AGILENT 87075C-012

Agilent 87075C
75-Ohm Multiport Test Sets
Data Sheet
Agilent 87075C
3 MHz to 1.3 GHz
This document describes the performance
and features of Agilent Technologies 87075C
75-Ohm multiport test sets, both as standalone
units and when combined with Agilent 75-Ohm
8712ET, 8712ES, 8714ET, or 8714ES network
analyzers. The following options are available:
• Option 006 (6 ports)
• Option 012 (12 ports)
For more information about these test sets,
please read the following documents:
• Agilent 87075C brochure,
literature number 5968-4766E
• Agilent 87075C Configuration Guide,
literature number 5968-4768E
Introduction
All specifications and characteristics apply over a
25 °C ± 5 °C range (unless otherwise stated) and
30 minutes after the instrument has been turned on.
Definitions
Specification: Warranted performance. Specifications
include guardbands to account for the expected
statistical distribution, measurement uncertainties,
and changes in performance due to environmental
conditions.
Characteristic: A performance parameter that the
product is expected to meet before it leaves the
factory, but is not verified in the field and is not
covered by the product warranty. A characteristic
includes the same guardbands as a specification.
Typical: Expected performance of an average unit.
A typical does not include guardbands. It is not
covered by the product warranty.
Nominal: A general, descriptive term that does not
imply a level of performance. It is not covered by
the product warranty.
Supplemental information: May include typical, nominal, or characteristic values.
Uncorrected (raw) performance: Indicates instrument
performance without error correction. The uncorrected (raw) performance affects the stability of a
calibration.
System performance: Performance of a complete
multiport test system, which includes an 87075C
test set and a 75-Ohm 8712ET/ES or 8714ET/ES
network analyzer.
Test Set Cal: The calibration of a multiport test system, requiring the connection of known calibration
standards to all of the ports that will be used for
measurements.
SelfCal: An automated system calibration that uses
calibration standards internal to the test set and
the most recent Test Set Cal data to calibrate the
test system.
Environmental specifications: Environmental specifications bound the external conditions for which
the specifications are valid. The environmental
specifications also bound the external conditions
the test set may be subject to without permanently
affecting performance or causing physical damage.
Table of contents
Calibration: The process of measuring known standards from a calibration kit to characterize the systematic (repeatable) errors of a network analyzer.
Corrected (residual) performance: Indicates performance after error correction (calibration). It is
determined by the quality of calibration standards
and how well “known” they are, plus system
repeatability, stability, and noise.
2
System performance, two-port calibration
System performance, T/R calibration
System performance, uncorrected
System performance, general
Test set input/output performance
Test set general information
Physical dimensions
Block diagrams
System features
3
5
8
9
12
13
14
15
16
System performance
Two-port calibration (user)
Agilent 85036B/E Type-N calibration kit
Specification1 (dB)
Parameter
3 MHz to 1.3 GHz
Directivity
47
Source match
37
Load Match
47
Reflection tracking
±0.1
Transmission tracking
±0.1
Transmission uncertainty (typical)2
Phase
Magnitude
Reflection uncertainty (typical)2
Magnitude
Phase
1.
These specifications apply under the following conditions: measurement uses the
“Fine” (15 Hz) bandwidth, no averaging, and isolation cal “on”; Test Set Cals use
the “Fine” (15 Hz) bandwidth, 16 averages, and assume an isolation calibration
has been performed; the test set must be used with a 75-Ohm 8712ES or 8714ES
network analyzer with firmware revision E.06.00 or later; the test set and the analyzer must have had their performance verified within the last year; both instruments must have warmed up for at least 30 minutes after turn-on; measurements
are made at an environmental temperature of 25 °C ± 5 °C and within ± 1 °C of
the last valid Test Set Cal.
2. These uncertainty curves only include the effects of the test port(s) within the
measurement path. The effect of the uncorrected match of test ports outside
the measurement path is ignored, and is dependent on the isolation between the
ports of the DUT that are within the measurement path and ports of the DUT that
are outside the measurement path.
3
System performance
Two-port calibration (user)
Agilent 85039B Type-F calibration kit
Specification1,2 (dB)
Parameter
3 MHz to 1.3 GHz
Directivity
35
Source match
28
Load Match
35
Reflection tracking
±0.1
Transmission tracking
±0.2
Transmission uncertainty (typical)3
Magnitude
Phase
Reflection uncertainty (typical)3
Magnitude
1. These specifications apply under the following conditions: measurement uses the
“Fine” (15 Hz) bandwidth, no averaging, and isolation cal “on”; Test Set Cals use
the “Fine” (15 Hz) bandwidth, 16 averages, and assume an isolation calibration
has been performed; the test set must be used with a 75-Ohm 8712ES or 8714ES
network analyzer with firmware revision E.06.00 or later; the test set and the
analyzer must have had their performance verified within the last year; both
instruments must have warmed up for at least 30 minutes after turn-on; measurements are made at an environmental temperature of 25 °C ± 5 °C and within
± 1 °C of the last valid Test Set Cal.
4
Phase
2. Measurements made using a DUT with center pins of the Type-F connectors
meeting the 0.77 to 0.86 mm limits.
3. These uncertainty curves only include the effects of the test port(s) within the
measurement path. The effect of the uncorrected match of test ports outside
the measurement path is ignored, and is dependent on the isolation between
the ports of the DUT that are within the measurement path and ports of the
DUT that are outside the measurement path.
System performance
Transmission/reflection (T/R) calibration (user)
Agilent 85036B/E Type-N calibration kit
Specification1 (dB)
Typical2 (dB)
Parameter
3 MHz to 1.3 GHz
3 MHz to 1.3 GHz
Directivity
Source match
Load match (reflection calibration)
Load match (transmission calibration)
Reflection tracking
Transmission tracking
40
35
20
15
±0.1
±0.1
—
—
25
20
—
—
1. These specifications are valid for reflection (one-port) and transmission
(enhanced-response) Test Set Cals, and apply under the following conditions:
measurement uses the “Fine” (15 Hz) bandwidth, no averaging, and isolation cal
“on”; Test Set Cals use the “Fine” (15 Hz) bandwidth, 16 averages, and assume an
isolation calibration has been performed; the test set must be used with a
75-Ohm 8712ET, 8712ES, 8714ET, or 8714ES network analyzer with firmware revision E.06.00 or later; the test set and the analyzer must have had their performance verified within the last year; both instruments must have warmed up for at
least 30 minutes after turn-on; measurements are made at an environmental
temperature of 25 °C ± 5 °C and within ± 1 °C of the last valid Test Set Cal.
2. Corrected system performance is changed to typical when the current temperature has drifted beyond ± 1 °C of the last valid Test Set Cal. Typicals are valid only
when the current temperature is within 25 °C ± 5 °C, and within ± 1 °C of the
most recent SelfCal.
5
System performance
Transmission/reflection (T/R) calibration (user)
Agilent 85036B/E Type-N calibration kit (continued)
Transmission uncertainty, enhanced-response calibration (typical)1
Magnitude, ES models
Magnitude, ET models with attenuator
Phase, ET models with attenuator
Magnitude, ET models without attenuator
Phase, ET models without attenuator
1. These uncertainty curves only include the effects of the test port(s) within the
measurement path. The effect of the uncorrected match of test ports outside the
measurement path is ignored, and is dependent on the isolation between the
ports of the DUT that are within the measurement path and ports of the DUT that
are outside the measurement path.
6
Phase, ES models
System performance
Transmission/reflection (T/R) calibration (user)
Agilent 85036B/E Type-N calibration kit (continued)
Reflection uncertainty, one-port calibration (typical)1
Magnitude, ES models
Phase, ES models
Magnitude, ET models with attenuator
Phase, ET models with attenuator
Magnitude, ET models without attenuator
Phase, ET models without attenuator
1. These uncertainty curves only include the effects of the test port(s) within the
measurement path. The effect of the uncorrected match of test ports outside the
measurement path is ignored, and is dependent on the isolation between the
ports of the DUT that are within the measurement path and ports of the DUT that
are outside the measurement path.
7
System performance
Uncorrected
Type-N
Parameter
Source match, ratioed
Load match, test port unselected 2
Load match, test port selected 3
Transmission tracking 4
Reflection tracking 4
Crosstalk 5, 8712ET/8714ET
Crosstalk 5, 8712ES/8714ES
Specification1 (dB)
Typical (dB)
3 MHz to 1.3 GHz
10
20
13
—
—
—
—
3 MHz to 1.3 GHz
15
25
18
±3.0
±3.0
72
72
1. These specifications apply under the following conditions: measurement uses the
“Fine” (15 Hz) bandwidth with narrowband detection and no averaging, and isolation cal “on”; Test Set Cals use the “Fine” (15 Hz) bandwidth, 16 averages, and
assume an isolation calibration has been performed; the test set must be used
with a 75-Ohm 8712ET, 8712ES, 8714ET or 8714ES network analyzer with firmware
revision E.06.00 or later; the test set and the analyzer must have had their performance verified within the last year; both instruments must have warmed up for at
least 30 minutes after turn-on; measurements are made at an environmental temperature of 25 °C ± 5 °C and within ± 1 °C of the last valid Test Set Cal.
2. This is the match of any test port that is unselected (not in the measurement
path). If the network analyzer is performing a reflection measurement with oneport calibration, then only one port on the test set is selected (the source port).
If the network analyzer is performing a transmission measurement or a reflection
measurement with two-port calibration, then only two ports on the test set are
selected (the source and load ports).
8
3. This is the match of the test set port that has been selected as the load port. The
network analyzer must be making a transmission measurement or a reflection
measurement with two-port calibration for a test port to be selected as the
load port.
4. The uncorrected tracking terms are defined as the deviation over the defined
frequency band, ignoring offset loss.
5. Crosstalk is computed by normalizing the result of an isolation measurement to a
through measurement between the two ports such that the path losses are taken
into account. Isolation is defined as the transmission signal measured between
any two ports of the test system when those two ports are terminated with
shorts.
System performance
General
Specification
Typical
System bandwidth
Network analyzer
Fine
15 Hz
Fine
15 Hz
Med Wide
4000 Hz
Wide
6500 Hz
3 MHz
to
1.3 GHz
3 MHz
to
1.3 GHz
3 MHz
to
1.3 GHz
3 MHz
to
1.3 GHz
System noise floor 1 (dBm)
8712ET/8714ET
8712ES/8714ES
–96
–87
8712ET, no attenuator
8712ET, with attenuator
8712ES
8714ET, no attenuator
8714ET, with attenuator
8714ES
65
63
61
65
60
58
–104
–79
–99
–75
System dynamic range 2 (dB)
82
81
79
79
78
76
82
81
75
79
78
72
–40
–42
44
43
42
41
40
39
1. Noise floor is defined as the rms value of the trace (in linear format) for a transmission measurement in CW mode, using the “Fine” bandwidth (15 Hz), the test
ports terminated in loads, 0 dBm at the test set source port, and no averaging.
This measurement ignores the effects of crosstalk. This is a system specification—the test set increases the network analyzer noise floor by adding loss to
the network analyzer measurement.
2. The system dynamic range is calculated as the difference between the receiver
minimum input (noise floor plus calibrated crosstalk) and the system’s maximum
output power. System dynamic range applies to transmission measurements only,
since reflection measurements are limited by directivity.
9
System performance
General (continued)
Specification
Supplemental information
System bandwidth
Network analyzer
Fine
15 Hz
Fine
15 Hz
Med Wide
4000 Hz
Wide
6500 Hz
Receiver dynamic range 1 (dB)
8712ET/8714ET
8712ES/8714ES
114
105
119
114
94
90
56
57
Dynamic accuracy (typical)2
Magnitude, ES models
Phase, ES models
Magnitude, ET models
Phase, ET models
1. The receiver dynamic range is calculated as the difference between the receiver
minimum input (noise floor) and the receiver maximum input. Receiver dynamic
range applies to transmission measurements only, since reflection measurements
are limited by directivity.
2. The reference power for dynamic accuracy is –20 dBm.
10
System performance
General (continued)
Network analyzer
Specification
Characteristic (dBm)
Output power (system maximum)
<1 GHz
>1 GHz
8712ET, no attenuator
8712ET, with attenuator
8712ES
8714ET, no attenuator
8714ET, with attenuator
8714ES
Parameter
Reflection calibration
—
—
—
—
—
—
7
6
4
2
1
0
3.5
2.5
0.5
0.5
–0.5
–2.5
Specification
Supplemental information
Port switching time1 (sec)
Transmission enhanced calibration
Two-port calibration
—
—
—
Reflection calibration
Transmission enhanced calibration
Two-port calibration
—
—
—
0.7
0.7
0.7
SelfCal time 2 (sec)
1 (typical)
3 (typical)
8 (typical)
1. This is the time required to switch to any new port configuration, and requires
that the Test Set Cal has been performed for 201 points, the new measurement
configuration is 201 points, and no new SelfCal occurs during switching.
2. This is the time required to perform the SelfCal for any single port configuration,
assuming that the Test Set Cal was performed for 201 points, the new measurement configuration is 201 points, and the measurement bandwidth is “Med Wide”
(4000 Hz). SelfCal times for other settings can be found in Table 4-1 of the Agilent
87050E/87075C User’s and Service Guide (87050-90026).
11
Test set input/output performance
Parameter
Frequency range
Maximum input power at 0.1 dB
compression 1
Input damage power
Specification
Supplemental
3 MHz to 1.3 GHz
—
RF input power
—
16 dBm (nominal)
—
20 dBm (characteristic)
Specification (dB)
Parameter
3 MHz to 1.3 GHz
Source match, test port2
12
Load match, test port unselected3
20
Load match, test port selected4
15
Interconnect match, reflection port5
12
Interconnect match, transmission port5
12
Insertion loss, reflection port to port-n6
7.5
Insertion loss, transmission port to port-n6
11.5
Tracking, reflection port to port-n6
—
Tracking, transmission port to port-n6
—
Crosstalk, uncalibrated, adjacent ports7
57
Crosstalk, uncalibrated, non-adjacent ports7
57
Typical (dB)
3 MHz to 1.3 GHz
16
25
20
18
18
6.5
10.5
1.5
1.5
72
72
1. Compression is defined for the test set, independent of the network analyzer.
2. This is the match of the test set port that has been selected as the source port.
The test set interconnect reflection port must be terminated with a load standard
from an 85036B/E calibration kit.
3. This is the match of any test port that is unselected (not in the measurement
path). Only the source port of the test set is selected when you make a reflection
measurement with one-port calibration. Only the source and load ports of the test
set are selected when you make a transmission measurement, or a reflection
measurement with two-port calibration.
4. This is the match of the test set port that has been selected as the load port.
A transmission or two-port measurement is required for a test port to be selected
as the load port. The test set interconnect transmission port must be terminated
with a load standard from an 85036B/E calibration kit.
12
5. This is the match of the test set interconnect ports (transmission and reflection
ports) with the test set in transmission, or two-port mode. The selected test set
test ports must be terminated with load standards from 85036B/E calibration kits.
6. The reflection and transmission ports of the test set are connected to the corresponding ports of the network analyzer. Port-n is any one of the test ports used to
connect to the device-under-test.
7. This is crosstalk of the test set measured between the test set’s interconnect
ports, with shorts on the selected test ports. Crosstalk is computed by normalizing the result of an isolation measurement to a through measurement between
the two ports such that the path losses are taken into account. Isolation is
defined as the transmission signal measured between any two ports of the test
system when these two ports are terminated with shorts.
Test set general information
Description
Specification
Supplemental information
Front panel ports
87075E, Option 006
87075E, Option 012
6
12
Type-N, female 75 Ohms (nominal)
Type-N, female 75 Ohms (nominal)
Real panel
Parallel in connector
Parallel out connector
Frequency
Input voltage, operating
Input power
—
—
Line power 1
50/60 Hz
100/120/220/240 Vac
—
25-pin D-subminiature female (DB-25)
25-pin D-subminiature female (DB-25)
—
—
<45 W (typical)
General environment
ESD
—
Dust
—
General
Operating Temperature
Altitude
Storage Temperature
Indoor use only
0 °C to + 55 °C
0 to 4.6 km (15,000 ft)
–40 °C to +70 °C
Cabinet dimensions
Height x width x depth
—
Minimize using static-safe work
procedures and an antistatic bench mat
(part number 9300-0797)
Minimize for optimum reliability
—
—
—
—
132.8 x 425 x 497 mm (nominal)
5.2 x 16.7 x 19.6 in (nominal)
Cabinet dimensions exclude front and
rear protrusions
Weight
Net
Shipping
—
—
8 kg (18 lb) (nominal)
11 kg (24 lb) (nominal)
1. A third-wire ground is required.
13
Physical dimensions
Physical dimensions of the Agilent 87075C
Option 006 multiport test set1
Physical dimensions of the Agilent 87075C
Option 012 multiport test set1
1. These dimensions exclude rear protrusions.
14
Block diagrams
Block diagram for the Agilent 8712ET and 8714ET
REAR PANEL
AUX Input
Block diagram for the Agilent 87075C
(only one test set port pair is shown)
External Detectors
Y
X
Y
Input B
RF
Source
ADC and Processor
Input B*
X
Incident
Reference
Input R*
Input A
Transmission
Input R
Reflected
CRT
FRONT PANEL
Reflection
(RF Out)
With Attenuator Option 1E1
Transmission
(RF In)
Device
Under
Test
Narrowband Detector
Broadband Detector
Block diagram for the Agilent 8712ES and 8714ES
REAR PANEL
AUX Input
External Detectors
Y
X
Y
Input B
RF
Source
Input B*
ADC and Processor
X
Input R
Reference
Input R*
Input A
CRT
FRONT PANEL
Narrowband Detector
Port 1
Port 2
Device
Under
Test
Broadband Detector
15
System features
Test set control
Control of the switches inside the test set and calibration of the test system can be accomplished
from the front panel of the network analyzer—an
external computer is not required. However, the
analyzers are fully programmable for use in automated test environments.
Measurement
Number of display measurements
Two measurement displays are available, with
independent control of display parameters including format type, scale per division, reference level,
reference position, and averaging. The displays can
share network analyzer sweep parameters, or, by
using alternate sweep, each measurement can have
independent sweep parameters including frequency
settings, IF bandwidth, power level, and number of
trace points. The instrument can display a single
measurement, or dual measurements on a split (two
graticules) or overlaid (one graticule) screen.
Measurement choices
• Narrowband
ET models: reflection (A/R), transmission (B/R),
A, B, R
ES models: S11 (A/R), S22 (B/R), S21 (B/R),
S12 (A/R), A, B, R
• Broadband
X, Y, Y/X, X/Y, Y/R*, power (B*, R*),
conversion loss (B*/R*)
Note: X and Y denote external broadband-detector
inputs; * denotes internal broadband detectors.
Formats
Log or linear magnitude, SWR, phase, group delay,
real and imaginary, Smith chart, polar, and impedance magnitude.
Trace functions
Current data, memory data, memory with current
data, division of data by memory.
Display annotations
Start/stop, center/span, or CW frequency, scale
per division, reference level, marker data, softkey
labels, warning and caution messages, screen titles,
time and date, and pass/fail indication.
Limits
Measurement data can be compared to any combination of line or point limits for pass/fail testing.
User-defined limits can also be applied to an amplitude- or frequency-reference marker. A limit16
test TTL output is available on the rear panel for
external control or indication. Limits are only
available with rectilinear formats.
Data markers
Each measurement channel has eight markers.
Markers are coupled between channels. Any one
of eight markers can be the reference marker for
delta-marker operation. Annotation for up to four
markers can be displayed at one time.
Marker functions
Markers can be used in absolute or delta modes.
Other marker functions include marker to center
frequency, marker to reference level, marker to
electrical delay, searches, tracking, and statistics.
Marker searches include marker to maximum,
marker to minimum, marker to target value, bandwidth, notch, multi-peak and multi-notch. The
marker-tracking function enables continuous update
of marker search values on each sweep. Marker
statistics enable measurement of the mean, peakto-peak, and standard deviation of the data
between two markers. For rapid tuning and testing of cable-TV broadband amplifiers, slope and
flatness functions are also available.
Storage
Internal memory
1.5 Mbytes (ET models) or 1 Mbyte (ES models) of
nonvolatile storage is available to store instrument
states, measurement data, screen images, and IBASIC
programs. Instrument states can include all control
settings, limit lines, memory data, calibration coefficients, and custom display titles. If no other data
files are saved in nonvolatile memory, between
approximately 20 and 150 instrument states can
be saved (depending on the model type and on
instrument parameters). Approximately 14 Mbytes
of volatile memory are also available for temporary storage of instrument states, measurement
data, screen images, and IBASIC programs.
Disk drive
Trace data, instrument states (including calibration data), and IBASIC programs can be saved on
floppy disks using the built-in 3.5-inch disk drive.
All files are stored in MS-DOS®-compatible format.
Instrument data can be saved in binary or ASCII
format (including Touchstone/.s2p format), and
screen graphics can be saved as PCX (bit-mapped),
HP-GL (vector), or PCL5 (printer) files.
NFS: See description under Control via LAN.
System features (continued)
Data hard copy
Hard copy prints can be made using PCL and PCL5
printers (such as HP DeskJet or LaserJet series
printers), or Epson-compatible graphics printers.
Single color and multicolor formats are supported.
Hard copy plots can be automatically produced
with HP-GL-compatible plotters such as the Agilent
7475A, or with printers that support HP-GL. The
analyzer provides Centronics (parallel), RS-232C,
GPIB, and LAN interfaces.
Automation
Controlling via GPIB
Interface: The GPIB interface operates to IEEE
488.2 and SCPI standard-interface commands.
Control: The analyzer can either be the system
controller, or pass bus control to another active
controller.
Data transfer formats:
• ASCII
• 32- or 64-bit IEEE 754 floating-point format
• Mass-memory-transfer commands allow file
transfer between external controller and
analyzer.
Control via LAN
The built-in LAN interface and firmware support
data transfer and control via direct connection to a
10 Base-T (Ethertwist) network. A variety of standard protocols are supported, including TCP/IP,
sockets, ftp, http, telnet, bootp, and NFS. The LAN
interface is standard.
SCPI: The analyzer can be controlled by sending
Standard Commands for Programmable Instruments (SCPI) within a telnet session or via a socket
connection and TCP/IP (the default socket port is
5025). The analyzer’s socket applications programming interface (API) is compatible with Berkeley
sockets, Winsock, and other standard socket APIs.
Socket programming can be done in a variety of
environments including C programs, Agilent VEE,
SICL/LAN, or a Java™ applet. A standard web
browser and the analyzer’s built-in web page can
be used to remotely enter SCPI commands via a
Java applet.
FTP: Instrument state and data files can be transferred via ftp (file-transfer protocol). An internal,
dynamic-data disk provides direct access to instru-
ment states, screen dumps, trace data, and operating parameters.
HTTP: The instrument’s built-in web page can be
accessed with any standard web browser using
http (hypertext transfer protocol) and the network
analyzer’s IP address. The built-in web page can be
used to control the network analyzer, view screen
images, download documentation, and link to other
sites for firmware upgrades and VXIplug&play
drivers. Some word processor and spreadsheet programs, such as Microsoft® Word 97 and Excel 97,
provide methods to directly import graphics and
data via a LAN connection using http and the network analyzer’s IP address.
SICL/LAN: The analyzer’s support for SICL (standard
instrument control library) over the LAN provides
control of the network analyzer using a variety of
computing platforms, I/O interfaces, and operating
systems. With SICL/LAN, the analyzer is controlled
remotely over the LAN with the same methods used
for a local analyzer connected directly to the computer via a GPIB interface. SICL/LAN protocol
also allows the use of Agiltent’s free VXIplug&play
driver to communicate with the multiport test
system over a LAN. SICL/LAN can be used with
Windows® 95/98/NT, or HP-UX.
NFS: The analyzer’s built-in NFS (network file system) client provides access to remote files and
directories using the LAN. With NFS, remote files
and directories (stored remotely on a computer)
behave like local files and directories (stored locally
within the analyzer). Test data taken by the network analyzer can be saved directly to a remote
PC or UNIX® directory, eliminating the need for a
remotely initiated ftp session. For Windows-based
applications, third-party NFS-server software must
be installed on the PC. NFS is fully supported in
most versions of UNIX.
Bootp: Bootstrap protocol (bootp) allows a network
analyzer to automatically configure itself at poweron with the necessary information to operate on
the network. After a bootp request is sent by the
analyzer, the host server downloads an IP and gateway address, and a subnet mask. In addition, the
analyzer can request an IBASIC file, which automatically executes after the transfer is complete.
For Windows-based applications, third-party
bootp-server software must be installed on the PC.
Bootp is fully supported in most versions of UNIX.
17
System features (continued)
Programming with IBASIC
As a standard feature, all Agilent 8712ET/ES
and 8714ET/ES network analyzers come with the
Instrument BASIC programming language
(IBASIC). IBASIC facilitates automated measurements and control of other test equipment, improving productivity. For simpler applications, you can
use IBASIC as a keystroke recorder to easily automate manual measurements. Or, you can use an
optional, standard PC keyboard to write custom
test applications that include:
•
•
•
•
•
Special softkey labels
Tailored user prompts
Graphical setup diagrams
Barcode-reading capability
Control of other test instruments via the GPIB,
serial, or parallel interfaces
Measurement calibration
Measurement calibration is a process that improves
measurement accuracy by using error-correction
arrays to remove systematic measurement errors.
The Test Set Cal and SelfCal features on your multiport test system increase the accuracy of your
measurements and significantly increase the test
efficiency of your work stations by eliminating
frequent and lengthy calibration procedures.
Test Set Cal
A Test Set Cal is a calibration that should be performed on a regular but relatively infrequent basis
(at least once a month is recommended). A Test Set
Cal requires connection of mechanical calibration
standards to all of the ports you will be using for
your measurements.
18
Test Set Cal for the 8712ET and 8714ET analyzers
The data collected by the analyzer during a
Test Set Cal always includes both transmission
(enhanced response) calibration data and reflection (one-port) calibration data. When making measurements after calibration, the analyzer automatically recalls and uses the correct set of calibration
data for the type of measurement chosen.
Test Set Cal for the 8712ES and 8714ES analyzers
The S-parameter network analyzers perform either
two-port calibration or enhanced response/oneport calibration. Choosing a two-port calibration
for the Test Set Cal removes the most systematic
errors, giving you the greatest measurement accuracy. Choosing the enhanced response/one-port
calibration allows faster measurement speeds, but
is not as accurate as full two-port calibration.
SelfCal
A SelfCal is an internal system calibration that
automatically executes in just a few seconds on
a regular, frequent basis (once per hour is recommended). A SelfCal does not require that you
remove your DUT or that you make any connections
of external calibration standards. The SelfCal uses
the results of the most recent Test Set Cal, along
with current measurements of internal, electronically switched, open, short, load, and through
standards. SelfCal removes the drift of the network
analyzer and multiport test set due to environmental variations.
System features (continued)
Other calibrations
Besides using a Test Set Cal, individual instrument
states and their corresponding calibrations can be
saved and recalled for use with specific measurement paths. For example, to improve measurement
throughput, one signal path could be tested using a
response calibration, while all other paths are tested
with a Test Set Cal using two-port calibration.
Note: the SelfCal feature is only supported with
Test Set Cals.
ET and ES models: reflection measurements
• One-port calibration
Compensates for frequency response, directivity,
and source-match errors. Requires short, open,
and load standards.
A variety of calibration types are available and
described below:
• Type-N 75-Ohm (Agilent 85036B/E)
• Type-F (Agilent 85039B)
ES models only
• Two-port calibration
Compensates for frequency response, source
and load match, and directivity errors while
making S-parameter measurements of transmission (S21, S12) and reflection (S11, S22).
Compensates for transmission crosstalk when
the Isolation on OFF softkey is toggled to ON.
Requires short, open, load, and through standards.
In addition, you can also describe the standards
for a user-defined calibration kit (for example,
open-circuit capacitance coefficients, offset-short
length, or through-standard loss).
Calibration kits
Data for several standard calibration kits are
stored in the instrument for use by the calibration
routines. They include:
For more information about calibration kits available from Agilent, consult the 87075C Configuration
Guide, literature number 5968-4768E.
Key network analyzer options
ET and ES models: transmission measurements
• Normalization
Provides simultaneous magnitude and phase
correction of transmission frequency-response
errors. Requires a through connection. Used for
both narrowband and broadband detection.
Does not support calibration interpolation.
• Response
Simultaneous magnitude and phase correction
of frequency response errors for transmission
measurements. Requires a through standard.
• Response and isolation
Compensates for frequency response and crosstalk errors. Requires a load termination
on both test ports and a through standard.
• Enhanced response
Compensates for frequency response and source
match errors. Requires short, open, load, and
through standards.
75 Ohms (Option 1EC)
Provides 75-Ohm system impedance.
Step attenuator (Option 1E1)
Adds a built-in 60 dB step attenuator to transmission/reflection (ET) models to extend the output
power range to –60 dBm. The attenuator is standard in S-parameter (ES) models.
19
www.agilent.com
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
JavaTM is a U.S. trademark of Sun Microsystems, Inc.
Microsoft® is a U.S. registered trademark of
Microsoft Corporation.
MS-DOS® is a U.S. registered trademark of Microsoft
Corporation.
Windows NT® is a U.S. registered trademark of
Microsoft Corporation.
Windows® and MS Windows® are U.S. registered
trademarks of Microsoft Corporation.
UNIX® is a registered trademark of the Open Group.
Remove all doubt
Our repair and calibration services will get
your equipment back to you, performing
like new, when promised. You will get
full value out of your Agilent equipment
throughout its lifetime. Your equipment
will be serviced by Agilent-trained technicians using the latest factory calibration
procedures, automated repair diagnostics
and genuine parts. You will always have the
utmost confidence in your measurements.
Agilent offers a wide range of additional
expert test and measurement services for
your equipment, including initial start-up
assistance on site education and training,
as well as design, system integration, and
project management.
For more information on repair and
calibration services, go to
www.agilent.com/find/removealldoubt
For more information on Agilent
Technologies’ products, applications
or services, please contact your local
Agilent office. The complete list is
available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0) 2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
07031 464 6333
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: March 24, 2009
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2009
Printed in USA, April 20, 2009
5968-4767E