EN

Load cell
DLRP L001
Features
• Passive load cell 0...10 kN
• Compact dimensions
• For compression
• Protection class IP 67
• Stainless steel
Order Code
Technical Data
Standard capacities
0...5000 N
0...10000 N
Sensitivity at FS
1 mV/V Combined error
Linearity
Hysteresis
< 0,5% FS
< 0,5% FS
< 0,5% FS
DLRP L001.
CL05 5 m cable length
CL10 10 m cable length
Load transmission
CO Compression (see drawing)
Measuring range
250 0...5000 N
310 0...10000 N
< ±0,06% /K
Combined error
C 0,5%
Temperature effect span < ±0,05% /K
Zero balance
< ±1% FS
Non-repeatability
< 0,1% FS
Creep error < 0,2% FS (after 30 min. with FS)
Sensitivity tolerance
< ±1% FS
Bridge resistance
Full bridge 350 Ω
Isolation resistance
> 3 GΩ
Excitation max.
7V
Signal polarity
unipolar (compression +1mV/V)
– static load
– dynamic load
Breaking load
150% FS
100% FS
220% FS
Deflection FS
0,05 mm typical
Protection class
IP 67
Cable
2 m, shielded, PUR
Load cell material
1.4542
Connection
FS = Full scale output
Combined error contains linearity, hysteresis and non-repeatability
2.4
CO/
Optional cable length
(2 m standard)
Compensated
0...+70 ˚C
temperature range
Operating temperature
-20...+70 ˚C
range
Storage temperature range -40...+85 ˚C
Temperature effect zero
.C
www.baumer.com
S80 4-pin connector series 712
W24 Cable, 4-wire, open cable end
DLRP L001
Dimensions (mm)
+
ø 8
SR40
2
5,5
14
17
Connector S35A, 4-pin
ø 32 h8
40
Electrical Connection
S80
W24
3
2
4
1
Bridge Circuit
EXC.+
brown
black
blue
white
SIG.+
Pin Color
Signal
+VS
1
+VS
brown
2
-VOUT
black
-VOUT
3
+VOUT
blue
GND
4
GND
white
Housing
+VOUT
SIG.-
EXC.-
Housing
Wiring
DLRP
Transducer
Precision amplifier/
power supply unit
+V S
+VOut
-VOut
Stabilized operating
voltage
Gain
+
-
Zero
GND
RL: >10 MΩ
VS: 5 - 24 VDC
www.baumer.com
2.5