FIDAMAT 6 Catalog Specification Sheet

© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
General information
■ Overview
■ Application
Areas of application
• Environmental protection
• Wastewater (in conjunction with a stripping device, verification
of the hydrocarbon content of liquids)
• TLV (Threshold Limit Value) monitoring at places of work
• Quality monitoring
• Process exhaust monitoring
• Ultra-pure gas measurements in media such as O2, CO2,
inert gases and cold sample gases
• Measurement of corrosive and condensing gases
• Process optimization
The FIDAMAT 6 gas analyzer is suitable for the determination of
the total hydrocarbon content in the air and high-boiling gas mixtures.
■ Benefits
The FIDAMAT 6 gas analyzer is distinguished by its wide range
of applications:
• In the presence of up to 100 % H2O vapor
• In ultra-pure gas applications
• With high-boiling components (up to 200 °C)
• In the presence of corrosive gases (with preliminary filter).
The FIDAMAT 6 exhibits:
• Extremely low cross-sensitivity to interfering gases
• Low consumption of combustion air
• Low influence of oxygen on measured value
The analyzer is additionally equipped with warning and fault
messages:
• For failure of combustion gas
• If the flame is extinguished
• To indicate pump and filter faults
Further applications
• Chemical plants
• Gas manufacturers (ultra-pure gas monitoring)
• Research and development
• Cement industry (measurement of emissions)
• Paint shops and dry-cleaning systems
• Refineries (tank farms, wastewater)
• Drying systems
• Solvent recovery systems
• Pharmaceutical industry
• Automotive industry (engine development, engine and transmission development and certification)
Special applications
Special applications
Special applications are available on request in addition to the
standard combinations, e.g. measuring range 0 to 100 %.
TÜV version
Measurement of flue gases according to 13th BlmSchV/17th
BlmSchV and TA Luft for oil, coal, gas, and waste as fuels.
Furthermore, the TÜV-approved versions of the FIDAMAT 6 comply with the requirements of EN 14956 and QAL 1 according to
EN 14181. Conformity of the analyzers with both standards is
TÜV-certified.
Determination of the analyzer drift according to EN 14181
(QAL 3) can be carried out manually or also with a PC using the
SIPROM GA maintenance and servicing software. In addition,
selected manufacturers of emission evaluation computers offer
the possibility for downloading the drift data via the analyzer’s
serial interface and to automatically record and process it in the
evaluation computer.
Siemens AP 01 · 2015
1/217
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
1
General information
■ Design
• 19" rack unit with 4 HU for installation
- in hinged frame
- in cabinets with or without telescopic rails
• Front plate can be swung down for servicing purposes
(laptop connection)
• Gas connections for sample gas inlet and outlet as well as
combustion gas and combustion air; pipe diameter 6 mm
or ¼"
• Gas and electrical connections at the rear
• Internal gas paths: stainless steel (mat. no. 1.4571)
Display and control panel
• Large LCD field for simultaneous display of:
- Measured value
- Status bar
- Measuring ranges
• Contrast of LCD panel adjustable using menu
• Permanent LED backlighting
• Washable membrane keyboard with five softkeys
• Menu-driven operation for parameterization, test functions,
adjustment
• User help in plain text
• Graphic display of concentration trend; programmable time
intervals
LED backlit graphic
display and
membrane keyboard
with noticeable click
Input and outputs
• One analog output for each measured component
• Two programmable analog inputs
• Six binary inputs freely configurable (e.g. for measurement
range switchover, processing of external signals from sample
preparation)
• Six relay outputs freely configurable (failure, maintenance request, maintenance switch, limit alarm, external solenoid
valves, measuring point switchover)
• Extension with eight additional binary inputs and eight additional relay outputs for autocalibration with up to four calibration gases
Communication
RS 485 present in basic unit (connection from the rear).
Options
• RS 485/RS 232 converter
• RS 485/Ethernet converter
• RS 485/USB converter
• Incorporation in networks via PROFIBUS DP/PA interface
• SIPROM GA software as service and maintenance tool
Status line for display
of analyzer status
(programmable)
Two code levels
according to NAMUR
(maintenance and
specialist level)
Operation with
menu control
using five softkeys
Display of
concentrations as
numbers and bargraph
Display of current
measuring ranges
Display of
start-of-scale and
full-scale values
ESC key
to abort inputs
Keypad to
enter values
INFO key
for help in plain text
CLEAR key
to delete inputs
ENTER key
to accept
input values
FIDAMAT 6, membrane keyboard and graphic display
1/218
Siemens AP 01 · 2015
MEAS key
to return to
measurement mode
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
General information
Designs – parts wetted by sample gas
Gas path
Material
Piping
Stainless steel, mat. no. 1.4571
Gas inlet
Stainless steel, mat. no. 1.4571
Gaskets
Graphite
Sample gas restrictor
Quartz
Auxiliary gas restrictors
Stainless steel, mat. no. 1.4571
Pump membrane
PTFE
Pump head
Stainless steel, mat. no. 1.4571
Detector
Nozzle
Quartz
FID housing
Stainless steel, mat. no. 1.4571
Gas path
Combustion
gas H2
Sample chamber
Pressure sensor
Mechan.
pressure
regulator
Solenoid
valve
Restrictor
Gas outlet
Combustion air
Pressure sensor
Restrictor
Damping
Mechan.
pressure
regulator
Pump 1
Sample gas
Filter
Zero gas
Calibration
gas
FIDAMAT 6 total hydrocarbon analyzer, gas path with pump and with connection for combustion air
Siemens AP 01 · 2015
1/219
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
1
General information
Combustion
gas H2
Sample chamber
Pressure sensor
Mechan.
pressure
regulators
Solenoid
valve
Restrictor
Gas outlet
Combustion air
Pressure sensor
Restrictor
Mechanical
pressure
regulators
Sample gas
Zero gas
Calibration gas
Purging
FIDAMAT 6 total hydrocarbon analyzer, gas path without pump and with connection for combustion air
1/220
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
General information
■ Function
Principle of operation
The FIDAMAT 6 carries out substance-specific measurements
and not component-specific measurements. It measures the total of all hydrocarbons in a sample gas, but with different weighting of the hydrocarbon molecules. To a first approximation, the
display is proportional to the number of C atoms in the respective molecule. However, there are fluctuations in practice. The
display deviation for the respective molecule is expressed by the
response factor.
The sample gas is supplied to the FIDAMAT 6 through overpressure or drawn in by the built-in diaphragm pump (optionally via
a heated line and an additional filter) and passed on to the flame
ionization detector via an obstruction-proof fused-silica restrictor.
In the detector, the hydrocarbons in the sample gas are burned
in an oxyhydrogen gas flame. Burning partially ionizes the proportion of organically-bound hydrocarbons. The released ions
are converted into an ionic current by the voltage present between two electrodes, and measured using a highly sensitive
amplifier. The current measured is proportional to the quantity of
organically-bound C atoms in the sample gas.
The FIDAMAT 6 provides various messages in the form of floating contacts:
• Maintenance request
E.g. sample gas flow (filter/pump)
Fan failure (advance warning for measuring accuracy)
The measured value remains unaffected.
• Fault
E.g. hydrogen, combustion air and sample gas pressures,
temperature, analyzer part and pump, fault in the electronics
(temperature).
The measured value may be influenced.
• Failure
In the event of failure of, for example, the electronics, power
supply, combustion gas, combustion air or sample gas, the
analyzer automatically shuts down (the combustion gas valve
is closed).
Note
The sample gases must be fed into the analyzers free of dust.
Condensation should be avoided. Therefore, the use of gas
modified for the measuring task is necessary in most application
cases.
A pressure regulator keeps the combustion gas pressure constant. The balanced system of pump, capillary tubes, and pressure regulator for combustion air ensures that the sample gas
pressure is kept constant.
Calibration
When the analyzer is switched on, ignition is carried out automatically when the setpoint temperature has been reached and,
for versions "with pump", the pump is also started up.
We recommend N2 as zero gas (at least 5.0; for measuring of hydrocarbons < 1 ppm: at least 6.0).
The calibrating interval should be adapted to the respective
measuring task.
The calibration gas should have a concentration of at least 60 %
of the leading measuring range. The concentration of residual
hydrocarbons must not exceed 0.1 ppm.
For pure gas measurement, use suitable additional gases.
Example:
200 V DC
Exhaust
gas
Combustion air
Combustion gas
1. Emission measurement
Measuring range
0 ... 50 mg C/m3
Zero gas
N2 (5.0 or better)
Calibration gas
21 ppm C3H8 in N2
(corresponds to 31.43 mg C/m3 at 20 °C)
2. Purity measurement in 100 % O2
Measuring range
0 ... 50 ppm C1
Zero gas
N2 (5.0 or better)
Calibration gas
At least 30 ppm CH4 in O2
OVEN
Sample gas
FIDAMAT 6, principle of operation
Siemens AP 01 · 2015
1/221
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
1
General information
Essential characteristics
• Four freely parameterizable measuring ranges, also with suppressed zero, all measuring ranges linear
• Galvanically isolated measured-value output 0/2/4 to 20 mA
(also inverted)
• Autoranging possible; remote switching is also possible
• Storage of measured values possible during adjustments
• Measuring range identification
• Measuring point switchover for up to 6 measuring points
• Measuring point identification
• Wide range of selectable time constants (static/dynamic noise
suppression); i.e. the response time of the device can be
adapted to the respective measuring task
• Easy handling thanks to menu-driven operation
• Low long-term drift
• Two control levels with their own authorization codes for the
prevention of accidental and unauthorized operator interventions
• Automatic, parameterizable measuring range calibration
• Operation based on the NAMUR recommendation
• Customer-specific analyzer options such as:
- Customer acceptance
- TAG labels
- Drift recording
• Wear-free, corrosion-proof filter housing
• No blocking of the sample gas capillaries through the use of a
quartz restrictor
• Purge function in the event of analyzer or power supply failure
(avoids build-up of toxic and corrosive substances in the device)
• Low consumption of combustion air
• Response factors comply with the minimum requirements in
accordance with German air purity guidelines and the Working Group of the German automotive Industry
• Simple handling using a numerical membrane keyboard and
operator prompting
Response factors (examples, mean values)
Substance
Mean response factor
n-butane
1.00
n-propane
1.00
n-heptane
1.00
Cyclohexane
1.08
Isopropanol
0.81
Toluene
1.06
Acetone
0.92
Ethyl acetate
0.76
Isobutyl acetate
0.83
Methane
1.06
Ethane
0.99
n-hexane
1.01
iso-octane
1.04
Ethine (acetylene)
0.91
Propene
0.84
Methanol
0.87
Ethanol
0.83
Ethanoic acid
1.13
Methyl acetate
0.67
Benzene
1.01
Ethyl benzene
0.96
p-xylene
1.03
Dichloromethane
1.13
Trichloroethene
1.01
Tetrachlorethene
1.07
Chloroform
0.72
Chlorobenzene
1.15
Cross-interferences (examples)1)
Interfering
component
Concentration of the
interfering component
Induced
cross-interference
O2 in N2
(21 vol. %)
< 0.3 mg/m3
(258
mg/m3)
< 0.15 mg/m3
NO in N2
(310
mg/m3)
< 0.5 mg/m3
NO2 in synth. air
(146 mg/m3)
< 0.1 mg/m3
CO in N2
(461 mg/m3)
< 0.15 mg/m3
CO2 in N2
(18 vol. %)
< 0.1 mg/m3
HCl in N2
mg/m3)
< 0.3 mg/m3
SO2 in N2
1)
1/222
Siemens AP 01 · 2015
(78
With measuring range 0 to 15 mg/m3.
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
19" rack unit
■ Technical specifications
General information
Measuring ranges
4, internally and externally switchable; manual and autoranging
possible
Smallest possible measuring span
0 ... 10 vpm
Largest possible measuring span
99.999 vpm*)
Concentration units
ppm, C1, C3, C6 or mgC/m3
Autoranging
Hysteresis, selectable
Measured-value display
Digital concentration display
(5 digits with floating point)
Resolution of digital display
0.1 % of measured value
Operating position
Conformity
Oven temperature
Adjustable, 100 ... 200 °C
Measuring response
(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample
gas flow and 25 °C ambient temperature)
Output signal fluctuation
< 0.75 % of the smallest possible
measuring range according to
rating plate, with electronic
damping constant of 1 s (corresponds to ± 0.25 % at 2 σ)
Zero point drift
< 0.5 %/month of the smallest
possible measuring span according to rating plate
Measured-value drift
< 1 %/week of the current measuring range
Front wall, vertical
Repeatability
CE mark in accordance with
EN 50081-1, EN 50082-2
< 1 % of the current measuring
range
Detection limit
0.1 ppm (version for ultra-pure
gas measurement: 50 ppb)
Linearity error
< 1 % of the current measuring
range
Design, enclosure
Degree of protection
IP20 according to EN 60529
Weight
Approx. 23 kg
Electrical characteristics
Influencing variables
(relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample
gas flow and 25 °C ambient temperature)
100 ... 120 V AC (nominal range
of use 90 ... 132 V), 48 ... 63 Hz or
200 ... 240 V AC (nominal range
of use 180 ... 264 V), 48 ... 63 Hz
Ambient temperature
< 1 %/10 K referred to smallest
possible span according to rating
plate
Atmospheric pressure
< 1 %/50 hPa
Power consumption
• Approx. 150 VA during operation,
• Approx. 350 VA during warm-up
phase
Sample gas pressure
< 2 % of the current measuring
range range/1 % pressure
change (within 600 ... 1 100 hPa)
Power supply
EMC
(Electromagnetic Compatibility)
In accordance with standard
requirements of NAMUR NE21
(08/98)
< 1 % of the current measuring
range with rated voltage ± 10 %
Position influence
< 1 % with < 15° inclination
Electrical safety
In accordance with EN 61010-1,
overvoltage category II
Analog output
0/2/4 ... 20 mA, isolated;
max. load 750 Ω
Fuse values
• 100 ... 120 V: 4.0T/250
• 200 ... 240 V: 2.5 T/250
Relay outputs
6, with changeover contacts,
freely parameterizable, e.g. for
measuring range identification;
load: 24 V AC/DC/1 A, potentialfree
• Without pump
< 2 000 hPa abs.
Analog inputs
• With integrated pump
600 ... 1 100 hPa
Sample gas flow
18 ... 60 l/h (0.3 ... 1 l/min)
Sample gas temperature
0 ... 200 °C
2, dimensioned for 0/2/4 to 20 mA
for external pressure sensor and
correction of influence of accompanying gas (correction of crossinterference)
Sample gas humidity
< 90 % RH (RH: relative humidity)
Binary inputs
6, designed for 24 V, floating,
freely parameterizable, e.g. for
measuring range switchover
Serial interface
RS 485
Options
AUTOCAL function with 8 additional binary inputs and relay outputs each, also with PROFIBUS
PA or PROFIBUS DP
Power supply
Electrical inputs and outputs
Gas inlet conditions
Permissible sample gas pressure
Dynamic response
Warm-up period
At room temperature,
approx. 2 ... 3 h
Delayed display (T90)
2 ... 3 s
Damping (electrical time constant)
0 ... 100 s, parameterizable
Dead time (purging time of the gas
path in the unit at 1 l/min)
With filter, 2 ... 3 s
Time for device-internal signal
processing
<1s
Climatic conditions
Permissible ambient temperature
5 ... 45 °C during operation,
-30 ... +70 °C during storage and
transportation
Permissible humidity
< 90 % RH (RH: relative humidity)
as annual average, during storage and transportation (must not
fall below dew point)
*) 100 % as special application
Siemens AP 01 · 2015
1/223
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
1
19" rack unit
FIDAMAT 6 with pump and heated oven, with combustion air connection
Gases
Operating pressure
Inlet pressure
Pump startup
Without
With
hPa (abs.)
Flow through FID
Flow through bypass
ml/min
ml/min
~ 25
—
hPa (abs.)
hPa (abs.)
Combustion gas
3 000 ... 5 000
2 000 ± 20
Combustion air
3 000 ... 5 000
1 420 ± 20
1 500
~ 320
~ 500
Sample gas
~ 1000
—
1 500 ± 2
~3
~ 1 000
Zero gas
3 500 ... 4 000
—
1 500 ± 2
~3
~ 1 000
Calibration gas
3 500 ... 4 000
—
1 500 ± 2
~3
~ 1 000
Flow through FID
Flow through bypass
ml/min
ml/min
FIDAMAT 6 without pump, with heated oven, with combustion air connection
Gases
Operating pressure
Inlet pressure
Sample/calibration gas
Without
With
hPa (abs.)
hPa (abs.)
hPa (abs.)
Combustion gas
3 000 ... 5 000
2 000 ± 20
~ 25
—
Combustion air
3 000 ... 5 000
1 480 ± 5
—
~ 320
~ 300
Sample gas
1 500 ... 2 000
—
1 500 ± 2
~3
~ 500
Zero gas
1 500 ... 2 000
—
1 500 ± 2
~3
~ 500
Calibration gas
1 500 ... 2 000
—
1 500 ± 2
~3
~ 500
The supply gases (combustion gas, combustion air) must have a degree of purity of 5.0 in order to guarantee correct measurements.
The degree of purity must be increased in the case of very small hydrocarbon concentrations (< 1 ppm).
1/224
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
19" rack unit
■ Selection and ordering data
FIDAMAT 6 gas analyzer
19" rack unit for installation in cabinets
Article No.
7MB2421- 7 7 7 7 7 - 7 7 A 7
Click on the Article No. for the online configuration in the PIA Life Cycle Portal.
Gas connections
Pipe with 6 mm outer diameter
Pipe with ¼" outer diameter
0
1
Version
Without pump, for sample gas with overpressure
With heated pump, for sample gas with atm. pressure
B
D
Combustion air feed
With connection for combustion air
A
Number of channels
1-channel version
1
Add-on electronics
Without
AUTOCAL function
• With 8 additional digital inputs/outputs
• With 8 digital inputs/8 digital outputs, PROFIBUS PA interface
• With 8 digital inputs/8 digital outputs, PROFIBUS DP interface
0
1
6
7
Power supply
100 ... 120 V AC, 48 ... 63 Hz
200 ... 240 V AC, 48 ... 63 Hz
0
1
Combustion gases
H2
A
Language (supplied documentation, software)
German
English
French
Spanish
Italian
Additional versions
0
1
2
3
4
Order code
Add "-Z" to Article No. and specify Order code
Telescopic rails (2 units)
A31
TAG labels (specific lettering based on customer information)
B03
Clean for O2 service (specially cleaned gas path)
Y02
Measuring range indication in plain text, if different from the standard setting
Y11
Special setting (only in conjunction with an application No.)
Y12
Extended special setting (only in conjunction with an application No.)
Y13
TÜV version acc. to 17th BlmSchV
Y17
Accessories
Article No.
RS 485/Ethernet converter
A5E00852383
RS 485/RS 232 converter
C79451-Z1589-U1
RS 485/USB converter
A5E00852382
AUTOCAL function each with 8 digital inputs/outputs
C79451-A3480-D511
AUTOCAL function 8 digital inputs/outputs each and PROFIBUS PA
A5E00057307
AUTOCAL function 8 digital inputs/outputs each and PROFIBUS DP
A5E00057312
Set of Torx screwdrivers
A5E34821625
Siemens AP 01 · 2015
1/225
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
19" rack unit
■ Dimensional drawings
Front view
Side view
&/($5
(6&
(17(5
0($6
86.5
177
101.6
s
,1)2
),'$0$7
9
465
483
M4
426.5
Top view
347
155.5
27.5
Rear view
FIDAMAT 6, 19“ unit, dimensions in mm
1/226
Siemens AP 01 · 2015
446
440
426.5
1
0
80
155
234.5 264.5
370 395
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
19" rack unit
■ Schematics
Pin assignment (electrical and gas connections)
&RQQHFWRU68%')56
0
*1'
9
0
*1'
5B/HYHO1
1&
5'7'1
5'7'3
5B/HYHO3
1&
1&
*1'
,WLVSRVVLEOHWRFRQQHFW
EXVWHUPLQDWLQJUHVLVWRUV
WRSLQVDQG
68%')FRQQHFWRU
0
0
0
0
*1'
1&
1&
$QDORJRXWSXW3
$QDORJRXWSXW1
1&
1&
$QDORJLQSXW3
$QDORJLQSXW1
$QDORJLQSXW3
$QDORJLQSXW1
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXWWR1
*1'
$QDORJRXWSXWVLVRODWHG
DOVRIURPHDFKRWKHU5/ืΩ
&RUUHFWLRQRISUHVVXUHRU
FURVVLQWHUIHUHQFH
&RUUHFWLRQRIFURVVLQWHUIHUHQFH
&RUUHFWLRQRIFURVVLQWHUIHUHQFH
$QDORJLQSXWV
QRQLVRODWHG
P$Ω
RU9
ORZUHVLVWDQFH
68%')FRQQHFWRU
0
*1'
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXWWR1
,VRODWHGYLDRSWRFRXSOHU
99
99
5HOD\
5HOD\
5HOD\
5HOD\
&RQWDFWORDG
PD[9$$&'&
UHOD\FRQWDFWVVKRZQ
UHOD\FRLOKDV]HURFXUUHQW
5HOD\
5HOD\
*1'
0
1RWH
$OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDO
EORFNVPXVWEHVKLHOGHGDQGUHVWDJDLQVW
WKHHQFORVXUHSRWHQWLDO
FIDAMAT 6, 19" rack unit, pin assignment
Siemens AP 01 · 2015
1/227
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
1
19" rack unit
&RQQHFWRU68%')RSWLRQ
0
*1'
1&
1&
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXW3
%LQDU\LQSXWWR1
,VRODWHG
YLDRSWRFRXSOHU
99
99
5HOD\
5HOD\
5HOD\
&RQWDFWORDG
PD[9$$&'&
IRUWKHUHOD\FRQWDFWV
VKRZQ
WKHUHOD\KDV]HURFXUUHQW
5HOD\ 5HOD\
5HOD\
5HOD\
0
&RQQHFWRU68%')
352),%86'3
5HOD\
*1'
RSWLRQDO
'*1'
&1751
&1753GLUHFWLRQFRQWURO
5['7['1$
5['7['3$
&RQQHFWRU68%'0
352),%863$
3$1
3$3
939
1RWH
$OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDOEORFNVPXVW
EHVKLHOGHGDQGUHVWDJDLQVWWKHHQFORVXUHSRWHQWLDO
FIDAMAT 6, 19" rack unit, pin assignment of the AUTOCAL board and PROFIBUS connectors
1/228
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
19" rack unit
Combustion gas
8
1
15-pin connector:
binary inputs and
analog inputs/outputs
Combustion air
2
3
4
5
6
7
9-pin
connector:
RS 485
9-pin interface
connector (option)
(e.g. PROFIBUS)
37-pin connector:
binary inputs and
relay outputs
(option board)
Power supply
and fuses
Sample gas inlet
Calibration
gas
Sample gas outlet
Zero gas
25-pin connector:
binary inputs and
relay outputs
Gas connections
1 ... 7 Swagelok, 6 mm or ¼ “
8
Stubs , 6 mm or ¼ “
FIDAMAT 6, gas connections and pin assignment, version with pump
Combustion air
Calibration gas
Combustion gas
8
Zero gas
2
3
4
5
6
7
9-pin
connector:
RS 485
9-pin interface
connector (option)
(e.g. PROFIBUS)
37-pin connector:
binary inputs and
relay outputs
(option board)
Power supply
and fuses
Sample inlet
Sample gas outlet
Gas connections
2 ... 7 Swagelok, 6 mm or ¼ “
8
Stubs , 6 mm or ¼ “
Flushing
Outlet
25-pin connector:
binary inputs and
relay outputs
15-pin connector:
binary inputs and
analog inputs/outputs
FIDAMAT 6, gas connections and pin assignment, version with pump
Siemens AP 01 · 2015
1/229
1
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
1
Documentation
■ Selection and ordering data
Operating instructions
Article No.
FIDAMAT 6
Gas analyzer for determination of
total hydrocarbon concentration
• German
A5E00221703
• English
A5E00222135
• French
A5E00222138
• Spanish
A5E00222141
• Italian
A5E00222144
FIDAMAT 6-G
Gas analyzer for determination of
total hydrocarbon content
• German
A5E00476038
• English
A5E00478463
• French
A5E00478466
• Spanish
A5E00478468
• Italian
A5E00478469
1/230
Siemens AP 01 · 2015
© Siemens AG 2015
Continuous Gas Analyzers, extractive
FIDAMAT 6
Suggestions for spare parts
■ Selection and ordering data
Article No. FIDAMAT 6
Description
2 years
(quantity)
5 years
(quantity)
With pump
Without pump
Analyzer unit
FI detector, complete
1
A5E00295816
A5E00295816
Sample gas path
Pump (KNF)
1
1
A5E00882121
Set of gaskets for pump (KNF)
4
10
A5E03792459
Filter, with gasket for sample gas
1
3
A5E00248845
Pressure regulators
1
1
A5E00248851
A5E00248851
Gasket for pressure regulator
1
2
A5E00295107
A5E00295107
Filter, complete (sample gas inlet, 6 mm)
1
A5E00295928
Filter, complete (sample gas inlet, ¼")
1
A5E00295976
Solenoid valve (1-way)
1
2
A5E00296562
Solenoid valve (2-way)
1
2
A5E00296565
A5E00296562
Gasket, PTFE, 1.5 mm (20 units)
1
2
C79451-A3040-D101
C79451-A3040-D101
Gasket, graphite, 0.5 to 1 mm (20 units)
1
2
C79451-A3040-D102
C79451-A3040-D102
Gasket, graphite, 1.5 mm (20 units)
1
2
C79451-A3040-D103
C79451-A3040-D103
Gasket, graphite, 3 mm (20 units)
1
2
C79451-A3040-D105
C79451-A3040-D105
Pressure ring, 1 mm (20 units)
1
C79451-A3040-D112
C79451-A3040-D112
Pressure ring, 1.5 mm (20 units)
1
C79451-A3040-D113
C79451-A3040-D113
Pressure ring, 3 mm (20 units)
1
A5E00295333
A5E00295333
Outer rings, 0.5 … 1 mm (20 units)
1
C79451-A3040-D121
C79451-A3040-D121
Outer rings, 1.5 … 3 mm (1/8") (20 units)
1
C79451-A3040-D122
C79451-A3040-D122
Electronics
Front plate
1
1
A5E00248790
A5E00248790
Adapter plate
1
1
A5E00248795
A5E00248795
Temperature fuse (retrofitting set)
1
2
A5E01040317
A5E01040317
A5E01040312
A5E01040312
Temperature fuse (spare part), from N1-V3-940
onwards
Fusible element, 230 V AC
2
3
A5E00248819
A5E00248819
Fusible element, 110 V AC
2
3
A5E00248822
A5E00248822
LC display
1
1
A5E00248920
A5E00248920
Cable, temperature sensor for oven
1
A5E00283770
A5E00283770
Cable, temperature sensor for analyzer part
1
A5E00283780
A5E00283780
Cable, magnetic distributor
1
A5E00283800
A5E00283800
Cable, heater for oven, 230 V AC
1
A5E00283817
A5E00283817
Cable, heater for oven, 110 V AC
1
A5E00295469
A5E00295469
Cable, electrode voltage, complete
1
A5E00284092
A5E00284092
Cable, signal cable
1
A5E00284094
A5E00284094
Cable, connecting cable (4-pole)
1
1
A5E00284095
A5E00284095
Cable, connecting cable (5-pole)
1
1
A5E00284096
A5E00284096
1
A5E00313839
A5E00313839
Axial-flow fan, 24 V DC
If the device was supplied with a specially cleaned gas path for high oxygen context ("Clean for O2 service"), please ensure that you
specify this when ordering spare parts. This is the only way to guarantee that the gas path will continue to comply with the special
requirements for this version.