GPM130 Datasheet

GPC130 Commercial/GPM130 Medical
130 Watt Global Performance Switchers
FEATURES:
• Wide-range ac input 85-264 Vac
• 2-year warranty
• Commercial Approved to UL1950, IEC950,
EN60950 and CSA22.2-234 L3
• Medical Approved to UL2601-1, IEC601-1 and
CSA22.2 No. 601
• Conducted EMI exceeds FCC Class B and CISPR 22
Class B (Commercial models) and CISPR 11 Class B
(Medical models)
•
marked to LVD
SPECIFICATIONS:
Ac Input
85-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at 120 Vac, 60 Hz with full rated output
load: 3.5 A
Hold-Up Time
20 ms minimum from loss of ac input at full load, nominal line
(115 Vac).
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode.
Measured with noise probe directly across output terminals of
the power supply.
Output Power
100 W continuous, 165 W peak; 130 W with 26 cfm air flow. Peak
ratings are for 60 s maximum duration, 10% duty cycle. During
peak load condition, output regulation may exceed total
regulation limits.
Transient Response
Main output—500µSec typical response time for return to within
0.5% of final value for a 50% load step change. ∆i/∆t<0.2 A/µs.
Maximum voltage deviation is 3.5%. Startup/shutdown overshoot
less than 3%.
Overload Protection
Fully protected against short circuit and output overload. Short
circuit protection is cycling type power limit on outputs 1 & 2;
foldback type on outputs 3 & 4. Recovery after fault is automatic.
See output ratings chart for additional notes or conditions.
Remote Sense
Provided as a standard feature on output 1.
Overvoltage Protection
Main outputs: 124% ± 12% typical.
Overload Protection
Factory set to begin power limiting at approximately 170 W.
Efficiency
70% at full rated load, nominal input voltage, depending on model
and load distribution.
EMI/EMC Compliance
All models include built-in EMI filtering to meet the following
emissions requirements:
Turn-on Time
Less than 1 second at 120 Vac, 25°C (inversely proportional to
input voltage and thermistor temperature).
Input Protection
Internal ac fuse provided. Designed to blow only if a catastrophic
failure occurs in the unit.
Inrush Current
Inrush is limited by internal thermistors. Inrush at 240 Vac under
cold start conditions will not exceed 34 A.
Temperature Coefficient
0.03%/°C typical on all outputs.
Thermal Shutdown
Provided as a standard feature. Designed to protect unit from
prolonged overtemperature.
Environmental
Designed for 0 to 50°C operation at full rated output power;
derate output current and total output power by 2.5% per °C
above 50°C. See Environmental and Packaging Specifications
on next page.
54
Power Fail
TTL- or CMOS-compatible output goes low (< 0.5V) 5 ms
before output voltage drops more than 4% below nominal
voltage upon loss of ac power. The signal is factory set to
trip on 84 to 94Vac brown-out depending upon incoming line
impedance and distortion. Other settings are available to the
user through adjustment of built-in potentiometer.
Voltage Adjustment
Built-in potentiometer adjusts voltage ±5% on outputs 1 & 2.
EMI SPECIFICATIONS
COMPLIANCE LEVEL
Conducted Emissions GPC130 EN55022 Class B; FCC Class B
Conducted Emissions GPM130 EN55022 Class B; FCC Class B
Static Discharge
EN61000-4-2, 6 kV contact, 8 kV air
RF Field Susceptibility
EN61000-4-3, 3 V/meter
Fast Transients/Bursts
EN61000-4-4, 2 kV, 5 kHz
Surge Susceptibility
EN61000-4-5, 1 kV diff., 2 kV com.
Commercial Leakage Current
0.7 mA 254 Vac @ 60 Hz input.
Commercial Safety
Approved to UL1950, CSA22.2 No. 234 Level 3, IEC950 and
EN60950. UL file #E135803 commercial; CSA #LR46516 all
models. The output(s) are intended for safety earthed Signal
Output and Intermediate Circuits only. All dc outputs are SELV
under normal and single fault conditions.
Medical Leakage Current
35 µA 254 Vac @ 60 Hz input.
Medical Safety
Approved to UL2601, CSA22.2 No. 601 Level 3 and IEC601. UL
file E116994; CSA #LR46516. The output(s) are intended for
safety earthed Signal Output and Intermediate Circuits only. The
output(s) are not acceptable for patient connection without
additional isolation. All dc outputs are SELV under normal and
single fault conditions.
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com
GPC130 Commercial/GPM130 Medical 130 Watt Multiple Output
Commercial
Model
Medical
Model
Output No.
Output
Output
Minimum
Output
Maximum (B)
Output
Maximum (C)
Output Peak
Noise P-P
Total
Regulation (A)
GPC130A
GPM130A
1
2
3
4
+5 V
+12 V
-12 V
+12 V
3A
0A
0A
0A
16 A
5A
1.2 A
1.2 A
20 A
5A
1.2 A
1.2 A
25 A
8A
1.5 A
1.5 A
50 mV
120 mV
120 mV
120 mV
2%
2%
3%
3%
GPC130B
GPM130B
1
2
3
4
+5V
+12 V
-5 V
-12 V
3A
0A
0A
0A
16 A
5A
1.2 A
1.2 A
20 A
5A
1.2 A
1.2 A
25 A
8A
1.5 A
1.5 A
50 mV
120 mV
50 mV
120 mV
2%
2%
3%
3%
GPC130C
GPM130C
1
2
3
4
+5 V
+12 V
-15 V
+15 V
3A
0A
0A
0A
16 A
5A
1.2 A
1.2 A
20 A
5A
1.2 A
1.2 A
25 A
8A
1.5 A
1.5 A
50 mV
120 mV
150 mV
150 mV
2%
2%
3%
3%
GPC130D
GPM130D
1
2
3
4
+5 V
+24 V
-12 V
+12 V
3A
0A
0A
0A
16 A
3.5 A
1.2 A
1.2 A
20 A
3.5 A
1.2 A
1.2 A
25 A
5A
1.5 A
1.5 A
50 mV
240 mV
120 mV
120 mV
2%
2%
3%
3%
GPC130E
GPM130E
1
+5 V
3A
16 A
20 A
25 A
50 mV
2
+24 V
0A
3.5 A
3.5 A
5A
240 mV
3
-15 V
0A
1.2 A
1.2 A
1.5 A
150 mV
4
+15 V
0A
1.2 A
1.2 A
1.5 A
150 mV
A.Total regulation is defined as the maximum deviation from the nominal voltage for all steady-state conditions of initial voltage setting, input line voltage and output load.
B. Unrestricted natural convection cooling.
C. Requires 26 cfm moving air.
2%
2%
3%
3%
GPC130/GPM130 MECHANICAL SPECIFICATIONS
Environmental
Specification
Operating
Non-operating
Temperature (A)
See individual specs
-40 to +85oC
0 to 95% RH
0 to 95% RH
20 gpk
40 gpk
Humidity (A)
Shock (B)
Altitude
Vibration (C)
-500 to 10,000 ft
-500 to 40,000 ft
1.5 grms, 0.003 g2/Hz
5 grms, 0.026 g2/Hz
A. Units should be allowed to warm up/operate under non-condensing conditions
before application of power.
B. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3
orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating.
C. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes,
total 6 shocks.