English User's Manual

User's Guide
800 Amp True RMS AC Clamp Meter
Model EX720
TEMP
AC TRMS CLAMP METER EX720
Introduction
Congratulations on your purchase of the EX720 Clamp DMM. Careful use of this meter will provide
many years of reliable service.
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the
manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages
may be present
Double insulation
SAFETY NOTES




Do not exceed the maximum allowable input range of any function.
Do not apply voltage to meter when resistance function is selected.
Set the function switch OFF when the meter is not in use.
Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS




Set function switch to the appropriate position before measuring.
When measuring volts do not switch to current/resistance modes.
Do not measure current on a circuit whose voltage exceeds 600V.
When changing ranges always disconnect the test leads from the circuit under test.
UL Note
A UL mark does not indicate that this product has been evaluated for accuracy.
Function
Maximum Input
A AC
800A AC
V DC, V AC
600V DC/AC
Resistance, Capacitance, Frequency, Diode Test
250V DC/AC
Temperature
60V DC, 24V AC
2
EX720-en-GB_v3.4 4/15
CAUTIONS

Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.

Always remove the test leads before replacing the battery or fuses.

Inspect the condition of the test leads and the meter itself for any damage before operating the
meter. Repair or replace any damage before use.

Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.

Always discharge capacitors and remove power from the device under test before performing
Diode, Resistance or Continuity tests.

Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty
of connection to the recessed electrical contacts. Other means should be used to ensure that
the terminals are not "live".

If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.

This device is not a toy and must not reach children’s hands. It contains hazardous objects as
well as small parts that the children could swallow. In case a child swallows any of them,
please contact a physician immediately

Do not leave batteries and packing material lying around unattended; they can be dangerous
for children if they use them as toys

In case the device is going to be unused for an extended period of time, remove the batteries
to prevent them from training

Expired or damaged batteries can cause cauterization on contact with the skin. Always,
therefore, use suitable hand gloves in such cases

See that the batteries are not short-circuited. Do not throw batteries into the fire.
3
EX720-en-GB_v3.4 4/15
Description
Meter Description
1.
Current clamp
2.
Clamp opening trigger
3.
Control buttons:

Data Hold

Mode

Peak

Backlight
4.
Backlit LCD Display
1
2
5.
COM negative input jack for black test lead
6.
Rotary function switch
7.
V· Ω·CAP·TEMP·Hz·
positive input jack for red lead
HOLD
Data Hold
Minus sign
Negative reading display
5
Measurement display digits
PEAK
Peak capture
AUTO
Auto Range mode
DC/AC
Direct Current / Alternating Current
BAT
Low battery
mV or V
Milli-volts or Volts (Voltage)
Ω
Ohms (Resistance)
A
Amperes (Current)
F
Farad (Capacitance)
Hz
Hertz (Frequency)
o
n, m, , M, k
)))
3
4
0 to 3999
F and C
6
AC TRMS CLAMP METER EX720
Display icons Description
o
TEMP
7
Fahrenheit and Celsius units (Temperature)
Unit of measure prefixes: nano, milli, micro, mega, and kilo
Continuity test
Diode test
4
EX720-en-GB_v3.4 4/15
Operation
NOTES: Read and understand all Warning and Caution statements in this operation manual prior to using this meter.
Set the function select switch to the OFF position when the meter is not in use.
AC Current Measurements
WARNING: Ensure that the test leads are disconnected from the meter before making current clamp measurements.
1.
Set the Function switch to the 800A, 400A, or
40A range. If the approx. range of the
measurement is not known, select the highest
range then move to the lower ranges if
necessary.
2.
Press the trigger to open jaw. Fully enclose only
one conductor. For optimum results, center the
conductor in the jaw.
3.
The clamp meter LCD will display the reading.
Correct
Incorrect
AC/DC Voltage Measurements
1.
Insert the black test lead into the negative COM terminal and the red test
lead into the positive V· Ω·CAP TEMP·Hz ·
terminal.
2.
Set the function switch to the V Hz position.
3.
Use the MODE button to select AC or DC Voltage.
4.
Connect the test leads in parallel to the circuit under test.
5.
Read the voltage measurement on the LCD display.
V
5
EX720-en-GB_v3.4 4/15
Resistance Measurements
1.
Insert the black test lead into the negative COM terminal and the red test lead
into the V·Ω·CAP·TEMP Hz·
positive terminal.
2.
Set the function switch to the Ωposition.
3.
Touch the test probe tips across the circuit or component under test.
4.
Read the resistance on the LCD display.
Capacitance Measurements
WARNING: To avoid electric shock, discharge the capacitor under test before measuring.
1.
Set the function switch to theCAP position.
2.
Insert the black test lead banana plug into the negative COM jack and the red
test lead banana plug into the V· Ω·CAP·TEMP Hz·
positive jack.
3.
Touch the test probe tips across the part under test. If “dISC” appears in the
display, remove and discharge the component.
4.
Read the capacitance value in the display.
5.
The display will indicate the proper decimal point and value.
Note: For very large values of capacitance measurement it can take several minutes
before the final reading stabilizes.
CAP
Frequency Measurements
1.
Insert the black test lead banana plug into the negative COM jack and the red
test lead banana plug into the V· Ω·CAP·TEMP Hz·
positive jack.
2.
Set the function switch to the V Hz Position.
3.
Press and hold the MODE button to select the Frequency (Hz) function. “k Hz”
will appear in the display.
4.
Touch the test probe tips across the part under test.
5.
Read the Frequency value on the display.
6.
The display will indicate the proper decimal point and value.
7.
Press and hold the MODE button again to return to the voltage mode.
Hz
6
EX720-en-GB_v3.4 4/15
Temperature Measurements
1.
Set the function switch to the TEMP position.
2.
Insert the Temperature Probe into the negative COM and the
V·Ω·CAP·TEMP·Hz·
positive jacks, observing polarity.
3.
Touch the Temperature Probe head to the device under test. Continue to
touch the part under test with the probe until the reading stabilizes.
4.
Read the temperature on the display. The digital reading will indicate the
proper decimal point and value.
5.
Use the MODE button to select F or C.
o
o
WARNING: To avoid electric shock, be sure the thermocouple probe has been removed before
changing to another measurement function.
Continuity Measurements
1.
Insert the black test lead into the negative COM terminal and the red test lead into the
V· Ω·CAP·TEMP Hz·
positive terminal.
2.
Set the function switch to the ))) position.
3.
Use the MODE button to select continuity “)))”. The display icons will change when the MODE
button is pressed.
4.
Touch the test probe tips across the circuit or component under test.
5.
If the resistance is < 40, a tone will sound.
Diode Test
1.
Insert the black test lead banana plug into the negative COM jack and the red test lead banana
plug into the V·Ω·CAP·TEMP Hz·
positive jack
2.
Turn the function switch to
position. Use the MODE button to select the diode function if
necessary (diode symbol will appear on the LCD when in Diode test mode)
3.
Touch the test probe tips to the diode or semiconductor junction under test. Note the meter
reading
4.
Reverse the test lead polarity by reversing the red and black leads. Note this reading
5.
The diode or junction can be evaluated as follows:

If one reading displays a value (typically 0.400V to 0.900V) and the other reading
displays OL, the diode is good.

If both readings display OL the device is open.

If both readings are very small or ‘0’, the device is shorted.
7
EX720-en-GB_v3.4 4/15
Data Hold
To freeze the LCD reading, press the HOLD button. While data hold is active, the HOLD icon
appears on the LCD. Press the HOLD button again to return to normal operation.
Peak Hold
The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture
negative or positive peaks as fast as 1 millisecond in duration.
1.
Turn the function switch to the A or V position.
2.
Allow time for the display to stabilize.
3.
Press and Hold the PEAK button until “CAL” appears in the display. This procedure will zero
the range selected.
4.
Press the PEAK button, Pmax will display.
5.
The display will update each time a higher positive peak occurs.
6.
Press the PEAK button again, Pmin will display. The display will now update and indicate the
lowest negative peak.
7.
To return to normal operation, press and hold the PEAK button until the Pmin or Pmax
indicator switches off.
Note: If the Function switch position is changed after a calibration the Peak Hold calibration must
be repeated for the new function selected.
LCD Backlight Button
The LCD is equipped with backlighting. Press the backlight button to turn the backlight on. Press
again to turn the backlight off.
Automatic Power OFF
In order to conserve battery life, the meter will automatically turn off after approximately 25 minutes.
To turn the meter on again, turn the function switch to the OFF position and then to the desired
function position.
8
EX720-en-GB_v3.4 4/15
Specifications
Function
Range &
Resolution
Accuracy (% of reading + digits)
AC Current
40.00A AC
± (2.5% + 10d)
50/60 Hz
400.0A AC
± (2.5% + 8d)
800A AC
± (2.8% + 5d)
AC Voltage
400.0mV AC
± (1.0% + 10d)
50/60Hz True RMS
4.000V AC
True RMS
40.00V AC
± (1.5% + 5d)
400.0V AC
DC Voltage
600V AC
± (2.0% + 5d)
400.0mV DC
± (0.8% + 2d)
4.000V DC
40.00V DC
± (1.5% + 2d)
400.0V DC
Resistance
600V DC
± (2.0% + 2d)
400.0Ω
± (1.0% + 4d)
4.000kΩ
40.00kΩ
± (1.5% + 2d)
400.0kΩ
4.000MΩ
Capacitance
± (2.5% + 3d)
40.00MΩ
± (3.5% + 5d)
4.000nF
± (5.0% + 30d)
40.00nF
± (5.0% + 20d)
400.0nF
4.000µF
± (3.0% + 5d)
40.00µF
Frequency
Temperature
400.0µF
± (4.0% + 10d)
4.000mF
± (10% + 10d)
40.00mF
unspecified
4.000kHz
± (1.5% + 2d)
Sensitivity: 100V (<50Hz); 50V (50 to 400Hz); 5V (401 to 4000Hz)
o
± (3%rdg + 9 F)
o
o
± (3%rdg + 5 C)
-4 to 1400 F
o
-20 to 760 C
9
EX720-en-GB_v3.4 4/15
General Specifications
Clamp jaw opening
Display
Continuity check
Diode test
Low Battery indication
Over-range indication
Measurement rate
PEAK
Temperature sensor
30mm (1.2") approx.
3-3/4 digits (4000 counts) backlit LCD
Threshold 40; Test current < 0.5mA
Test current of 0.3mA typical;
Open circuit voltage < 3VDC typical
‘BAT’ is displayed
‘OL’ display
2 readings per second, nominal
Captures peaks >1ms
Type K thermocouple
Input Impedance
AC bandwidth
AC response
10M (VDC and VAC)
50 to 400Hz (AAC and VAC)
True rms (AAC and VAC)
Crest Factor
3.0 in 40A and 400A ranges, 1.4 in 800A range (50/60Hz and 5% to
100% of range)
Operating Temperature
Storage Temperature
Operating Humidity
5 C to 40 C (41 F to 104 F)
o
o
o
o
-20 C to 60 C (-4 F to 140 F)
o
o
o
Max 80% up to 31 C (87 F) decreasing linearly to 50% at 40 C
o
(104 F)
<80%
2000meters (7000ft). maximum.
One (1) 9V Battery
After approx. 25 minutes
229 x 80 x 49mm (9.0 x 3.1 x 2.0"); 303g (10.7 oz).
For indoor use and in accordance with the requirements for double
insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage
Category III 600V and Category II 1000V, Pollution Degree 2.
Storage Humidity
Operating Altitude
Battery
Auto power OFF
Dimensions & Weight
Safety
o
o
o
o
10
EX720-en-GB_v3.4 4/15
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input
terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If
the meter is not to be used for 60 days or more, remove the battery and store it separately.
Battery Replacement
1.
Remove the Phillips head screw that secures the rear battery door
2.
Open the battery compartment
3.
Replace the 9V battery
4.
Secure the battery compartment
Never dispose of used batteries or rechargeable batteries in household waste.
As consumers, users are legally required to take used batteries to appropriate
collection sites, the retail store where the batteries were purchased, or wherever
batteries are sold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end-oflife devices to a designated collection point for the disposal of electrical and electronic equipment.
Other Battery Safety Reminders
o Never dispose of batteries in a fire. Batteries may explode or leak.
o Never mix battery types. Always install new batteries of the same type.
Copyright © 2013‐2015 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form ISO‐9001 Certified www.extech.com 11
EX720-en-GB_v3.4 4/15