HP N5230C

Agilent
4-Port PNA-L
Microwave Network Analyzer
N5230C
300 kHz to 13.5, 20 GHz
Data Sheet
Note:
Specification information in
this document is also available
within the PNA-L network analyzer’s internal Help system.
Table of Contents
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Corrected System Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Table 1. System dynamic range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Table 2. Extended dynamic range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
N5230C Option 140/145/146/240/245/246
Corrected system performance with 3.5mm connectors . . . . . . . . . . . 6
Table 3. 85052B Calibration kit N5230C
Option 140/145/146/240/245/246 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Table 4. N4433A Electronic calibration module N5230C
Option 140/145/146/240/245/246 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Table 5. Uncorrected system performance . . . . . . . . . . . . . . . . . . . . . . . 10
Table 6. Test port output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Table 7: Test port input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Table 8. Dynamic accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Table 9. Test port input (group delay) . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Table 10. Miscellaneous information . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 11. Front panel information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 12. Rear panel information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table 13. Analyzer environment and dimensions . . . . . . . . . . . . . . . . .
23
23
24
25
Measurement Throughput Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Table 14. Typical cycle time (ms) for measurement completion . . . . . . 26
Table 15. Cycle time vs IF bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Table 16. Cycle time vs number of points . . . . . . . . . . . . . . . . . . . . . . . . . 28
Table 17. Data transfer time (ms). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Specifications: Front-Panel Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Table 18: Measurement receiver inputs, 0.1 dB typical compression .
Table 19: Reference receiver input at max. specified output power . .
Table 20: Reference output at max. specified output power. . . . . . . . .
Table 21: Source outputs at max. specified output power . . . . . . . . . .
Table 22: Coupler inputs, insertion loss of coupler thru. . . . . . . . . . . .
Table 23: Coupler outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
29
29
29
30
30
30
Test Set Block Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
N5230C Option 140 or 240
(standard test set and standard power range) network analyzer. . . 31
N5230C Option 145 or 245
(configurable test set and extended power range) network analyzer . . 31
N5230C Option 146 or 246
(configurable test set, extended power range, and internal second
source) network analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Web Resources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2
This is a subset of technical specifications for the N5230C Options 140, 145, 146, 240, 245,
and 246 network analyzers.
• Option 140, 300 kHz to 13.5 GHz, 4-port with standard test set and standard power range
• Option 145, 300 kHz to 13.5 GHz, 4-port with configurable test set and extended power
range
• Option 146, 300 kHz to 13.5 GHz, 4-port with configurable test set and extended power
range with internal second source
• Option 240, 300 kHz to 20 GHz, 4-port with standard test set and standard power range
• Option 245, 300 kHz to 20 GHz, 4-port with configurable test set and extended power
range
• Option 246, 300 kHz to 20 GHz, 4-port with configurable test set and extended power
range with internal second source
To view or print the N5230C technical specifications, visit our web site at
www.agilent.com/find/pnal
This N5230C document provides technical specifications for the following calibration
kit and ECal module only: 85052B and N4433A. Please download our free Uncertainty
Calculator from www.agilent.com/find/na_calculator to generate the curves for
your calibration kit and PNA setup.
Definitions
All specifications and characteristics apply over a 25 °C ±5 °C range
(unless otherwise stated) and 90 minutes after the instrument has been turned
on.
Specification (spec.): Warranted performance. Specifications include
guardbands to account for the expected statistical performance distribution,
measurement uncertainties, and changes in performance due to
environmental conditions.
Characteristic (char.): A performance parameter that the product is expected
to meet before it leaves the factory, but that is not verified in the field and is not
covered by the product warranty. A characteristic includes the same guardbands
as a specification.
Typical (typ.): Expected performance of an average unit which does not include
guardbands. It is not covered by the product warranty.
Nominal (nom.): A general, descriptive term that does not imply a level of
performance. It is not covered by the product warranty.
Calibration: The process of measuring known standards to characterize a
network analyzer's systematic (repeatable) errors.
Corrected (residual): Indicates performance after error correction (calibration). It
is determined by the quality of calibration standards and how well "known" they
are, plus system repeatability, stability, and noise.
Uncorrected (raw): Indicates instrument performance without error correction. The
uncorrected performance affects the stability of a calibration.
Standard: When referring to the analyzer, this includes no options unless
noted otherwise.
3
Corrected System Performance
The specifications in this section apply for measurements made with the N5230C
Options 140, 145, 146, 240, 245, and 246 analyzers with the following conditions:
• 10 Hz IF bandwidth
• No averaging applied to data
• Isolation calibration with an averaging factor of 8
Table 1. System dynamic range at test port1
Standard configuration and standard power range (Options 140, 240)
Description
300 kHz to 10 MHz2
10 MHz to 4 GHz2
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
Specification (dB)
at test port
120
118
115
107
107
103
Typical (dB)
at test port
111
128
129
127
119
119
116
Configurable test set and extended power range (Options 145, 245)
Configurable test set, extended power range, and internal second source
(Options 146, 246)
Description
300 kHz to 10 MHz2
10 MHz to 4 GHz2
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
Specification (dB)
at test port
120
118
113
105
105
98
Typical (dB)
at test port
111
128
128
125
117
117
115
1. The system dynamic range is calculated as the difference between the noise floor and the specified
source maximum output power. The effective dynamic range must take measurement uncertainties
and interfering signals into account.
2. May be degraded typically by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz due
to spurious receiver residuals. Methods are available to regain the full dynamic range.
4
Table 2. Extended dynamic range1
Configurable test set and extended power range (Options 145, 245)
Configurable test set, extended power range, and internal second source
(Options 146, 246)
Description
300 kHz to 10 MHz2
10 MHz to 4 GHz2
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
Specification (dB) at
direct receiver
access input
Typical (dB) at
direct receiver
access input
124
136
134
129
121
121
114
1. The direct receiver access input extended dynamic range is calculated as the difference between
the direct receiver access input noise floor and the source maximum output power. The effective
dynamic range must take measurement uncertainties and interfering signals into account. This
setup should only be used when the receiver input will never exceed its compression or damage
level. When the analyzer is in segment sweep mode, it can have predefined frequency segments
which will output a higher power level when the extended dynamic range is required (i.e. devices
with high insertion loss), and reduced power when receiver compression or damage may
occur (i.e. devices with low insertion loss). The extended range is only available in one-path
transmission measurements.
2. May be degraded typically by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz
due to spurious receiver residuals. Methods are available to regain the full dynamic range.
5
N5230C Option 140/145/146/240/245/246
Corrected system performance with 3.5 mm connectors1
Note: For any Sii reflection measurement:
Sjj = 0
For any Sij transmission measurement:
Sji = Sij when Sij ≤ 1
Sji = 1/Sij when Sij ≥ 1
Skk = 0 for all k
Table 3. 85052B Calibration kit
N5230C
• Option 140 or 240 standard test set and standard power range
• Option 145 or 245 configurable test set and extended power range
• Option 146 or 246 configurable test set and extended power range with internal
second source
Applies to the N5230C Option 140/145/146/240/245/246 analyzers, 85052B
(3.5mm) calibration kit, 85131F flexible test port cable set, and a full 4-port
calibration. Also applies to the following condition: Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature.
Directivity
Source match
Load match
Reflection
tracking
Transmission
10 to
500 MHz
48
40
48
±0.003
(+0.01/°C)
±0.017
500 MHz to
2 GHz
48
40
48
±0.003
(+0.01/°C)
±0.017
tracking
(+0.01/°C)
(+0.01/°C)
Description
Specification (dB)
2 to
8 to
8 GHz
13.5 GHz
44
44
33
31
44
44
±0.003
±0.006
(+0.02/°C)
(+0.03/°C)
±0.062
±0.125
(+0.02/°C)
(+0.03/°C)
13.5 to
20 GHz
44
31
44
±0.006
(+0.03/°C)
±0.125
(+0.03/°C)
Transmission uncertainty (specifications)
Magnitude
N5230C Option 140/145/146/240/245/246 with 85052B
10
■ 10 to 500 MHz
Uncertainty (dB)
♦ 500 MHz to 2 GHz
▲ 2 to 8 GHz
● 8 to 20 GHz
1
0.1
S11 = S22 = 0
Source power = -5 dBm
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90
Transmission coefficient (dB)
1. From 300 kHz to 10 MHz, performance is characterized as “typical”. To generate these typical values,
please download our free Uncertainty Calculator from www.agilent.com/find/na_calculator.
6
Table 3. 85052B Calibration kit (continued)
N5230C
• Option 140 or 240 standard test set and standard power range
• Option 145 or 245 configurable test set and extended power range
• Option 146 or 246 configurable test set and extended power range with internal
second source
Phase
N5230C Option 140/145/146/240/245/246 with 85052B
100
Uncertainty (degrees)
■
10 to 500 MHz
♦ 500 MHz to 2 GHz
▲ 2 to 8 GHz
● 8 to 20 GHz
10
1
S11 = S22 = 0
Source power = -5 dBm
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90
Transmission coefficient (dB)
Reflection uncertainty (specifications)
Magnitude
N5230C Option 140/145/146/240/245/246 with 85052B
0.05
Uncertainty (linear)
■
10 to 500 MHz
♦ 500 MHz to 2 GHz
0.04
▲ 2 to 8 GHz
● 8 to 20 GHz
0.03
S21 = S12 = 0
Source power = -5 dBm
0.02
0.01
0
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Phase
N5230C Option 140/145/146/240/245/246 with 85052B
10
Uncertainty (deg)
■
10 to 500 MHz
♦ 500 MHz to 2 GHz
8
▲ 2 to 8 GHz
● 8 to 20 GHz
6
S21 = S12 = 0
Source power = -5 dBm
4
2
0
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
7
N5230C Option 140/145/146/240/245/246
Corrected system performance with 3.5 mm connectors1 (continued)
Table 4. N4433A Electronic calibration module
N5230C
• Option 140 or 240 standard test set and standard power range
• Option 145 or 245 configurable test set and extended power range
• Option 146 or 246 configurable test set and extended power range with internal
second source
Applies to the N5230C Option 140/145/146/240/245/246 analyzers, N4433A
electronic calibration module, 85131F flexible test port cable set, and a full 4-port
calibration. Also applies to the following condition: Environmental temperature
23° ±3 °C, with < 1 °C deviation from calibration temperature.
10 to
500 MHz to
Description
500 MHz
2 GHz
Directivity
52
52
Source match
42
42
Load match
45
45
Reflection
±0.060
±0.060
tracking
(+0.01/°C)
(+0.01/°C)
Transmission
±0.045
±0.039
tracking
(+0.01/°C)
(+0.01/°C)
Transmission uncertainty (specifications)
Specification (dB)
2 to
8 to
8 GHz
13.5 GHz
47
45
39
31
41
35
±0.090
±0.040
(+0.02/°C)
(+0.03/°C)
±0.055
±0.127
(+0.02/°C)
(+0.03/°C)
13.5 to
20 GHz
45
31
35
±0.180
(+0.03/°C)
±0.160
(+0.03/°C)
Magnitude
N5230C Option 140/145/146/240/245/246 full 2-port2 cal using N4433A
10
Uncertainty (dB)
■
10 to 500 MHz
♦ 500 MHz to 2 GHz
▲ 2 to 8 GHz
● 8 to 20 GHz
1
0.1
S11 = S22 = 0
Source power = -5 dBm
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90
Transmission coefficient (dB)
Phase
100
N5230C Option 140/145/146/240/245/246 full 2-port2 cal using N4433A
Uncertainty (degrees)
■ 10 to 500 MHz
♦ 500 MHz to 2 GHz
▲ 2 to 8 GHz
● 8 to 20 GHz
10
1
S11 = S22 = 0
Source power = -5 dBm
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90
Transmission coefficient (dB)
1. From 300 kHz to 10 MHz, performance is characterized as “typical”. To generate these typical values,
8
please download our free Uncertainty Calculator from www.agilent.com/find/na_calculator.
2. All of the curves are for 2-port calibrations. Multiport uncertainties are currently not supported.
Table 4. N4433A Electronic calibration module (continued)
N5230C
• Option 140 or 240 standard test set and standard power range
• Option 145 or 245 configurable test set and extended power range
• Option 146 or 246 configurable test set and extended power range with internal
second source
Reflection uncertainty (specifications)
Magnitude
N5230C Option 140/145/146/240/245/246 with N4433A
0.05
■
Uncertainty (linear)
10 to 500 MHz
♦ 500 MHz to 2 GHz
0.04
▲ 2 to 8 GHz
● 8 to 20 GHz
0.03
S21 = S12 = 0
Source power = -5 dBm
0.02
0.01
0
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Phase
N5230C Option 140/145/146/240/245/246 with N4433A
10
■
8
Uncertainty (deg)
10 to 500 MHz
♦ 500 MHz to 2 GHz
▲ 2 to 8 GHz
● 8 to 20 GHz
6
S21 = S12 = 0
Source power = -5 dBm
4
2
0
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
9
Table 5. Uncorrected system performance1
Directivity
Specifications
Options 140, 145, 146,
240, 245, 246
300 kHz to 10 MHz
10 MHz to 1 GHz
1 to 3 GHz
3 to 5 GHz
5 to 11.5 GHz
11.5 to 13.5 GHz
13.5 to 20 GHz
–28 dB
–25 dB
–20 dB
–17 dB
–15 dB
–15 dB
Source match
300 kHz to 10 MHz
10 MHz to 1 GHz
1 to 3 GHz
3 to 5 GHz
5 to 10.5 GHz
10.5 to 11.5 GHz
11.5 to 13.5 GHz
13.5 to 20 GHz
–12 dB
–12 dB
–12 dB
–12 dB
–10 dB
–8 dB
–8 dB
Load match
300 kHz to 10 MHz
10 MHz to 1 GHz
1 to 3 GHz
3 to 5 GHz
5 to 11.5 GHz
11.5 to 13.5 GHz
13.5 to 16 GHz
16 to 20 GHz
–20 dB
–20 dB
–18 dB
–12 dB
–7 dB
–7 dB
–7.5 dB
Crosstalk2
300 kHz to 5 MHz
5 to 10 MHz
10 to 45 MHz
45 MHz to 4 GHz
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
–
–
–
–
–
–
–
–
–
Typicals
Options 140, 145, 146,
240, 245, 246
–23 dB
–8 dB
–9 dB
–70 dB
–100 dB
–110 dB
–122 dB
–123 dB
–120 dB
–115 dB
–115 dB
–110 dB
1. Specifications apply over environmental temperature of 25 °C ±5 °C with less than 1 °C variation
from calibration temperature.
2. Measurement conditions: normalized to a thru, measured with two shorts, 10 Hz IF bandwidth,
averaging factor of 8, alternate mode source power set to the lesser of the maximum power out
or the maximum receiver power.
10
Table 6. Test port output1
Specifications
Description
Typicals
Options Options Options Options
140, 240
145, 146, 245, 246
140, 240
Frequency range
Options 140, 145, 146 300 kHz to 13.5 GHz
Options 240, 245, 246 300 kHz to 20 GHz
Nominal power
–5 dBm
–8 dBm
145, 146, 245, 246
Preset power; attenuator switch
point 10 dB below nominal power
Frequency resolution
1 Hz
CW accuracy
±1 ppm
Frequency stability
–
–
Description
–
–
Specifications
Options
Options
140, 240
145, 146, 245, 246
±0.05 ppm. –10° to 70° C
±0.1 ppm/yr maximum
Typicals
Options
140, 240
Options
145, 146, 245, 246
Power level accuracy
Variation from nominal power in range 0
300 kHz to 10 MHz
±1.0 dB
±1.0 dB
10 MHz to 2 GHz
±1.0 dB
±1.0 dB
2 to 10.5 GHz
±1.5 dB
±1.5 dB
10.5 to 13.5 GHz
±2.5 dB
±2.5 dB
13.5 to 20 GHz
±2.5 dB
±2.5 dB
Max leveled power
300 kHz to 10 MHz
+8 dBm
+8 dBm
10 MHz to 4 GHz
+8 dBm
+8 dBm
+12 dBm
+11 dBm
4 to 6 GHz
+6 dBm
+6 dBm
+10 dBm
+9 dBm
6 to 10.5 GHz
+3 dBm
+1 dBm
+8 dBm
+6 dBm
10.5 to 13.5 GHz
0 dBm
–2 dBm
+5 dBm
+3 dBm
13.5 to 15 GHz
0 dBm
–2 dBm
+5 dBm
+3 dBm
15 to 20 GHz
–3 dBm
–8 dBm
+2 dBm
–1 dBm
Power level linearity
Specified on Port 1 only. Ports 2, 3, 4 performance is Typical. Test is at the nominal power level.
300 kHz to 10 MHz
±2.0 dB
±2.0 dB
10 MHz to 1 GHz
±2.0 dB
±2.0 dB
1 to 13.5 GHz
±1.5 dB
±1.5 dB
13.5 to 20 GHz
±1.5 dB
±1.5 dB
Power sweep range (ALC)
ALC range starts at maximum-leveled power and decreases by the dB amount specified here.
300 kHz to 10 MHz
35 dB
35 dB
10 MHz to 4 GHz
33 dB
33 dB
4 to 6 GHz
31 dB
31 dB
6 to 10.5 GHz
28 dB
26 dB
10.5 to 13.5 GHz
25 dB
23 dB
13.5 to 15 GHz
25 dB
23 dB
15 to 20 GHz
22 dB
17 dB
Power resolution
0.01 dB
0.01 dB
11
Table 6. Test port output1 (Continued)
Description
Typicals
Options
Options
140, 240
145, 146, 245, 246
Power range
300 kHz to 10 MHz
–27 to +8 dBm
–87 to +8 dBm
10 to 45 MHz
–27 to +12 dBm
–87 to +11 dBm
45 MHz to 4 GHz
–27 to +12 dBm
–87 to +11 dBm
4 to 6 GHz
–27 to +10 dBm
–87 to +9 dBm
6 to 10.5 GHz
–27 to +8 dBm
–87 to +6 dBm
10.5 to 13.5 GHz
–27 to +5 dBm
–87 to +3 dBm
13.5 to 15 GHz
–27 to +5 dBm
–87 to +3 dBm
15 to 20 GHz
–27 to +2 dBm
–87 to –1 dBm
Power settings
Minimum power setting
–30 dBm
–90 dBm
Maximum power setting
+20 dBm
+20 dBm
Harmonics (2nd or 3rd) at maximum output power 2
In-band source harmonics
300 kHz to 10 MHz
–17 dBc
10 MHz to 1 GHz
–17 dBc
1 to 13.5 GHz
–20 dBc
13.5 to 20 GHz
–20 dBc
Non-harmonic spurious (at nominal output power 2)
300 kHz to 20 GHz
–50 dBc for offset frequency > 1 kHz
Typical performance
Phase noise (Nominal power at test port)
10 kHz Offset
300 kHz to 10 MHz
–86 dBc/Hz
10 MHz to 1.5 GHz
–86 dBc/Hz
1.5 to 3.125 GHz
–83 dBc/Hz
3.125 to 6.25 GHz
–77 dBc/Hz
6.25 to 12.5 GHz
–71 dBc/Hz
12.5 to 13.5 GHz
–65 dBc/Hz
13.5 to 20 GHz
–65 dBc/Hz
100 kHz Offset
–86 dBc/Hz
–91 dBc/Hz
–91 dBc/Hz
–85 dBc/Hz
–79 dBc/Hz
–73 dBc/Hz
–73 dBc/Hz
1 MHz Offset
–95 dBc/Hz
–95 dBc/Hz
–95 dBc/Hz
–89 dBc/Hz
–83 dBc/Hz
–77 dBc/Hz
–77 dBc/Hz
1. For Options 140/145/240/245, performance specified on Port 1 only; Ports 2, 3, and 4 performance
is typical. For Options 146/246, performance is specified on Ports 1 and 3 only; Ports 2 and 4
performance is typical.
2. Spurious signal levels are valid for a 10 MHz span centered on the carrier frequency. Spurious
signals up to -15 dBc may exist outside the span. These signals do not affect the accuracy of the
network analyzer measurements.
12
Table 7: Test port input
Description
Specification
Options 140, 145, 146, 240, 245, 246
Typicals
Options 140, 145, 146, 240, 245, 246
Test port noise floor
Total average (rms) noise power calculated as the mean value of a linear magnitude trace expressed in dBm.
10 Hz IF bandwidth
300 kHz to 10 MHz
< –103 dBm
10 to 500 MHz
< –112 dBm
< –116 dBm
500 MHz to 4 GHz
< –112 dBm
< –120 dBm
4 to 10.5 GHz
< –112 dBm
< –119 dBm
10.5 to 13.5 GHz
< –107 dBm
< –114 dBm
13.5 to 15 GHz
< –107 dBm
< –114 dBm
15 to 20 GHz
< –106 dBm
< –114 dBm
1 KHz IF bandwidth
300 kHz to 10 MHz
< –83 dBm
10 to 500 MHz
< –92 dBm
< –96 dBm
500 MHz to 4 GHz
< –92 dBm
< –100 dBm
4 to 10.5 GHz
< –92 dBm
< –99 dBm
10.5 to 13.5 GHz
< –87 dBm
< –94 dBm
13.5 to 15 GHz
< –87 dBm
< –94 dBm
15 to 20 GHz
< –86 dBm
< –94 dBm
Direct receiver access input noise floor (Options 145, 146, 245, 246)
Total average (rms) noise power calculated as the mean value of a linear magnitude trace expressed in dBm.
10 Hz IF bandwidth
300 kHz to 10 MHz
< –119 dBm
10 to 500 MHz
< –128 dBm
< –132 dBm
500 MHz to 4 GHz
< –128 dBm
< –136 dBm
4 to 10.5 GHz
< –128 dBm
< –135 dBm
10.5 to 13.5 GHz
< –123 dBm
< –130 dBm
13.5 to 15 GHz
< –123 dBm
< –130 dBm
15 to 20 GHz
< –122 dBm
< –130 dBm
1 KHz IF bandwidth
300 kHz to 10 MHz
< –99 dBm
10 to 500 MHz
< –108 dBm
< –112 dBm
500 MHz to 4 GHz
< –108 dBm
< –116 dBm
4 to 10.5 GHz
< –108 dBm
< –115 dBm
10.5 to 13.5 GHz
< –103 dBm
< –110 dBm
13.5 to 15 GHz
< –103 dBm
< –110 dBm
15 to 20 GHz
< –102 dBm
< –110 dBm
Specification
Options 140, 145, 146, 240, 245, 246
Compression level (at +8 dBm except as noted)
Power
Compression
300 kHz to 10 MHz
10 to 50 MHz
+8 dBm
0.35 dB
50 MHz to 1 GHz
+8 dBm
0.35 dB
1 to 8 GHz
+8 dBm
0.25 dB
8 to 12.5 GHz
+8 dBm
0.30 dB
12.5 to 13.5 GHz
+8 dBm
0.55 dB
13.5 to 20 GHz
+8 dBm
0.55 dB
Test port compression at 0.1 dB
300 kHz to 10 MHz
–
–
10 MHz to 1 GHz
–
–
1 to 12.5 GHz
–
–
12.5 to 13.5 GHz
–
–
13.5 to 20 GHz
–
–
Description
Typicals
Options 140, 145, 146, 240, 245, 246
Power
+5 dBm
Compression
0.10 dB
+5 dBm
+9 dBm
+10 dBm
+9 dBm
+9 dBm
13
Table 7. Test port input (Continued)
Description
Specifications
Options 140, 240 Options 145, 245 Options 146, 246
Trace noise magnitude
Ratioed measurement, nominal power at test port.
100 kHz IF bandwidth
300 kHz to 10 MHz
10 MHz to 10.5 GHz
0.006 dB rms
0.008 dB rms
10.5 to 13.5 GHz
0.010 dB rms
0.014 dB rms
13.5 to 20 GHz
0.010 dB rms
0.014 dB rms
600 kHz IF bandwidth
300 kHz to 10 MHz
–
10 MHz to 10.5 GHz
–
10.5 to 13.5 GHz
–
13.5 to 20 GHz
–
100 kHz IF bandwidth
Measured at maximum specified power
300 kHz to 10 MHz
–
10 MHz to 2 GHz
–
2 to 10.5 GHz
–
10.5 to 13.5 GHz
–
13.5 to 20 GHz
–
Trace noise phase
Ratioed measurement, nominal power at test port.
100 kHz IF bandwidth
300 kHz to 10 MHz
10 MHz to 10.5 GHz
0.05° rms
0.07° rms
10.5 to 13.5 GHz
0.08° rms
0.10° rms
13.5 to 20 GHz
0.08° rms
0.10° rms
600 kHz IF bandwidth
300 kHz to 10 MHz
–
10 MHz to 10.5 GHz
–
10.5 to 13.5 GHz
–
13.5 to 20 GHz
–
100 kHz IF bandwidth
Measured at maximum specified power
300 kHz to 10 MHz
–
10 MHz to 2 GHz
–
2 to 10.5 GHz
–
10.5 to 13.5 GHz
–
13.5 to 20 GHz
–
Stability magnitude
Stability as defined as a ratio measurement made at the test port.
300 kHz to 10 MHz
–
10 MHz to 2 GHz
–
2 to 4 GHz
–
4 to 13.5 GHz
–
13.5 to 16 GHz
–
16 to 19 GHz
–
19 to 20 GHz
–
14
0.016 dB rms
0.038 dB rms
0.038 dB rms
0.130° rms
0.250° rms
0.250° rms
Typicals
Options 140, 240 Options 145, 245 Options 146, 246
0.015 dB rms
0.004 dB rms
0.007 dB rms
0.007 dB rms
0.030 dB rms
0.005 dB rms
0.009 dB rms
0.009 dB rms
0.050 dB rms
0.013 dB rms
0.026 dB rms
0.026 dB rms
0.015 dB rms
0.013 dB rms
0.017 dB rms
0.017 dB rms
0.030 dB rms
0.015 dB rms
0.023 dB rms
0.023 dB rms
0.050 dB rms
0.032 dB rms
0.063 dB rms
0.063 dB rms
0.005 dB rms
0.001 dB rms
0.002 dB rms
0.006 dB rms
0.006 dB rms
0.010 dB rms
0.003 dB rms
0.003 dB rms
0.009 dB rms
0.009 dB rms
0.012 dB rms
0.004 dB rms
0.004 dB rms
0.023 dB rms
0.023 dB rms
0.110° rms
0.025° rms
0.050° rms
0.050° rms
0.180° rms
0.035° rms
0.060° rms
0.060° rms
0.280° rms
0.090° rms
0.170° rms
0.170° rms
0.110° rms
0.080° rms
0.120° rms
0.120° rms
0.180° rms
0.100° rms
0.160° rms
0.160° rms
0.300° rms
0.200° rms
0.430° rms
0.430° rms
0.040° rms
0.007° rms
0.012° rms
0.040° rms
0.040° rms
0.050° rms
0.012° rms
0.015° rms
0.060° rms
0.060° rms
0.075° rms
0.013° rms
0.030° rms
0.150° rms
0.150° rms
±0.015 dB/°C
±0.010 dB/°C
±0.015 dB/°C
±0.020 dB/°C
±0.020 dB/°C
±0.025 dB/°C
±0.030 dB/°C
±0.015 dB/°C
±0.010 dB/°C
±0.015 dB/°C
±0.020 dB/°C
±0.020 dB/°C
±0.025 dB/°C
±0.030 dB/°C
Table 7. Test port input (Continued)
Description
Specifications
Typicals
Options 140, 240 Options 145, 245 Options 146, 246
Options 140, 240 Options 145, 245 Options 146, 246
Stability phase
Stability as defined as a ratio measurement made at the test port.
300 kHz to 10 MHz
–
10 to 45 MHz
–
45 to 500 MHz
–
500 MHz to 2 GHz
–
2 to 4 GHz
–
4 to 8 GHz
–
8 to 13.5 GHz
–
13.5 to 16 GHz
–
16 to 20 GHz
–
Reference level magnitude
Range
Resolution
Reference level phase
Range
Resolution
Damage input level
Test port 1, 2, 3, and 4
Receivers R, A, B, C, D
Source out (reference)
Source out (test ports)
Coupler thru
Coupler arm
±0.360°/°C
±0.020°/°C
±0.030°/°C
±0.050°/°C
±0.100°/°C
±0.150°/°C
±0.300°/°C
±0.300°/°C
±0.350°/°C
±200 dB
.001 dB
±200 dB
.001 dB
±200 dB
.001 dB
±500°
.01°
±500°
.01°
±500°
.01°
–
–
–
–
–
–
–
+27 dBm
or ±16 VDC
±0.360°/°C
±0.020°/°C
±0.030°/°C
±0.070°/°C
±0.150°/°C
±0.250°/°C
±0.500°/°C
±0.500°/°C
±0.650°/°C
+27 dBm or ±16 VDC
+15 dBm or ±16 VDC
+27 dBm or ±16 VDC
+27 dBm or ±16 VDC
+27 dBm or ±16 VDC
+15 dBm or ±0 VDC
15
Table 8. Dynamic Accuracy (specification)
Accuracy of the test port input power reading relative to the reference input power level.
Dynamic accuracy is verified with the following measurements:
• Compression over frequency
• IF linearity at a single frequency of 1.195 GHz using a reference level of –20 dBm
for an input power range of 0 to –110 dBm
Dynamic Accuracy 0.045 GHz
Magnitude
N5230C Option 140/145/146/240/245/246
10
■
-10 dBm at 0.045 GHz
Accuracy (dB)
♦ -20 dBm at 0.045 GHz
▲ -30 dBm at 0.045 GHz
● -40 dBm at 0.045 GHz
1
0.1
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
Test port power (dBm)
Phase
N5230C Option 140/145/146/240/245/246
100
Accuracy (degrees)
■
-10 dBm at 0.045 GHz
♦ -20 dBm at 0.045 GHz
▲ -30 dBm at 0.045 GHz
● -40 dBm at 0.045 GHz
10
1
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
Test port power (dBm)
16
-80
-90 -100 -110 -120
Table 8. Dynamic Accuracy (continued)
Dynamic Accuracy 0.500 GHz
Magnitude
N5230C Option 140/145/146/240/245/246
10
▄ -10 dBm at 0.500 GHz
Accuracy (dB)
♦ -20 dBm at 0.500 GHz
▲ -30 dBm at 0.500 GHz
z -40 dBm at 0.500 GHz
1
0.1
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
Test port power (dBm)
Phase
N5230C Option 140/145/146/240/245/246
100
▄ -10 dBm at 0.500 GHz
Accuracy (degrees)
♦ -20 dBm at 0.500 GHz
▲-30 dBm at 0.500 GHz
● -40 dBm at 0.500 GHz
10
1
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
Test port power (dBm)
17
Table 8. Dynamic Accuracy (continued)
Dynamic Accuracy 1 to 2 GHz
Magnitude
N5230C Option 140/145/146/240/245/246
10
■ -10 dBm (1 to 2 GHz)
Accuracy (dB)
♦ -20 dBm (1 to 2 GHz)
▲ -30 dBm (1 to 2 GHz)
● -40 dBm (1 to 2 GHz)
1
0.1
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
Test port power (dBm)
Phase
N5230C Option 140/145/146/240/245/246
100
Accuracy (degrees)
■ -10 dBm (1 to 2 GHz)
♦ -20 dBm (1 to 2 GHz)
▲ -30 dBm (1 to 2 GHz)
● -40 dBm (1 to 2 GHz)
10
1
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
Test port power (dBm)
18
-80
-90 -100 -110 -120
Table 8. Dynamic Accuracy (continued)
Dynamic Accuracy 2 to 12.5 GHz
Magnitude
N5230C Option 140/145/146/240/245/246
10
▄ -10 dBm (2 to 12.5 GHz)
Accuracy (dB)
♦ -20 dBm (2 to 12.5 GHz)
▲-30 dBm (2 to 12.5 GHz)
● -40 dBm (2 to 12.5 GHz)
1
0.1
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
Test port power (dBm)
Phase
N5230C Option 140/145/146/240/245/246
100
Accuracy (degrees)
▄ -10 dBm (2 to 12.5 GHz)
♦ -20 dBm (2 to 12.5 GHz)
▲ -30 dBm (2 to 12.5 GHz)
● -40 dBm (2 to 12.5 GHz)
10
1
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
Test port power (dBm)
19
Table 8. Dynamic Accuracy (continued)
Dynamic Accuracy 12.5 to 13.5 GHz
Magnitude
N5230C Option 140/145/146/240/245/246
10
▄ -10 dBm (12.5 to 13.5 GHz)
Accuracy (dB)
♦ -20 dBm (12.5 to 13.5 GHz)
▲ -30 dBm (12.5 to 13.5 GHz)
● -40 dBm (12.5 to 13.5 GHz)
1
0.1
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
Test port power (dBm)
Phase
N5230C Option 140/145/146/240/245/246
100
Accuracy (degrees)
▄ -10 dBm (12.5 to 13.5 GHz)
♦ -20 dBm (12.5 to 13.5 GHz)
▲ -30 dBm (12.5 to 13.5 GHz)
● -40 dBm (12.5 to 13.5 GHz)
10
1
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
Test port power (dBm)
20
-80
-90 -100 -110 -120
Table 8. Dynamic Accuracy (continued)
Dynamic Accuracy 13.5 to 20 GHz
Magnitude
N5230C Option 240/245/246
10
▄ -10 dBm (13.5 to 20 GHz)
Accuracy (dB)
♦ -20 dBm (13.5 to 20 GHz)
▲ -30 dBm (13.5 to 20 GHz)
● -40 dBm (13.5 to 20 GHz)
1
0.1
0.01
10
0
-10
-20
-30
-40
-50
-60
-70
-80
-90 -100 -110 -120
-80
-90 -100 -110 -120
Test port power (dBm)
Phase
N5230C Option 240/245/246
100
Accuracy (degrees)
▄ -10 dBm (13.5 to 20 GHz)
♦ -20 dBm (13.5 to 20 GHz)
▲ -30 dBm (13.5 to 20 GHz)
● -40 dBm (13.5 to 20 GHz)
10
1
0.1
10
0
-10
-20
-30
-40
-50
-60
-70
Test port power (dBm)
21
Table 9. Test port input (group delay)
Description
Aperture (selectable)
Maximum aperture
Range
Maximum delay
Supplemental information (typ.)
(frequency span)/(number of points –1)
20% of frequency span
0.5 x (1/minimum aperture)
Limited to measuring no more than 180° of phase change within the
minimum aperture
See graph below. (Char.)
Accuracy
The following graph shows characteristic group delay accuracy with full 2-port calibration and
a 10 Hz IF bandwidth. Insertion loss is assumed to be < 2 dB and electrical length to be ten meters.
For any Sij group delay measurement, Sii = 0, Sij = 0, Skl = 0 for all kl ≠ ij
Group delay (typical)
N5230C Option 140/145/146/240/245/246 with 85052B
100
Accuracy (nsec)
10
1
0.1
Frequency = 1 GHz
S11 = 0; S21 = 1; S12 = 0; S22 = 0
0.01
IF bandwidth = 10 Hz; Average factor = 1
Cal power = -5 dBm; Meas power = -5 dBm; Electrical length = 10 m
0.001
0.01
0.1
1
10
100
Aperture (MHz)
In general, the following formula can be used to determine the accuracy, in seconds, of specific
group delay measurement:
±Phase Accuracy (deg)/[360° Aperture (Hz)]
Depending on the aperture and device length, the phase accuracy used is either incremental
phase accuracy or worst case phase accuracy.
22
General Information
Table 10. Miscellaneous information
Description
System IF bandwidth range
CPU
Supplemental information
1 Hz to 600 kHz, nominal
Intel® 1.1 GHz Pentium® M with 1 GByte RAM
Table 11. Front panel information
Description
RF connectors
Type
Center pin recession
Display
Size
Refresh rate
Display range
Magnitude
Phase
Polar
Display resolution
Magnitude
Phase
Marker resolution
Magnitude
Phase
Polar
Supplemental information
Options 140, 145, 146, 240, 245, 246: 3.5 mm (male), 50 ohm (nominal)
0.002 in. (characteristic)
21.3 cm (8.4 in) diagonal color active matrix LCD;
1024 (horizontal) X 768 (vertical) resolution
A display is considered faulty if:
• A complete row or column of "stuck" or "dark" pixels.
• More than six "stuck on" pixels (but not more than three green) or
more than 0.002% of the total pixels are within the LCD specifications.
• More than twelve "dark" pixels (but no more than seven of the same
color) or more than 0.004% of the total pixels are within the LCD
specifications.
• Two or more consecutive "stuck on" pixels or three or more consecutive
"dark" pixel (but no more than one set of two consecutive dark pixels)
"Stuck on" of "dark" pixels less than 6.5 mm apart (excluding
consecutive pixels)
60 Hz
±500 dB (at 20 dB/div), max
±500°, max
10 pUnits, min
1000 Units, max
0.001 dB/div, min
0.01°/div, min
0.001 dB, min
0.01°, min
0.01 mUnit, min; 0.01°,min
23
Table 12. Rear panel information
Description
Trigger inputs/outputs
10 MHz Reference in
Connector
Input frequency
Input level
Input impedance
10 MHz Reference out
Connector
Output frequency
Signal type
Output level
Output impedance
Harmonics
VGA Video output
Connector
Test set IO
Supplemental information
BNC (f), TTL/CMOS compatible
BNC, female
10 MHz ± 10 ppm, typical
–15 to +20 dBm, typical
200 Ω, nom.
BNC, female
10 MHz ± 1 ppm, typical
Sine Wave, typical
+10 dBm ± 4 dB into 50 Ω, typical
50 Ω, nominal
< –40 dBc, typical
15-pin mini D-Sub; Drives VGA compatible monitors
25-pin D-Sub connector, female, available for external test set control
Aux IO
25-pin D-Sub connector, male, analog and digital I/O
Handler IO
36-pin parallel I/O port; all input/output signals are default set to
negative logic; can be reset to positive logic via GPIB command
GPIB
Two ports: dedicated controller and dedicated talker/listener
24-pin D-sub (Type D-24), female; compatible with IEEE-488.
USB Port
1 port on front panel and 4 ports on rear panel.
LAN
10/100BaseT Ethernet, 8-pin configuration; auto selects between the
two data rates
Line power (single phase)
Frequency, voltage
Maximum
50/60 Hz/400 Hz for 100-120 V, 50/60 Hz for 220-240 V, power supply
is auto switching
350 Watts
Note: Option H08 and Option H11 specifications are not provided in this N5230C specifications document.
24
Table 13. Analyzer environment and dimensions
Description
General environmental
EMC
Safety
Operating environment
Temperature
Error-corrected temperature range
Relative humidity
Altitude
Non-operating storage environment
Temperature
Cabinet dimensions
Supplemental information
Complies with European EMC directive 2004/108/EC
• IEC/EN 61326-1:2005
• CISPR Pub 11 Group 1, class A
• AS/NZS CISPR11:2004
• ICES/NMB-001
Complies with European Low Voltage Directive 2006/95/EC
• IEC/EN 61010-1:2001
• Canada: CSA C22.2 No. 61010-1:2004
• USA: UL 61010-1:2004
0 to +40 °C
Instrument powers up and displays no error messages within this
temperature range (except for "source unleveled" error message that
may occur at temperatures outside the specified performance
temperature range of 25 ± 5 °C).
23 °C ± 3 °C with less than 1 °C deviation from calibration temp.
Type tested 0 to 95% at 40 °C, non-condensing
0 to 4600 m (15,000 ft.)
–40 to +70 °C
Excluding front and rear panel hardware and feet
As shipped - includes front panel connectors,
rear panel bumpers, and feet.
As shipped plus handles
As shipped plus rack-mount flanges
As shipped plus handles and rack-mount flanges
Weight
Net
N5230C
Shipping
N5230C
Height
267 mm
10.5 in
280 mm
11 in
280 mm
11 in
280 mm
11 in
280 mm
11 in
Width
426 mm
16.75 in
435 mm
17.10 in
458 mm
18 in
483 mm
19 in
483 mm
19 in
Depth
427 mm
16.8 in
470 mm
18.5 in
501 mm
19.7 in
470 mm
18.5 in
501 mm
19.7 in
24.9 kg (55 lb), nominal
36.3 kg (80 lb), nominal
25
Measurement Throughput Summary
Table 14. Typical cycle time1 (ms) for measurement completion
Number of Points
201
401
801
1601
Start 8 GHz, stop 18 GHz, 600 kHz IF bandwidth
Uncorrected
21.148 21.743
23.01
25.198
4-Port cal
74.597 82.296
296.5
307.75
Start 300 kHz, stop 10 GHz, 600 kHz IF bandwidth
Uncorrected
19.814 22.801 24.973
29.01
4-Port cal
69.752 85.111 100.125 129.347
Start 300 kHz, stop 20 GHz, 600 kHz IF bandwidth
Uncorrected
32.575
34.7
39.237 43.155
4-Port cal
121.254 133.626 157.506 179.223
Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth
Uncorrected
38.083 51.816 55.488
56.36
4-Port cal
143.271 201.814 215.056 230.133
Start 300 kHz, stop 10 GHz, 100 kHz IF bandwidth
Uncorrected
37.03
42.532 45.122 46.729
4-Port cal
137.431 162.37 194.13 192.182
Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth
Uncorrected
44.98
69.408 87.161 92.475
4-Port cal
169.041 268.877 343.898 369.526
Start 8 GHz, stop 18 GHz, 50 kHz IF bandwidth
Uncorrected
42.171
70.09
88.702 90.981
4-Port cal
157.107 271.791 351.517 368.02
Start 300 kHz, stop 10 GHz, 50 kHz IF bandwidth
Uncorrected
43.713
61.41
66.878 69.373
4-Port cal
163.58 238.267 259.687 279.816
Start 300 kHz, stop 20 GHz, 50 kHz IF bandwidth
Uncorrected
48.673 80.798 124.605 147.303
4-Port cal
184.429 313.392 493.142 587.548
16,001
54.836
538.646
67.733
480.711
69.625
487.779
184.154
934.161
198.683
906.768
198.792
914.963
371.611
1532.609
385.04
1580.761
388.46
1587.839
1. Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with
DISPLAY:ENABLE OFF. Add 21 ms for display on. Data for one trace (S11) measurement.
26
Table 15. Cycle Time vs IF Bandwidth
Applies to the preset condition (201 points, correction off) except for the following changes:
• CF = 10 GHz
• Span = 100 MHz
• Display off (add 21 ms for display on)
Description
IF Bandwidth (Hz)
600,000
360,000
280,000
200,000
150,000
100,000
70,000
50,000
30,000
20,000
15,000
10,000
7000
5000
3000
2000
1500
1000
700
500
300
200
100
30
10
1
Typical performance
Trace noise (dB rms)
Cycle time (ms)1
3.13
0.00544
3.21
0.00602
3.17
0.00321
3.17
0.00259
3.19
0.00207
4.05
0.00155
4.99
0.00144
6.41
0.00121
8.78
0.00094
12.07
0.00080
14.91
0.00069
26.02
0.00052
34.54
0.00047
45.87
0.00044
69.91
0.00032
99.69
0.00029
128.18
0.00010
215.62
0.00009
291.58
0.00006
397.36
0.00007
633.86
0.00000
930.15
0.00000
1824.19
0.00000
6001.70
0.00000
17899.79
0.00000
178391.58
0.00000
1. Cycle time includes sweep and retrace time.
27
Table 16. Cycle time vs number of points
Applies to the preset condition (correction off) except for the following changes:
• CF = 10 GHz
• Span = 100 MHz
• Display off (add 21 ms for display on)
IF Bandwidth (Hz)
30,000
100,000
600,000
Number of points
3
11
51
101
201
401
801
1,601
6,401
16,001
3
11
51
101
201
401
801
1,601
6,401
16,001
3
11
51
101
201
401
801
1,601
6,401
16,001
Cycle time (ms)1
2.88
3.50
3.91
5.29
8.75
15.66
29.46
57.73
221.04
549.71
2.87
2.82
2.86
2.96
4.02
6.23
10.65
19.49
70.96
173.78
2.84
2.84
2.87
3.03
3.14
3.51
4.22
6.22
19.35
45.12
1. Cycle time includes sweep and retrace time.
Table 17. Data transfer time (ms)
Number of points
201
401
1601
SCPI over GPIB (program executed on external PC)
32-bit floating point
6
10
33
64-bit floating point
10
18
65
ASCII
53
105
406
CPI over SIU/LAN or TCP/IP socket (program executed in the analyzer)
32-bit floating point
1
2
2.5
64-bit floating point
2
2
3
ASCII
11
20
73
COM (program executed in the analyzer)
32-bit floating point
< 0.2
0.2
0.3
Variant type
0.6
1
3.2
DCOM over LAN (program executed on external PC)
32-bit floating point
< 0.8
1
1.6
Variant type
1.9
3
8.9
16,001
313
622
4032
7
10
720
0.9
32
7.5
82
Note: Specifications for recall and sweep speed are not provided for the N5230C analyzers.
28
Specifications: Front-Panel Jumpers
Table 18: Measurement receiver inputs (rcvr A In, rcvr B In, rcvr C In, rcvr D In)
0.1 dB Typical compression
Description
Maximum input level
300 kHz to 10 MHz
10 MHz to 1 GHz
1 to 12.5 GHz
12.5 to 13.5 GHz
13.5 to 20 GHz
Damage level
N5230C
Maximum DC level
N5230C
Typical
Options 145, 146, 245, 246
–11 dBm
–7 dBm
–6 dBm
–7 dBm
–7 dBm
+15 dBm
±16 V
Table 19: Reference receiver input (rcvr in) at maximum specified output power
Description
Maximum input level
300 kHz to 10 MHz
10 to 45 MHz
45 to 500 MHz
500 MHz to 4 GHz
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
Damage level
N5230C
Maximum DC level
N5230C
Typical
Options 145, 146, 245, 246
–15 dBm
–15 dBm
–15 dBm
–15 dBm
–16 dBm
–20 dBm
–21 dBm
–21 dBm
–27 dBm
+15 dBm
±16 V
Table 20: Reference output (source out) at maximum specified output power
Description
Maximum output level
300 kHz to 10 MHz
10 to 45 MHz
45 to 500 MHz
500 MHz to 4 GHz
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
Damage level
N5230C
Maximum DC level
N5230C
Typical
Options 145, 146, 245, 246
–15 dBm
–15 dBm
–15 dBm
–15 dBm
–15 dBm
–20 dBm
–21 dBm
–21 dBm
–27 dBm
+27 dBm
±16 V
29
Table 21: Source outputs (port 1 source out, port 2 source out, port 3 source out, port 4 source out)
at maximum specified output power
Description
Maximum output level
300 kHz to 10 MHz
10 to 45 MHz
45 to 500 MHz
500 MHz to 4 GHz
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
Damage level
N5230C
Maximum DC level
N5230C
Typical
Options 145, 146, 245, 246
+10 dBm
+10 dBm
+10 dBm
+10 dBm
+9 dBm
+4 dBm
+1 dBm
+1 dBm
–4 dBm
+27 dBm
±16 V
Table 22: Coupler inputs (port 1 cplr thru, port 2 cplr thru, port 3 cplr thru, port 4 cplr thru)
Insertion loss of coupler thru
Description
Insertion loss to test port
300 kHz to 10 MHz
10 to 45 MHz
45 to 500 MHz
500 MHz to 4 GHz
4 to 6 GHz
6 to 10.5 GHz
10.5 to 13.5 GHz
13.5 to 15 GHz
15 to 20 GHz
Damage level
N5230C
Maximum DC level
N5230C
Typical
Options 145, 146, 245, 246
1.5 dB
1.5 dB
1.5 dB
2.0 dB
2.5 dB
2.5 dB
3.0 dB
3.0 dB
3.0 dB
+27 dBm
±16 V
Table 23: Coupler outputs (port 1 cplr arm, port 2 cplr arm, port 3 cplr arm, port 4 cplr arm)
Description
Damage level
N5230C
Maximum DC level
N5230C
30
Typical
Options 145, 146, 245, 246
+15 dBm
0V
Test Set Block Diagrams
N5230C Option 140 or 240
standard test set and
standard power range
PORT 4
D
PORT 3
Switch
C
Source
B
Reference
receiver
PORT 2
R
A
PORT 1
N5230C Option 145 or 245
configurable test set,
extended power range
Source out
Coupler thru
PORT 4
Coupler arm
Receiver D in
D
Source out
Coupler thru
PORT 3
Switch
Source
Coupler arm
Receiver C in
C
60 dB
B
Reference
receiver
Receiver B in
Coupler arm
PORT 2
Coupler thru
Source out
R
A
Receiver A in
Coupler arm
PORT 1
Coupler thru
Source out
Source out
Receiver R in
N5230C Option 146 or 246
configurable test set,
extended power range,
and internal second source
Source 2
Source 1
60 dB
60 dB
R
A
Test port 1
B
Test port 2
C
D
Test port 3
Test port 4
31
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the products
and applications you select.
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
Remove all doubt
www.agilent.com
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised.
You will get full value out of your Agilent
equipment throughout its lifetime. Your
equipment will be serviced by Agilenttrained technicians using the latest
factory calibration procedures, automated
repair diagnostics and genuine parts. You
will always have the utmost confidence
in your measurements.
Agilent offers a wide range of additional
expert test and measurement services
for your equipment, including initial
start-up assistance, onsite education
and training, as well as design, system
integration, and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
For more information on Agilent
Technologies’ products, applications
or services, please contact your local
Agilent office. The complete list is
available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
07031 464 6333
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: July 2, 2009
Product specifications and descriptions
in this document subject to change
without notice.
Intel® and Pentium® are US registered trademarks of
Intel Corporation.
© Agilent Technologies, Inc. 2008, 2009
Printed in USA, September 1, 2009