© Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 General information ■ Overview ■ Application Areas of application • Environmental protection • Wastewater (in conjunction with a stripping device, verification of the hydrocarbon content of liquids) • TLV (Threshold Limit Value) monitoring at places of work • Quality monitoring • Process exhaust monitoring • Ultra-pure gas measurements in media such as O2, CO2, inert gases and cold sample gases • Measurement of corrosive and condensing gases • Process optimization The FIDAMAT 6 gas analyzer is suitable for the determination of the total hydrocarbon content in the air and high-boiling gas mixtures. ■ Benefits The FIDAMAT 6 gas analyzer is distinguished by its wide range of applications: • In the presence of up to 100 % H2O vapor • In ultra-pure gas applications • With high-boiling components (up to 200 °C) • In the presence of corrosive gases (with preliminary filter). The FIDAMAT 6 exhibits: • Extremely low cross-sensitivity to interfering gases • Low consumption of combustion air • Low influence of oxygen on measured value The analyzer is additionally equipped with warning and fault messages: • For failure of combustion gas • If the flame is extinguished • To indicate pump and filter faults Further applications • Chemical plants • Gas manufacturers (ultra-pure gas monitoring) • Research and development • Cement industry (measurement of emissions) • Paint shops and dry-cleaning systems • Refineries (tank farms, wastewater) • Drying systems • Solvent recovery systems • Pharmaceutical industry • Automotive industry (engine development, engine and transmission development and certification) Special applications Special applications Special applications are available on request in addition to the standard combinations, e.g. measuring range 0 to 100 %. TÜV version Measurement of flue gases according to 13th BlmSchV/17th BlmSchV and TA Luft for oil, coal, gas, and waste as fuels. Furthermore, the TÜV-approved versions of the FIDAMAT 6 comply with the requirements of EN 14956 and QAL 1 according to EN 14181. Conformity of the analyzers with both standards is TÜV-certified. Determination of the analyzer drift according to EN 14181 (QAL 3) can be carried out manually or also with a PC using the SIPROM GA maintenance and servicing software. In addition, selected manufacturers of emission evaluation computers offer the possibility for downloading the drift data via the analyzer’s serial interface and to automatically record and process it in the evaluation computer. Siemens AP 01 · 2015 1/217 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 1 General information ■ Design • 19" rack unit with 4 HU for installation - in hinged frame - in cabinets with or without telescopic rails • Front plate can be swung down for servicing purposes (laptop connection) • Gas connections for sample gas inlet and outlet as well as combustion gas and combustion air; pipe diameter 6 mm or ¼" • Gas and electrical connections at the rear • Internal gas paths: stainless steel (mat. no. 1.4571) Display and control panel • Large LCD field for simultaneous display of: - Measured value - Status bar - Measuring ranges • Contrast of LCD panel adjustable using menu • Permanent LED backlighting • Washable membrane keyboard with five softkeys • Menu-driven operation for parameterization, test functions, adjustment • User help in plain text • Graphic display of concentration trend; programmable time intervals LED backlit graphic display and membrane keyboard with noticeable click Input and outputs • One analog output for each measured component • Two programmable analog inputs • Six binary inputs freely configurable (e.g. for measurement range switchover, processing of external signals from sample preparation) • Six relay outputs freely configurable (failure, maintenance request, maintenance switch, limit alarm, external solenoid valves, measuring point switchover) • Extension with eight additional binary inputs and eight additional relay outputs for autocalibration with up to four calibration gases Communication RS 485 present in basic unit (connection from the rear). Options • RS 485/RS 232 converter • RS 485/Ethernet converter • RS 485/USB converter • Incorporation in networks via PROFIBUS DP/PA interface • SIPROM GA software as service and maintenance tool Status line for display of analyzer status (programmable) Two code levels according to NAMUR (maintenance and specialist level) Operation with menu control using five softkeys Display of concentrations as numbers and bargraph Display of current measuring ranges Display of start-of-scale and full-scale values ESC key to abort inputs Keypad to enter values INFO key for help in plain text CLEAR key to delete inputs ENTER key to accept input values FIDAMAT 6, membrane keyboard and graphic display 1/218 Siemens AP 01 · 2015 MEAS key to return to measurement mode © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 General information Designs – parts wetted by sample gas Gas path Material Piping Stainless steel, mat. no. 1.4571 Gas inlet Stainless steel, mat. no. 1.4571 Gaskets Graphite Sample gas restrictor Quartz Auxiliary gas restrictors Stainless steel, mat. no. 1.4571 Pump membrane PTFE Pump head Stainless steel, mat. no. 1.4571 Detector Nozzle Quartz FID housing Stainless steel, mat. no. 1.4571 Gas path Combustion gas H2 Sample chamber Pressure sensor Mechan. pressure regulator Solenoid valve Restrictor Gas outlet Combustion air Pressure sensor Restrictor Damping Mechan. pressure regulator Pump 1 Sample gas Filter Zero gas Calibration gas FIDAMAT 6 total hydrocarbon analyzer, gas path with pump and with connection for combustion air Siemens AP 01 · 2015 1/219 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 1 General information Combustion gas H2 Sample chamber Pressure sensor Mechan. pressure regulators Solenoid valve Restrictor Gas outlet Combustion air Pressure sensor Restrictor Mechanical pressure regulators Sample gas Zero gas Calibration gas Purging FIDAMAT 6 total hydrocarbon analyzer, gas path without pump and with connection for combustion air 1/220 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 General information ■ Function Principle of operation The FIDAMAT 6 carries out substance-specific measurements and not component-specific measurements. It measures the total of all hydrocarbons in a sample gas, but with different weighting of the hydrocarbon molecules. To a first approximation, the display is proportional to the number of C atoms in the respective molecule. However, there are fluctuations in practice. The display deviation for the respective molecule is expressed by the response factor. The sample gas is supplied to the FIDAMAT 6 through overpressure or drawn in by the built-in diaphragm pump (optionally via a heated line and an additional filter) and passed on to the flame ionization detector via an obstruction-proof fused-silica restrictor. In the detector, the hydrocarbons in the sample gas are burned in an oxyhydrogen gas flame. Burning partially ionizes the proportion of organically-bound hydrocarbons. The released ions are converted into an ionic current by the voltage present between two electrodes, and measured using a highly sensitive amplifier. The current measured is proportional to the quantity of organically-bound C atoms in the sample gas. The FIDAMAT 6 provides various messages in the form of floating contacts: • Maintenance request E.g. sample gas flow (filter/pump) Fan failure (advance warning for measuring accuracy) The measured value remains unaffected. • Fault E.g. hydrogen, combustion air and sample gas pressures, temperature, analyzer part and pump, fault in the electronics (temperature). The measured value may be influenced. • Failure In the event of failure of, for example, the electronics, power supply, combustion gas, combustion air or sample gas, the analyzer automatically shuts down (the combustion gas valve is closed). Note The sample gases must be fed into the analyzers free of dust. Condensation should be avoided. Therefore, the use of gas modified for the measuring task is necessary in most application cases. A pressure regulator keeps the combustion gas pressure constant. The balanced system of pump, capillary tubes, and pressure regulator for combustion air ensures that the sample gas pressure is kept constant. Calibration When the analyzer is switched on, ignition is carried out automatically when the setpoint temperature has been reached and, for versions "with pump", the pump is also started up. We recommend N2 as zero gas (at least 5.0; for measuring of hydrocarbons < 1 ppm: at least 6.0). The calibrating interval should be adapted to the respective measuring task. The calibration gas should have a concentration of at least 60 % of the leading measuring range. The concentration of residual hydrocarbons must not exceed 0.1 ppm. For pure gas measurement, use suitable additional gases. Example: 200 V DC Exhaust gas Combustion air Combustion gas 1. Emission measurement Measuring range 0 ... 50 mg C/m3 Zero gas N2 (5.0 or better) Calibration gas 21 ppm C3H8 in N2 (corresponds to 31.43 mg C/m3 at 20 °C) 2. Purity measurement in 100 % O2 Measuring range 0 ... 50 ppm C1 Zero gas N2 (5.0 or better) Calibration gas At least 30 ppm CH4 in O2 OVEN Sample gas FIDAMAT 6, principle of operation Siemens AP 01 · 2015 1/221 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 1 General information Essential characteristics • Four freely parameterizable measuring ranges, also with suppressed zero, all measuring ranges linear • Galvanically isolated measured-value output 0/2/4 to 20 mA (also inverted) • Autoranging possible; remote switching is also possible • Storage of measured values possible during adjustments • Measuring range identification • Measuring point switchover for up to 6 measuring points • Measuring point identification • Wide range of selectable time constants (static/dynamic noise suppression); i.e. the response time of the device can be adapted to the respective measuring task • Easy handling thanks to menu-driven operation • Low long-term drift • Two control levels with their own authorization codes for the prevention of accidental and unauthorized operator interventions • Automatic, parameterizable measuring range calibration • Operation based on the NAMUR recommendation • Customer-specific analyzer options such as: - Customer acceptance - TAG labels - Drift recording • Wear-free, corrosion-proof filter housing • No blocking of the sample gas capillaries through the use of a quartz restrictor • Purge function in the event of analyzer or power supply failure (avoids build-up of toxic and corrosive substances in the device) • Low consumption of combustion air • Response factors comply with the minimum requirements in accordance with German air purity guidelines and the Working Group of the German automotive Industry • Simple handling using a numerical membrane keyboard and operator prompting Response factors (examples, mean values) Substance Mean response factor n-butane 1.00 n-propane 1.00 n-heptane 1.00 Cyclohexane 1.08 Isopropanol 0.81 Toluene 1.06 Acetone 0.92 Ethyl acetate 0.76 Isobutyl acetate 0.83 Methane 1.06 Ethane 0.99 n-hexane 1.01 iso-octane 1.04 Ethine (acetylene) 0.91 Propene 0.84 Methanol 0.87 Ethanol 0.83 Ethanoic acid 1.13 Methyl acetate 0.67 Benzene 1.01 Ethyl benzene 0.96 p-xylene 1.03 Dichloromethane 1.13 Trichloroethene 1.01 Tetrachlorethene 1.07 Chloroform 0.72 Chlorobenzene 1.15 Cross-interferences (examples)1) Interfering component Concentration of the interfering component Induced cross-interference O2 in N2 (21 vol. %) < 0.3 mg/m3 (258 mg/m3) < 0.15 mg/m3 NO in N2 (310 mg/m3) < 0.5 mg/m3 NO2 in synth. air (146 mg/m3) < 0.1 mg/m3 CO in N2 (461 mg/m3) < 0.15 mg/m3 CO2 in N2 (18 vol. %) < 0.1 mg/m3 HCl in N2 mg/m3) < 0.3 mg/m3 SO2 in N2 1) 1/222 Siemens AP 01 · 2015 (78 With measuring range 0 to 15 mg/m3. © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 19" rack unit ■ Technical specifications General information Measuring ranges 4, internally and externally switchable; manual and autoranging possible Smallest possible measuring span 0 ... 10 vpm Largest possible measuring span 99.999 vpm*) Concentration units ppm, C1, C3, C6 or mgC/m3 Autoranging Hysteresis, selectable Measured-value display Digital concentration display (5 digits with floating point) Resolution of digital display 0.1 % of measured value Operating position Conformity Oven temperature Adjustable, 100 ... 200 °C Measuring response (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) Output signal fluctuation < 0.75 % of the smallest possible measuring range according to rating plate, with electronic damping constant of 1 s (corresponds to ± 0.25 % at 2 σ) Zero point drift < 0.5 %/month of the smallest possible measuring span according to rating plate Measured-value drift < 1 %/week of the current measuring range Front wall, vertical Repeatability CE mark in accordance with EN 50081-1, EN 50082-2 < 1 % of the current measuring range Detection limit 0.1 ppm (version for ultra-pure gas measurement: 50 ppb) Linearity error < 1 % of the current measuring range Design, enclosure Degree of protection IP20 according to EN 60529 Weight Approx. 23 kg Electrical characteristics Influencing variables (relating to sample gas pressure 1 013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient temperature) 100 ... 120 V AC (nominal range of use 90 ... 132 V), 48 ... 63 Hz or 200 ... 240 V AC (nominal range of use 180 ... 264 V), 48 ... 63 Hz Ambient temperature < 1 %/10 K referred to smallest possible span according to rating plate Atmospheric pressure < 1 %/50 hPa Power consumption • Approx. 150 VA during operation, • Approx. 350 VA during warm-up phase Sample gas pressure < 2 % of the current measuring range range/1 % pressure change (within 600 ... 1 100 hPa) Power supply EMC (Electromagnetic Compatibility) In accordance with standard requirements of NAMUR NE21 (08/98) < 1 % of the current measuring range with rated voltage ± 10 % Position influence < 1 % with < 15° inclination Electrical safety In accordance with EN 61010-1, overvoltage category II Analog output 0/2/4 ... 20 mA, isolated; max. load 750 Ω Fuse values • 100 ... 120 V: 4.0T/250 • 200 ... 240 V: 2.5 T/250 Relay outputs 6, with changeover contacts, freely parameterizable, e.g. for measuring range identification; load: 24 V AC/DC/1 A, potentialfree • Without pump < 2 000 hPa abs. Analog inputs • With integrated pump 600 ... 1 100 hPa Sample gas flow 18 ... 60 l/h (0.3 ... 1 l/min) Sample gas temperature 0 ... 200 °C 2, dimensioned for 0/2/4 to 20 mA for external pressure sensor and correction of influence of accompanying gas (correction of crossinterference) Sample gas humidity < 90 % RH (RH: relative humidity) Binary inputs 6, designed for 24 V, floating, freely parameterizable, e.g. for measuring range switchover Serial interface RS 485 Options AUTOCAL function with 8 additional binary inputs and relay outputs each, also with PROFIBUS PA or PROFIBUS DP Power supply Electrical inputs and outputs Gas inlet conditions Permissible sample gas pressure Dynamic response Warm-up period At room temperature, approx. 2 ... 3 h Delayed display (T90) 2 ... 3 s Damping (electrical time constant) 0 ... 100 s, parameterizable Dead time (purging time of the gas path in the unit at 1 l/min) With filter, 2 ... 3 s Time for device-internal signal processing <1s Climatic conditions Permissible ambient temperature 5 ... 45 °C during operation, -30 ... +70 °C during storage and transportation Permissible humidity < 90 % RH (RH: relative humidity) as annual average, during storage and transportation (must not fall below dew point) *) 100 % as special application Siemens AP 01 · 2015 1/223 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 1 19" rack unit FIDAMAT 6 with pump and heated oven, with combustion air connection Gases Operating pressure Inlet pressure Pump startup Without With hPa (abs.) Flow through FID Flow through bypass ml/min ml/min ~ 25 — hPa (abs.) hPa (abs.) Combustion gas 3 000 ... 5 000 2 000 ± 20 Combustion air 3 000 ... 5 000 1 420 ± 20 1 500 ~ 320 ~ 500 Sample gas ~ 1000 — 1 500 ± 2 ~3 ~ 1 000 Zero gas 3 500 ... 4 000 — 1 500 ± 2 ~3 ~ 1 000 Calibration gas 3 500 ... 4 000 — 1 500 ± 2 ~3 ~ 1 000 Flow through FID Flow through bypass ml/min ml/min FIDAMAT 6 without pump, with heated oven, with combustion air connection Gases Operating pressure Inlet pressure Sample/calibration gas Without With hPa (abs.) hPa (abs.) hPa (abs.) Combustion gas 3 000 ... 5 000 2 000 ± 20 ~ 25 — Combustion air 3 000 ... 5 000 1 480 ± 5 — ~ 320 ~ 300 Sample gas 1 500 ... 2 000 — 1 500 ± 2 ~3 ~ 500 Zero gas 1 500 ... 2 000 — 1 500 ± 2 ~3 ~ 500 Calibration gas 1 500 ... 2 000 — 1 500 ± 2 ~3 ~ 500 The supply gases (combustion gas, combustion air) must have a degree of purity of 5.0 in order to guarantee correct measurements. The degree of purity must be increased in the case of very small hydrocarbon concentrations (< 1 ppm). 1/224 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 19" rack unit ■ Selection and ordering data FIDAMAT 6 gas analyzer 19" rack unit for installation in cabinets Article No. 7MB2421- 7 7 7 7 7 - 7 7 A 7 Click on the Article No. for the online configuration in the PIA Life Cycle Portal. Gas connections Pipe with 6 mm outer diameter Pipe with ¼" outer diameter 0 1 Version Without pump, for sample gas with overpressure With heated pump, for sample gas with atm. pressure B D Combustion air feed With connection for combustion air A Number of channels 1-channel version 1 Add-on electronics Without AUTOCAL function • With 8 additional digital inputs/outputs • With 8 digital inputs/8 digital outputs, PROFIBUS PA interface • With 8 digital inputs/8 digital outputs, PROFIBUS DP interface 0 1 6 7 Power supply 100 ... 120 V AC, 48 ... 63 Hz 200 ... 240 V AC, 48 ... 63 Hz 0 1 Combustion gases H2 A Language (supplied documentation, software) German English French Spanish Italian Additional versions 0 1 2 3 4 Order code Add "-Z" to Article No. and specify Order code Telescopic rails (2 units) A31 TAG labels (specific lettering based on customer information) B03 Clean for O2 service (specially cleaned gas path) Y02 Measuring range indication in plain text, if different from the standard setting Y11 Special setting (only in conjunction with an application No.) Y12 Extended special setting (only in conjunction with an application No.) Y13 TÜV version acc. to 17th BlmSchV Y17 Accessories Article No. RS 485/Ethernet converter A5E00852383 RS 485/RS 232 converter C79451-Z1589-U1 RS 485/USB converter A5E00852382 AUTOCAL function each with 8 digital inputs/outputs C79451-A3480-D511 AUTOCAL function 8 digital inputs/outputs each and PROFIBUS PA A5E00057307 AUTOCAL function 8 digital inputs/outputs each and PROFIBUS DP A5E00057312 Set of Torx screwdrivers A5E34821625 Siemens AP 01 · 2015 1/225 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 19" rack unit ■ Dimensional drawings Front view Side view &/($5 (6& (17(5 0($6 86.5 177 101.6 s ,1)2 ),'$0$7 9 465 483 M4 426.5 Top view 347 155.5 27.5 Rear view FIDAMAT 6, 19“ unit, dimensions in mm 1/226 Siemens AP 01 · 2015 446 440 426.5 1 0 80 155 234.5 264.5 370 395 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 19" rack unit ■ Schematics Pin assignment (electrical and gas connections) &RQQHFWRU68%')56 0 *1' 9 0 *1' 5B/HYHO1 1& 5'7'1 5'7'3 5B/HYHO3 1& 1& *1' ,WLVSRVVLEOHWRFRQQHFW EXVWHUPLQDWLQJUHVLVWRUV WRSLQVDQG 68%')FRQQHFWRU 0 0 0 0 *1' 1& 1& $QDORJRXWSXW3 $QDORJRXWSXW1 1& 1& $QDORJLQSXW3 $QDORJLQSXW1 $QDORJLQSXW3 $QDORJLQSXW1 %LQDU\LQSXW3 %LQDU\LQSXW3 %LQDU\LQSXWWR1 *1' $QDORJRXWSXWVLVRODWHG DOVRIURPHDFKRWKHU5/ืΩ &RUUHFWLRQRISUHVVXUHRU FURVVLQWHUIHUHQFH &RUUHFWLRQRIFURVVLQWHUIHUHQFH &RUUHFWLRQRIFURVVLQWHUIHUHQFH $QDORJLQSXWV QRQLVRODWHG P$Ω RU9 ORZUHVLVWDQFH 68%')FRQQHFWRU 0 *1' %LQDU\LQSXW3 %LQDU\LQSXW3 %LQDU\LQSXW3 %LQDU\LQSXW3 %LQDU\LQSXWWR1 ,VRODWHGYLDRSWRFRXSOHU 99 99 5HOD\ 5HOD\ 5HOD\ 5HOD\ &RQWDFWORDG PD[9$$&'& UHOD\FRQWDFWVVKRZQ UHOD\FRLOKDV]HURFXUUHQW 5HOD\ 5HOD\ *1' 0 1RWH $OOFDEOHVWRWKHFRQQHFWRUVRUWHUPLQDO EORFNVPXVWEHVKLHOGHGDQGUHVWDJDLQVW WKHHQFORVXUHSRWHQWLDO FIDAMAT 6, 19" rack unit, pin assignment Siemens AP 01 · 2015 1/227 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 1 19" rack unit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rack unit, pin assignment of the AUTOCAL board and PROFIBUS connectors 1/228 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 19" rack unit Combustion gas 8 1 15-pin connector: binary inputs and analog inputs/outputs Combustion air 2 3 4 5 6 7 9-pin connector: RS 485 9-pin interface connector (option) (e.g. PROFIBUS) 37-pin connector: binary inputs and relay outputs (option board) Power supply and fuses Sample gas inlet Calibration gas Sample gas outlet Zero gas 25-pin connector: binary inputs and relay outputs Gas connections 1 ... 7 Swagelok, 6 mm or ¼ “ 8 Stubs , 6 mm or ¼ “ FIDAMAT 6, gas connections and pin assignment, version with pump Combustion air Calibration gas Combustion gas 8 Zero gas 2 3 4 5 6 7 9-pin connector: RS 485 9-pin interface connector (option) (e.g. PROFIBUS) 37-pin connector: binary inputs and relay outputs (option board) Power supply and fuses Sample inlet Sample gas outlet Gas connections 2 ... 7 Swagelok, 6 mm or ¼ “ 8 Stubs , 6 mm or ¼ “ Flushing Outlet 25-pin connector: binary inputs and relay outputs 15-pin connector: binary inputs and analog inputs/outputs FIDAMAT 6, gas connections and pin assignment, version with pump Siemens AP 01 · 2015 1/229 1 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 1 Documentation ■ Selection and ordering data Operating instructions Article No. FIDAMAT 6 Gas analyzer for determination of total hydrocarbon concentration • German A5E00221703 • English A5E00222135 • French A5E00222138 • Spanish A5E00222141 • Italian A5E00222144 FIDAMAT 6-G Gas analyzer for determination of total hydrocarbon content • German A5E00476038 • English A5E00478463 • French A5E00478466 • Spanish A5E00478468 • Italian A5E00478469 1/230 Siemens AP 01 · 2015 © Siemens AG 2015 Continuous Gas Analyzers, extractive FIDAMAT 6 Suggestions for spare parts ■ Selection and ordering data Article No. FIDAMAT 6 Description 2 years (quantity) 5 years (quantity) With pump Without pump Analyzer unit FI detector, complete 1 A5E00295816 A5E00295816 Sample gas path Pump (KNF) 1 1 A5E00882121 Set of gaskets for pump (KNF) 4 10 A5E03792459 Filter, with gasket for sample gas 1 3 A5E00248845 Pressure regulators 1 1 A5E00248851 A5E00248851 Gasket for pressure regulator 1 2 A5E00295107 A5E00295107 Filter, complete (sample gas inlet, 6 mm) 1 A5E00295928 Filter, complete (sample gas inlet, ¼") 1 A5E00295976 Solenoid valve (1-way) 1 2 A5E00296562 Solenoid valve (2-way) 1 2 A5E00296565 A5E00296562 Gasket, PTFE, 1.5 mm (20 units) 1 2 C79451-A3040-D101 C79451-A3040-D101 Gasket, graphite, 0.5 to 1 mm (20 units) 1 2 C79451-A3040-D102 C79451-A3040-D102 Gasket, graphite, 1.5 mm (20 units) 1 2 C79451-A3040-D103 C79451-A3040-D103 Gasket, graphite, 3 mm (20 units) 1 2 C79451-A3040-D105 C79451-A3040-D105 Pressure ring, 1 mm (20 units) 1 C79451-A3040-D112 C79451-A3040-D112 Pressure ring, 1.5 mm (20 units) 1 C79451-A3040-D113 C79451-A3040-D113 Pressure ring, 3 mm (20 units) 1 A5E00295333 A5E00295333 Outer rings, 0.5 … 1 mm (20 units) 1 C79451-A3040-D121 C79451-A3040-D121 Outer rings, 1.5 … 3 mm (1/8") (20 units) 1 C79451-A3040-D122 C79451-A3040-D122 Electronics Front plate 1 1 A5E00248790 A5E00248790 Adapter plate 1 1 A5E00248795 A5E00248795 Temperature fuse (retrofitting set) 1 2 A5E01040317 A5E01040317 A5E01040312 A5E01040312 Temperature fuse (spare part), from N1-V3-940 onwards Fusible element, 230 V AC 2 3 A5E00248819 A5E00248819 Fusible element, 110 V AC 2 3 A5E00248822 A5E00248822 LC display 1 1 A5E00248920 A5E00248920 Cable, temperature sensor for oven 1 A5E00283770 A5E00283770 Cable, temperature sensor for analyzer part 1 A5E00283780 A5E00283780 Cable, magnetic distributor 1 A5E00283800 A5E00283800 Cable, heater for oven, 230 V AC 1 A5E00283817 A5E00283817 Cable, heater for oven, 110 V AC 1 A5E00295469 A5E00295469 Cable, electrode voltage, complete 1 A5E00284092 A5E00284092 Cable, signal cable 1 A5E00284094 A5E00284094 Cable, connecting cable (4-pole) 1 1 A5E00284095 A5E00284095 Cable, connecting cable (5-pole) 1 1 A5E00284096 A5E00284096 1 A5E00313839 A5E00313839 Axial-flow fan, 24 V DC If the device was supplied with a specially cleaned gas path for high oxygen context ("Clean for O2 service"), please ensure that you specify this when ordering spare parts. This is the only way to guarantee that the gas path will continue to comply with the special requirements for this version. Siemens AP 01 · 2015 1/231 1