RPD-60 60W Dual Output Medical Type series ■ Features : 4"×2" miniature size Universal AC input/Full range Low leakage current <100μ A Protections: Short circuit / Over load / Over voltage Cooling by free air convection Medical safety approved (2 x MOPP between primary to secondary) UL60950-1/IEC60950-1/EN60950-1 ITE safety approved Fixed switch frequency at 100KHz Suitable for BF application with appropriate system consideration 3 years warranty .. SPECIFICATION MODEL OUTPUT INPUT RPD-60B RPD-60A OUTPUT NUMBER CH1 CH2 CH1 CH2 DC VOLTAGE 5V 12V 5V 24V RATED CURRENT 5A 2A 3.5A 1.5A CURRENT RANGE 0.5 ~ 5.5A 0.1 ~ 2.2A 0.5 ~ 3.85A 0.1 ~ 1.65A RATED POWER 49W PEAK LOAD(10sec.) 53.5W Note.4 53.9W 58.85W RIPPLE & NOISE (max.) Note.2 80mVp-p 80mVp-p 80mVp-p 100mVp-p VOLTAGE TOLERANCE Note.3 +3,-2% ±6.0% +3,-2% +8,-4% LINE REGULATION ±0.5% ±1.0% ±0.5% ±1.0% LOAD REGULATION ±1.5% ±2.0% ±1.5% ±2.0% SETUP, RISE TIME 300ms, 15ms/230VAC HOLD UP TIME (Typ.) 70ms/230VAC VOLTAGE RANGE 90 ~ 264VAC FREQUENCY RANGE 47 ~ 63Hz EFFICIENCY (Typ.) 78% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 1.1A/115VAC 0.7 A/230VAC COLD START 60A/230VAC 30A/115VAC LEAKAGE CURRENT 300ms, 15ms/115VAC at full load 14ms/115VAC at full load 127 ~ 370VDC 82% Note.8 Earth leakage current < 150 μA/264VAC , Touch current < 100μ A/264VAC OVERLOAD PROTECTION OVER VOLTAGE WORKING TEMP. WORKING HUMIDITY 115 ~ 150% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed CH1: 5.75 ~ 6.75V Protection type : Shut down o/p voltage, re-power on to recover -20 ~ +65℃ (Refer to "Derating Curve") 20 ~ 90% RH non-condensing -40 ~ +85℃, 10 ~ 95% RH ±0.03%/℃ (0 ~ 45℃) ENVIRONMENT STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT VIBRATION SAFETY STANDARDS 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes UL60950-1, TUV EN60950-1, ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved SAFETY & ISOLATION LEVEL WITHSTAND VOLTAGE EMC Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC (Note 5) OTHERS ISOLATION RESISTANCE EMC EMISSION I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH Compliance to EN55011(CISPR11),EN55022 (CISPR22) Class B, EN61000-3-2,-3 EMC IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A MTBF 677.8K hrs min. MIL-HDBK-217F (25℃) 101.6*50.8*29mm (L*W*H) 0.15Kg; 96pcs/15.4Kg/0.89CUFT DIMENSION PACKING NOTE 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power. 5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time. 7. Heat Sink HS1,HS2 can not be shorted. 8. Touch current was measured from primary input to DC output. File Name:RPD-60-SPEC 2015-07-17 RPD-60 60W Dual Output Medical Type Mechanical Specification Unit:mm 101.6 3.175 95.25 50.8 44.45 FS1 CN1 series 3.175 AC FUSE T2.5/250V 3 2 1 HS2 1 2 3 4 5 6 FS2 AC FUSE T2.5/250V CN2 HS1 M1 M2 3 max. 29 4-ψ3.3 DC Output Connector (CN2) : JST B6P-VH or equivalent AC Input Connector (CN1) : JST B3P-VH or equivalent Pin No. 1 2 3 Assignment AC/N No Pin AC/L Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent Pin No. 1,2 3,4 5 6 : Grounding Required Assignment V1 COM V2 NC Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent 1.HS1,HS2 cannot be shorted. 2.M1 is safety ground. For better EMC performance, Please secure an electrical connection between M1,M2 and chassis grounding. Block Diagram fosc : 100KHz RECTIFIERS & FILTER EMI FILTER I/P POWER SWITCHING RECTIFIERS & FILTER V2 RECTIFIERS & FILTER +V1 COM DETECTION CIRCUIT FG O.L.P. CONTROL O.V.P. Derating Curve Output Derating VS Input Voltage 100 100 90 80 80 70 LOAD (%) LOAD (%) 60 40 20 -20 0 10 20 30 40 50 AMBIENT TEMPERATURE (℃) 60 65 70 (HORIZONTAL) 60 50 40 90 95 100 115 120 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz File Name:RPD-60-SPEC 2015-07-17