CTE9000 / CTU9000 Series OEM pressure transmitters for industrial media FEATURES · 0...0.35 to 0...35 bar, 0...5 to 0...500 psi gage1 or absolute10 · 0...10 V, 0.5...4.5 V, 0...5 V, 1...6 V or 4...20 mA output · Field interchangeable · For industrial use · All welded stainless steel diaphragm construction MEDIA COMPATIBILITY Wetted materials: stainless steel 1.4404 (316)9 Housing: stainless steel, protection class IP 64 (according to DIN EN 60529) respectively NEMA 4X1 ELECTRICAL CONNECTION SPECIFICATIONS11,12 Maximum ratings Supply voltage (reverse polarity protection) CTE(M)/CTU9...0 12...32 V CTE(M)/CTU9...1, ...6, ...7 9...32 V CTE(M)/CTU9...42 9...32 V 0...10 V, 0.5...4.5 V, 0...5 V, 1...6 V output +Vs sensor Maximum load current (source) CTE(M)/CTU9...0, ...1, ...6, ...7 12 (9) to 32 V 1 Vout 2 -Vs 1 mA V GND 4 Temperature limits Storage Operating Compensated Humidity limits Vibration (5 to 500 Hz) Mechanical shock -55 to 100°C -40 to 100°C 0 to 70°C December 2008 / 509 4...20 mA output +Vs 0 - 98 %RH 10 gRMS 50 g sensor 1 N/C 1000 mbar 15 psi 2 x rated pressure 9 to 32 V 2 V -Vs RL 4 3 Proof pressure CTEM9350... CTU9005... all others case 3 GND case 3 1/6 www.sensortechnics.com CTE9000 / CTU9000 Series OEM pressure transmitters for industrial media COMMON PERFORMANCE CHARACTERISTICS Characteristics Operating pressure Min. Typ. Max. CTEM9350... 0 0.35 CTE9001... CTE9N01... CTE9P01... CTE9002... CTE9003... CTE9005... CTE9010... CTE9016... CTE9020... CTE9035... 0 -1 -1 0 0 0 0 0 0 0 1 1 0 2 3 5 10 16 20 35 CTU9005... 0 5 CTU9015... CTU9N15... CTU9P15 CTU9030... CTU9050... CTU9100... CTU9300... CTU9500... 0 -15 -15 0 0 0 0 0 15 15 0 30 50 100 300 500 Thermal effects (0 to 70°C)4 Offset Span ±0.02 ±0.02 Thermal effects (-40 to 0°C, 70 to 100°C) Offset Span ±0.03 ±0.03 Non-linearity and hysteresis (BSL)5 ±0.2 Repeatability ±0.1 Long term stability6 ±0.1 ±0.05 ±0.05 Unit bar p si %FSO/°C ±0.5 %FSO Output noise (0 < f < 1 kHz) ±0.04 Response time (10 to 90 %) 1 Power supply rejection December 2008 / 509 Offset CTE(M)/CTU9...4 all others ±0.05 ±0.002 Span CTE(M)/CTU9...4 all others ±0.08 ±0.002 ms %FSO/V 2/6 www.sensortechnics.com CTE9000 / CTU9000 Series OEM pressure transmitters for industrial media INDIVIDUAL PERFORMANCE CHARACTERISTICS 0...10 V output (VS = 15 V, RL > 100 kΩ, tamb = 25°C) Characteristics Zero pressure offset 7 Full scale span Min. Typ. Max. CT...9N... 4.9 5 5.1 all others -0.1 0 0.1 CT...9N... 4.9 5 5.1 all others 9.9 10 10.1 Unit V 25 Ω 4 9.5 mA Min. Typ. Max. Unit CT...9N... 2.45 2.5 2.55 all others 0.45 0.5 0.55 CT...9N... 1.95 2 2.05 all others 3.95 4 4.05 Output impedance Current consumption (no load) 0.5...4.5 V output (VS = 15 V, RL > 100 kΩ, tamb = 25°C) Characteristics Zero pressure offset Full scale span7 V 25 Ω 4 8.5 mA Min. Typ. Max. Unit CT...9N... 2.45 2.5 2.55 all others -0.05 0 0.05 CT...9N... 2.45 2.5 2.55 all others 4.95 5.0 5.05 Output impedance Current consumption (no load) 0...5 V output (VS = 15 V, RL > 100 kΩ, tamb = 25°C) Characteristics Zero pressure offset Full scale span7 25 Ω 4 8.5 mA Unit Output impedance Current consumption (no load) V 1...6 V output (VS = 15 V, RL > 100 kΩ, tamb = 25°C) Characteristics Zero pressure offset 7 Full scale span Min. Typ. Max. CT...9N... 3.45 3.5 3.55 all others 0.95 1 1.05 CT...9N... 2.45 2.5 2.55 all others 4.95 5.0 5.05 V 25 Ω 4 8.5 mA Min. Typ. Max. Unit CT...9N... 11.8 12.0 12.2 all others 3.8 4.0 4.2 CT...9N... 7.8 8.0 8.2 all others 15.8 16.0 16.2 Output impedance Current consumption (no load) 4...20 mA output (VS = 15 V, RL = 100 Ω, tamb = 25°C) Characteristics Zero pressure offset 7 Full scale span Power consumption (IL = 20 mA) December 2008 / 509 260 mA mW 3/6 www.sensortechnics.com CTE9000 / CTU9000 Series OEM pressure transmitters for industrial media Specification notes: 1. IP 64 protection is given when the connector is locked. For proper function the gage port is vented to the atmosphere through the connector/cable assembly. Thus the cable end must have access to the ambient pressure. 2. The minimum supply voltage is directly proportional to the load resistance seen by the transmitter. For more details see the load limitation diagram. 3. Proof pressure is the maximum pressure which may be applied without causing damage to the sensing element. 4. Thermal effects tested and guaranteed from 0 to 70°C relative to 25°C. All specifications shown are relative to 25°C. 5. Non-linearity refers to the Best Straight Line fit measured for offset, full scale span and 1/2 full scale span. 6. Long term stability is the change in output after one year or 1 million pressure cycles. 7. Span is the arithmetic difference in transmitter output signal measured at zero pressure and the maximum operating pressure. 8. Tests are in accordance with EN61000-6-2, April 1999. 9. When using devices with optional nickel plated fittings, consider the media compatibility of the fittings also. 10. Available for pressure ranges from 1 bar (15 psi) absolute upwards only. 11. CE-labelling is in accordance with 89/336/EEC. 12. The pressure transmitters must not be used as safety accessories according to article 1, 2.1.3 of the directive 97/23/EC. OUTLINE DRAWING NOM 85.0 (3.35) 65.0 (2.56) vent hole¹ 4 G 1/8" (BSP) 1 3 M12 x 1 21.8 (0.86) diam. 6.5 (0.26) Detail A 2 Detail A 4.0 (0.16) diam. Output Pin o-ring grove for o-ring 11.8 x 1.8 (0.456 x 0.071) (ZA004029) mass: 82 g December 2008 / 509 1 2 3 4 0...10 V 0.5...4.5 V 0...5 V, 1...6 V +VS Vout case -Vs 4...20 mA +VS NC case -Vs dimensions in mm (inches) 4/6 www.sensortechnics.com CTE9000 / CTU9000 Series OEM pressure transmitters for industrial media RECOMMENDED ACCESSORY (not included in delivery) ZK000101: ZP000112-B: Connector/cable assembly (different cable lenghts available) Mating Connector (without cable) NOM 2 m PIN CONNECTION (other cable lenghts on request) 36.5 NOM 60 Shield Pin Flying lead end 1 Brown 2 Green 3 White and shield 4 Yellow Vent tube Note: For proper function of all gage devices the vent tube must have access to the ambient atmosphere. M12x1 Ø 20 OPTIONAL PRESSURE FITTINGS (brass, nickel plated) B Male fittings dimensions in mm LOAD LIMITATION 4...20 mA output version 1400 Ohm 1200 C A D 1000 800 G F Female fittings 400 200 E H 0 B 8 (0.315) 12 (0.472) 11.5 (0.453) 16 (0.630) 12.5 (0.492) 8.5 (0.335) 12.5 (0.492) 10 (0.394) 14 (0.551) December 2008 / 509 C 13 (0.512) 5.5 (0.217) 5 (0.197) 7 (0.276) 11 (0.433) 5 (0.197) 7 (0.276) 13 (0.512) 6 (0.236) 5 9 16 20 25 30 32 V supply voltage Dimensions in mm (inches) Dimensions in mm (inches) A 1/8" BSPT 1/4" BSPT 3/8" BSPT 1/2" BSPT 1/8" BSP 1/4" BSP 3/8" BSP 1/8" NPT 1/4" NPT op e ra ratin ng g e 600 D (Hex.) 14 (9/16") 14 (9/18") 17 (11/16") 22 (7/8") 14 (9/16") 19 (3/4") 22 (7/8") 17 (11/16") 22 (7/8") E 1/8" BSP 1/4" BSP 3/8" BSP 1/2" BSP F 5 (0.197) 7 (0.276) 6 (0.236) 18 (0.707) G 15 (0.591) 20 (0.787) 20 (0.787) 23 (0.906) H (Hex.) 14 (9/16") 17 (11/16") 22 (7/8") 24 (15/16") 5/6 www.sensortechnics.com CTE9000 / CTU9000 Series OEM pressure transmitters for industrial media ELECTROMAGNETIC CAPABILITY8 Test conditions Criterion Interference Radiated, radio frequency EN61000-4-3: Grade 3, 10 V/m, 80 to 1000 MHz A <1 %FSO electromagnetic field immunity (RFI) 80 % AMC (1 kHz) Electrical fast transient / EN61000-4-4: Grade 3, ±2 kV B <1 %FSO burst immunity (EFT) Electrostatic discharge immunity EN61000-4-2: Grade 4, ±8 kV, contact discharge B <1 %FSO test (ESD) Immunity to conducted disturbances EN61000-4-6: Grade 3, 0.15 to 80 MHz A <1 %FSO induced by radio-frequency fields 10 V, 80 % AMC (1 kHz) ORDERING INFORMATION CTx E: U: M 9xxx x x x bar calibration psi calibration Output signal 0: 0...10 V 1: 1...6 V 4: 4...20 mA 6: 0.5...4.5 V 7: 0...5 V mbar-ranges (only) CTE9000 series 350: 0 to 350 mbar (gage only) 001: 0 to 1 bar N01: -1 to +1 bar (gage only) P01: -1 to 0 bar (gage only) 002: 0 to 2 bar 003: 0 to 3 bar 005: 0 to 5 bar 010: 0 to 10 bar 016: 0 to 16 bar 020: 0 to 20 bar 035: 0 to 35 bar CTU9000 series 005: 0 to 5 psi (gage only) 015: 0 to 15 psi N15: -15 to +15 psi (gage only) P15: -15 to 0 psi (gage only) 030: 0 to 30 psi 050: 0 to 50 psi 100: 0 to 100 psi 300: 0 to 300 psi 500: 0 to 500 psi Pressure mode G: gage pressure 1 A: absolute pressure (available for pressure ranges from 1 bar /15 psi absolute upwards only) Note: Fitting size9 D: 1/8" BSPT E: 1/4" BSPT F: 3/8" BSPT G: 1/2" BSPT male, male, male, male, brass, brass, brass, brass, nickel nickel nickel nickel plated plated plated plated K: L: M: N: 1/8" 1/4" 1/8" 1/4" NPT male, brass NPT male, brass NPT male, SS 1.4305 (303) NPT male, SS 1.4305 (303) P: Q: R: S: G G G G 1/8" 1/4" 3/8" 1/2" (BSP) (BSP) (BSP) (BSP) male, male, male, male, U: V: W: X: G G G G 1/8" 1/4" 3/8" 1/2" (BSP) (BSP) (BSP) (BSP) female, female, female, female, Y: G 1/8" (BSP) male, SS 1.4404 (316) no optional fitting brass, brass, brass, brass, nickel nickel nickel nickel brass, brass, brass, brass, plated plated plated plated nickel nickel nickel nickel plated plated plated plated Other pressure ranges and options are widely available. Please contact your nearest Sensortechnics sales representative. Sensortechnics reserves the right to make changes to any products herein. Sensortechnics does not assume any liability arising out of the application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others. December 2008 / 509 6/6 www.sensortechnics.com