Bolt-in Switched and Fuse Filtered - 66-67-BSF Series, For Medical or General Purpose Applications

Switched and Fused Filtered
Power
Entry Modules
For Medical or General Purpose Applications
66-67-BSF/66-67-SSF Series
Features
Tested and found to be
IAW VDE 0565 Part 3
Metric and North American fuse holders available
Fuse holder and a double pole power ON/OFF switch
provides a convenient/compact package
■ Suitable for products that must conform
to FCC and FTZ requirements
■ Meets over voltage category II of IEC 664
and complies with IEC 950
■ Provides susceptibility protection without the leakage
current associated with line-to-ground capacitors
■ Designed to meet requirements for non-patient
and patient care equipment
■ Metal case provides effective EMI shielding
■ Easy access fuse drawer - space for spare fuse
■ Flange-mounted or snap-in styles available
for quick mounting
■ Operating temperature: -25°C to +85°C
(including temperature rise, see graph on page PF41)
■
■
Specifications
Model*
66-XXX-020-0-12
66-XXX-020-0-04
66-XXX-020-0-14
66-XXX-020-0-06
66-XXX-020-1-12
66-XXX-020-1-04
66-XXX-020-1-14
66-XXX-020-1-06
66-XXX-020-4-12
66-XXX-020-4-04
66-XXX-020-4-14
66-XXX-020-4-06
66-XXX-040-0-12
66-XXX-040-0-04
66-XXX-040-0-14
66-XXX-040-0-06
66-XXX-040-1-12
66-XXX-040-1-04
66-XXX-040-1-14
66-XXX-040-1-06
66-XXX-040-4-12
66-XXX-040-4-04
66-XXX-040-4-14
66-XXX-040-4-06
66-XXX-060-0-12
66-XXX-060-0-04
66-XXX-060-0-14
66-XXX-060-0-06
66-XXX-060-1-12
66-XXX-060-1-04
66-XXX-060-1-14
66-XXX-060-1-06
66-XXX-060-4-12
66-XXX-060-4-04
66-XXX-060-4-14
66-XXX-060-4-06
Rated Rated Leakage
Voltage Current Current
Capacitance
50/60Hz
(Max.) CY
CX1
CX2
0.22uF
.075mA 330pF
0.47uF
0.22uF
2A
.01mA NONE
0.47uF
0.22uF
0.1mA 470pF
0.47uF
0.22uF
.075mA 330pF
0.47uF
250
VAC
0.22uF
4A
.01mA NONE
0.47uF
0.22uF
.01mA 470pF
0.47uF
0.22uF
.075mA 330pF
0.47uF
0.22uF
6A
.01mA NONE
0.47uF
0.22uF
0.1mA 470pF
0.47uF
Temp.
Induct.
(L1)
Applications
■
■
■
■
■
■
Digital equipment (including
switching power supplies)
General purpose filter for
susceptibility or high frequency
“clean up” applications
Circuit Diagram
L
L
CX1
CX2
CY1
E
E
LOAD
L1
R
LINE
Note: CY1 and CY2 capacitors
omitted on 66/67
B(S)F-XXX-1-X Filters
CY2
N
N
Rise
(Max.)
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF 10.5mH 40°C
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
4.2mH
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
45°C
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
1.6mH
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
Note: Test Voltage: 1500VAC one minute, line to ground
Insulation Resistance: 300 MΩ min. at 500VDC
Voltage Drop: 1V max. at rated current
Medical equipment
Industrial equipment
Telecommunications equipment
Measuring and testing instruments
Model*
67-XXX-020-0-12
67-XXX-020-0-04
67-XXX-020-0-14
67-XXX-020-0-06
67-XXX-020-1-12
67-XXX-020-1-04
67-XXX-020-1-14
67-XXX-020-1-06
67-XXX-020-4-12
67-XXX-020-4-04
67-XXX-020-4-14
67-XXX-020-4-06
67-XXX-040-0-12
67-XXX-040-0-04
67-XXX-040-0-14
67-XXX-040-0-06
67-XXX-040-1-12
67-XXX-040-1-04
67-XXX-040-1-14
67-XXX-040-1-06
67-XXX-040-4-12
67-XXX-040-4-04
67-XXX-040-4-14
67-XXX-040-4-06
67-XXX-060-0-12
67-XXX-060-0-04
67-XXX-060-0-14
67-XXX-060-0-06
67-XXX-060-1-12
67-XXX-060-1-04
67-XXX-060-1-14
67-XXX-060-1-06
67-XXX-060-4-12
67-XXX-060-4-04
67-XXX-060-4-14
67-XXX-060-4-06
Rated Rated Leakage
Voltage Current Current
Capacitance
50/60Hz
(Max.) CY
CX1
CX2
0.22uF
0.04mA 330pF
0.47uF
0.22uF
2A
.005mA NONE
0.47uF
0.22uF
0.05mA 470pF
0.47uF
0.22uF
0.04mA 330pF
0.47uF
125
VAC
0.22uF
4A
.005mA NONE
0.47uF
0.22uF
0.05mA 470pF
0.47uF
0.22uF
0.04mA 330pF
0.47uF
0.22uF
6A
Weight: 130g
Inlet: Compatible with IEC-320
B(S) = Bolt-in terminals or (Snap-in terminals)
.005mA NONE
0.47uF
0.22uF
0.05mA 470pF
0.47uF
Temp.
Induct.
(L1)
Rise
(Max.)
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF 10.5mH 40°C
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
4.2mH
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
45°C
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
1.6mH
NONE
0.22uF
NONE
0.22uF
NONE
0.22uF
* Substitute BSF or SSF for XXX
BSF - Bolt-In Switched and Fused
SSF - Snap-In Switched and Fused
API TECHNOLOGIES • 8061 Avonia Rd. • Fairview, PA 16415 • Ph: 814-474-1571 • Fax: 814-474-3110 • eis.apitech.com
API TECHNOLOGIES’ SPECTRUM CONTROL GmbH • Hansastrasse 6 • 91126 Schwabach, Germany • Phone: (49)-9122-795-0 • Fax: (49)-9122-795-58
PF40
Temperature Characteristics
120
Current (%)
100
Switched and Fused Filtered
Power
Entry Modules
For Medical or General Purpose Applications
80
60
40
20
0
66/67-BSF Series
.248
(6.3)
1.313 (33.4)
1.101 (28)
1.526 (38.8)
2.556
(65)
.248
(6.3)
2.340
(59.5)
Ø.142
(3.6)
2 - .142
(3.6) Hole
R .059
(1.5)
2.340
(59.5)
2.348
(59.7)
R .079
(2.0)
Dimensions in inches (mm)
Common Mode
1
3
Filter
2
4
50Ω
Level
Meter
50Ω
Oscillator
50Ω
80
80
70
70
Insertion Loss (dB)
Insertion Loss (dB)
90
60
2 Amp
40
30
2 Amp
50
40
30
4 Amp
6 Amp
10
10
0
0.1
0.2
0.5
1
2
5
Frequency (MHz)
10
20
0
0.1
50
90
80
80
70
70
Insertion Loss (dB)
90
60
2 Amp
50
40
30
20
1
2
5
Frequency (MHz)
10
20
50
2 Amp
60
50
40
4 Amp
30
6 Amp
10
10
0
0.1
0.5
20
6 Amp
4 Amp
0.2
0.075 and 0.1 mA Leakage Current
0.075 and 0.1 mA Leakage Current
Insertion Loss (dB)
50Ω
Level
Meter
50Ω
60
20
6 Amp
4 Amp
50Ω
0.01 mA Leakage Current
0.01 mA Leakage Current
90
20
* General Tolerance
± .039 (1.0)
* Metal Case
Panel cutout dimension
mount from back side
1.16 +.02
-0
(29.5 +0.5
-0 )
A
2.430 (61.8±0.1)
2.446 (62.2±0.1)
2.462 (62.6±0.1)
2.470 (62.8±0.1)
2.473 (62.9±0.1)
2.477 (63.0±0.1)
Normal Mode
1
3
Filter
2
4
50
Thickness of Panel
.031 (0.8t)
.039 (1.0t)
.047 (1.2t)
.059 (1.5t)
.071 (1.8t)
.079 (2.0t)
A
2.399 (61)
Oscillator
50Ω
2.273 (57.8)
t= .031 (0.8)
Panel cutout
dimension
mount from
back side
1.573
1.16 +.02
-0
(40.±0.2) (29.5 +0.5)
-0
90
Fast-on tab #250
t= .031 (0.8)
1.133
(28.8)
Fast-on Tab #250
2.273 (57.8)
1.573 (40±0.2)
80
1.133
(28.8)
1.514 (38.8)
2.556
(65)
2.348
(59.7)
30
40
50
60
70
Ambient Temperature (°C)
66/67-SSF Series
Dimensions
1.947 (49.5)
1.101 (28)
20
0.2
0.5
1
2
5
Frequency (MHz)
10
20
50
0
0.1
0.2
0.5
1
2
5
Frequency (MHz)
10
20
50
API TECHNOLOGIES • 8061 Avonia Rd. • Fairview, PA 16415 • Ph: 814-474-1571 • Fax: 814-474-3110 • eis.apitech.com
API TECHNOLOGIES’ SPECTRUM CONTROL GmbH • Hansastrasse 6 • 91126 Schwabach, Germany • Phone: (49)-9122-795-0 • Fax: (49)-9122-795-58
PF41