AGILENT 85070E

Agilent
85070E Dielectric Probe Kit
200 MHz to 50 GHz
Technical Overview
New!
• Performance probe! Combines
rugged, high temperature and
frequency performance in a slim
design. Perfect for your most
demanding applications!
Features
• Measures complex permittivity
over a broad frequency range
• Results can be viewed in a variety
of formats: er', er", loss tangent,
and Cole-Cole
• Software runs on a PC, or internally
on the PNA Series of network
analyzer and guides user through
easy calibration and measurement
• Data is easily shared with other
Windows® based programs or
through the user programmable
Component Object Model (COM)
interface
Swept High-Frequency Dielectric Measurements
The Agilent Technologies 85070E
Dielectric Probe Kit determines the
dielectric properties, or complex
permittivity, of many materials.
Because a materials dielectric
properties are determined by its
molecular structure, if the molecular
structure changes, so will the
dielectric properties. Measuring
them can indirectly measure other
properties that are also correlated
to the molecular structure, and can
be a valuable alternative when the
property of interest is difficult to
measure directly.
Measurements are made by simply
immersing the probe into liquids
or semi-solids – no special
fixtures or containers are required.
Measurements are non-destructive
and can be made in real time.
These important features allow
the Dielectric Probe Kit to be used
in process analytic technologies.
The complete system is based on a
network analyzer, which measures
the material’s response to RF or
microwave energy. The probe transmits
a signal into the material under test
(MUT). Depending on the Agilent
network analyzer and probe used,
frequencies can extend from 200 MHz
to 50 GHz.
The included software controls the
network analyzer and guides the user
through easy setup and measurement
steps. In seconds, it calculates and
displays complex permittivity in a
variety of formats, including dielectric
constant, dielectric loss factor, loss
tangent or Cole-Cole.
2
Software displays dielectric constant and loss factor as a function of frequency.
Easy Data Analysis Display
The split screen window and marker
aids in data analysis. Simply click on a
point on the chart or table to activate
and move the marker.
Connect to other programs
Data charts and tables can easily be
copied and pasted into any Windowsbased application for further analysis
or report generation. The component
object model (COM) interface allows
the measurement to be setup, triggered,
and read from a user written program.
This is valuable for analyzing material
changes over time. Example Visual
Basic® and C++ projects are included
to aid program development.
Display data in chart form, table form, or both.
3
Calibration Refresh Reduces Drift Errors
The new automated Electronic
Calibration Refresh feature recalibrates
the system automatically, in seconds,
just before each measurement is made.
This virtually eliminates cable instability
and system drift errors. Processes can
now be monitored over long time
periods, including tests that vary MUT
temperature and pressure over time.
How it works:
The Agilent Electronic Calibration
module (ECal) microwave ports are
connected in line between the probe
and the network analyzer test port
cable. The ECal module communication
port is connected either to the PC or
PNA Series network analyzer running
the 85070E software. The software
guides the user through a normal
“three standard” calibration, (usually
open, short, water), performed at the
end of the probe. This calibration is
then transferred to the ECal module.
The ECal module remains in line
and a complete ECal calibration is
automatically performed before each
measurement. Errors due to test port
cable movement are removed by the
new calibration.
ECal module connected in line
Water with and without Electronic Calibration Refresh
This measurement shows the effects
of system drift and cable instability
on a dielectric measurement of water
and the improvement with Electronic
calibration refresh. Both measurements
were made 24 hours after the original
calibration. The lighter colored, noisier,
trace was made before the Electronic
Calibration refresh was turned on.
The darker, smoother, trace shows the
improvement made after the Electronic
Calibration refresh was turned on.
4
For systems without an ECal module,
a simpler, "one standard" refresh
calibration feature is also available,
which can reduce the effects of
system drift over time or temperature.
After the initial "three standard" probe
calibration is performed, the calibration
can be refreshed at any time with the
connection of a single standard. Any
one of the three calibration standards
can be defined as the refresh standard.
New Options Allow You to Configure Kit to Meet Your Needs
Now three probe designs to choose from
New! Performance Probe
Combines rugged, high temperature,
and frequency performance in a
slim design.
This probe features rugged,
high temperature and frequency
performance in a slim design, perfect
for your most demanding applications.
The probe is sealed on both the probe
tip and the connector end, which make
it our most rugged probe. The probe
withstands a wide –40 ºC to +200 ºC
temperature range, which allows
measurements versus frequency
and temperature.
The probe can be autoclaved, so it
is perfect for applications in the food,
medical, and chemical industries
where sterilization is a must. The
slim design allows it to fit easily in
fermentation tanks, chemical reaction
chambers, or other equipment with
small apertures.
The small diameter also allows it to
be used with smallest sample sizes
of all Agilent’s probes. It is useful for
measuring liquid, semi-solid, as well
as flat surfaced solid materials. The
Performance Form Probe Kit comes
complete with a calibration short.
Frequency range 500 MHz to 50 GHz.
2.4mm male connector
Performance Probe Kit
Borosilicate
glass seal
150
1.6
2.4 mm male
connector
Hermetically sealed on both ends
(all dimensions in mm)
Stainless steel
T316L body
9.5
Nickel plated
tungsten center
conductor
Stainless steel
T304L tip
5
High temperature probe
Survives corrosive chemicals and
high temperatures
Frequency range: 200 MHz to 20 GHz.
3.5 mm male connector.
Rugged in design, this probe features
a hermetic glass-to-metal seal, which
makes it resistant to corrosive or
abrasive chemicals. The probe
withstands a wide –40 ºC to +200 ºC
temperature range, which allows
measurements versus frequency and
temperature. The large flange makes
it easier to measure flat surfaced solid
materials, in addition to liquids and
semi-solids. The 3.5 mm aperture
has a larger sensing volume than
our other probes.
High Temperature Probe Kit
9.0
Hole
10.2 min.
Mounting flange
max thickness
15.5
Stainless
steel
Inconel
17.4
3.5 mm
connector
46.7
(1.84)
16.8
Grip nut
M10X1.0 thread
All Dimensions in mm (inches)
Glass
19.0
(0.75)
Slim Form Probe
Smaller diameter fits into tight spaces.
Frequency range 500 MHz to 50 GHz.
2.4 mm male connector.
This probe features a slim design, which
allows it to fit easily in fermentation
tanks, chemical reaction chambers, or
other equipment with small apertures.
The slim design also allows it to be
used with smaller sample sizes. Best
used for liquids and soft semi-solids.
For castable solids, the probe is
economical enough to be cast into the
material and left in place. Because of
the consumable nature of this design,
these probes are offered in sets of
three. The Slim Form probe kit comes
with a sealed Slim Form holder that
adapts 2.2 mm outer diameter to 10 mm
inner diameter bracket included in the
kit as well as commercially available
“Midi” sized adapters and bushings.
6
Slim Form Probe measurement to 50 GHz
ø 2.2 mm
Finish, Nickel 100 µ inches
200 mm
2.4 mm Male
connector
Three Cables to Choose From
Accessories
20 GHz Flexible Cable
Cable to choose when tempertature
performance is not critical. SMA female
connectors connect to High Temperature
probe. When connecting to Slim Form
probe, an additional adapter may be
needed (see configuration guide).
Probe Stand
The probe stand has a 13 x 7 inch
porcelain base and 24 inch high by
0.5 inch diameter metal support. This
stand works with mounting bracket
and ECal holder included in the
standard kit. It is highly recommended
to stabilize measurement setup.
50 GHz Flexible Cable
Cable to choose for high frequency
applications. 2.4 mm female connectors
connect directly to Slim Form
Performance probes.
20 GHz High Temperature Cable
Use with High Temperature or
Performance probes for high
temperature applications from –40 °C
to +200 °C. SMA female connectors
connect directly to High Temperature
probe. Adapter included in kit
connects High Temperature Cable
to Performance probe.
Software Menu Items
File
Save or recall measurement setups or
previous measurement results. Print
copies of the measurement results in
a tabular or graphical format.
Edit
Copy the measurement results to the
clipboard. Either graph or the tabular
listing can be copied. This allows your
measurement results to be pasted
into other applications.
View
Select the section you want to view.
Selections include the toolbar, status
bar, table of the measurement data,
and chart of the measurement data.
Calibration
Select the frequency range, number
of points, linear or log sweep.
Guided calibration sequence; choice
of calibration materials or userspecified; refresh calibration for
single standard or ECal; recalibration
versus temperature; automatic refresh
on or off.
Measure
Trigger a measurement.
Chart
Select the format to be displayed on
the chart. Choices include er’, er’’,
loss tangent, and Cole-Cole. Set
Graticule scale factors or “autoscale”.
Select from linear, semi-log, or log-log
representations.
Display
Display current measurement data;
save/display up to 3 memory traces;
compare data to reference trace with
trace math. Turn the marker on or off.
Preferences
Select your preference of fonts, colors,
and annotations used to plot and list
the measurement data.
Help
On-line help including the product
manual.
Toolbar
Provides single click access to the
most important menu items.
Table
Choose between a tabular formatting
of real and imaginary or real and loss
tangent
7
Performance Characteristics
Specifications describe the warranted
performance over the temperature
range 0 to 55 °C. Supplemental
characteristics are intended to provide
information useful in applying the
instrument, by giving typical but
non-warranted performance parameters.
These are denoted as “typical,”
“nominal,” or “approximate.”
Probe Characteristics Table
Frequency Range
(nominal)
Performance Probe
500 MHz to 50 GHz
Maximum limited by MUT properties
(285-j125)GHz
<
√|er*|
Slim Form Probe
500 MHz to 50GHz
Maximum limited by MUT properties
(285-j125)GHz
<
√|er*|
High Temperature Probe
200 MHz to 20 GHz with network analyzer
10 MHz to 3 GHz with E4991A Impedance
analyzer with option 10.
Maximum limited by MUT properties
110GHz
<
√|er*|
Temperature Range
Temperature Slew Rate
Immersable length
(approximate)
Connector
Repeatability and
resolution
Material under test
assumptions
Sample size
requirements
–40 to +200 °C
< 10 degrees/minute
140 mm
0 to +125 °C
< 10 degrees/minute
200 mm
–40 to +200 °C
< 10 degrees/minute
35 mm
2.4 mm male
2.4 mm male
3.5 mm male
Two to four times better than accuracy Two to four times better than accuracy Two to four times better than accuracy
Material is “infinite” in size,
non-magnetic (µr* = 1),
isotropic (uniform orientation),
and homogeneous
(uniform composition)2 .
Solids have a single, smooth,
flat 3 surface with gap-free
contact at the probe face.
Minimum 5 mm insertion and
1 mm around tip of probe
Liquid or soft semi-solid.
Material is “infinite” in size,
non-magnetic (µr* = 1),
isotropic (uniform orientation), and
homogeneous (uniform composition)2.
Material is “infinite” in size,
non-magnetic (µr* = 1),
isotropic (uniform orientation),
and homogeneous (uniform composition)2 .
Solids have a single, smooth,
flat 3 surface with gap-free contact
at the probe face.
Minimum 5 mm insertion and
5 mm around tip of probe
Diameter: > 20 mm
20
Thickness: >
(√|er*|) mm
Granule size4: < 0.3 mm
Expected Value
requirements
Accuracy (typical)1
Maximum recommended er’: < 100
Minimum recommended loss
tangent > 0.05
Not recommended for low loss
(loss tangent < 0.5) materials
with e’ > 5
Dielectric constant, er’ = er’
±0.05|er*|. er”= er”±0.05|er*|
Maximum recommended er’: < 100
Minimum recommended loss
tangent > 0.05
Not recommended for low loss
(loss tangent < 0.5) materials
with e’ > 5
Dielectric constant, er’ = er’
±0.05|er*|. er”= er”±0.05|er*|
Maximum recommended er’: < 100
Minimum recommended loss
tangent > 0.05
Not recommended for low loss
(loss tangent < 0.5) materials
with er’ > 5
Dielectric constant, er’ = er’
±0.05|er*|. er”= er”±0.05|er*|
1. Practical frequency range, accuracy and resolution depend on properties of the MUT. Value indicates typical accuracy at 23 ±3 °C, not including
effects of probe contact and cable flexure.
2. If the material is not homogeneous, the result is an average value weighted by the intensity of the E-field, which is highest at the center conductor
of the probe tip.
3. Sample must be as flat as the probe face, which is lapped to ±100 µ inches.
4. Measurement repeatability for granular materials is dependent on density variation.
8
Configuration Guide
Choose one of our
suggested configurations
High Temperature Configuration
85070E:
• Performance Probe Kit, Option 050,
or High-Temperature Probe Kit,
Option 020
• High Temperature Cable, Option 002
• Probe Stand, Option 001
• USB Security Key, Option UL8
• Recommend 20 GHz PNA Series
network analyzer.
Rugged High Frequency Configuration
85070E:
• Performance Probe, Option 050
• Probe Stand, Option 001
• USB Security Key, Option UL8
• Recommend 50 GHz PNA Series
network analyzer and 50 GHz
ECal module
Slim Form Probe Kit (Option 030)
Economy Configuration
85070E
• Slim Form Probe Kit, Option 030
• 20 GHz Flexible Cable, Option 022
• USB Security Key, Option UL8
• Recommend ENA-L or PNA-L Series
network analyzer
High Temperature Probe Kit (Option 020)
Performance Probe Kit (Option 050)
9
Configuration Guide - continued
Or, customize your own
Probes – Choose one or all
Additional available parts
Dielectric Probe Kit, 85070E
Includes:
• Dielectric Probe Software
application on CD-Rom
• 1 mounting bracket to connect probes
to Option 001 Probe Stand or similar
stand
• 1 10 mm holder to connect
performance probe or slim form
holder to mounting bracket
• 1 3 mm hex key for 10 mm holder
and Ecal holder screws
• 1 ECal holder to connect ECal
module to mounting bracket
• 1 Type-N male to 3.5 mm male
adapter, 1250-1743
• 1 3.5 mm male to 2.4 mm female
adapter 11901D
• 1 foam lined walnut box.
Performance Dielectric Probe Kit,
Option 050
Includes:
• 1 Performance Dielectric Probe
• 1 Calibration Short
• 8710-2036 High-Temperature
Dielectric Probe
• 85070-60003 Shorting block and
clamp for high temperature probe
• 85070-60004 Short for Slim Form
Probes
• 85070-60007 Slim Form Probe holder
• 85070-60008 ECal Holder
• 85070-60009 Set of three Slim Form
Probes
• 85070-60010 Performance Probe
• 85070-60012 Short for Performance
Probe
• 85070-60011 10mm Holder for
Performance Probe and Slim Form
probe
• 8120-6286 High Temperature Cable
• 8120-6192 20GHz Flexible Cable
• 8121-1290 50GHz Flexible Cable
• 9301-1298 Probe Stand
• 1250-3449 Connector Saver for
Slim Form Probe
High-Temperature Dielectric Probe
Kit, Option 020
Includes:
• 1 High Temperature Probe
• 1 Calibration Short
Slim Form Probe Kit, Option 030
Includes:
• 3 Slim Form probes
• 1 connector saver
• 1 Calibration short
• 1 10 mm dia sealed probe holder.
• 6 O-rings
Slim Form probe replenishment Kit,
Option 033
Contains 3 extra Slim Form Probes
Cables (Optional, choose any or all)
• High Temperature Cable, Option 002
• 20 GHz Flexible Cable, Option 022
• 50 GHz Flexible Cable, Option 032
Accessory (Optional, highly recommended)
• Probe Stand, Option 001
Security Key (Must choose one)
• Parallel Hardware Key, Option UL7
(required for Windows NT® 4.0)
• USB Hardware Key, Option UL8
10
Adapter Selection Guide
Some configurations may need extra
adapters. The Agilent adapter part
numbers are charted below.
Network analzyer port to cable
Network analyzer port connector
Type-N female
3.5 mm male
2.4 mm male
High Temperature 20 GHz Cable
1250-1743 (included in kit)
None needed
11901D (included in kit)
Flexible 20 GHz Cable
1250-1743 (included in kit)
None needed
11901D (included in kit)
Flexible 50 GHz Cable
11903A
11901C
None needed
High Temperature 20 GHz Cable
None needed
11901D (included in kit)
Flexible 20 GHz Cable
None needed
11901D (included in kit)
Flexible 50 GHz Cable
11901C
None needed
Probe to cable
Probe
High Temperature Probe
Slim Form Probe
Adapters needed when using automated Electronic Calibration Refresh
Ecal module to cable
ECal Module Connector
Type-N male
Type-N female
3.5 mm male
3.5 mm female
2.4 mm male
2.4 mm female
High Temperature 20 GHz Cable
(3.5 mm female)
1250-1750
1250-1743 (included in kit)
None needed
1250-1748
11901D (included in kit)
11901C
Flexible 20 GHz Cable
(3.5 mm female)
1250-1750
1250-1743 (included in kit)
None needed
1250-1748
11901D (included in kit)
11901C
Flexible 50 GHz Cable
(2.4 mm female)
11903C
11903A
11901C
11901A
None needed
11900A
High Temperature Probe
1250-1745
1259-1744
83059B
None needed
11901B
11901C
Performance and Slim Form Probe
11903B
11903D
11901B
11901D (included in kit)
11900B
None needed
Ecal module to probe
ECal Module Connector
Type-N male
Type-N female
3.5 mm male
3.5 mm female
2.4 mm male
2.4 mm female
Compatible ECal modules 1
ECal module requires USB connection
to PC or PNA Series network analyzer
N469xA series
8509xC series
Compatible network
analyzers
8753ET/ES
8719ET/ES
8720ET/ES
8722ET/ES
8712ET/ES
8714ET/ES
8510C
All PNA Series
All ENA Series
Older Agilent network analyzers may
be compatible. Please download free
trial demo from our Web site to
determine compatibility:
www.agilent.com/find/materials
1.
Optional, needed for Automated Electronic Calibration Refresh
PC requirements
• Windows® 98, 2000, ME, XP, or
Windows NT® 4.0*
• CD-Rom drive to load software
Software can be run directly on PNA
series network analyzers or interfaced
over LAN
All other network analyzers require a
GPIB interface card with a compatible
driver (Agilent SICL or National
Instruments 488.2M)
* Windows NT® 4.0* requires Option UL7
Parallel Security Key
11
Web resources
Visit Agilent Web sites for additional
product and literature information.
Materials test:
www.agilent.com/find/materials
PNA Series network analyzers:
www.agilent.com/find/pna
Electronic Calibration (ECal):
www.agilent.com/find/ecal
RF and microwave test accessories:
www.agilent.com/find/accessories
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
www.agilent.com
For more information on Agilent Technologies’
products, applications or services, please
contact your local Agilent office. The
complete list is available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
0820 87 44 11
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
01805 24 6333**
**0.14 €/minute
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom
44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus