Catalog

OVERVIEW
The C10066-10 Immunochromato-Reader makes rapid, high
sensitivity, quantitative measurements of color intensities for
red/blue based immunochromatographic reagents.
Since measurement data is saved as a CSV file, it can be analyzed on commercially available spreadsheet software, and
calibration curve and time-course graphs then easily drawn.
The C10066-10 Immunochromato-Reader is an optimal tool
for R&D work as well as quality control of immunochromatography reagent kits.
APPLICATIONS
●Immunochromatography Reagent
• Development
• Quality control
Immunochromato-Reader
FEATURES
1High-sensitivity measurement with high
repeatability
The C10066-10 Immunochromato-Reader acquires highly reliable
data with high measurement reproducibility. The C10066-10 in this
way provides powerful support for developing quantitative reagent kit.
Measurements can be made even in low color regions where
visual checks are extremely difficult, so the C10066-10 will prove a
powerful tool for developing high-sensitivity reagent kit.
POWER
12
H A M
A M A
T S U
2Red-based color line and blue-based color line can
be measured with a single unit.
3
4
* PC must be prepared by the user.
567
Supports a wide range of reagents such as for reagent development.
* The two colors cannot be measured simultaneously.
TPMZC0001EA
3Compatible with different
reagent housing
configurations
The supplied flexible tray fits various
reagent kit housings just by adjusting its holders. This flexible tray can
also be used with dip type reagent
kit. (sold separately)
4Color sample provided to check
the C10066-10 main unit makes
daily and periodic
inspections
easy.
Dropping
Section
Reagent kit
Housing
Flexible
Tray
Color sample (supplied)
for checking main unit
TPMZC0002EB
5Automatic color line position detection and
color intensity calculation (mABS)
6Calibration curve function can
be calculated
Color lines are automatically
detected to calculate the color
intensity (absorbance). Highly
reliable data is acquired by a
fully automatic procedure from
the start of measurement to
calculation of absorbance.
Color intensity profile data can
also be stored as a record of
the coloration state.
Calculated calibration curve function converts absorbance data to concentrations on a graph.
●Calibration Curve Graph
7Lot measurement and time-course
measurement modes selectable
Lot measurement mode saves single lot data lists as
CSV files that can then be processed with commercially available spreadsheet software.
Time-course measurement mode monitors absorbance fluctuations that show how the color changes
over time after dropping a sample onto a reagent kit.
This allows finding the difference in reaction speed
between material lots.
●Lot Measurement Data List
●Time-course Measurement Graph
CHARACTERISTICS
DIMENSIONAL OUTLINE (Unit: mm)
●Sensitivity Characteristics (C10066-10 vs. spectroscope)
235
1000
100
POWER
10
TOP VIEW
1
Red-based color line
Blue-based color line
0
0
1
10
100
1000
10000
92
C10066-10 ABSORBANCE (mABS)
10000
215
R=0.997
TPMZB0001EB
USB
FRONT VIEW
RS-232C
6V
REAR VIEW
SPECTROSCOPE ABSORBANCE (mABS)
TPMZA0001EB
SPECIFICATIONS
MAIN UNIT
Parameter
Input Voltage (AC)
Power Consumption
Interface
Light Source
Light Detection
Measurement Object
Dimensions (W × H × D)
Weight
Max.
Description / Value
100 to 240
3
USB 2.0 (cable supplied) or RS-232C (dedicated cable optional)
Green LED / Red LED
Silicon photodiode
Red-based color line / Blue-based color line
215 × 92 × 235 (excluding projecting parts)
1.6
Unit
V
VA
—
—
—
—
mm
kg
Description
Windows® 2000 / XP / Vista / 7 (32 bit / 64 bit)
Lot measurement, time-course measurement (30 second interval / 60 second interval)
Calibration curve function, concentration conversion
Unit
—
—
—
Value
6
3
0.8 to 1.2
3 × 20
5
800
3 % CV (at +25 °C, 20 mABS)
3 % CV (at 20 mABS)
+15 to +30
Below 70
-20 to +45
Below 70
Unit
Lines
mm
mm
mm
mABS
mABS
—
—
°C
%
°C
%
Description / Value
35 × 8 × 95 (See "Accessories" on next page for details.)
10 × 90 (See "Sold separately" on next page for details.)
Unit
mm
mm
SOFTWARE (for operation on PC)
Parameter
Compatible OS
Program
Measurement
Analysis
MAIN UNIT
Parameter
Number of Measurement Lines
Measurement Line Interval (center to center)
Measurement Line Width
Measurable Range
Max.
Min.
Max.
Min.
Sensitivity (absorbance)
Max.
Accuracy (measurement reproducibility, difference between units) Max.
Temperature Characteristics (absorbance fluctuations with temperature) Max.
Ambient Operating Temperature
Ambient Operating Humidity
Storage Temperature
Storage Humidity
COMPATIBLE REAGENT SIZE
Parameter
Reagent Kit Housing (W × H × D)
Dip Type Reagent Kit Strip (W × D)
Immunochromato-Reader
ACCESSORIES
●Flexible Tray (Type No.: A10791)
●Color Sample for Checking Main Unit
(Type No.: A10792)
Dropping Section
3
Measurable
Range
(3 × 20)
45
20
50
Positioning Jig
Holder Tray
Height Adjusting
Plates (4)
Flexible
Tray
●Dedicated AC Adapter (Type No.: A10795)
15 15
●USB Cable (Type No.: A10796)
●CD-ROM
(includes software and instruction manual)
●Setup Guide
* Cable length: AC side 1.8 m
DC side 1.5 m
* Cable length: 1.5 m
SOLD SEPARATELY
●Dedicated RS-232C Cable (Type No.: A10794)
●Dip Type Holder (Type No.: A10793)
Dropping Section
3
45
20
45
Measurable
Range
(3 × 20)
10
* Cable length: 1.5 m;
General-purpose RS-232C cables cannot be used.
●Custom Holders
Custom holders that fit customers' reagent kit housings may be available.
* This unit is not a medical device. Do not use it for medical diagnosis and treatment.
Windows® is a registered trademark of Microsoft Corporation in the United States and / or other countries.
Other product and software names mentioned herein may be either registered trademarks or trademarks of their respective owners.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2013 Hamamatsu Photonics K.K.
HAMAMATSU PHOTONICS K.K.
www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Division
314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected]
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected]
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected]
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: [email protected]
North Europe: Hamamatsu Photonics Norden AB: Thorshamnsgatan 35 SE-164 40 Kista, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected]
TPMZ1006E02
Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: [email protected]
APR. 2013 IP
China: Hamamatsu Photonics (China) Co., Ltd.: 1201 Tower B, Jiaming Center, 27 Dongsanhuan Road North, Chaoyang District, Beijing 100020, China, Telephone: (86)10-6586-6006, Fax: (86)10-6586-2866 E-mail: [email protected]