MEANWELL RPS

RPS-60
60W Single Output Medical Type
series
Features :
4"x2" miniature size
Universal AC input/Full range
Low leakage current<200uA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
UL60950-1/IEC60950-1/EN60950-1 ITE safety approved
No load power consumption<0.75W
Fixed switch frequency at 100KHz
3 years warranty
EN60601-1
IEC60601-1
SPECIFICATION
MODEL
RPS-60-3.3
RPS-60-5
RPS-60-12
RPS-60-15
RPS-60-24
DC VOLTAGE
3.3V
5V
12V
15V
24V
48V
RATED CURRENT
10A
10A
5A
4A
2.5A
1.25A
CURRENT RANGE
0 ~ 11A
0 ~ 11A
0 ~ 5.5A
0 ~ 4.4A
0 ~ 2.75A
0 ~ 1.375A
RATED POWER
33W
50W
60W
60W
60W
60W
55W
66W
66W
66W
66W
RIPPLE & NOISE (max.) Note.2 80mVp-p
80mVp-p
120mVp-p
150mVp-p
240mVp-p
300mVp-p
VOLTAGE ADJ. RANGE
4.75 ~ 5.5V
2.0%
11.4 ~ 13.2V
2.0%
13.5 ~ 16.5V
2.0%
22.8 ~ 27.6V
1.0%
45.6 ~ 52.8V
1.0%
PEAK LOAD(10sec.)
OUTPUT
Note.4 36.3W
VOLTAGE TOLERANCE Note.3
INPUT
RPS-60-48
3.1 ~ 3.6V
2.0%
LINE REGULATION
0.5%
0.5%
0.5%
0.5%
0.5%
0.5%
LOAD REGULATION
1.0%
1.0%
1.0%
1.0%
1.0%
1.0%
SETUP, RISE TIME
500ms, 30ms/230VAC
HOLD UP TIME (Typ.)
50ms/230VAC
VOLTAGE RANGE
90 ~ 264VAC
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY (Typ.)
74%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.8A/115VAC
COLD START 60A/230VAC
LEAKAGE CURRENT
For earth <200uA / 264VAC, For patient <100uA/264VAC
OVER LOAD
PROTECTION
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
500ms, 30ms/115VAC at full load
13ms/115VAC at full load
127 ~ 370VDC
79%
83%
84%
86%
28.4 ~ 32.4V
55.2 ~ 64.8V
30A/115VAC
115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8 ~ 5V
5.75 ~ 6.75V
13.8 ~ 16.2V
17.25 ~ 20.25V
Protection type : Shut down o/p voltage, re-power on to recover
-20 ~ +70
85%
1 A/230VAC
(Refer to output load derating curve)
20 ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 45 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY STANDARDS
WITHSTAND VOLTAGE
UL60950-1, TUV EN60950-1, UL60601-1, TUV EN60601-1, IEC60601-1 approved
HARMONIC CURRENT
Compliance to EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
SAFETY & ISOLATION RESISTANCE
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC/ 25 / 70% RH
EMC
EMI CONDUCTION & RADIATION Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B
(Note 5)
EMS IMMUNITY
MTBF
OTHERS
DIMENSION
PACKING
NOTE
353.6Khrs min. MIL-HDBK-217F (25
101.6*50.8*29mm (L*W*H)
0.15Kg; 96pcs/15.4Kg/0.89CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. Heat Sink HS1,HS2 can not be shorted.
File Name:RPS-60-SPEC 2011-03-31
RPS-60
60W Single Output Medical Type
Mechanical Specification
series
Unit:mm
101.6
3.175
95.25
50.8
44.45
FS1
3.175
HS2
AC FUSE
T2.5/250V
3
2
1
CN1
SVR1
1
2
3
4
FS2
AC FUSE
T2.5/250V
CN2
HS1
3 max.
29
4- 3.3
DC Output Connector (CN2) : JST B4P-VH or equivalent
AC Input Connector (CN1) : JST B3P-VH or equivalent
Assignment
AC/N
No Pin
AC/L
Pin No.
1
2
3
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Pin No.
1,2
3,4
Assignment
+V
-V
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
HS1,HS2 can not be shorted
Block Diagram
fosc : 100KHz
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
RECTIFIERS
&
FILTER
POWER
SWITCHING
+V
-V
DETECTION
CIRCUIT
FG
O.L.P.
CONTROL
O.V.P.
Output Derating VS Input Voltage
Output Derating
Ta=25
100
100
90
80
80
70
LOAD (%)
LOAD (%)
60
40
20
-20
0
10
20
30
40
50
AMBIENT TEMPERATURE ( )
60 65 70 (HORIZONTAL)
60
50
40
90
95
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-60-SPEC 2011-03-31