34201A

34201A
±1 g to ±2 g Triaxial Accelerometer
Zero g Bias Stability ±2 mg
Very Low Noise 110 µg/Hz
Measurement Specialties, Inc.
2236 N. Cleveland-Massillon Rd.
Akron, Ohio 44333-1255 USA
(330) 659-3312 www.meas-spec.com
precisionsensors.meas-spec.com
Parameter
Specifications*
Min
Range
Measurement Full Scale
Sensitivity
At 25°C, Option R002
Drift Tmin to Tmax
Zero g Bias Level
At 25°C
Drift Tmin to Tmax
Alignment
Deviation from Ideal Axes
Transverse Sensitivity
Nonlinearity
Frequency Response
Noise Density
Self Test Pull-up Resistor
Temperature Sensor
Sensitivity
+0°C Bias Level
Outputs
Output Voltage Swing, R001, R1.5
Output Voltage Swing, R002
Power Supply (Vs)
Input Voltage Limits
Input Voltage - Operating
Input Current
Rejection Ratio
Temperature Range (Ta)
Mass
Shock Survival
Data subject to change without notice
†Scale linearly with range option Rnnn
Typical
Max
Units
±2.0
g
1000†
±0.3
mV/g
%
2.5
±2
±0.15
0.25
±0.2
0
110
5
±6
±0.5
±1.25
2100
6.45
509
V
mg
degrees
%
% FSR
Hz
µg/√Hz
kΩ
mV/°C
mV
0.05
0.55
4.95
4.8
V
V
-20
+8.5
13
>120
-40
35
-3500
+36
+36
V
V
mA
dB
°C
grams
g
+85
+3500
*within one year of calibration
Three-Year Limited Warranty
Measurement Specialties, Inc. warrants this product against defects
in materials and workmanship for a period of three years from date of
shipment. During the warranty period, Measurement Specialties will, at
its option, either repair or replace products which prove to be defective.
Exclusions
This warranty does not apply to defects or damage resulting from:
•Improper installation or maintenance
•Customer supplied software or interfacing
•Unauthorized modification or misuse
•Operation outside of the product specifications
Limitations
The warranty set forth above is exclusive and no other warranty,
whether written or oral, is expressed or implied. Measurement
Specialties specifically disclaims the implied warranties of
merchantability and fitness for a particular purpose.
Some jurisdictions do not allow limitations on how long an implied
warranty lasts, so the above limitation or exclusion may not apply to
you. However, any implied warranty of merchantability or fitness is
limited to the three-year duration of this written warranty.
Warranty Service
To obtain service during the warranty period, products must be returned,
transportation prepaid, to Measurement Specialties.
34201A Rev. 8/18/13
How to Use the 34201A
►Connect
►
a +8.5 to +36 V DC power source to pin 8, ground
to pin 9.
Accelerometer Connections
Mounting Instructions
►Use
►
two 3.0 mm x 0.5 mm thread machine screws to attach
to one of the three 34201A mounting faces [recommended
torque 5 lbf-in (0.56 N.m)]. DO NOT thread a screw more
than 2.5 mm (0.100 inch or 5 turns) into any 34201A mounting
face as this can damage the accelerometer and will VOID the
warranty.
►A
► light film of oil or silicone grease can be used to enhance
contact between two mounting surfaces.
►A
► locking compound should be used on all threads to prevent
screws loosening due to vibrations.
Pin
Description
1
2
3
4
5
6
7
8
9
Accelerometer 1 Brown
Signal common Red
Accelerometer 2 Orange
+5V Output
Yellow
Accelerometer 3 Green
Temperature
Blue
Self Test-L
Violet
Power
Grey
Ground*
White
Cable Wire
}
}
}
}
twisted pair
twisted pair
twisted pair
twisted pair
*connected to braided shield at D-shell pin
CAUTION
Improper wiring, particularly Power, may cause
permanent damage and void the warranty.
Pins other than +Vs should not exceed 5 V
►Connect
►
a differential or floating input A/D converter
to each signal and signal common pins (pins 1 and 2 for
Accelerometer 1, pins 3 and 2 for Accelerometer 2, pins 5 and
2 for Accelerometer 3 and pins 6 and 2 for Temperature). For
best results, use a converter with an input impedance greater
than 1 MΩ and common mode rejection ratio greater than
60 dB. Single-ended converters are not recommended, but
if used, leave pin 2 unconnected to eliminate ground loops.
A positive acceleration (marked on the accelerometer label)
causes a positive change on the corresponding output.
►Keep
►
cable lengths between the accelerometer and A/D
converter as short as practical. Use shielded twisted pair
wiring, especially near electric motors or other noise sources.
►As
► a rule of thumb, sample frequency should be at least
10 times the signal bandwidth when linear interpolation is
used for signal reconstruction. Built-in or external low-pass
filtering can be used to lower the noise floor and eliminate
spurious peak measurements from short-duration shock
impulses with insignificant energy content.
►A
► logic 0 on the self-test input will add to all three accelerometer outputs (see calibration certificate for amount). An
output that changes significantly more or less (±20%) than
the certificate amount indicates a defective channel, shorted
pin or broken wire. The self-test signal is approximate and
should not be used for calibration. If the self-test feature will
not be used, pin 7 may be left disconnected.