PMMK130S-5U-A” represents
a single output 5VDC U-Channel supply with 3.96 pitched pins (5 Input; 10 Output) . See the Mechanical Dimensions on pages 3 through 8 in this
specification
2. All “U” options are not covered by UL Certification unless provided with more than 10 CFM forced air cooling applied 6-10cm above T1.
3. Derating curves acceptable up tp 40°C for UL certified products
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNE.COM
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
INPUT SPECIFICATIONS
Input Voltage / Frequency Range
Inrush Current, typ:
Input Current
Power Factor (cold start)
Leakage Current
90-264VAC / 47-63 Hz
60A @ 230VAC Input *
2.3~0.8A (90-264VAC)
PF > 0.96 @ 230VAC
<0.3mA @ 264VAC, 50Hz
See Selection Chart
130 Watts
1%
0.5% nom / 1% max
0.5% nom / 1% max
See Selection Chart
<4S, typ.
20mS, typ.
30mS, typ.
Auto recovery *
130% max., Auto recovery *
150% max., Auto recovery *
10mS
<10% of Nominal O/P VDC
GENERAL SPECIFICATIONS
Safety UL/cUL: UL-EU: CB: CE: Isolation (Note 3)
Insulation Resistance
EMI
Efficiency
series
ENVIRONMENTAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Voltage and Current
Maximum Output Power
Preset Accuracy (Note 6)
Line Regulation (Note 5)
Load Regulation (20% to FL)
Ripple/Noise (Notes 1,2,7)
Turn On
Hold Up Time
Rise Time
Short Circuit Protection
Over Voltage Protection
Over Current Protection
Transient Response
Over Shoot/Under Shoot
PMMK130
ANSI/AAMI ES60601-1 3rd ed.
CSA C22.2 No. 60601-1 3rd ed.
EN60601-1 3rd ed.
IEC60601-1 3rd ed.
EN60601-1-2
5656VDC I/P - O/P *
2828 VDC I/P - GND *
707VDC O/P - GND *
≥ 20MΩ (500VDC, 1S I/P-O/P)
CISPR EN55011 class B
See Selection Chart
Oper. Temperature (Note 4)
Storage Temperature
Relative Humidity
EMS
Harmonics:
Fluctuations:
ESD:
RS:
EFT:
Surge:
CS:
Dips:
MTBF
Vibration
Drop Test
-10°C to +70°C (See Derate Curve)
-25°C to +85°C
10 to +90%, non-cond *
IEC61000-3-2 Class A
IEC61000-3-3
IEC61000-4-2, 6KV Contact, 8KV Air
IEC61000-4-3 FR: 80MHz-2.5GHz
Field Strength: 3V/M
IEC61000-4-4 2KV on AC Line
IEC61000-4-5 1KV (L-L);
2KV (L, N-PE)
IEC61000-4-6 3V (EMF)
IEC61000-4-11 95% 250CY,
70% 25CY, 40% 5CY, 5% 0.5CY
197,000 Hrs min. MIL-HDBK-217F
4G Pk, 50~500Hz, 3 Axes, 30 min* Test Height 70cm *
PHYSICAL SPECIFICATIONS
Size Open Frame:
U-Channel:
Weight Open Frame:
U-Channel:
76.2 x 127 x 35.5mm (3” x 5” x 1.4”)
127 x 80.4 x 38mm
(5” x 3.17” x 1.50”)
10.25 oz (291g)
13.38 oz (380g)
NOTES:
1. All measurements should be made directly at the terminals of the
power supply.
2. Ripple and Noise depend upon output voltage a specified per
particular model.
3. Isolation for up to 1min duration.
4. Specified for free-air convection cooling.
5. Line regulation measured from 90 - 264VAC/100VAC minimum
required for full load start.
6. Preset accuracy measured at nominal load, 120VAC input.
7. O/P noise measured directly at pins/terminals @ nominal load, 0uF by pass and 47uF electrolytic, pk-pk@ 20MHz bandwidth.
Special EMC Notice: The end application shall provide an earth ground
connection for conducted & radiated emission performance. Astrodyne is
not liable for the end application without extra earth connection to
PMMK130S Open Frame Power Supply.
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
MECHANICAL DIMENSIONS
Open Frame (PMMK130S-xx)
5.00in
127mm
.20in
5mm
4.55in
115.6mm
.22in
5.7mm
.14in
4_ 3.5mm
Vo
Vo
3.00in
76.2mm
Vo
2.61in
66.2mm
COM
FG
COM
ACN
COM
ACL
1.3in
33mm
.10in
2.5mm
WWW.ASTRODYNE.COM
Input connector : 3P / 8.25mm pitch TERMINAL BLOCK
.06in
1.6mm
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
MECHANICAL DIMENSIONS
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
MECHANICAL DIMENSIONS - FAN PLACEMENT RECOMMENDATION
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
MECHANICAL DIMENSIONS
Open Frame (PMMK130S-xx-A)
5.00in
127mm
4.55in
115.6mm
.20in
5mm
.22in
5.7mm
.14in
4_ 3.5mm
3.00in
76.2mm
Vo
Vo
Vo
Vo
Vo
COM
COM
COM
COM
COM
2.61in
66.2mm
FG
ACN
ACL
1.4in
35.5mm
1.24in
31.4mm
.10in
2.5mm
WWW.ASTRODYNE.COM
Input connector : 5P / 3.96mm pitch MOLEX 41791-0005
Output connector : 10P / 3.96mm pitch MOLEX 41791-0010
.06in
1.6mm
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
MECHANICAL DIMENSIONS
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
MECHANICAL DIMENSIONS - FAN PLACEMENT RECOMMENDATION
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
DERATE CURVES
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
1 3 0 W O p e n F r a me and U-Channel Medical Switchers with PFC
PMMK130
series
DERATE CURVES
WWW.ASTRODYNE.COM
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458