REFERENCE INFORMATION Configuration Sheet CUSTOMER/COMPANY DATE CONTACT/TITLE PHONE SUBMITTED BY REP AGENCY AND SALESPERSON PHONE INSTRUMENT TRANSMITTER MODEL SENSOR MODEL REMOTE CABLE (if applicable) QUANTITY PROCESS DATA APPLICATION Low Conductivity (<50 µS/cm)? No Yes High Solids Content (>35%)? No Yes The following table shows flow ranges for various sensor sizes. Optimal performance takes place between 5–50% of the maximum flow rate. Calibrations are standard for the ranges below. Size (in) Size (DN) Min (USG/min) Max (USG/min) Min (m3/h) Max (m3/h) Min (I/s) Max (I/s) ½" 15 0.29 28.5 0.065 6.5 0.018 1.8 ¾" 20 0.53 52.8 0.12 12 0.033 3.3 1" 25 0.79 79.3 0.18 18 0.05 5 1¼" 32 1.32 132 0.3 30 0.083 8.3 1½" 40 1.98 198.1 0.45 45 0.125 12.5 2" 50 3.17 317 0.72 72 0.2 20 2½" 65 5.28 528.3 1.2 120 0.333 33.3 3" 80 7.93 792.5 1.8 180 0.5 50 4” 100 12.33 1232.7 2.8 280 0.778 77.8 5" 125 18.93 1893.2 4.3 430 1.194 119.4 6" 150 28.62 2861.8 6.5 650 1.806 180.6 8" 200 50.63 5062.6 11.5 1150 3.194 319.4 10" 250 79.25 7925.2 18 1800 5 500 12" 300 110.95 11095.2 25.2 2520 7 700 14" 350 154.07 15406.5 35 3500 9.72 972 16" 400 198.13 19812.9 45 4500 12.5 1250 18" 450 252.00 25200 57.2 5720 15.9 1590 20" 500 317.01 31700.7 72 7200 20 2000 24" 600 440.32 44032.2 100 10000 27.78 2778 28" 700 616.42 61641.9 140 14000 38.89 3889 32" 800 792.52 79251.6 180 18000 50 5000 FACTORY CONFIGURATION Flow Units (check one) 4 mA value 20 mA value I/h m3/min USG/s kg/h* I/min m3/s UKG/h t/h* I/s USG/h UKG/min f3/s If not specified, MAGNETROL will use: m3/h USG/min UKG/s MGD 4mA = 0; 20 mA = Max *Density (g/l) Flow Empty Pipe Detection 705 Enterprise Street • Aurora, Illinois 60504–8149 • 630-969-4000 • Fax 630-489-9489 • www.magnetrol.com Yes No BULLETIN: 62-340.2 EFFECTIVE: April 2015 SUPERSEDES: April 2014