Flow Measurement
SITRANS F C
Flow sensor MC2
■ Overview
The main applications of the Coriolis flowmeter can be found in all
industries, such as:
3
SITRANS F C MC2 is available in sizes DN 100 and DN 150 (4"
and 6").
The MC2 sensor is suitable for accurate mass flow measurement
of a variety of liquids.
The sensor offers superior performance in terms of flow accuracy, turn-down ratio and density accuracy and delivers true
multi-parameter measurements i.e.: mass flow, volume flow,
density, temperature and fraction flow.
The very compact sensor construction makes installation and
commissioning of even the largest sizes very straight forward
and easy.
■ Benefits
• High accuracy better than 0.15 % of mass flow rate
• Large dynamic turn-down ratio
• Densitometer performance available through density accuracy better than 0.001 g/cm3
• Space-saving split-flow sensor design facilitating low pressure loss
• Parallel S-tube design and optimal oriented inductive sensors
enhances accuracy and turn-down ratio.
• Self-draining in both horizontal and vertical position
• Rigid enclosure design reduces the influence from pipeline vibration and thermal stress
• 4-wire Pt100 temperature measurement ensures optimum accuracy on mass flow, density and fraction flow
• SENSORPROM enables true "plug & play" - installed and commissioned in less than 10 minutes.
• Safe Ex design
• Sensor pipe available in high-quality stainless steel
AISI 316Ti/1.4571 or Hastelloy C4/2.4610 offering optimum
corrosion resistance.
• CIP cleanability for food and beverage and pharmaceutical
applications
Chemical and pharma
Detergents, bulk chemicals, pharmaceuticals, acids, alkalis
Food and beverage
Dairy products, beer, wine, softdrinks, Plato/Brix, fruit juices and
pulps, bottling, CO2 dosing, CIPliquids
Oil and gas
Liquid measurement, furnace control, test separators, LPG, oil bunkering
Water and waste water
Dosing of chemicals for water
treatment
The wide variety of combinations and versions from the modular
system means that ideal adaptation is possible to each measuring task. MC2 ist not recommended for gas applications.
■ Design
The MC2 sensor consists of 2 parallel measuring pipes, welded
directly onto a flow-splitter at each end to eliminate a direct coupling to the process connectors and significantly reduce effects
from external vibrations.
The flow-splitters are welded onto a rigid sensor housing which
acts as a mechanical low-pass filter.
The sensor is available in 2 material configurations, AISI 316L/
1.4436 or Hastelloy C4/2.4610 with a wide variety of process
connections.
The enclosure is made of stainless steel AISI 304/1.4301 with an
encapsulation grade of IP67/NEMA 4.
The sensor is Ex-approved.
It can be installed in horizontal or vertical position, and is selfdraining in both positions.
The MC2 Ex version sensor is based on a different Ex concept
than MASS 6000. Therefore the MC2 Ex version sensor can only
be connected to MASS 6000 IP67, MASS 6000 19" or SIFLOW
FC070 standard versions, which have to be remote mounted in
the safe area. MASS 6000 Ex d, MASS 6000 19" Ex and SIFLOW
FC070 Ex can not be used with MC2 Ex sensors.
■ Application
Coriolis mass flowmeters are suitable for measuring all liquids.
The measurement is independent of changes in process conditions/parameters such as temperature, density, pressure, viscosity, conductivity, and flow profile.
Due to this versatility the meter is easy to install and the Coriolis
flowmeter is recognized for its high accuracy in a wide turndown
ratio which is paramount in many applications.
MC2 sensors are not designed or approved for flow measurement of gaseous process media.
The product is manufactured by ABB Automation Products
GmbH and distributed by Siemens.
3/230
Siemens FI 01 · 2016
Hazardous area
Zone 1 + 2
Safe area
© Siemens AG 2016
Flow Measurement
SITRANS F C
Flow sensor MC2
Avoid cross talk
■ Function
The measuring principle is based on the Coriolis effect. See
“System information Coriolis mass flowmeters”.
■ Integration
Installation guidelines MC2 DN 100 and DN 150
Installation of sensor
Rigid mounting brackets must be used when installing the sensor. The brackets must be installed as close to the sensor as
possible, attached to the piping outside the process connections.
The optimal installation orientation is a vertical installation with an
upward flow as shown in the following figure. This has the advantage that any solids contained in the fluid will settle downward
and gas bubbles will move upward out of the meter tube when
the flow rate is zero. Additionally, it is easy to drain the meter
tube. Deposits can thereby be avoided.
3
Installation in a drop line
Mount with reduction (A) or orifice (B) to prevent partially draining (min. back pressure: 0.2 bar).
Vertical orientation:
A
B
Vertical installation self-draining (upward flow)
α
Horizontal orientation, self-draining
min. 2°
Installation in a drop line
Avoid vibrations
Siemens FI 01 · 2016
3/231
© Siemens AG 2016
Flow Measurement
SITRANS F C
Flow sensor MC2
■ Technical specifications
Inside pipe diameter
mm (inch) 43.1 (1.69)
76.1 (2.99)
Pipe wall thickness
mm (inch) 2.6 (0.10)
3.2 (0.13)
Fraction e.g. Brix
°Brix
602 000
(1 327 181)
10 0
0.5 ... 3.5 (0.18 ... 0.126)
DN
g/cm3
(lb/inch3)
p [mbar]
1000
100
(6 ˝
)
Density
203 500
(448 640)
Pressure drop
0 ... 100
(on request)
Not possible
DN
Mass flow measurkg/h
ing range at pressure (lb/h)
drop of 2 bar (29 psi)
at 1 g/cm3
(0.036 lb/inch3)
3
150 (6)
(4˝ )
100 (4)
150
Versions (mm (inch))
Temperature
Standard-version
-50 ... +200 °C (-58 ... +392 °F)
Ex-version
-50 ... +200 °C (-58 ... +392 °F)
Liquid pressure
measuring pipe
Stainless steel
(DIN 2413, 20 °C
(68 °F))
bar (psi)
40 (580)
40 (580)
Materials
Measuring pipe
Stainless steel AISI 316Ti/1.4571
or Hastelloy C4/2.4610
IP67
Enclosure
Enclosure material/
connection box
AISI 304 (1.4301)/aluminum,
max. pressure 40 bar (580 psi)
Process connections
See dimensional drawings
Electrical connections
Screw terminals, M 20
5 x 2 x 0.35 mm2 twisted and
screened in pairs, ext. Ø 12 mm
Cable
10, 25, 75 or 150 m
(32.8, 82, 246 or 492 ft.)
Cable length
Ex-version
ATEX
II 1/2 G Ex e m d [ia] IIC T6 ... T2
II 2 D Ex tD A21 IP6X
T115°C ... Tfluid
Weight approx.
kg (lb)
91 (201)
261 (573)
For accuracy specifications see „System information Coriolis
mass flowmeters“.
3/232
Siemens FI 01 · 2016
10
1
10
100
1000
10000
0DVVIORZ>NJPLQ@
© Siemens AG 2016
Flow Measurement
SITRANS F C
Flow sensor MC2
Selection and Ordering data
Article No. Ord. code
SITRANS F C flow sensors MC2
7 ME 4 3 0 0 7 77 7 7 - 777 7 77 7
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Article No.
7 ME 4 3 0 0 - 77777 - 7777
Dairy MLFB example
MC2 sensor
1D
Sensor size DN 100. AISI 316Ti/1.4571
Nominal pressure: Clamps
DIN 11851, DN 100, PN 25
F
43
Nominal diameter
Stainless steel AISI 316Ti/1.4571
DN 100
DN 150
1D
1E
DN 100
Nominal pressure
PN 40
Class 150
Class 300
Clamps/screwed-connections
2D
A
C
D
F
Flange ASME/ANSI B16.5
3“ (class 150/300/600)
4“ (class 150/300)
6“ (class 150/300)
22
23
24
32
33
34
Dairy clamp connection DIN 32676
(ISO 2852) Tri-clamp
81 mm clamp (PN 10)
100 mm clamp (PN 10)
52
53
Ex-approval and cable gland
Non-Ex, M20 x 1.5
Ex, M20 x 1.5
Cable
No cable
Calibration
(Density and fraction calibrations must be
ordered as PVR)
Standard
1
Order code
Please add “-Z“ to Article No. and specify Order
code(s) and plain text.
42
43
Configuration
Flow (0.15% of rate) and density
(5 kg/m3 [0.31 lb/ft3])
Flow (0.15% of rate) and density
(1 kg/m3 [0.06 lb/ft3])
A
Additional information
Dairy screwed connection to DIN 11851
DN 80 (PN 25)
DN 100 (PN 25)
Aseptic nut flange DIN 11864-2 form A for
pipes dimensioned by DIN 11866
DN 80 (3“)
DN 100 (4“)
3
A
Selection and Ordering data
Process connections
Flange EN 1092-1
DN 80 (PN 40, PN 100)
DN 100 (PN 40)
DN 150 (PN 40)
1
Configuration/calibration type: flow
and density (5 kg/m3 [0.31 lb/ft3])
Without Ex appproval
No cable
Standard calibration
Hastelloy C4/2.4610
Pressure testing certificate PED: 2014/68/EU
C11
Material certificate EN 10204-3.1
C12
Material certificate according to NACE
C16
Tag name plate, stainless steel
Y17
Tag name plate, plastic self-adhesive
Y18
Customer-specified, matched pair (5 x 2)
On request
Customer-specified calibration (5 x 2)
On request
Customer-specified, matched pair (10 x 1)
On request
Customer-specified calibration (10 x 1)
On request
Operating instructions for SITRANS F C MC2
This device is shipped with ABB documentation and an installation/connection instruction in four languages (Article No. A5E34730442).
All literature is available to download for free, in a range of languages, at
www.siemens.com/processinstrumentation/documentation
63
64
1
5
A
B
A
1
Siemens FI 01 · 2016
3/233
© Siemens AG 2016
Flow Measurement
SITRANS F C
Flow sensor MC2
■ Dimensional drawings
Remote design, flanged construction, DIN EN/ANSI
$
*
)ODQJHV
(1
$16,%
,62
)
3
)ORZGLUHFWLRQ
/
%
Process L
connec- [mm (inch)]
tion size
inch DN inch DN DIN 11864-2
form A
4
100 3
80 1618 (63.70)
4
100 1463 (57.60)
6
150 N/A
6
150 6
150 N/A
G1)
F
B
A
Weight
[mm
[mm
[mm
[kg (lb)]
[mm
(inch)] (inch)] (inch)] (inch)]
Meter
size
1)
EN 1092-1
PN 40
1640 (64.57)
1480 (58.27)
1778 (69.92)
2040 (80.31)
EN 1092-1
PN 100
1680 (66.14)
1530 (60.24)
N/A
N/A
ANSI B16.5
CL 150
1660 (65.35)
1500 (59.06)
1806 (71.10)
2070 (81.50)
ANSI B16.5
CL 300
1680 (66.14)
1520 (59.84)
1826 (71.89)
2090 (82.28)
ANSI B16.5
CL 600
1702 (67.01) 500
1568 (61.73) (19.69)
N/A
N/A
613
(24.13)
215
(8.46)
131
(5.16)
170
(6.69)
285
190
(11.22) (7.84)
260
(9.84)
84 (185)
91 (201)
120 (265)
260 (573)
For Ex add 54 mm
Remote design, food industry fittings, DIN 11851
Food industry fittings DIN 11851
Ø80
Ø80
Threaded
stubs
Threaded stubs
F
A
G
R
Flow direction
L
B
Meter size
Process connection size
inch
4
inch
3
4
1)
DN
100
DN
80
100
For Ex add 54 mm
3/234
g
Siemens FI 01 · 2016
Rd 110 x 1/6
Rd 130 x ¼
L [mm
(inch)]
g [mm
(inch)]
1618 (63.70) 8 (0.31)
1463 (57.60) 10 (0.39)
L
G1) [mm
(inch)]
F [mm
(inch)]
500 (19.69)
215 (8.46) 131 (5.16) 170 (6.69) 401 (15.79)
314 (12.36)
B [mm
(inch)]
A [mm
(inch)]
R [mm
(inch)]
Weight
[kg (lb)]
82 (180)
86 (190)
© Siemens AG 2016
Flow Measurement
SITRANS F C
Flow sensor MC2
Remote design, Tri-clamp DIN 32676 (ISO 2852)
80
80
G
R
F
A
3
Flow direction
L
B
Dimensions in mm (inch)
L
G1)
F
B
A
R
[mm (inch)] 3 [mm (inch)] [mm (inch)] [mm (inch)] [mm (inch)] [mm (inch)]
Weight
[kg (lb)]
80
1598 (62.91)
440 (17.32)
71 (157)
100
1448 (57.01)
365 (14.37)
69 (152)
Meter size
Process connection
size
inch
DN
inch
DN
4
100
3
4
1)
500 (19.69) 215 (8.46)
131 (5.16)
170 (6.69)
For Ex add 54 mm
Siemens FI 01 · 2016
3/235
© Siemens AG 2016
Flow Measurement
SITRANS F C
Flow sensor MC2
The max. allowable operating pressure is a function of the process connection type, the fluid temperature, the bolts and the
gaskets.
80
PN 40
20
PN 16
0
-50
Housing as secondary containment
• Max. 40 bar
Pressure Equipment Directive 97/23/EG
• Conformity evaluation category III, fluid group 1
Material strength for process connections
Size
PSmax.
at 20 °C
(68 °F)
TSmax.
0
50
100
150
7HPSHUDWXUH>r&@
DIN-Flanges stainless steel AISI 316Ti/1.4571 to DN 100 (4”)
Corrosion resistance of measuring pipe material to measuring
medium has to be considered.
TSmin.
120
100
80
60
DN
inch
bar (psi g) °C (°F)
100
4
25 (363)
140 (284) -40 (-40)
Tri-Clamp acc. 100
DIN 32676
4
10 (145)
120 (248) -40 (-40)
Thread acc.
DIN 11851
100
40
Pressure Rating
• PN 16, PN 40
Class 150, Class 300
Process
connection
120
60 PN 63
3UHVVXUH>EDU@
3
3UHVVXUH>EDU@
Pressure/temperature curves
Process Connections
• Flanges EN 1092-1/ANSI B16.5
• Tri-Clamp DIN 32676 (ISO 2852)
- DN 100: Series 1
• Food Industry fittings DIN 11851
°C (°F)
40
20
300 lb
150 lb
0
-50
0
50
100
150
7HPSHUDWXUH>r&@
ASME-Flanges stainless steel AISI 326Ti/1.4571 to DN 100 (4”)