3255-50 DIGITAL HiTESTER

3255-50
D I G I TA L H i T E S T E R
Field measuring instruments
Combine with
CLAMP ON PROBE
for AC current measurement
Combine with
CLAMP ON PROBE
for AC current measurement
Actual-size
http://www.hioki.co.jp/
CAT III 600V
CAT II 1000V
HIOKI company overview, new products, environmental considerations and
other information are available on our website.
For industrial power lines
S afety M easures
● Dust-proof and water-proof construction (IP54)
● Polarized terminal design prevents incorrect insertion of test leads
● Internal current-limiting resistor and fuse protect
E nhanced C onvenience
● Safe and simple
AC load current measurement by optional CLAMP-ON PROBE
against operating errors and catastrophic failures
0.44 A / 1000 V Fuse
9010, 9018
(shut-off capacity 10 kA)
3255-50
Protection
device
Current limiting resistor
In addition to internal
(5 Ω)
circuitry to protect against +
connection errors during
Tester
circuit
use, the current-limiting
COM
resistor and fast-blow fuse
protect against short-circuit conditions in the unlikely event of
catastrophic failure of the tester circuit: short-circuit current is
limited until interruption by the fuse, rated for up to 10 kA.
*If voltage is applied to the test leads while in the resistance measurement mode, as
long as it is within the measurement category (CAT III 600 V, or CAT II 1000 V),
protective components in the tester circuitry prevent the fuse from opening.
■ 3255-50 specifications (Accuracy at 23˚C±5˚C (73˚F±9˚F), 80% rh or less)
10.00
A
20.00
A
50.0
A
100.0
A
200.0
500
A
A
1000
A
420.0
Ω
± 1.0 %rdg. ± 8 dgt.
4.200 k Ω
± 0.7 %rdg. ± 4 dgt.
42.00 k Ω
420.0 k Ω
4.200 MΩ
42.00 MΩ
Continuity
(420.0 Ω)
Diode
Main unit
Accuracy
± 2.0 % rdg.
± 4 dgt.
(50Hz to 500Hz)
Add the accuracy
of clamp sensor
± 1.0 %rdg. ± 4 dgt.
± 2.0 %rdg. ± 4 dgt.
± 5.0 %rdg. ± 4 dgt.
Input impedance
1000
V ± 0.7 % rdg. ± 1.5 % rdg.
± 4 dgt.
± 4 dgt.
V
9010, 9018 (10 to 500 A)
V ±0.5%rdg.± 4dgt. ±1.1%rdg.± 4dgt.
Range table of clamp on probe
42.00
420.0
Appox.
10 MΩ
Appox.
11 MΩ
Appox.
10 MΩ
Appox.
10 MΩ
3.4 Vmax.
Appox.
0.7 V
Appox.
0.5 V
A built-in buzzer sounds when the resistance Open-circuit terminal voltage
value is less than 45 Ω ± 35 Ω
(±1.0 %rdg. ± 8 dgt.)
Judgment
(0.15 V to 2.00 V)
3.4 Vmax.
Open terminal voltage/current
3.4 V/850 µA max.
Overload
protection
Current limiting resistor: 5 Ω, Fuse: 0.44 A / 1000 VAC/DC (shut-off capacity 10 kA)
420.0 mV ± 1.0 % rdg. ± 2.0 % rdg.
± 4 dgt.
± 4 dgt.
4.200 V
Notes
1000 V DC/AC rms(sin) or 107 V·Hz, 1 minute
Accuracy
AC 50Hz to 500Hz
Input impedance
DC
Open-circuit terminal voltage
Resistance
AC current Clamp sold separately AC / DC voltage
Range
Set the clamp range to match
the range of the 3255 -50
● Option
9010
CLAMP ON PROBE
R a n g e
Accuracy(45 - 66 Hz)
Frequency characteristic
Output voltage
Max. rated voltage to earth
Meas. conductor dia.
Dimensions and mass
9018
AC 10/20/50/100/200/500 A
±1.5 %rdg. ±0.1% f.s.
±3 % f.s.
40 Hz to 1 kHz: ±6 %
40 Hz to 3 kHz: ±1 %
AC 0.2 V f.s.
600 V rms (insulated conductor)
φ 46 mm, 50 × 20 mm busbar
74W × 184H × 37D mm, 420 g(9010), 410 g(9018)
Lead length approx. 3 m
● AC Measurement: average value ● Ancillary functions:
automatic and manual range, data hold, auto power save, battery
life warning ● Display: 4199 max. (1099 for 1000 V range) ●
Sampling rate: 2.5 samples/sec ● Applicable standards:
Safety - EN61010 CAT III 600V, CAT II 1000 V, EMC EN61326, Dust-proof and water-proof structure - EN60529 IP54
● Ambient temperature / humidity: 0 to 40˚C (32˚F to 104˚F) 80%
rh (no condensation) ● Storage temperature/ humidity range: -20
to 40˚C (-4˚F to 104˚F) 80% rh / 40 to 60˚C (104˚F to 140˚F) 70% rh(no
condensation) ● Power source: R03 manganese battery✕2 or
LR03 alkaline battery ✕2 ● Continuous operation: In DC
voltage approx. 200 hours (with manganese batteries), approx. 500
hours or more (with alkaline batteries) ● Dimensions and Mass:
Approx. 70 W ✕145 H ✕ 31 D mm, approx. 210g (Approx 2.8˝ W ✕
5.8˝ H ✕1.3˝ D, 7.5 oz.)(including batteries)
cord length 70 cm
Includes 9207-10 TEST LEAD and 9371 CARRYING CASE
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@ hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@ hiokiusa.com
Shanghai Representative Office :
1704 Shanghai Times Square Office
93 Huaihai Zhong Road
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090, 0092
FAX +86-21-6391-0360
E-mail: [email protected]
All information correct as of Oct 3, 2003. All specifications are subject to change without notice.
325550E2-3XB-05U