EN

Charge Amplifier DACU 800
Features
• Multi-range charge amplifier for industrial
application
• 14 selectable ranges
• 13 fixed ranges 100 pC - 1'000'000 pC
• 1 variable range 100 pC - 1'000'000 pC
• Adjustable limit value with switching output
• Peak value
• RS 232 serial interface
Electrical Data
Order Code
Voltage supply15...35 VDC
DACU 800-0.1-1K0BS
Current draw< 70 mA
Measuring range± 100...1'000'000 pC
Output signal± 10 V
Included
Characteristic curve< 1% FS
deviation
• Mounting screws 2 pcs. M4 x 16
Linearity< 0,02% FS
Output offset< ± 5 mV
Noise voltage< 5 mVpp (0,1 Hz...100 kHz)
< 30 mVpp at 100 pC range
Dimensions (mm)
Output impedance10 Ω
Reset operate offsetelectronically compensated
Drift< 0,03 pC/s (2)
Frequency range (-3 dB)0...20 kHz (1)
Control input± 5 V...± 45 V, galv. separated
Switching outputmax. 45 V, max. 100 mA
galv. separated
(1)
@ 100 pC...1'000'000 pC; > 2 kHz @ 1'000'000 pC
(2)
DACU at least 30 min. attached to operational voltage
All specifications at ambient temperature (23°C ±2°C)
Mechanical Data
Control connection25 pin D-Sub
Sensor connectionBNC male
Enclosure materialAluminum die cast
Environmental Conditions
Operating temp. range-5...+60 °C
Storage temperature-20...+80 °C
Protection classIP 40
EMC
EN 61000-6-2 immunity
EN 61000-6-4 emission
9.4
www.baumer.com
DACU 800
Electrical Connections D-Sub 25
Measuring Range Selection
Pin
Function
Range 1
Signal out
3
2 10
pC/10V
2
Peak out
0 0 00
1'000'000
3
Level 1 (In or Out)
0 0 01
500'000
4
Level 0 (In or Out)
0 0 10
200'000
5
Range 3
0 0 11
100'000
6
Range 2
0 1 00
50'000
7
Range 1
0 1 01
20'000
8
Range 0
0 1 10
10'000
9
Supply GND
Measuring range
0 1 11
5'000
10+Supply
1 0 00
2'000
11
Code 2
1 0 01
1'000
12
Code 0
1 0 10
500
13
Code Supply +
1 0 11
200
14
Signal GND
1 1 00
100
15
Alarm 1
16
Alarm 0
1 1 01
1 1 10
1 1 11
100'000...1'000'000
10'000...100'000
100...10'000
17RX
18TX
19
Com Logic Input
20Operate
21
80% Test
22
Supply GND
Accessories
Connecting cable with open end
23NC
Code 1
25
Com Logic Output (Alarm)
L
24
9
50
Control Elements
Range
Adjustment of variable range
L0
Adjustment of value 1. The limit switch voltage
can be set with the potentiometer and has to
be measured at pin 4. Alternatively a voltage can
be supplied as well.
L1
Adjustment limit value 2. ditto, pin 3 only
www.baumer.com
Length(L) Order code
5 m
DZCS 05/DACU 8
9.5