MP150 datasheet

150W Open Frame Medical Power Supply
MP-150 series
Features:
Universal AC input / Full range with active PFC
4"x 2" footprint
Medical safety approval: ANSI / AAMI ES 60601-1 /
TUV EN 60601-1; IEC 60601-1 / 2xMOPP
100W convection cooling and 150W with 10.5 CFM forced air
Protection: Short circuit, Over load, Over voltage
High operation temperature up to 70°C
Built-in fan supply
Withstand 2G vibration test
3 years warranty
MEDICAL
MP-150-12
MODEL
DC Voltage Range
Rated Current
Current Range (Convection)
Current Range (20 CFM Fan)
Rated Power (Convection)
Rated Power (20 CFM Fan)
Output
Input
Ripple & Noise (Max.)
Voltage Adj. Range
Voltage Tolerance
Line Regulation
Load Regulation
Setup, Rise Time
Hold Up Time
Voltage Range
Frequency Range
Efficiency (Typ.) at 230VAC
AC Current (Typ.)
Inrush Current (Typ.)
Leakage Current
Over Load
Protection
Over Voltage
Environment
Working Temp.
Working Humidity
Storage Temp. & Humidity
Temp. Coefficient
Vibration
Safety Standards
Safety & EMC
Others
Note
MP-150-15
E324816
MP-150-48
MP-150-24
15V
12V
24V
12.5A
10A
6.25A
0 ~ 8.3A
0 ~ 6.7A
0 ~ 4.2A
0 ~ 12.5A
0 ~ 10A
0 ~ 6.25A
99.6W
100.5W
100.8W
150W
150W
150W
120mVp-p
150mVp-p
240mVp-p
±10%
±2.0%
±2.0%
±2.0%
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
2000ms, 50ms / 230VAC, 3500ms, 50ms / 115VAC at full load
16ms / 230VAC, 16ms / 115VAC at full load
90 ~ 264VAC, 127 ~ 373VDC
47 ~ 63Hz
90%
91%
90%
1.72A / 100VAC, 0.73A / 240VAC
Cold Start 30A / 115VAC, 60A / 230VAC
For earth < 275uA / 264VAC, For patient < 100uA / 264VAC
Above 110% rated output power
Protection type: latch-style (Recovery after reset AC power ON)
115 ~ 140% rated output voltage
Protection type: latch-style (Recovery after reset AC power ON)
-20 ~ 70°C (Refer to output load de-rating curve)
20 ~ 90% RH non-condensing
-40 ~ +85°C, 10 ~ 95% RH
±0.03% / °C (0 ~ 50°C)
48V
3.125A
0 ~ 2.1A
0 ~ 3.125A
100.8W
150W
480mVp-p
±2.0%
±1.0%
±1.0%
91%
10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes
Meets UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99
Meets TUV (EN 60601-1; IEC 60601-1); MOPP approval
I/P-O/P: 4KVAC, I/P-FG: 1.8KVAC, O/P-FG: 1.5KVAC
Withstand Voltage
Isolation Resistance
I/P-O/P: 100M Ohms / 500VDC
EMI Conduction & Radiation
Certified EN 55011
Certified EN 61000-3-2; EN 61000-3-3
Harmonic Current
Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11
EMS Immunity
XXX K HRS Certified MIL-HDBK-217F
MTBF
Free Air convection for 100W, with 20 CFM Fan for 150W
Cooling
Dimension (WxHxD)
50.8x32.1x101.6 mm / 2.00x1.26x4.00 inch
Packing
XXX kg
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
3. Tolerance: includes set up tolerance, line regulation and load regulation.
4. Line regulation is measured from low line to high line at rated load.
5. Load regulation is measured from 0% to 100% rated load.
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed
that it still meets EMC directives.
*Note:The specifications are subject to change without notice.
REV. A2
14/02/20
1
150W Open Frame Medical Power Supply
MP-150 series
Mechanical Specification:
Unit : mm
AC Input Connector (CN1): JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC / N
N.C
AC / L
Mating Housing / Contact
JST VHR-3N
or equivalent
JST SVH-21T-P1.1
or equivalent
External Fan Power Connector (CN2): JST B2B-XH-A or equivalent
Pin No.
1
2
Assignment
GND
+12V
Mating Housing / Contact
JST XHP-2 JST SXH-001T-P0.6
or equivalent
or equivalent
10.5CFM
FAN
DC Output Connector (CN3): JST B6P-VH or equivalent
Pin No.
1,2,3
4,5,6
Assignment
+VO
-VO
50mm
AIR flow
direction
Mating Housing / Contact
JST VHR-6N JST SVH-41T-P1.1
or equivalent
or equivalent
Block Diagram:
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
PFC
CIRCUIT
POWER
SWITCHING
RECTIFIERS
&
FILTER
+V
-V
Fan
RECTIFIERS
&
FILTER
+V
-V
Output
DETECTION
CIRCUIT
O.L.P.
FG
CONTROL
O.V.P.
De-rating Curve:
Load V.S Temp.
Load V.S I/P Voltage
Ta=25°C
With 10.5CFM Free Air Convection
100
100
90
Load (%)
Load (%)
80
66
60
70
60
Free Air Convection
40
80
50
20
-20
40
0
10
20
30
40
50
Ambient Temperature (°C)
60
70
90
95
100
115
120
140
160
180
264Vac
Input Voltage (Vac) 60Hz
*Note:The specifications are subject to change without notice.
REV. A2
14/02/20
2