NI PXI-2529

High-Density Multiconfiguration
Matrix Modules
NI PXI-2529, NI PXIe-2529, NI SCXI-1129 NEW!
■
■
8 matrix configurations
Switch capacity
150 VDC, 150 Vrms CAT I
1 A switching/2 A carry
32,000-step scan list for
deterministic scanning
Fully software programmable
Effortless matrix expansion
Multiple-module synchronization
with hardware triggers
Electromechanical relays
Available in both PXI and PXI Express
versions for optimal slot placement
NI PXI-2529/PXIe-2529
128-crosspoint matrix
4x32 and 8x16 2-wire
matrix configurations
NI SCXI-1129
256-crosspoint matrix
Six 2-wire matrix configurations
■
■
■
■
■
■
■
■
■
Operating Systems
Windows XP/2000/NT
■
Recommended Software
LabVIEW
LabWindows™/CVI
Measurement Studio
for Visual C++
NI Switch Executive
■
■
■
■
Other Compatible Software
Visual Basic
ANSI C/C++
■
■
Driver Software (included)
NI-SWITCH
■
■
■
■
■
■
Overview
configurations by simply changing the terminal block. For example, you can
The NI PXI-2529, PXIe-2529, and SCXI-1129 are high-density matrix relay
modules built for applications with high-channel counts. The modules are
two-wire matrices that are configurable with front mounting terminal blocks to
achieve many different matrix configurations. Table 1 provides a complete list of
possible configurations. Expanding the matrix channel count is as easy as adding
more modules. With the SCXI-1129, you can pass analog signals between two
or more switch modules via a high-voltage backplane (HVAB), matrix expansion
create an 8x32 two-wire matrix by using the NI SCXI-1335 terminal block with an
SCXI-1129 module. You can easily transform the same SCXI-1129 module into a
16x16 two-wire matrix by simply replacing the SCXI-1335 with an NI SCXI-1336
terminal block.
The NI 2529 modules use latching relays in the matrices to ensure that the
state of each matrix remains unchanged in the event of accidental power loss.
This feature prevents instruments and units under test from switching during
cables, or matrix expansion plugs. Using these connections, you can instantly
potentially damaging conditions. All relays open when power returns.
expand your matrix without external wiring. The NI 2529 modules are designed
Automatic Scanning
to work well with both low- and high- voltage levels. They use relays with low
thermal offset to ensure accurate low-voltage measurements. These relays can
switch up to 150 Vrms or 150 VDC.
the use of scanning. Scanning improves throughput by downloading a list of
up to 32,000 connections to the switch and cycling through the list using an
Matrix Operation
The NI 2529 modules are general-purpose,
two-wire matrices used to route differential
event (trigger) without any interruption from the host processor. You can most
efficiently implement scanning by mating an NI 2529 or SCXI-1129 with an
instrument such as the NI PXI-4070 6½-digit FlexDMM, which issues a trigger
signals. Through software, you can control
after each measurement.
the matrices to connect any row to any
Safety Disconnect
column. One of the key advantages of
these modules is their adaptability to
your switching needs. You can configure
Figure 1. The matrix expansion plug
simplifies expansion.
NI 2529 modules can maximize throughput in automated test applications with
the SCXI-1129 in six different matrix
configurations and an NI 2529 in two
The SCXI-1129 incorporates a safety interlock that disconnects the SCXI
high-voltage analog bus from the matrix if the front terminal block is removed.
This arrangement prevents potentially hazardous voltages from appearing on
exposed front panel connector pins.
High-Density Multiconfiguration Matrix Modules
NI SCXI-1129
NI 2529
Terminal Block
Configuration (Row by Column)
Column Expansion Via Row Connection
Row Expansion Via Column Connection
SCXI-1333
Quad 4x16
External wires
With matrix expansion plug.
Examples: quad 8x16, quad 16x16
SCXI-1334
4x64
With HVAB or matrix expansion cable.
Examples: 4x128, 4x256
With matrix expansion plug.
Examples: 8x64,16x64, 32x64
SCXI-1335
8x32
With matrix expansion cable (2 cables required).
Examples: 8x64, 8x128
With matrix expansion plug.
Examples: 16x32, 32x32
SCXI-1336
16x16
External wires
With matrix expansion plug.
Examples: 32x16, 64x16
SCXI-1337
Dual 8x16
External wires
External wires
SCXI-1339
Dual 4x32
External wires
External wires
TB-2634
4x32
External wires
External wires
TB-2635
8x16
External wires
External wires
Table 1. NI 2529 and SCXI-1129 Configuration and Matrix Expansion
Expansion with the SCXI-1129
SCXI Analog Routing
Matrix modules can serve as building blocks for creating configurations well
You can increase the flexibility of your switch system with the addition of the
beyond the size of a single module. Connecting the rows of two modules
high-voltage analog bus (HVAB). Using the HVAB, you can easily route high-
doubles the column count. Likewise, connecting the columns of two modules
voltage or low-voltage analog signals between multiple switch modules or
doubles the row count. For example, you can expand a 4x64 matrix, made up of
instruments without the need for external wiring (assuming it is safe to do so).
one SCXI-1129 module and one SCXI-1334 terminal block, to a 4x128, 4x192,
For example, you can use the HVAB to effortlessly create a 4x256 matrix from
4x256, and so forth, by connecting the columns of multiple SCXI-1129 modules
four SCXI-1129 modules in the 4x64 matrix configuration (SCXI-1334 terminal
via the terminal block. To expand the columns of the SCXI-1129, you can use
block). Without the HVAB, the 4x256 matrix requires external wiring or cables.
expansion cables. Expansion cables connect four differential signals. Hence,
The HVAB works with the SCXI-1127, SCXI-1128, and SCXI-1129 modules.
you must use two matrix expansion cables for each expansion with the
SCXI-1335 because it has eight rows. You can expand the row count of the
SCXI-1129 with expansion plugs, which connect the columns of two adjacent
switch modules as shown in Figure 1. With the expansion plugs, you can
Signal Connections
You can choose from several solutions for your signal connections:
■
for easy connections
create the 128x16 matrix in less than five minutes. For further information
on matrix expansion, refer to the tutorial, Matrix Switch Expansion Guide,
Terminal blocks provide screw terminals, solder cups, or headers
■
A shielded backshell features a 180-pin connector housing
to accommodate custom cables on the SCXI-1129
at http://zone.ni.com/devzone/cda/tut/p/id/3628.
■
Mass interconnect solutions from industry leaders
SCXI Relay Control
Contact your local National Instruments sales representative for information
Every SCXI switch system requires an external switch controller. The switch
on these products.
controller uses the digital communications bus on the SCXI chassis to control
the switch circuitry. An NI 4021 switch controller, available for both PCI and
PXI, is the recommended SCXI switch controller.
Software
All National Instruments PXI and SCXI switch modules are shipped with
NI-SWITCH, an IVI-compliant driver that offers complete functionality for
all switch modules. For additional assistance in configuring, programming,
and managing higher-channel-count switching systems, NI Switch Executive
software provides an easy-to-use, intelligent switch management and visual
routing environment.
BUY ONLINE at ni.com or C ALL 800 813 36 93 (U.S.)
2
High-Density Multiconfiguration Matrix Modules
Specifications
Environment
Specifications subject to change without notice. Visit ni.com/manuals for
the latest full specifications.
Input Characteristics
Maximum switching voltage.................
Operating temperature
NI 2529 .............................................
0 to 55 °C
NI SCXI-1129 ....................................
0 to 50 °C
Storage temperature .............................
-20 to 70 °C
150 Vrms or 150 VDC
Relative humidity ..................................
5 to 85% noncondensing
(channel-to-channel and
Pollution degree ....................................
2
channel-to-ground)
Approved at altitudes up to 2,000 m.
Maximum switching current .................
1 A (per channel)
Maximum carry current .........................
2 A (per channel)
Ordering Information
Maximum module current
NI 2529 .............................................
8A
NI PXI-2529 ....................................................................................778739-01
NI SCXI-1129 ....................................
5A
NI PXIe-2529 ..................................................................................780587-29
Maximum switching power ..................
37.5 VA, 30 W (per channel)
NI SCXI-1129 ..................................................................................776572-29
Path resistance......................................
<1 Ω
Includes NI-SWITCH driver software.
Accessories
Transfer Characteristics
Thermal EMF .........................................
<9 μV
Typical single crosspoint bandwidth
1 row – 1 column ..............................
>10 MHz
Typical crosstalk
10 kHz ...........................................
≤80 dB
100 kHz .........................................
≤65 dB
1 MHz ...........................................
≤50 dB
NI SCXI-1333 (quad 4x16) ..............................................................777687-33
NI SCXI-1334 (4x64) .......................................................................777687-34
NI SCXI-1336 (16x16) .....................................................................777687-36
NI SCXI-1337 (dual 8x16) ...............................................................777687-37
NI SCXI-1339 (dual 4x32) ...............................................................777687-39
NI matrix expansion plugs .............................................................778364-01
NI matrix expansion cables
Dynamic Characteristics
40 cm ...................................................................................185440-0R4
Maximum operating speed
110 crosspoints/s .............................
NI 2529
125 crosspoints/s .............................
NI SCXI-1129
4 ms max
Expected relay life
Mechanical .......................................
NI TB-2635 (8x16)...........................................................................778839-01
NI SCXI-1335 (8x32) .......................................................................777687-35
50 Ω system
Relay operate time (at 20 °C) ...............
NI TB-2634 (4x32)...........................................................................778840-01
5x107 operations
Electrical
30 VDC, 1 A.......................................
105 operations
30 VDC, 100 mA................................
5x10 operations
150 Vrms, 0.25 A .................................
105 operations
5
75 cm .................................................................................185440-0R75
NI Switch Executive
Development system......................................................................778546-01
Deployment engine ........................................................................778548-00
BUY NOW
For complete product specifications, pricing, and accessory information,
call 800 813 3693 (U.S.) or go to ni.com/switches.
Physical
NI 2529
Dimensions .......................................
2 by 10 by 16 cm
(0.8 by 3.9 by 6.4 in.)
I/O connector ....................................
100-pin HDI
NI SCXI-1129
Dimensions .......................................
3.0 by 17.2 by 20.3 cm
(1.18 by 6.9 by 8.0 in.)
I/O connector ....................................
180-pin HDI
BUY ONLINE at ni.com or C ALL 800 813 36 93 (U.S.)
3
High-Density Multiconfiguration Matrix Modules
Safety Compliance
■
IEC 61010-1, EN 61010-1
■
UL 61010-01, CSA 61010-1
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for
electrical equipment for measurement, control, and laboratory use:
■
EN 61326 (IEC 61326): Class A emissions; Basic immunity
■
EN 55011 (CISPR 11): Group 1, Class A emissions
■
AS/NZS CISPR 11: Group 1, Class A emissions
■
FCC 47 CFR Part 15B: Class A emissions
■
ICES-001: Class A emissions
Note: For EMC compliance, operate this product according to the documentation.
CE Compliance
This product meets the essential requirements of applicable European Directives
as follows:
■
2006/95/EC; Low-Voltage Directive (safety)
■
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
BUY ONLINE at ni.com or C ALL 800 813 36 93 (U.S.)
4
NI Services and Support
your needs around the globe and through
PL
AN
MA
IN
NI has the services and support to meet
IN
TA
the application life cycle – from planning
and development through deployment and
SERVICE
NEEDS
ongoing maintenance. We offer services
P
DE
EL
OP
and service levels to meet customer
LO
Y
requirements in research, design, validation,
V
DE
and manufacturing.
Visit ni.com/services.
Local Sales and Technical Support
In offices worldwide, our staff is local to the country, giving you access to
engineers who speak your language. NI delivers industry-leading technical
support through online knowledge bases, our applications engineers,
and access to 14,000 measurement and automation professionals within NI
Developer Exchange forums. Find immediate answers to your questions at
ni.com/support.
We also offer service programs that provide automatic upgrades to your
application development environment and higher levels of technical support.
Training and Certification
Visit ni.com/ssp.
NI training is the fastest, most certain route to productivity with our products.
Hardware Services
NI training can shorten your learning curve, save development time, and reduce
maintenance costs over the application life cycle. We schedule instructor-led
System Assurance Programs
courses in cities worldwide, or we can hold a course at your facility. We also
NI system assurance programs are designed to make it even easier for you
offer a professional certification program that identifies individuals who have
to own an NI system. These programs include configuration and deployment
high levels of skill and knowledge on using NI products.
services for your NI PXI, CompactRIO, or Compact FieldPoint system. The NI Basic
Visit ni.com/training.
System Assurance Program provides a simple integration test and ensures that
Professional Services
your system is delivered completely assembled in one box. When you configure
your system with the NI Standard System Assurance Program, you can select
Our NI Professional Services team is composed of NI applications and systems
from available NI system driver sets and application development environments
engineers and a worldwide National Instruments Alliance Partner program of
to create customized, reorderable software configurations. Your system arrives
more than 600 independent consultants and integrators. Services range from
fully assembled and tested in one box with your software preinstalled. When
start-up assistance to turnkey system
you order your system with the standard program, you also receive system-
integration. Visit ni.com/alliance.
specific documentation including a bill of materials, an integration test report,
a recommended maintenance plan, and frequently asked question documents.
OEM Support
Finally, the standard program reduces the total cost of owning an NI system
We offer design-in consulting and product integration assistance if you want to
online product advisors at ni.com/advisor to find a system assurance program
use our products for OEM applications. For information about special pricing and
services for OEM customers, visit ni.com/oem.
by providing three years of warranty coverage and calibration service. Use the
to meet your needs.
Calibration Services
NI recognizes the need to maintain properly calibrated devices for highaccuracy measurements. We provide manual calibration procedures, services
to recalibrate your products, and automated calibration software specifically
designed for use by metrology laboratories. Visit ni.com/calibration.
Repair and Extended Warranty
NI provides complete repair services for our products. Express repair and
advance replacement services are also available. We offer extended warranties
to help you meet project life-cycle requirements. Visit ni.com/services.
ni.com
■
800 813 3693
National Instruments
■
[email protected]
©2009 National Instruments. All rights reserved. CompactRIO, CVI, FieldPoint, FlexDMM, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com, and SCXI are trademarks
of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Other product and
company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National Instruments and has no agency, partnership, or
joint-venture relationship with National Instruments.
2009-11147-101-D