© Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 ■ Overview The pressure drop through the sensor is a function of the properties of the fluid and the flow rate. A pressure loss and accuracy calculator can be found on the Siemens Internet site www.siemens.com/fc430/sizer The preferred flow direction is indicated by an arrow on the sensor. Flow in the direction of the arrow will be measured as positive. The flow direction can be adjusted at the transmitter to compensate for reverse installation. Installation orientation The optimal installation orientation is vertical with the flow upwards. This ensures that suspended solids or bubbles are completely pushed through the sensor. A drain valve below the sensor will allow the pipe and sensor to drain completely. Supports In order to support the weight of the flowmeter and to ensure reliable measurements when external effects exist (e.g. plant vibrations), the sensor should be installed in rigidly supported pipelines. The flow measuring principle is based on the Coriolis Effect. The FCS400 sensor’s measuring tubes are energized by an electromechanical driver circuit which oscillates them at their resonance frequency. Two pick-ups are placed symmetrically upstream and downstream of the central driver. When a process fluid passes through the sensor, the Coriolis Effect will act on the vibrating tubes and cause deflection which can be measured as a phase shift between pick-ups 1 and 2. The phase shift is proportional to the mass flow rate. The amplitude of the driver is automatically regulated to ensure a stable output from both of the pickups. The temperatures of the sensor tubes and frame are measured with high precision to provide compensation for changes with temperature in the measuring properties. The sensor signals are analyzed for flow, density and fluid temperature in the sensor front end. The digital signal is controlled to conform to high Safety Integrated Level (SIL) and sent digitally to the transmitter via standard cable. The FCT030 further calculates total mass and volume, fraction, dosing control and many other functions. The front-end module has a process noise filter, which can be used to improve the meter’s performance when installation and application conditions are not ideal. Typical interferences from process conditions such as pump pulsations, mechanical vibrations, oscillating valves can be reduced considerably. ■ Integration The SITRANS FCS400 Massflow sensor is suitable for both indoor and outdoor installation and meets the requirements of Protection Class IP67/NEMA 4X. Optionally the sensor can be ordered with hazardous certification to Zone 1 + 20/21 (ATEX, IECEx, EAC Ex, FM, CSA, NEPSI, INMETRO) or Class I + II + III Div. 1 (FM). The flowmeter is bidirectional and can be installed in any orientation. The sensor is self-draining in many positions, with vertical mounting preferred. It is important to ensure that the sensor tubes are always completely filled with homogeneous fluid; otherwise measuring errors may occur. Suitable fluids are clean liquids, pastes, light slurries or gases. Condensing vapours, aerated liquids or slush are not recommended. The materials in contact with the process medium must be evaluated for corrosion and erosion resistances for long sensor life. Supports or hangers should be installed symmetrically and stress-free in close proximity to both of the process connections. Shut-off devices To conduct a system zero adjustment, secure shut-off devices are required in the pipeline. Where possible, shut-off devices should be installed both upstream and downstream of the flowmeter. System design • The sensor design consists of process connections, inlet and outlet manifolds mounted in a stiff frame and two parallel tubes equally sharing the process medium flow. The meter is protected in a pressure-rated stainless steel enclosure with two purge ports to support a pressure guard in non-Ex applications. • The sensing tubes are curved in the CompactCurve shape which gives high sensitivity and low pressure loss. The CompactCurve shape was selected to ensure that the smallest flows are measured with optimal signal to noise ratio. • Vibration mode separation creates a controlled measuring environment only within the CompactCurve part of the tubes. As a result the sensor has high immunity to plant vibration while avoiding large mass balancing of the meter components. • The 15° slope of the CompactCurve shape ensures secure self-draining when the sensor axis is mounted vertically or up to 10° off vertical. • The sensor frame is designed to conduct plant vibrations directly through the sensor body to adjacent pipeline while providing isolation of the metering section from the vibration. Careful mounting of the pipeline with regard to minimizing vibration at the meter will ensure a secure measurement environment. ■ Installation guidelines • The mass flowmeter does not require any flow conditioning or straight inlet pipe sections. Care should be exercised however to ensure that any upstream valves, gates, sight glasses etc. do not cavitate and are not set into vibration by the flow. • It is always preferred to place the flowmeter upstream of any control valve (what goes in, comes out) or other pipeline component which may cause flashing, cavitation or vibrations. • The presence of gas bubbles in the fluid may result in erroneous measurements, particularly in the density measurement. Therefore the flowmeter should not be installed at the lowest pressure point in the liquid piping system or where vapour can collect. Install the meter in pipeline sections with high pressure to maintain system pressure and compress any bubbles. Siemens FI 01 · 2016 3/151 3 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 3 • Drop lines downstream from the flow sensor should be avoided to prevent the meter tube from draining during flowing conditions. A back-pressure device or orifice is recommended to ensure that flow does not separate within the flow sensor but the metering section remains at positive pressure at all times while there is flow. • The flowmeter should not come into contact with any other objects. Avoid making attachments to the housing except for the pressure guard components (if required). • When the connecting pipeline is larger than the sensor size, suitable standard reducers may be installed. A selection of oversize and undersize connections can be ordered - refer to the sizes tables below. • The flow sensor may be supported at the junction between process connection and the manifold, but should not be used to support adjacent piping. Ensure that the piping is also supported on both sides so that connection stresses are neutral. • If strong vibrations exist in the pipeline, they should be damped using elastic pipeline elements. The damping devices must be installed outside the supported flowmeter section. Direct connection of flexible elements to the sensor should be avoided. • Make sure that any dissolved gases, which are present in many liquids, do not outgas. The back pressure at the outlet should be at least 0.2 bar (3 psi) above the vapour pressure of the process fluid. • Assure that operation below the vapour pressure cannot occur particularly for fluids with low latent heat of vaporisation. • The sensor should not be installed in the vicinity of strong electromagnetic fields, e.g. near motors, pumps, variable frequency drives, transformers etc. • When operating meters on a common mounting base the sensors should be mounted and spaced separate from each other to avoid cross-talk and other vibration interferences. • When operating meters in interconnected pipelines the pipes should be decoupled to prevent cross talk. Remote system cabling The system is designed so that standard instrumentation cable with four cores and overall screen or two screened pairs can be used, or cable sets can be ordered with the flowmeter. The cable can be ordered in various set lengths and terminated in the field. The maximum design length for the sensor cable is 200 m (656 ft), limited to 150 m (492 ft) for Ex applications with Class IIC gases. Data transmission speed and process variable update rates may be affected by the cable characteristics. For best results, choose a cable with the following electrical characteristics: Property Unit Value Resistance [/km] 59 Characteristic impedance [] 100 @ 1 MHz Insulation resistance [M/km] 200 Maximum voltage [V] 300 The flowmeter system applies maximum 15 V DC in operation and is certified intrinsically safe. The complete system is insulation tested to 1500 V in production. Cabling solutions which can be ordered with the flowmeter are as follows: 1. High performance plugged cable using M12 plugs into prepared sockets 2. Cable glands for either metric or NPT threaded terminal housings. 3. Plain cable in set lengths to be passed through flexible and rigid conduit (not supplied) for metric or NPT threaded terminal housings Cable for items 1, 2 and 3 are available either gray for standard applications or light blue for Ex applications to identify the circuit as intrinsically safe. Insulation and heating For applications where pipeline insulation is required for personnel protection or process temperature maintenance, the SITRANS FCS400 flow sensor may also be insulated. The form and material of insulation is not prescribed and entirely depends on the practices at the application location or plant. Insulation must not be crowded around the sensor pedestal but shaped at a 45° cone to allow the pedestal to radiate excess heat and maintain a suitable working temperature within the front-end housing. Where trace heating is employed, an electric heating jacket can be ordered as an accessory. It is shaped to the sensor body and controlled from a weatherproof setpoint device. The jacket can heat the sensor enclosure up to 200 °C (392 °F). However further insulation is also recommended for personnel protection or low loss temperature maintenance. Calibration To ensure accurate measurement all flowmeters must be initially calibrated. The calibration of each SITRANS FCS400 coriolis sensor is conducted at SIEMENS flow facilities accredited according to ISO/IEC 17025 by DANAK. A calibration certificate for every sensor is stored on the SensorFlash SD card. The accreditation body DANAK has signed the ILAC MRA agreement (International Laboratory Accreditation Corporation - Mutual Recognition Arrangement). Therefore the accreditation ensures international traceability and recognition of the test results in 39 countries worldwide, including the US (NIST traceability). 3/152 Siemens FI 01 · 2016 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 ■ Technical specifications Flow sensor FCS400 Parameter Unit Value Process pressure range [barg (psi)] 316L: 0 ... 100 (0 ... 1450) Hastelloy C22: 0 ... 160 (0 ... 2321) Process temperature range [°C (°F)] -50 ... +200 (-58 ... +392) Ambient temperature range [°C (°F)] -40 ... +60 (-40 ... +140) Transport temperature range [°C (°F)] -40 ... +70 (-40 ... +158) Density range [kg/m3 (lb/ft3)] 1 ... 5000 (0.062 ... 312.2) Process media Fluid group 1 (suitable for dangerous fluids) Form Light slurry, liquid and non-condensing gas 3 No. of process values • Primary process values • Mass flow • Density • Process medium temperature • Derieved process values • Volume flow • Corrected volume flow (with reference density) • Fraction A:B • Fraction % A:B Performance specifications Sensor Parameter Unit DN 15 DN 25 DN 50 DN 80 Max. zero point error [kg/h (lb/min)] 0.2 (0.007) 2.0 (0.072) 7.5 (0.276) 18 (0.66) Qmin [kg/h (lb/min)] 20 (0.735) 200 (7.35) 750 (27.6) 900 (33.1) Qnom [kg/h (lb/min)] 3 700 (136.0) 11 500 (422.6) 52 000 (1 911) 136 000 (4 997) Qmax [kg/h (lb/min)] 6 400 (235.2) 17 700 (650.4) 70 700 (2598) 181 000 (6651) Linearity error mass flow [%] ± 0.1 ± 0.1 ± 0.1 ± 0.1 Repeatability mass flow [%] ± 0.05 ± 0.05 ± 0.05 ± 0.05 Density accuracy standard calibration [kg/m3 (lb/ft3)] ± 5 (± 0.31) ± 5 (± 0.31) ± 5 (± 0.31) ± 5 (± 0.31) Density accuracy extended calibration [kg/m3 (lb/ft3)] ± 0.5 (± 0.031) ± 0.5 (± 0.031) ± 0.5 (± 0.031) ± 0.5 (± 0.031) Temperature error [°C (°F)] ± 0.5 (± 0.9) ± 0.5 (± 0.9) ± 0.5 (± 0.9) ± 0.5 (± 0.9) Pressure/temperature curves Process pressure (bar) Metric flange ratings, EN 1092-1 120 100 ANSI Flange Ratings per ASME B16.5 Process pressure (bar) With two major exceptions, the pressure rating of the flow sensors is independent of the process medium temperature. Design rules for flange connections in both the EN1092-1 and ASME B16.5 standards dictate pressure derating with increasing temperature. The charts below show the effect of process medium temperature on the pressure ratings for the flanges within the FCS400 product program. 120 100 80 Class 600 60 40 Class Graphic 300 follows end of the week 20 Class 150 0 -50 PN 100 0 80 60 PN 63 40 50 100 150 200 Process temperature (°C) ASME B16.5 flanged sensors Graphic PN 40 follows end of the week 20 PN 16 0 -50 0 50 100 150 200 Process temperature (°C) EN1092-1 flanged sensors Siemens FI 01 · 2016 3/153 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 Sensor variants SITRANS FCS400 sensors are available in three main variants: Standard, hygienic and NAMUR. A wide range of process connections is available for the FCS400 sensors. The available combinations of type, sensor size and connection size are shown in the tables below. 12-VCO-4 Quick connect JIS B2200:2004/10K JIS B2200:2004/20K JIS B2200:2004/40K ● SMS 1145 Hygineic screwed DIN 11864-2A Aseptic flanged ● ISO 2853 Hygienic screwed DIN 11864-1A Aseptic secrewed ● ISO 2852 Hygienic clamped DIN 32676 Hygienic Tri-clamp DIN 11851 Hygienic screwed ASME B1.20.1 NPT pipe thread ISO 228-1 G pipe thread ANSI B16.5-2009, class 600 ANSI B16.5-2009, class 300 ANSI B16.5-2009, class 150 EN 1092-1 D Nut, PN 100 EN 1092-1 D Nut, PN 63 EN 1092-1 D Nut, PN 40 EN 1092-1 B1, PN 100 EN 1092-1 B1, PN 63 EN 1092-1 B1, PN 16 Connection Sensor 3 EN 1092-1 B1, PN 40 Standard sensors O O O O O O O Standard: 7ME461.-... DN 15 (½") DN 6 (¼") DN 10 ● ● O ● DN 25 (1") ● O ● O O O ● DN 20 (¾") ● ● ● ● ● ● ● ● O ● ● ● ● O ● ● O ● ● O O O DN 15 (½") DN 25 (1") O ● O O O O ● ● ● ● ● ● ● O O ● DN 40 (1½") O ● O ● O O O DN 50 (2") ● ● O ● O O O DN 65 (2½") O ● O O DN 80 (3") ● ● O ● O O O DN 100 (4") ● O O O DN 40 (1½") ● ● DN 32 (1¼") DN 50 (2") O 8") DN 15 (½") DN 25 (1") O (3/ O ● O O O ● ● ● ● O ● ● ● ● ● ● DN 25 (1") ● ● O O O ● ● ● ● ● O O O O ● ● ● ● ● O O O O O ● ● ● DN 65 (2½") DN 80 (3") DN 50 (2") • Combinations shown ● are Mainstream products with delivery time of up to 15 days depending on the combination and production stock levels. • Combinations shown O are Sidestream products with delivery from 45 to 90 days. Not all components are held in production stock for Sidestream products. 3/154 Siemens FI 01 · 2016 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 ISO 2853 0.8 μm screwed ISO 2852 0.8 μm clamped DIN 11864-2 0.8 μm flanged Connection DIN 11864-1 0.8 μm screwed The aseptic flanges offered for FCS400 conform with the standard DIN 11864-2A BF-A. The flange fitted to the sensor is therefore the back flange and the seal is an O-ring. DIN 32676 0.8 μm Tri-clamp The hygienic sensors all have maximum internal surface roughness < 0.8 µm and are EHEDG and 3A approved. Hygienic sensors are offered with process connection conforming to various international quick-connect clamps or threaded connectors. Pressure ratings are according to the relevant standard and the sensor size. Maximum pressure in the hygienic program is PN 40. DIN 11851 0.8 μm screwed Aseptic flanged process connections Sensor Hygienic sensor variants Hygienic: 7ME462.-... DN 15 (½") DN 6 (¼") O DN 15 (½") ● ● ● Ring OD d11 Bolt Circle Bolt d5 holes Flange diameter d10 10 13 x 1.5 10 22.4 37 4 x 9 54 15 19 x 1.5 16 28.4 42 4 x 9 59 20 23 x 1.5 20 32.4 47 4 x 9 64 25 29 x 1.5 26 38.4 53 4 x 9 70 32 35 x 1.5 32 47.7 59 4 x 9 76 40 41 x 1.5 38 53.7 65 4 x 9 82 50 53 x 1.5 50 65.7 77 4 x 9 94 65 70 x 2.0 66 81.7 95 8 x 9 107 80 85 x 2.0 81 97.7 112 8 x 11 113 ● O ● ● ● ● O O DN 15 (½") DN 25 (1") ● DN 32 (1¼") O ● ● ● ● DN 40 (1½") DN 50 (2") Bore d1 ● DN 20 (¾") DN 25 (1") Size DN Pipe DIN 11864-2A BF-A flange dimensions DN 10 (3/8") DN 25 (1") The flange dimensions in the FCS400 program are as follows: DN 25 (1") DN 40 (1½") O DN 50 (2") ● O ● O O ● ● ● ● ● ● ● ● ● ● DN 65 (2½") DN 80 (3") DN 50 (2") DN 65 (2½") ● DN 80 (3") ● DN 100 (4") • Combinations shown ● are Mainstream products with delivery time of up to 15 days depending on the combination and production stock levels. • Combinations shown O are Sidestream products with delivery from 45 to 90 days. Not all components are held in production stock for Sidestream products. Siemens FI 01 · 2016 3/155 3 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 NAMUR sensor variants ● ● ● ISO 2853 Hygienic screwed ● ISO 2852 Hygienic clamped ● DIN 11864-2A Aseptic flanged O DIN 11864-1A Aseptic secrewed ASME B1.20.1 NPT pipe thread O DIN 32676 Hygienic Tri-clamp ISO 228-1 G pipe thread ANSI B16.5-2009, class 600 ANSI B16.5-2009, class 300 ANSI B16.5-2009, class 150 EN 1092-1 D Nut, PN 100 EN 1092-1 D Nut, PN 63 EN 1092-1 D Nut, PN 40 EN 1092-1 B1, PN 100 EN 1092-1 B1, PN 63 EN 1092-1 B1, PN 40 EN 1092-1 B1, PN 16 Connection Sensor 3 DIN 11851 Hygienic screwed The NAMUR variants have build-in lengths according to NAMUR recommendation NE 132. The recommendations of NE 132 are stated for sensors with flanges the same size as the sensor nominal size, and for flanges to EN1092-1 PN 40 with B1 flange facing. For couplings of other standards such as ASME B16.5 Class 150, the overall length incorporates the difference in length between standard EN and ASME flanges. NAMUR variants are offered with flange and pipe thread connections according to EN, ISO oand ASME standards, as shown in the table below. ● ● ● ● O O NAMUR: 7ME471.-... DN 15 (½") DN 6 (¼") DN 10 (3/8") DN 15 (½") O O ● O ● O O O DN 20 (¾") DN 25 (1") DN 25 (1") O ● O ● ● O ● ● O ● ● ● O DN 15 (½") DN 25 (1") O ● O O O ● O ● ● ● DN 32 (1¼") DN 80 (3") ● ● ● ● DN 40 (1½") O ● O ● O O O DN 50 (2") O ● O ● O O O DN 65 (2½") O O O O O ● O O O ● O ● ● O ● ● ● O ● ● DN 25 (1") O ● O O ● ● ● ● ● ● ● ● ● ● O ● ● ● DN 50 (2") DN 65 (2½") O ● O O DN 80 (3") O ● O ● DN 100 (4") O O O O • Combinations shown ● are Mainstream products with delivery time of up to 15 days depending on the combination and production stock levels. • Combinations shown O are Sidestream products with delivery from 45 to 90 days. Not all components are held in production stock for Sidestream products. 3/156 ● O O DN 40 (1½") DN 50 (2") ● Siemens FI 01 · 2016 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 ■ Dimensional drawings ' & 3 % $ Sensor A B C Weight [DN] [inch] [mm] [inch] [mm] [inch] [mm] [inch] [kg] [lb] 15 ½ 90 3.54 280 11.0 90 3.54 4.6 10.1 25 1 123 4.84 315 12.4 90 3.54 7.9 17.4 50 2 187 7.36 390 15.4 90 3.54 25.7 56.7 80 3 294 11.6 504 19.8 90 3.54 66.5 147 SITRANS FCS400, dimensions in mm (inch), weights in kg (lb), for a EN 1092 PN 40 flanged version. The build-in length D depends on the flange. Siemens FI 01 · 2016 3/157 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 Overall length The overall length (build-in length (D)) of each sensor depends on the connection standard and the pressure rating. The tables below summarize the dimensions available at the time of publishing. Please contact Siemens for further information about our desired process connection specification. Standard: 7ME461.-... 3 Sensor DN 15 (½") Connection DN 6 (¼") DN 25 (1") 265 265 360 EN 1092-1 B1, PN 40 265 265 360 EN 1092-1 B1, PN 63 265 EN 1092-1 B1, PN 100 270 ANSI B16.5, class 150 270 270 360 ANSI B16.5, class 300 270 270 ANSI B16.5, class 600 270 285 265 ANSI B1.20.1 NPT pipe thread 265 DIN 11851 Hygienic screwed DN 80 (3") DN 10 DN 15 DN 20 DN 25 DN 25 DN 32 DN 40 DN 40 DN 50 DN 65 DN 80 DN 100 (3/8") (½") (¾") (1") (1") (1¼") (1½") (1½") (2") (2½") (3") (4") EN 1092-1 B1, PN 16 ISO 228-1 GH pipe thread DN 50 (2") 275 265 610 915 840 840 610 610 915 840 840 360 610 610 915 915 915 360 610 610 915 915 915 365 620 915 875 360 380 620 915 875 360 380 620 915 875 840 840 365 270 265 610 620 365 265 193 360 DIN 32676-C Hygienic clamp 265 265 360 DIN 11864-1 Aseptic screwed 265 265 360 DIN 11864-2 Aseptic flange 265 265 ISO 2852 Hygienic clamp 620 360 265 610 360 360 265 ISO 2853 Hygienic screwed 365 274 360 360 610 610 875 610 875 620 610 875 610 610 840 610 860 274 SITRANS FCS400, overall length (D), dimensions in mm Sensor DN 15 (½") Connection DN 6 (¼") DN 25 (1") 10.43 10.43 14.17 EN 1092-1 B1, PN 40 10.43 10.43 14.17 EN 1092-1 B1, PN 63 10.43 EN 1092-1 B1, PN 100 10.63 ANSI B16.5, class 150 10.63 10.63 14.17 ANSI B16.5, class 300 10.63 10.63 ANSI B16.5, class 600 10.63 11.22 10.43 ANSI B1.20.1 NPT pipe thread 10.43 DIN 11851 Hygienic screwed 10.83 10.43 24.02 36.02 33.07 33.07 24.02 24.02 36.02 33.07 33.07 14.17 24.02 24.02 36.02 36.02 36.02 14.17 24.02 24.02 36.02 36.02 36.02 14.37 24.41 36.02 34.45 14.17 14.96 24.41 36.02 34.45 14.17 14.96 24.41 36.02 34.45 33.07 33.07 7.60 14.17 10.43 10.43 14.17 DIN 11864-1 Aseptic screwed 10.43 10.43 14.17 DIN 11864-2 Aseptic flange 10.43 10.43 SITRANS FCS400, overall length (D), dimensions in inch 3/158 Siemens FI 01 · 2016 14.17 10.43 10.43 24.41 14.37 10.43 ISO 2852 Hygienic clamp 14.37 14.37 10.63 10.43 24.02 DIN 32676-C Hygienic clamp ISO 2853 Hygienic screwed DN 80 (3") DN 10 DN 15 DN 20 DN 25 DN 25 DN 32 DN 40 DN 40 DN 50 DN 65 DN 80 DN 100 (3/8") (½") (¾") (1") (1") (1¼") (1½") (1½") (2") (2½") (3") (4") EN 1092-1 B1, PN 16 ISO 228-1 GH pipe thread DN 50 (2") 24.41 14.17 24.02 14.17 10.78 14.17 14.17 10.78 24.02 24.02 34.45 24.02 34.45 24.41 24.02 34.45 24.02 24.02 33.07 24.02 33.86 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 Hygienic 0.8 m: 7ME462.-... Sensor DN 15 (½") Connection DN 10 (3/8") DN 15 (½") DIN 11851 Hygienic screwed 265 265 DN 25 (1") DN 20 (¾") DN 25 (1") 265 DN 25 (1") DN 32 (1¼") 360 360 360 DN 40 (1½") DN 50 (2") DN 80 (3") DN 40 (1½") DN 50 (2") DN 65 (2½") DN 80 (3") 610 610 840 840 DIN 32676-C Hygienic clamp 265 DIN 11864-1 Aseptic screwed 265 360 DIN 11864-2 Aseptic flange 265 360 620 610 ISO 2852 Hygienic clamp 265 360 ISO 2853 Hygienic screwed 265 360 360 610 875 610 875 610 875 610 840 610 860 SITRANS FCS400, overall length, dimensions in mm Sensor DN 15 (½") Connection DN 10 (3/8") DN 15 (½") DIN 11851 Hygienic screwed 10.43 10.43 DN 25 (1") DN 20 (¾") DN 25 (1") DN 25 (1") DN 32 (1¼") 7.60 14.17 14.17 DIN 32676-C Hygienic clamp 10.43 DIN 11864-1 Aseptic screwed 10.43 14.17 DIN 11864-2 Aseptic flange 10.43 14.17 10.43 14.17 ISO 2852 Hygienic clamp 10.43 14.17 ISO 2853 Hygienic screwed 10.43 14.17 DN 40 (1½") DN 50 (2") DN 80 (3") DN 40 (1½") DN 50 (2") DN 65 (2½") DN 80 (3") 24.20 24.20 33.07 33.07 14.17 24.20 34.45 24.20 34.45 24.41 24.20 34.45 24.20 24.20 33.07 24.20 33.86 SITRANS FCS400, overall length, dimensions in inch Siemens FI 01 · 2016 3/159 3 © Siemens AG 2016 Flow Measurement SITRANS F C Flow sensor SITRANS FCS400 NAMUR: 7ME471.-... Sensor DN 15 (½") Connection DN 6 (¼") DN 25 (1") DN 80 (3") EN 1092-1 B1, PN 16 510 510 600 715 715 915 915 915 EN 1092-1 B1, PN 40 510 510 600 715 715 915 915 915 EN 1092-1 B1, PN 63 510 600 715 715 915 915 915 600 715 715 915 915 915 EN 1092-1 B1, PN 100 3 DN 50 (2") DN 10 DN 15 DN 20 DN 25 DN 25 DN 32 DN 40 DN 40 DN 50 DN 65 DN 80 DN 100 (3/8") (½") (¾") (1") (1") (1¼") (1½") (1½") (2") (2½") (3") (4") EN 1092-1 D, PN 16 510 600 715 715 915 EN 1092-1 D, PN 40 510 915 600 715 715 EN 1092-1 D, PN 63 600 715 715 ANSI B16.5, class 150 600 915 ANSI B16.5, class 300 600 915 600 915 ANSI B16.5, class 600 ISO 228-1 GH pipe thread 510 ANSI B1.20.1 NPT pipe thread 510 DIN 11851 Hygienic screwed 915 510 510 510 DIN 32676-C Hygienic clamp 510 DIN 11864-1 Aseptic screwed 510 600 510 600 600 715 600 715 915 915 715 600 715 DIN 11864-2 Aseptic flange ISO 2852 Hygienic clamp 510 600 ISO 2853 Hygienic screwed 510 600 715 715 915 715 SITRANS FCS400, overall length, dimensions in mm Sensor DN 15 (½") Connection DN 6 (¼") DN 25 (1") DN 50 (2") DN 80 (3") DN 10 DN 15 DN 20 DN 25 DN 25 DN 32 DN 40 DN 40 DN 50 DN 65 DN 80 DN 100 (3/8") (½") (¾") (1") (1") (1¼") (1½") (1½") (2") (2½") (3") (4") EN 1091-1 B1, PN 16 20.08 20.08 23.62 28.15 28.15 36.02 36.02 36.02 EN 1091-1 B1, PN 40 20.08 20.08 23.62 28.15 28.15 36.02 36.02 36.02 EN 1091-1 B1, PN 63 20.08 23.62 28.15 28.15 36.02 36.02 36.02 23.62 28.15 28.15 36.02 36.02 36.02 EN 1091-1 B1, PN 100 EN 1092-1 D, PN 16 20.08 23.62 28.15 28.15 36.02 EN 1092-1 D, PN 40 20.08 23.62 28.15 28.15 36.02 EN 1092-1 D, PN 63 23.62 28.15 28.15 ANSI B16.5, class 150 23.62 36.02 ANSI B16.5, class 300 23.62 36.02 ANSI B16.5, class 600 23.62 36.02 ISO 228-1 GH pipe thread 20.08 ANSI B1.20.1 NPT pipe thread 20.08 DIN 11851 Hygienic screwed 36.02 20.08 20.08 20.08 DIN 32676-C Hygienic clamp 20.08 DIN 11864-1 Aseptic screwed 20.08 23.62 20.08 23.62 23.62 28.15 23.62 28.15 36.02 36.02 28.15 23.62 28.15 DIN 11864-2 Aseptic flange ISO 2852 Hygienic clamp 20.08 23.62 ISO 2853 Hygienic screwed 20.08 23.62 SITRANS FCS400, overall length, dimensions in inch 3/160 Siemens FI 01 · 2016 28.15 28.15 28.15 36.02 © Siemens AG 2016 Flow Measurement SITRANS F C Flowmeter - Accessories/Spare parts Spare parts - transmitter FCT030 Article No. Description Article No. Display and keypad assembly with firewire connection to the transmitter module1) A5E03548971 Mounting bracket - FCT030; in painted aluminum for pipe or wall mounting of transmitter FCT030 remote version. Including lock ring, pressure pads and seal cap A5E03906091 Sensor interface (Compact). Front end flow calculator and process detection. SIL 3 approved1) A5E03549142 M12 option for sensor housing in stainless steel. Prewired and potted to replace M12 socket in DSL housing A5E03906095 A5E03549098 M12 option - remote - in painted aluminum. Prewired and potted replacement M12 connection for FCT030 transmitter remote version A5E03906104 Sensor interface (Remote); barrier unit for high speed digital communication and Ex ib power supply to remote front end DSL module Remote terminal house M20 A5E03906112 Display lid in painted aluminum with Ex glass plate and o-ring seal A5E03549344 Remote terminal house NPT - in painted aluminum for sensor cable termination at FCT030 transmitter remote version. Pre-wired and potted A5E03906130 Transmitter cassette (active) with SIL approved 4 ... 20 mA output and HART 7.21) A5E03549357 Transmitter cassette (passive) with SIL approved 4 ... 20 mA output and HART 7.21) A5E03549383 Bag of loose spare parts; including cable strain relief components, mounting tool, seals and gasket, assorted screws and washers, hex cap nut, blind plugs, and o-rings A5E03549396 Power supply 240 V AC, 47 ... 63 Hz 24 ... 90 V DC A5E03549413 Blind lid in painted aluminum with o-ring seal A5E03549429 Description 3 Spare parts - sensor FCS400 I/O assembly Advise Order code F00 to F97 from Selection and Ordering data2) A5E03939114 SensorFlash (micro SD card) A5E03915258 Description Article No. Blind lid in painted aluminum with o-ring seal A5E03549295 Frontend cassette Spare part frontend cassette for remote version of FC430 and cassette for FC4101) A5E03549191 Sensor housing metric A5E03549313 Sensor housing NPT in painted aluminum A5E03906080 Bag of loose parts for sensor; including cable strain relief components, washer, seals, o-rings, and assorted screws A5E03549324 1) The system firmware bundle must be stated in the "Remark" field when ordering, to ensure compatibility of the system. The FW revision is found on the product label for FC430 and FC410. Further for FC430 the firmware bundle can be found in the local display in the menu items 3.1.10. e.g. "2.02.01-02" 2) The I/O configuration must be stated in the "Remark" field. The I/O configuration is found in the F option of the ordering code. e.g. code “F40” for ordering Ch2 Active Current/Freq/Pulse, Ch3 Active Current/Freq/Pulse, Ch4 Active Input Siemens FI 01 · 2016 3/181