NI SCXI-1190

1.3 GHz and 4 GHz Multiplexers
1.3 GHz and 4 GHz Multiplexers
NI SCXI-1190, NI PXI-2590, NI SCXI-1191, NI PXI-2591
• 4x1 unterminated multiplexer
• 50 characteristic impedance
• High-bandwidth
electromechanical relays
• Excellent insertion loss, VSWR,
and isolation specifications
• Fully software programmable
SCXI-1190 (quad 4x1),
PXI-2590 (single 4x1)
• 1.3 GHz bandwidth
• SMB direct connectivity
• 1 A at 24 VDC capacity
SCXI-1191 (quad 4x1),
PXI-2591 (single 4x1)
• 4 GHz bandwidth
• SMA direct connectivity
• 330 mA at 30 VDC capacity
Operating Systems
• Windows 2000/NT/XP
Recommended Software
• LabVIEW
• LabVIEW Real-Time Module
(PXI Modules only)
• LabWindows/CVI
• Measurement Studio
• NI Switch Executive
Other Compatible Software
• Visual Basic
• C/C++
Driver Software (included)
• NI-SWITCH
Compliance
• CE
Overview
Switches
The National Instruments PXI-2590 and PXI-2591 are 4x1 highfrequency unterminated multiplexing switches. The NI SCXI-1190 and
NI SCXI-1191 perform the same functions as the PXI-2590 and
PXI-2591 but also offer a quad 4x1 multiplexing configuration. The
switches are capable of switching signals from DC to 1.3 GHz
(SCXI-1190 and PXI-2590) and DC to 4 GHz (SCXI-1191 and
PXI-2591). The characteristic impedance of the channels is 50 .
For the SCXI-1190 and the PXI-2590, the maximum voltage rating
of the switches is 24 Vrms with a maximum current of 1 A. For the
SCXI-1191 and the PXI-2591, the maximum voltage rating is 30 Vrms
with a maximum current of 330 mA. All are well suited for
applications that require the routing of high-frequency signals inside
automated test equipment (ATE) systems because they can switch
signals with a very low insertion loss. In addition, the excellent
voltage standing-wave ratio (VSWR) and isolation parameters make
these modules the perfect choice in a system geared to highfrequency applications as shown in Table 1.
Ordering Information
NI SCXI-1190 ................................................................776572-90
NI PXI-2590 ..................................................................777987-01
NI SCXI-1191 ................................................................776572-91
NI PXI-2591 ..................................................................778339-01
Includes switch module and NI-SWITCH driver software.
For information on extended warranty and value added
services, see page 20.
See page 499 for accessory and cable information.
BUY ONLINE!
Visit ni.com/products and enter pxi2590, pxi2591, scxi1190,
and/or scxi1191.
Module
SCXI-1190, PXI-2590 (@ 1.3 GHz)
SCXI-1190, PXI-2590 (@ 2 GHz)
SCXI-1191, PXI-2591 (@ 4 GHz)
Table 1. Parameters of High-Frequency Switching Modules
Extended Features and Specifications
National Instruments switch modules are built with a number of core
features that are covered in detail in the Switch Overview section.
For additional information about the PXI-2590, PXI-2591, SCXI-1190, and
SCXI-1191, including software, certifications and compliance, relay
control, etc., please see page 20. For detailed specifications, please
see page 504.
492
Insertion Loss
< 1.5 dB
< 3 dB
< 0.9 dB
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • [email protected] • ni.com
VSWR
< 1.5
< 2.5
< 1.5
Specifications
Switch Specifications
Specifications (continued)
2
SCXI-1190, PXI-2590
1.9
Input Characteristics
1.8
1.7
VSWR
1.6
1.5
8x1
1.4
32x1
1.3
16x1
1.2
1.1
0
0
100
200
300
400
500
600
Frequency (MHz)
700
800
900
1000
Input Characteristics
Maximum switching voltage............................
Maximum switching current ............................
Maximum carry current....................................
Simultaneous channels at maximum current
PXI-2593 ...................................................
SCXI-1193 ..................................................
DC path resistance
Initial .........................................................
End of life ..................................................
150 V, CAT I
0.5 A (per channel)
1 A (per channel)
2, maximum
4, maximum
<1.0 Ω
≥2.0 Ω
Dynamic Characteristics
Maximum scan rate ......................................... 100 operations/s
Expected relay life:
Mechanical ............................................... 5x107 operations
Electrical .................................................... 3x105 operations (30 V, 0.3 ADC resistive)
Physical Characteristics
Relay type.........................................................
Relay contact material .....................................
Dimensions
PXI-2593.....................................................
SCXI-1193 ..................................................
I/O connectors
PXI-2593.....................................................
SCXI-1193 ..................................................
Trigger connectors............................................
Electromechanical, latching
Silver palladium and gold
10 by 16 by 4 cm (3.9 by 6.3 by 1.6 in.)
3.0 by 17.3 by 19.6 cm (1.2 by 6.7 by 7.6 in.)
18 MCX jacks
36 MCX jacks
2 SMB jacks
Environment
Operating temperature.....................................
Storage temperature ........................................
Relative humidity .............................................
Pollution degree ...............................................
Approved altitudes ..........................................
0 to 50 °C
- 20 to 70 °C
5 to 85% noncondensing
2
Up to 2,000 m
Switches
Safety
504
This product is designed to meet the requirements of the following standards of safety for electrical
equipment for measurement, control and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 3111-1, UL 61010B-1
• CAN/CSA C22.2 No. 1010.1
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE Marking, as follows:
Low-voltage Directive (safety) .................. 73/23/EEC
Electromagnetic Compatibility
Directive (EMC).......................................... 89/336/EEC
Number of channels per bank..........................
Number of banks
SCXI-1190 ..................................................
PXI-2590.....................................................
Input voltage
Channel-to-earth........................................
Maximum switching voltage
AC ............................................................
DC ............................................................
Maximum switching capacity per channel
AC ............................................................
DC ............................................................
Maximum switching power per channel .........
Contact resistance (initial) ...............................
Contact material...............................................
4
4
1
24 VDC
24 Vrms
24 VDC
1 Arms at 24 Vrms
1 A at 24 VDC
24 W
100 mΩ maximum
Gold-clad silver
RF Performance Characteristics
Characteristic impedance (Z0 ).........................
Insertion loss
100 MHz ....................................................
500 MHz ....................................................
1.3 GHz ......................................................
2 GHz .........................................................
VSWR
100 MHz ....................................................
500 MHz ...................................................
1.3 GHz ......................................................
2 GHz .........................................................
Isolation
500 MHz ....................................................
1.3 GHz ......................................................
2 GHz .........................................................
Rise time ..........................................................
Signal delay......................................................
Maximum RF carry power at 900 MHz............
50 Ω
<0.4 dB
<0.9 dB
<1.5 dB
<3 dB
<1.15
<1.35
<1.5
<2.5
>60 dB
>50 dB
>30 dB
<300 ps
<3 ns
10 W
Dynamic Characteristics
Relay operate time (at 20 °C)
Typical........................................................ 15 ms
Relay release time (at 20 °C)
Typical........................................................ 15 ms
Expected life
Mechanical (no load)................................. 5x106 operations
Electrical at maximum switching capacity 105 operations
Caution: Exceeding the maximum switching capability decreases expected life.
Power Requirement
+5 VDC ............................................................ 600 mA maximum (all relays closed)
Physical
Dimensions
SCXI-1190 .................................................. 17.3 by 19.6 by 3.0 cm (6.81 by 7.70 by 1.19 in.)
PXI-2590..................................................... 10 by 16 by 4 cm (3.9 by 6.3 by 1.6 in.)
I/O connectors .................................................. 5 SMB female per bank
Environment
Operating temperature.....................................
Storage temperature ........................................
Relative humidity .............................................
Pollution degree ...............................................
Approved altitudes ..........................................
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • [email protected] • ni.com
0 to 50 °C
- 20 to 70 °C
5 to 85% noncondensing
2
Up to 2,000 m
Switch Specifications
The total input power for all relays combined should not exceed the preceding specifications.
When using multiple relays, refer to table below for adjusted maximum input power
SCXI-1191, PXI-2591
Input Characteristics
Number of channels per bank .........................
Number of banks..............................................
SCXI-1191 ..................................................
PXI-2591.....................................................
Input voltage
Channel-to-ground.....................................
Maximum switching voltage
AC ............................................................
DC ............................................................
Maximum switching capacity per channel
AC. ............................................................
DC ............................................................
Maximum switching power per channel .........
Contact on resistance (initial) ..........................
Contact material ..............................................
Active Relays
1
2
4
8
4
4
4
1
30 Vrms/VDC
30 Vrms
30 VDC
8 to 12.4 GHz
60 W
30 W
15 W
7.5 W
12.4 to 18 GHz
50 W
25 W
12.5 W
6.25 W
RF Performance Characteristics
Characteristic impedance (Z0) .........................
Insertion loss at:
≤4 GHz .......................................................
4 to 8 GHz ..................................................
8 to 12.4 GHz .............................................
12.4 GHz to 18 GHz ...................................
VSWR at:
≤1 GHz .......................................................
1 to 4 GHz ..................................................
4 to 8 GHz ..................................................
8 to 12.4 GHz .............................................
12.4 to 18 GHz ...........................................
Isolation:
≤1 GHz .......................................................
1 to 4 GHz ..................................................
4 to 8 GHz ..................................................
8 to 12.4 GHz .............................................
12.4 to 18 GHz ...........................................
0.33 A at 30 Vrms
0.33 A at 30 VDC
10 W
200 mΩ maximum
Gold
50 Ω
<0.6 dB
<0.9 dB
<1.3
<1.5
>60 dB
>55 dB
10 W
50 Ω
≤0.2 dB
≤0.3 dB
≤0.4 dB
≤0.5 dB
≤1.1
≤1.15
≤1.25
≤1.35
≤1.5
≥85 dB
≥80 dB
≥70 dB
≥65 dB
≥60 dB
Dynamic Characteristics
Dynamic Characteristics
Relay operate time (at 20 °C)
Typical........................................................ 15 ms
Relay release time (at 20 °C)
Typical........................................................ 15 ms
Expected life
Mechanical (no load)................................. 5x106 operations
Electrical at 5 W........................................ 5x106 operations (to 3 GHz, 50 Ω, VSWR maximum 1.2)
Caution: Exceeding these electrical parameters decreases expected life.
Relay operate time (at 20 °C)
Typical........................................................ 15 ms
Relay release time (at 20 °C)
Typical........................................................ 15 ms
Expected life
Mechanical (no load)................................. 5x106 operations
Electrical (maximum switching capacity).. 105 operations
Caution: Exceeding the maximum switching capability decreases expected life.
Power Requirement
Power Requirement
+5 VDC ............................................................ 800 mA, maximum
+5 VDC ........................................................... 950 mW maximum (all relays closed)
Physical
Physical
Dimensions
SCXI-1191 .................................................. 17.3 by 19.6 by 3.0 cm (6.81 by 7.70 by 1.19 in.)
PXI-2591..................................................... 10 by 16 by 4 cm (3.9 by 6.3 by 1.6 in.)
I/O connector.................................................... 5 SMA female per bank
Environment
Operating temperature..................................... 0 to 50 °C
Storage temperature ........................................ -20 to 70 °C
Relative humidity ............................................. 5 to 85% noncondensing
Input Characteristics
8 SPDT (latching)
30 Vrms/VDC
30 VDC
2 ADC
120 W
80 W
60 W
50 W
Dimensions....................................................... 17.3 by 19.6 by 3.0 cm (6.81 by 7.70 by 1.19 in.)
I/O connector.................................................... 24 SMA female
Environment
Operating temperature..................................... 0 to 50 °C
Storage temperature ........................................ -20 to 70 °C
Relative humidity ............................................. 5 to 85% noncondensing
SCXI-1163R
Input Characteristics
Number of relays..............................................
Relay type.........................................................
Maximum switching voltage
AC ............................................................
DC ............................................................
Maximum switching capacity ..........................
Commonmode isolation ...................................
On-resistance ...................................................
Output capacitance ..........................................
Leakage current................................................
Transfer rate in serial mode1
(1 word = 32 bits) ......................................
Relay set time ..................................................
Relay reset time ...............................................
Power-on state .................................................
32 organized as 8 optically isolated banks of 4 relays each
Normally open (Form A). solid-state relays
240 VAC
240 VDC
200 mA
250 Vrms between banks, and bank to ground
8Ω
110 pF at 50 V, 1 MHz
1 µA maximum
Switches
SCXI-1192
Number of relays..............................................
Input voltage
Channel-to-ground.....................................
Maximum switching voltage............................
Maximum carry current per channel................
Maximum continuous carry power per channel
1 to 3 GHz ..................................................
3 to 8 GHz ..................................................
8 to 12.4 GHz .............................................
12.4 to 18 GHz ...........................................
3 to 8 GHz
80 W
40 W
20 W
10 W
Contact resistance (initial) ............................... 100 mΩ maximum
Contact material............................................... Gold plate
RF Performance Characteristics
Characteristic impedance (Z0 ) ........................
Insertion loss
2.5 GHz ......................................................
4 GHz .........................................................
VSWR
2.5 GHz ......................................................
4 GHz .........................................................
Isolation
2.5 GHz ......................................................
4 GHz .........................................................
Maximum RF carry power at 900 MHz............
1 to 3 GHz
120 W
60 W
30 W
15 W
Specifications
Specifications (continued)
750 words/s
0.6 ms
0.1 ms
Relays open
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • [email protected] • ni.com
505