ASM-SENSOR PD-TDC

PRODIS-TDC
Digital Process Meter
for Time-of-Flight Sensors
· For Sensors with pulse output:
· Automatic sensor parameter upload at power-on
– no calibration necessary
· Integrated sensor supply
· 6-digit LED display
· RS-232 interface
· Optional 4 comparator outputs
· Easy programming
Description
PRODIS-TDC displays measurement values of magnetostrictive position sensors,
programmable scaling is provided to display units like mm or inch. The request signal will be
sent to the sensor periodically. The time between the start pulse and the stop pulse will be
counted by a high-resolution time-digital-converter (TDC). When used with POSICHRON
sensors the calibration parameters of the sensor will be uploaded to the meter at power-on.
This avoids a calibration when the sensor has been changed. A tare function or
programming lock can be activated with two control terminals.
Sensor excitation is supplied from the meter. Optional comparator functions with 4 NPN
open-collector outputs are available, additional 2 of them have relay output.
Display
Specifications
Excitation voltage/current
Sampling rate
Sensor excitation
Start pulse input
Measurement pulse input
Control inputs
Comparator outputs (option) Relay
NPN
Connection
Operating temperature
Storage temperature
Weight
Protection class
Humidity
Safety of equipment
Electromagnetic compatibility
6 - digit, 7-segment-LED, 14 mm high, decimal point
programmable
24 V DC ±10%/150 mA, residual ripple 1%ss
85-250 V AC/180 mA max.
1 … 5 ms, adjustable
24 V DC/300 mA or 5 V DC/500 mA
INIT, INIT (RS422)
STSP, (RS422)
2 control inputs, active low
250 V AC/5 A, 30 V DC/5 A
24 V max./50 m A to GND
Terminal strip 12-pole, excitation 3-pole
-10 ... +40 °C
-20 ... +85 °C
24 VDC: 250 g approx., 230 VAC: 400 g approx.
Front IP60, rear IP40
Max. 95% R.H, non condensing
Directive 73/23/EWG: DIN EN 61010:1994-03
Directive 89/336/EWG
Order code PRODIS-TDC
PD-TDC
Model Name
Excitation Voltage
24VDC = 24 V DC
230VAC = 85...250 V AC
Comparator (option)
REL2
Order example: PD - TDC - 24VDC
50
CAT-PC-E-04
Tel.: +49-8123-986-0 Fax: +49-8123-986-500 www.asm-sensor.com
ASM GmbH
PRODIS-TDC
Digital Process Meter
for Time-of-Flight Sensors
Programmable
Parameters
/ value range
Divisor, Multiplier, offset, limit values
Other programmable parameters
0 up to 999999
Wave speed, average function, edge evaluation,
use with several magnets, difference evaluation,
decimal point position, display brightness
Connector X1
Pin no.
1
2
3
4
5/6
7
8
9
10
11
12
Signals
Wiring basic
equipment
Sensor excitation +UB
Sensor excitation 0 V (GND)
Control input terminal1, active low
Control input terminal2, active low
Not used
INIT output request pulse
INIT output request pulse
STSP input measurement pulse
STSP input measurement pulse
Do not connect!
GND
PD-TDC-24VDC
Connector X2
Pin no.
13
14
Excitation +24 V
Excitation 0 V (GND)
PD-TDC-230VAC
13, 15
14
Excitation
Protective ground
5 .......1
RS-232
1 ................................12 13 14 15
PD-TDC-24VDC
Rear view without
comparator outputs
X1
X2
13 14 15
PD-TDC-230VAC
X1
Level
RS-232 interface
ASM GmbH
X2
RS-232: ±8 V, galvanically isolated
Data format
1 startbit, 8 data bits, 1 stopbit, no parity
Transmission rate
9600 Baud
Signals
Connector X3 Pin no.
DSUB Pin no.
TxD
17
2
RxD
16
3
GND
18
5
Tel.: +49-8123-986-0 Fax: +49-8123-986-500 www.asm-sensor.com
CAT-PC-E-04
51
PRODIS-TDC
Digital Process Meter
for Time-of-Flight Sensors
24 23 22 21 20 19 18 17 16
30 29 28 27 26 25
X3
Rear view with
comparator outputs
Comparator
Comparator function
(option)
Comparator 1
Comparator 2
Comparator 3
Comparator 4
Signals
RS-232
X4
Comparator output
NPN
Connector X3
collector
pin no.
NPN1
NPN2
NPN3
NPN4
NPN GND
NPN UB (+24V)
Sensor circuit
20
21
Relay
Connector X4
pin no.
Relay 1
NO
NC
Common
Relay2
NO
NC
Common
LED
LED1
25
27
26
LED2
28
30
29
22
23
24
19
Cable
PRODIS-TDC
PRODIS-TDC generates a request pulse via the INIT line. The position sensor responds
with a START and a STOP signal.
Position value x = t(x) · vs
To calculate the position value
the time interval t(x) must be
multiplicated with the conversion factor vs on the type label.
When ASM POSICHRONâ
sensors are used the conversion factor vs will be uploaded
at power-on.
52
CAT-PC-E-04
Tel.: +49-8123-986-0 Fax: +49-8123-986-500 www.asm-sensor.com
ASM GmbH
PRODIS-TDC
Digital Process Meter
for Time-of-Flight Sensors
Outline drawing
LED1
LED2
Control panel cutout
acc. to DIN 43700
PD - TDC - 24VDC
PD - TDC - 230VAC
Dimensions informative only
For guaranteed dimensions consult factory!
ASM GmbH
Tel.: +49-8123-986-0 Fax: +49-8123-986-500 www.asm-sensor.com
CAT-PC-E-04
53