MP-05 - cotek

5W Open Frame Medical Power Supply
MP05 series
Features:
Universal AC input / Full range
Medical safety approval: ANSI / AAMI ES 60601-1 /
TUV EN 60601-1: 2006; IEC 60601-1 / 2xMOPP
Isolation class Ⅱ
Green design, No-load power consumption < 0.3W
Protection: Short circuit, Over load, Over voltage
Brown-out (Low AC Input Voltage)
High operation temperature up to 70°C
Withstand 2G vibration test
3 years warranty
MEDICAL
MODEL
Output
Input
DC Voltage Range
Rated Current
Current Range
Rated Power
Ripple & Noise (Max.)
Voltage Adj. Range
Voltage Tolerance
Line Regulation
Load Regulation
Setup, Rise Time
Hold Up Time
Voltage Range
Frequency Range
Efficiency (Typ.) at 230VAC
AC Current (Typ.)
Inrush Current (Typ.)
Over Load
Protection
Over Voltage
Environment
Working Temp.
Working Humidity
Storage Temp. & Humidity
Temp. Coefficient
Vibration
Safety Standards
Safety & EMC
Others
Note
Withstand Voltage
Isolation Resistance
EMI Conduction & Radiation
Harmonic Current
EMS Immunity
MTBF
Dimension (WxHxD)
Packing
MP05-05
MP05-12
E324816
MP05-15
MP05-24
5V
15V
12V
24V
0.33A
0.42A
0.21A
1A
0 ~ 0.33A
0 ~ 1A
0 ~ 0.42A
0 ~ 0.21A
5W
5.04W
4.95W
5.04W
100mVp-p
150mVp-p
150mVp-p
240mVp-p
±10%
±2%
±1%
±1%
±1%
±0.5%
±0.5%
±0.5%
±0.5%
±1%
±0.5%
±0.5%
±0.5%
100ms, 25ms / 230VAC, 100ms, 25ms / 115VAC at full load
100ms / 230VAC, 25ms / 115VAC at full load
85 ~ 264VAC, 125 ~ 373VDC
47 ~ 63Hz
71%
75%
75%
77%
0.13A / 115VAC, 0.08A / 230VAC
Cold Start 25A / 115VAC, 45A / 230VAC
Above 110% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is removed
115 ~ 145% rated output voltage
Protection type: latch-off mode
-20 ~ 70°C (Refer to de-rating curve)
20 ~ 90% RH non-condensing
-40 ~ +85°C, 10 ~ 95% RH
±0.03% / °C (0 ~ 50°C)
10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes
Certified UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99
Certified TUV (EN 60601-1; IEC 60601-1); MOPP approval
I/P-O/P: 4KVAC
I/P-O/P: 100M Ohms / 500VDC
Certified EN 55011
Certified EN 61000-3-2; EN 61000-3-3
Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11
217.4K HRS Certified MIL-HDBK-217F
45x19.1x57.9 mm / 1.771x0.752x2.2798 inch
0.033kg; 144pcs / 5.75kg / 1.43CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
3. Tolerance: includes setup time tolerance, line regulation and load regulation.
4. Line regulation is measured from low line to high line at rated load.
5. Load regulation is measured from 0% to 100% rated load.
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed
that it still meets EMC directives.
7. Length of setup time is measured at first cold start. Turning ON/OFF the power supply continuously may increase the setup time.
REV. B5
14/04/23
1
MP05 series
5W Open Frame Medical Power Supply
Mechanical Drawings:
19.1 [0.752]
Unit : mm[inch]
4.5 [0.177]
3.5 [0.138]
1.8 [0.070]
1.05 [0.041]
DC OUTPUT
20.04 [0.789]
+V
AC/N
AC/L
5
[0.196]
-V
12.47
[0.490]
C2
17.5 [0.689]
45 [1.771]
AC INPUT
5.33
[0.209]
47 [1.8504]
57.9 [2.279]
Block Diagram:
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
RECTIFIERS
&
FILTER
POWER
SWITCHING
PWM
CONTROL
+V
-V
DETECTION
CIRCUIT
De-rating Curve:
Load V.S Temp.
Load V.S I/P Voltage
Ta=25°C
100
100
90
Load (%)
Load (%)
80
60
80
70
60
40
50
20
-20
40
0
10
20
30
40
50
Ambient Temperature (°C)
60
70
90
100
110
120
264Vac
Input Voltage
REV. B5