LEM LH1040

True RMS AC/DC Clamp-on Multimeters
LH40 Series
•
•
•
•
•
•
•
•
•
Three models: 200A, 1,000A and 2,000A
AC and DC Amps, Volts, Ohms, Diode and
Continuity Test
Large jaw capacity takes 50mm ∅ or 2 x 30mm ∅
cables
True RMS measurement of complex waveforms
and analysis of AC and DC current components
Auto-ranging and Auto-zeroing
Excellent accuracy, even for distorted or nonsinusoidal A and V
Display-Hold and Max-Hold (surge) modes for
convenience in use
Analogue outputs
oscilloscope
for
recorder,
logger
or
Conformance to IEC1010 and EMC standards
High Accuracy
IEC 1010 and EMC Conformance
Advanced jaw design means that 40-series accuracy is little
affected by external magnetic fields or off-centre conductor
positioning. Utilisation of Hall effect technology ensures a broad
flat frequency response resulting in optimum accuracy even when
harmonics are present.
IEC 1010 safety features including a tactile barrier and special
jaw design provide the user with confidence when making
measurements in hazardous voltage areas.
Conformance to EMC standards ensures high reliability through
reduced susceptibility to electromagnetic interference.
True RMS measurement
By using True RMS measurement techniques the 40-series
instruments avoid errors (up to 50%) which can occur when nonsinusoidal waveforms created by today’s complex loads are
measured using traditional average-reading techniques. True
RMS measurements are available for AC, DC (A, V) and AC+DC
(A).
SPECIFICATION
Model
NON-INVASIVE CURRENT MEASUREMENT
Ranges (auto-ranging)
Measurement methods
Resolution
Accuracy
Crest Factor
Maximum measurable load
Maximum overload
ANALOGUE OUTPUT (2)
Switchable Instantaneous or RMS
(100 ms) output
Analogue output accuracy
LH240
40A, 200A
10mA (40A range)
100mA (200A range)
200A DC or AC peak
5mV/A
ELMES • HEME • NORMA
LH1040
400A, 1,000A
DC or AC True RMS, or DC only
100mA (400A range)
1A (1,000A range)
± 1.3 % of reading ± 3 digits (1)
6 maximum for True RMS measurements
1,000A DC or AC peak
10,000 Amps
1mV/A
± 1.3% of reading ± 1mV (1)
LH2040
400A, 2,000A
100mA (400A range)
1A (2,000A range)
2000A DC or AC peak
0.5mV/A
Voltage Measurement
Mechanical Data:
Methods of measurement
AC True RMS or DC (4)
Ranges (auto-ranging)
400V, 600V
Maximum overload
Accuracy
Resolution
Dimensions (H x W x D):
251 x 98 x 52 mm
9.88 x 3.86 x 2.05 inches
1,000V
Weight:
500gm / 1.1 lbs
±1% of reading ± 3 digits (1)
Jaw Capacity:
100mV (400V range)
Jaw Opening
1V (600V range)
Crest Factor
6 for V < 1,000 V peak
Input impedance
1 MΩ
1 x 50 mm/ 2.0-inch ∅ cables or
2 x 30mm/ 1.2-inch ∅ cables
55mm/ 2.2 inches
Environmental Data:
Operating Temperature:
0 oC to 50 oC (32 oF to 122 oF)
Resistance, Continuity and Diode Testing
Temperature Coefficient
± 0.1% of reading per °C
Ω ranges (auto-ranging)
400 Ω, 4 kΩ
(Current)
± 0.06% of reading per °F
Ω resolution
0.1 Ω (400 Ω range)
Storage Temperature:
-20o to 60o (-4 oF to 140 oF)
Ω accuracy
± 1% of reading ± 3 digits (1)
Safety:
Continuity sounder
Toggled on & off by )))) button;
All models comply with IEC1010-1, 600V working,
Installation category III, Pollution degree 2.
1 Ω (4 kΩ range)
(Ω range only)
Sounds when resistance < 50Ω
Input protection, Ω &
To 600 V, DC or sine wave RMS
diode test
Diode-test 3.2V max. open
Reads forward-biased diode
Circuit, 0.3mA short-circuit
Voltage to 2,000 mV
Diode-test accuracy
± 1 % of reading ± 2 digits (1)
Frequency Response
Maximum Safe Voltages:
Current measurement
600V AC RMS or DC between
(bare conductors)
uninsulated conductor & ground
Voltage measurement
600V AC RMS or DC between
input terminals or between live
terminal & local ground.
LEM
(Measurements and Analogue Output)
AC only True RMS V & A
15Hz to 1kHz (3,4)
DC True RMS A
DC + 15Hz to 1KHz (3,4)
DC V and A
DC only
Display
Size and type
4000-count LCD
Status indication
Low Battery, Data
(MAX Hold), AC, DC,
12mm / 0.5-inch characters
Hold,Pk
Diode Test, Ω, )))) (Continuity)
Refresh rate
3 times per second
Power Supply
Battery type
LEM offers a comprehensive range of electrical test products to
ensure the safe and efficient operation of electrical equipment
and installations. Other products from the LEM group includes
systems and analysers for power quality and power quality
monitoring and transducers for current and voltage
measurement. LEM provides complete measurement solutions
in current, voltage and power quality.
Notes:
1.
All accuracies stated at 23°C ± 1°C (73.4 ± 1.8°F)
2.
Analogue output is form plug-in adaptor with
standard BNC output socket
3.
At stated accuracy, extend to 5kHz for 3dB point
4.
True RMS measurements taken over 100ms
9 V Alkaline: MN1604, PP3
IEC 6LR61 or equivalent
Battery life
Typically 40 hours
Printed in Austria
Technical modifications reserved
Publication A 99000640
LEM NORMA GmbH
Export department
Palmersstrasse 2
A-2351 WIENER NEUDORF
TEL:
+43(0)2236 691 0
FAX:
+43(0)2236 63 080
E-mail: [email protected]
BELGIUM & LUXEMBURG
LEM Belgium sprl-bvba
Route de Petit-Roeulx, 95
B-7090 BRAINE-LE-COMTE
TEL:
+32(0)67 55 01 14
FAX:
+32(0)67 55 01 15
E-mail: [email protected]
GREAT BRITAIN & IRELAND
LEM UK LTD
Geneva Court
1 Penketh Place
West Pimbo
Skelmersdale, Lancashire WN8 9QX
TEL: +44(0)1 695 72 07 77
FAX: +44(0)1 695 50 704
E-mail: [email protected]
NAFTA
LEM Instruments Inc.
23822 Hawthorne Boulevard #100
US-TORRANCE, CA 90505
TEL: +1 310 373 09 66
FAX: +1 310 373 90 56
E-mail: [email protected]
NETHERLANDS
LEM Nederland B.V.
Rijzendeweg 5
NL-4634 TV WOENSDRECHT
TEL: +31(0)16 46 154 62
FAX: +31(0)16 46 166 06
E-mail: [email protected]