EMERSON-NETWORKPOWER DPS54-M

Embedded Power for
Business-Critical Continuity
Rev. 10.01.08_123
DPS50-M
of 2
DPS50-M
60 Watts
Medical
Total Power:
Input Voltage:
# of Outputs:
30 - 60 Watts
90 - 264 VAC
127 - 300 VDC
Single
RoHS
Electrical Specifications
Input
Special Features
•Medical safety approvals
•CE Mark EMC & LVD
•Universal AC input
•Fully regulated output
•Overload and short circuit
protection
•Internal remote sense to the
output plug
•Constant voltage
•High efficiency
•High MTBF
•IEC320 AC receptacle, 3 pin
(type C14)
•Built in EMI filter
(CISPR 22 Class B)
•LED power good indicator
•Input power <74 watts
•Complies with EN61000-3-2
•Meets Energy Star
•WEEE & CeC iV compliant
Safety
UL
CSA
VDE
CE UL 60601-1
CSA-C22.2 no.601.1-M90
07501/EN60601-1 (IEC601)
EMC & LVD
Input ra­nge
Frequency
Inrush current
Input current
Efficiency
EMI/RFI
Safety ground
leakage current
90-264 VAC (wide range)
127 - 300 VDC
47-440 Hz
<60A @ 230 VAC, cold start @ 25C
<2.0A rms @ 115 VAC; <1A @ 230 VAC
80% @ full load typical, @115VAC
FCC Part 15, Subpart B Class B & EN55022 (CISPR 22) Class B
275µA max. @ 50/60 Hz, 264 VAC input
Output
Maximum Power (Po)
Hold-up time
Overvoltage protection
Overload protection
Cable/connector
60W (DPS52, 30W)
10ms. typical at full load @ 115VAC
20% - 45% above nominal output maximum
Latching type, recycle AC to reset.
Output short circuit protection auto recover
Overload protected @ 120 - 150% above maximum rating
DC cable with 2.5mm center plug
DC plug center +v
DC plug outer -v
Environmental Specifications
Operating temperature:
Storage temperature:
Electromagnetic
susceptibility:
Humidity:
Vibration:
MTBF demonstrated:
0° to +40°C ambient. Derate each output as 2.5% per degree from
40°C to 60°C. -20°C start up
-40°C to +85°C
Designed to meet EN61000-4, -2, -3, -4, -5, -6, -8, -11, level 3
Operating; non-condensing 10% to 90% RH
Three orthogonal axes, 3 sweeps per axe at 1 oct/min. 2G peak 8Hz
to 500Hz, operational
>550,000 hours at full load and 25°C ambient conditions
Embedded Power for
Business-Critical Continuity
­Americas
Ordering Information
Model
Number
Maximum
Power
Output
Voltage
Maximum
Load
Peak Load1
Regulation2
Ripple
P/P (PARD)3
DPS52-M
DPS53-M
DPS54-M
DPS55-M
DPS58-M
30W
60W
60W
60W
60W
5V
12V
15V
24V
48V
6A
5A
4A
2.5A
1.25A
6.5A
5.4A
4.3A
2.7A
1.35A
±2%
±2%
±2%
±2%
±2%
50mV
120mV
150mV
240mV
480mV
1. Peak-to-peak with 20MHz bandwidth and 10µF tantalum capacitor in parallel with a 0.1 µF capacitor.
2. At 25ºC including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings.
3. Peak-to-peak with 20MHz bandwidth and 10µF (tantalum capacitor) in parallel with a 0.1µF capacitor at rated
line voltage and load ranges.
Mechanical Drawing
5810 Van Allen Way
Carlsbad, CA 92008
USA
Telephone:+1 760 930 4600
Facsimile: +1 760 930 0698
Rev. 10.01.08_123
DPS50-M
of 2
Europe (UK)
Waterfront Business Park
Merry Hill, Dudley
West Midlands, DY5 1LX
United Kingdom
Telephone:+44 (0) 1384 842 211
Facsimile: +44 (0) 1384 843 355
Asia (HK)
2.39"
(60.7)
14/F, Lu Plaza
2 Wing Yip Street
Kwun Tong, Kowloon
Hong Kong
Telephone:+852 2176 3333
Facsimile: +852 2176 3888
5.24"
(133.0)
INDICATOR LED
For global contact, visit:
DC OUTPUT CABLE
www.powerconversion.com
1.62"
(41.0)
techsupport.embeddedpower
@emerson.com
AC INLET
78.7"
(2000±15)
MODEL LABEL
While every precaution has been taken to ensure
accuracy and completeness in this literature, Emerson
Network Power assumes no responsibility, and disclaims
all liability for damages resulting from use of this
information or for any errorsINDICATOR
or omissions.
LED
Emerson Network Power.
The global leader in enabling
business-critical continuity.
AC Power
Notes:
1.Specifications subject to change without notice.
2.All dimensions in inches (mm), tolerance is ± 0.02” (±0.5mm)
3.Warranty: 2 year
4.Weight: 1.06 lb./ .48 kg
5.AC input power cord sold separately.
6.Specifications are for convection rating at factory settings
at 115VAC input, 25°C unless otherwise stated
Connectivity
DC Power
Embedded Computing
3D VIEW Power
Embedded
Monitoring
Outside Plant
Power Switching & Controls
Precision Cooling
Racks & Integrated Cabinets
Services
Surge Protection
EmersonNetworkPower.com
Emerson Network Power and the Emerson
Network Power logo are trademarks and
service marks of Emerson Electric Co. ©2008 Emerson Electric Co.