HP L4411A

Agilent L4411A
‘Low profile’ 6½ Digit Multimeter
Leading the Industry in High-Performance
System Test
Product Overview
Agilent L4411A
6½-Digit HighPerformance DMM
• 50,000 readings/sec @ 4½ digits
direct to PC
• 10,000 readings/sec @ 5½ digits
direct to PC
• 1,000 readings/sec @ 6½ digits
direct to PC
• Analog level triggering
• Programmable pre/post triggering
• LAN extensions for instruments
(LXI), USB & GPIB standard
• 30 PPM 1 year basic DC accuracy
• DCV, ACV, DCI, ACI, 2-wire and
4-wire resistance, frequency,
period, and diode test
• Capacitance & temperature
measurements
• Expanded measurement ranges
A new standard for modular
system DMMs
The L4411A 6½ digit high performance
DMM expands Agilent’s industry
leading offering of LXI system
products. For the test system integrator looking for the next generation
modular DMM, this new meter offers
the industry’s best measurement
speed and throughput, a reduced size
(1 rack unit high), superior measurement performance, and a choice of
computer interfaces, including LXI,
providing high performance, easy
to use, economical I/O. A simple
display, including ‘latest reading’
and LAN address, allows the system
integrator to quickly integrate and
debug the test system. And finally,
the DMM comes with a compatibility
mode, requiring little-to-no code
change to upgrade your test system
with next generation capabilities.
Dramatic system performance
Whether it’s raw reading speed or
fast system throughput, the L4411A
sets a new benchmark in performance.
Using a new A/D technology, the
L4411A achieves an impressive
50,000 readings a second at 4½ digits,
and can stream readings to your
computer at this same speed!
Transactional I/O (single reading—
measurement and PC transfer time) is
3x faster than other popular modular
DMMs, significantly enhancing your
test throughput. Triggering is fast and
precise, with both trigger latency and
trigger jitter less than 1 µs, while bus
query response is less than 500 µs.
ACV measurements are faster as
well thanks to a digital measurement
technique that additionally improves
accuracy at high and low frequencies.
LXI—Class C
LAN Extensions for Instruments (LXI)
provides the next generation I/O for
system applications requiring high
throughput. Transfer rates of over
250,000 readings/sec are attainable
ensuring even the most data intensive
measurements are fast, without the
overhead cost of an instrument
mainframe. LXI provides a built-in
Graphical Web Interface that allows
you to interactively control the DMM
without the hassle of programming,
making it great for debugging your
system. The L4411A DMM is LXI —
Class C compliant.
Enhanced measurement
capabilities
The L4411A offers temperature and
capacitance capabilities in addition to
those measurements you have come
to expect such as DCV, ACV, DCI,
ACI, 2-wire and 4-wire resistance,
frequency, period and diode test. You
also get offset compensated Ohms,
allowing you to accurately measure
resistance in the presence of voltages. Measurement ranges have been
expanded as well; for example, DC
and AC current ranges now go down
to 100 µA, resulting in 100 pA resolution. Real-time math and statistics are
included, and a peak-detect capability
allows you to capture peaks as short
as 20 µs.
System integration
When deciding on your next system
DMM you can’t go wrong with the
L4411A. Choose from LAN (LXI),
USB or GPIB interfaces, all standard
on the L4411A, to connect to your
computer. The 1U size is perfect
for space constrained applications
like aerospace/defense depot test.
Concerned about the viability of your
existing software programs? This
new DMM responds to standard
commands for programmable instrumentation (SCPI). Additionally there is
a 34401A/E1412A emulation mode to
ensure the easiest upgrade possible,
virtually eliminating costly software
and documentation changes. The
autoranging power supply allows
you to connect to any input power
without selecting input voltages or
changing fuses. Agilent’s I/O Library
Suite ships with the L4411A to help
you quickly establish an error-free
connection between your PC and
instrument. It provides robust instrument control and works with the
software development environment
you choose.
DMM dimensions (mm)
212.3
363.2
Companion LXI switch modules
Need a switch to route your signal
to the L4411A? Consider Agilent’s
LXI switch modules. Choose from a
40-channel armature relay multiplexer
(L4421A), a dual/quad 4x8 reed relay
matrix (L4433A) or a 32-channel Form
A/C general purpose relay switch
module (L4437A). Additionally, Agilent
has LXI DAC, digital I/O and multifunction modules to help complete
your test system requirements. All
from the leader in LXI instrumentation, Agilent Technologies.
Built to last
Our new DMM was designed to high
standards of ruggedness and reliability. From the robust, compact package
to careful selection of components
and conservative circuit design, this
meter is built to last. Calculated mean
time between failure (MTBF) is in
excess of 100,000 hours. Backed by
a 1-year warranty and a worldwide
network of service centers, you can
buy with confidence.
Go to the web
For the latest information on these
or other Agilent DMMs, go to
www.agilent.com/find/dmm
2
40.9
Accessories included:
Test report, power cord, LAN
cross-over interface cable.
Product reference CD-ROM
(34410-13601)
• Software
◦ IntuiLink software
◦ IVI-COM driver
◦ LabView driver
◦ Example programs
• Online documentation
◦ Programmer’s reference
◦ Getting started guide
◦ User’s guide
◦ Service guide
◦ Localized manuals
Agilent I/O Libraries CD-ROM
(E2094-60003)
Accuracy Specifications ± (% of reading + % of range) 1
Function
Range 3
DC voltage
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V 4
100.0000 mV
to 750.000 V
True RMS
AC voltage 5
Resistance 6
DC current
True RMS
AC current 7
Frequency
or period
Capacitance 8
100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
1.000000 GΩ
100.0000 µA
1.000000 mA
10.00000 mA
100.0000 mA
1.000000 A
3.000000 A
100.0000 µA
to 3.00000 A
100 mV
to 750 V
1.0000 nF
10.000 nF
100.00 nF
1.0000 µF
10.000 µF
Temperature 9
RTD
-200 °C to 600 °C
Thermistor -80 °C to 150 °C
1.0000 V
Diode test 10
Frequency,
test current or
burden voltage
24 hour 2
Tcal ±1°C
0.0030 + 0.0030
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0006
3 Hz – 5 Hz
0.50 + 0.02
5 Hz – 10 Hz
0.10 + 0.02
10 Hz – 20 kHz
0.02 + 0.02
20 kHz – 50 kHz
0.04 + 0.025
50 kHz – 100 kHz
0.10 + 0.040
100 kHz – 300 kHz
1.00 + 0.250
1 mA
0.0030 + 0.0030
1 mA
0.0020 + 0.0005
100 µA
0.0020 + 0.0005
10 µA
0.0020 + 0.0005
5 µA
0.0020 + 0.0010
500 nA
0.0100 + 0.0010
500 nA || 10 MΩ
0.200 + 0.001
500 nA || 10 MΩ
2.000 + 0.001
< 0.03 V
0.010 + 0.020
< 0.30 V
0.007 + 0.006
< 0.03 V
0.007 + 0.020
< 0.30 V
0.010 + 0.004
< 0.80 V
0.050 + 0.006
< 2.0 V
0.100 + 0.020
3 Hz – 5 kHz
0.10 + 0.04
5 kHz – 10 kHz
0.20 + 0.04
3 Hz – 5 Hz
0.070 + 0.000
5 Hz – 10 Hz
0.040 + 0.000
10 Hz – 40 Hz
0.020 + 0.000
40 Hz – 300 kHz
0.005 + 0.000
500 nA
0.50 + 0.50
1 µA
0.40 + 0.10
10 µA
0.40 + 0.10
10 µA
0.40 + 0.10
100 µA
0.40 + 0.10
1 mA
0.06 °C
0.08 °C
0.002 + 0.010
1. Specifications are for 90 minute warm-up and 100 PLC.
2. Relative to calibration standards.
3. 20% overrange on all ranges, except DCV 1000 V, ACV 750 V, DCI and
ACI 3 A ranges.
4. For each additional volt over ± 500 V add 0.02 mV of error.
5. Specifications are for sinewave input > 0.3% of range and > 1 mVrms.
Add 30 µV error for frequencies below 1 kHz. 750 VAC range limited to 8 x 107
Volts-Hz. For each additional volt over 300 Vrms add 0.7 mVrms of error.
6. Specifications are for 4-wire resistance measurements, or 2-wire using
Math Null. Without Math Null, add 0.2 Ω additional error in 2-wire resistance
measurements.
90 day
Tcal ±5°C
1 year
Tcal ±5°C
Temperature coefficient/°C
0 °C to (Tcal -5°C)
(Tcal +5°C) to 55°C
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0006
0.50 + 0.03
0.10 + 0.03
0.04 + 0.02
0.08 + 0.025
0.20 + 0.040
1.20 + 0.250
0.008 + 0.004
0.007 + 0.001
0.007 + 0.001
0.010 + 0.001
0.030 + 0.001
0.600 + 0.001
6.000 + 0.001
0.0050 + 0.0035
0.0035 + 0.0007
0.0030 + 0.0005
0.0040 + 0.0060
0.0040 + 0.0060
0.50 + 0.03
0.10 + 0.03
0.045 + 0.02
0.090 + 0.025
0.200 + 0.040
1.200 + 0.250
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.012 + 0.001
0.040 + 0.001
0.800 + 0.001
8.000 + 0.001
0.040 + 0.025
0.030 + 0.006
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.120 + 0.020
0.10 + 0.04
0.20 + 0.04
0.070 + 0.000
0.040 + 0.000
0.020 + 0.000
0.006 + 0.000
0.50 + 0.50
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.050 + 0.025
0.050 + 0.006
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.150 + 0.020
0.10 + 0.04
0.20 + 0.04
0.070 + 0.000
0.040 + 0.000
0.020 + 0.000
0.007 + 0.000
0.50 + 0.50
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.40 + 0.10
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.010 + 0.003
0.008 + 0.003
0.005 + 0.002
0.010 + 0.0025
0.020 + 0.0040
0.120 + 0.0200
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1000 + 0.0001
1.0000 + 0.0001
0.0020 + 0.0030
0.0020 + 0.0005
0.0020 + 0.0020
0.0020 + 0.0005
0.0050 + 0.0010
0.0050 + 0.002
0.015 + 0.006
0.030 + 0.006
0.005 + 0.000
0.005 + 0.000
0.001 + 0.000
0.001 + 0.000
0.05 + 0.05
0.05 + 0.01
0.01 + 0.01
0.01 + 0.01
0.01 + 0.01
0.06 °C
0.08 °C
0.008 + 0.020
0.06 °C
0.08 °C
0.010 + 0.020
0.003 °C
0.002 °C
0.0010 + 0.0020
7. Specifications are for sinewave input > 1% of range and > 10 µArms.
Frequencies > 5 kHz are typical for 1 A and 3 A ranges.
8. Specifications are for 1-hour warm-up using Math Null. Additional errors may
occur for non-film capacitors.
9. For total measurement accuracy, add temperature probe error.
10. Accuracy specifications are for the voltage measured at the input terminals only.
1 mA test current is typical. Variation in the current source will create some
variation in the voltage drop across a diode junction.
3
A-to-D converter noise performance
Resolution
(ppm of range) 1
Normal mode
rejection (dB) 2
Readings/second 4
0.001
0.002
0.006
0.02
0.06
0.2
1
2
10
100
30
15
6
3
1.5
0.7
0.3
0.2
0.1
0.03
0
0
0
0
0
0
55
110 3
110 3
110 3
50,000
25,000
10,000
3,000
1,000
300
60 (50)
30 (25)
6 (5)
0.6 (0.5)
100
RMS noise (ppm of range)
Integration time
(NPLC)
10
1
0.1
0.01
0.001
0.01
0.1
1
10
100
1. Resolution is defined as the typical DCV 10 V range RMS noise.
Auto-zero on for NPLC ≥ 1. See manual for additional noise characteristics.
2. Normal mode rejection for power line frequency ± 0.1%.
3. For power-line frequency ± 1% 75 dB and for ± 3% 55 dB.
4. Maximum rate with auto-zero off for 60 Hz and (50 Hz) operation.
System reading and throughput rates
DMM memory to PC (maximum reading rate out of memory) 1
Drawing – Path B
LAN (VXI-11)
readings/sec
LAN (sockets)
readings/sec
4,000
89,000
47,000
8,500
265,000
154,000
7,000
110,000
60,000
8,500
270,000
160,000
ASCII
32-bit binary
64-bit binary
Direct I/O measurements (single reading – measure and I/O time) 1
Drawing – Path C
Function
Resolution
(NPLC)
GPIB
msec
DCV/2-wire
0.001
resistance
ACV/
Fast filter
frequency
1 ms gate
USB 2.0
LAN
LAN
msec (VXI-11) (sockets)
msec
msec
(readings/sec)
Drawing–Path A or C
2.6
2.9
4.6
3.2
50,000
10.0
10.0
10.0
10.0
500
System performance
DCV/2-wire
resistance
ACV/frequency
Range change Auto-range
LAN/GPIB
(msec) 3
2
(msec)
Path ‘A’
DMM
input
ADC
Maximum reading
rate into memory
or to direct I/O
1. ½ scale input signal, immediate trigger, trigger delay 0, auto-zero off, auto-range off, no math, null off,
60 Hz line frequency. See manual for performance on other functions.
Function
change
(msec) 1
System reading architecture
Maximum
external
trigger rate
Maximum
internal
trigger rate
22
3.9/2.6
7.5
5,000/s
50,000/s
37
6.5/6.4
19
500/s
500/s
1. Time to change from 2-wire Resistance to this specified function, or DCV to 2-wire resistance using the
SCPI ”FUNC” command
2. Time to change from one range to the next higher range, ≤ 10 V, ≤ 10 MΩ
3. Time to automatically change one range and be ready for the new measurement, ≤ 10 V, ≤ 10 MΩ
4
Path ‘C’
L4411A/
34410A/
DMM memory 34411A
50 k readings DMM
non-volatile
PC
GPIB
USB 2.0
readings/sec
LAN
GPIB
readings/sec
USB
Reading format
Path ‘B’
Measurement Characteristics
DC voltage
Measurement
method:
Linearity:
(10 V range)
Input
resistance:
Frequency and period
Continuously integrating multi-slope IV A/D
converter
0.0002% of reading
+ 0.0001% of range
0.1 V, 1 V, 10 V ranges
100 V, 1000 V ranges
Measurement
method:
Reciprocal-counting technique. AC-coupled
input using the AC voltage measurement
function.
Input impedance: 1 MΩ ± 2% in parallel with < 150 pF
Input protection: 750 Vrms all ranges
10 MΩ or > 10 GΩ
(selectable)
10 MΩ ±1% (fixed)
Capacitance
Measurement
method:
Connection type:
Input bias current: < 50 pA at 25 °C
Input protection: 1000 V
DC CMRR:
140 dB 1
Temperature
True RMS AC voltage
Thermistor:
RTD:
Measurement
method:
Crest factor:
AC-coupled True RMS measurement.
Digital sampling with anti-alias filter.
No additional error for crest factors < 10.
Limited by peak input and 300 kHz bandwidth.
Peak input:
300% of range or 1100 V
Overload
Will select higher range if peak input
ranging:
overload is detected during auto range.
Overload is reported in manual ranging.
AC CMR:
70 dB 2
Maximum input: 400 Vdc, 1100 Vpk
Input impedance: 1 MΩ ± 2% in parallel with < 150 pF
Input protection: 750 Vrms all ranges
Response time:
Continuity
threshold:
Input protection:
Maximum readings/second
Digits
4.5
Function 3
DCV
50 k
2-wire Ω
50 k
DCI
50 k
Frequency
500
Period
500
Filter setting
fast
ACV
500
ACI
500
Selectable 2-wire or 4-wire.
Current source referenced to LO input.
Selectable on the 100 Ω, 1 kΩ, and 10 kΩ
ranges
10% of range per lead for 100 Ω, 1 kΩ. 1 kΩ
per lead on all other ranges
1000 V on all ranges
Current shunt:
Maximum input:
Input protection:
5.5
10 k
10 k
10 k
90
90
med
150
150
6.5
1k
1k
1k
10
10
slow
50
50
Additional specifications
Resolution:
Overall bandwidth,
DCV & DCI:
Triggering:
Timebase
resolution:
Trigger jitter:
200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
3 A, 250 V fuse
True RMS AC current
Measurement
method:
300 samples/sec
Fixed at 10 Ω
Operating characteristics
DC current
Current shunt:
2.2 kΩ, 5 kΩ, and 10 kΩ
α = 0.00385
Ro from 49 Ω to 2.1 kΩ
Diode test
Resistance
Measurement
method:
Offset
compensation:
Max. lead
resistance
(4-wire):
Input protection:
Current input with measurement of
resulting ramp.
2-wire
AC-coupled true RMS measurement.
Directly coupled to the fuse and shunt.
Digital sampling with anti-alias filter.
200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
The peak value of the DC + AC current must
be < 300% of range. The RMS current must
be < 3 A including the DC current content.
3 A, 250 V fuse
See table on page 4
15 kHz typical @ 20 µs aperture (-3 dB)
Pre/Post, Analog Level, Int/Ext, Pos/Neg
19.9524 µs 0.01% accuracy
2 µs (p-p), 20 µs (p-p) when pre-triggered
1. For 1 kΩ unbalanced in LO lead, ± 500 V peak maximum
2. For 1 kΩ unbalanced in LO lead and < 60 Hz, ± 500 V peak maximum
3. Maximum rate for DCV, DCI, and resistance functions
(using zero settling delay, autozero off, manual range)
5
Ordering Information
Spurious-free dynamic range & signal to noise
distortion ratio
Agilent L4411A Multimeter
Function
DCV
Accessories included
Range
1V
10 V 1
100 V
Spur-free
-75 dB
-70 dB
-75 dB
SNDR
60 dB
60 dB
60 dB
Test report, power cord, LAN cross-over interface cable.
Product reference CD-ROM (34410-13601):
1. 10 V range: 2 V (p-p) <signal< 16 V (p-p)
• Software
◦ IntuiLink software
◦ IVI-COM driver
◦ LabView driver
◦ Example programs
• Online documentation
◦ Programmer’s reference
◦ Getting started guide
◦ User’s guide
◦ Service guide
◦ Localized manuals
Triggering and memory
Reading hold
sensitivity:
Samples per
trigger:
Trigger delay:
External trigger:
1% of reading
1 to 1,000,000
0 to 3600 sec (20 µs step size)
Programmable edge, low-power TTL
compatible
Delay:
< 1 µs
Jitter:
< 1 µs
Max rate:
5,000/sec
Min pulse width: 1 µs
Voltmeter
3 V logic output, 2 µs pulse with
complete:
programmable edge
Nonvolatile
50,000 readings
memory:
Volatile memory: 1,000,000 readings
Sample timer:
Range:
0 to 3600 s (20 µs step sizes)
Jitter:
< 100 ns
Agilent I/O Libraries CD-ROM (E2094-60003)
Options
Opt. STD
Opt. 001
Opt. ABA
Opt. 0B0
Opt. A6J
General specifications
Power supply:
Power line
frequency:
Power
consumption:
Operating
environment:
Storage
temperature:
Weight:
Dimensions:
Safety:
90 V – 264 V @ 45-66 Hz
90 V – 134 V @ 360 – 440 Hz
Automatically sensed at power-on
Rear measurement terminals ONLY
Front measurement terminals ONLY
Printed manual set (English)
Delete printed manual set
(full documentation included on CD ROM)
ANSI Z540 compliant calibration
Agilent accessories
Y1133A
Y1160A
50 VA peak (18 W average)
Full accuracy for 0 °C to 55 °C,
80% R.H. at 40 °C non-condensing
-40 °C to 70 °C
11059A
11060A
11062A
34171B
34172B
34330A
E2308A
1.9 kg (4.25 Lbs)
(W x H x D) 40.9 mm x 212.3 mm x 363.2 mm
IEC 61010-1, EN 61010-1, UL 61010-1,
CAN/CSA-C22.2 No. 61010-1, refer to
Declarations of Conformity for current
revisions. Measurement CAT II 300V, CAT I
1000V. Pollution Degree 2
EMC:
IEC 61326, EN 61326, CISPR 11, ICES-001,
AS/NZS 2064.1, refer to Declaration of
Conformity for current revisions.
Vibration & shock: MIL-T-28800E, Type III, Class 5 (Sine only)
LXI compliance:
LXI Class C, ver. 1.0
Warranty:
1 year
6
Measurement & trigger cable kit
Rack mount kit for L4400A series instruments
racks 1 or 2 instruments side-by-side on
sliding tray
Kelvin probe set
Surface mount device (SMD) test probes
Kelvin clip set
Input terminal connector (sold in pairs)
Input calibration short (sold in pairs)
30 A current shunt
5 kΩ thermistor probe
www.agilent.com
www.agilent.com/find/L4411A
www.agilent.com/find/dmm
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
Agilent
Open
www.agilent.com/find/open
Agilent Open simplifies the process
of connecting and programming
test systems to help engineers
design, validate and manufacture
electronic products. Agilent offers
open connectivity for a broad range
of system-ready instruments, open
industry software, PC-standard I/O
and global support, which are
combined to more easily integrate
test system development.
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
For information regarding self
maintenance of this product, please
contact your Agilent office.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
www.lxistandard.org
LXI is the LAN-based successor to
GPIB, providing faster, more efficient
connectivity. Agilent is a founding
member of the LXI consortium.
Product specifications and descriptions
in this document subject to change
without notice.
For more information on Agilent Technologies’ products, applications or services,
please contact your local Agilent office.
The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
07031 464 6333**
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus