PHIHONG PSAM30R-240

WWW.PHIHONG.COM
30 Watt Universal 2-Wire Input Medical Adapter
PSAM30R Series Adapter Meets CEC and EISA Requirements
V071146
Features
•
•
•
Double Insulated
< 100uA Leakage Current
Class B EMI
•
•
•
No Load Power Draw < 0.5W at 115VAC
No Load Power Draw < 0.75W at 230VAC
EISA Energy Efficient Compliance
•
Patient Hoists
•
AS/NZS 3260 & 4417
Applications
•
•
Dental Curing Devices
Portable Dental Equipment
Safety Approvals
•
•
cUL/UL60601-1
CE
Mechanical Characteristics
•
•
•
•
Length: 100mm (3.94in)
Width: 63mm (2.48in)
Height: 35.7mm (1.41in)
Weight: 200g (7.05oz.)
Output Specifications
Model
DC Output
Voltage
PSAM30R-120
+12V
PSAM30R-240
+24V
Load
Min.
Max.
0A
2.5A
0A
1.25A
Ripple (1)
P-P (max.)
120mV
Regulation
Line
Load
±5%
240mV
±5%
Efficiency Level
(1) Measured with using 12-inch twisted pair terminated with a 10uF capacitor and 0.1uF ceramic in parallel.
Interchangeable clips sold separately.
Phihong reserves the right to make changes without further notice. Please consult Phihong USA and visit
www.phihong.com for the most up-to-date specifications.
PSAM30R Series Characteristics
WWW.PHIHONG.COM
INPUT:
AC Input Voltage Rating
Short-Circuit Protection
100 to 240VAC
Output can be shorted without damage
AC Input Voltage Range
90 to 264VAC
ENVIRONMENTAL
Temperature
AC Input Frequency
Operation
Non-operation
0 to +40°C
-25 to +85°C
47 to 63Hz
Relative Humidity
Input Current
1.0A (rms) max. @ 115VAC, 60Hz
0.5A (rms) max. @ 230VAC, 50Hz
Operation
Non-operation
10 to 90% Non-Condensing
10 to 95% Non-Condensing
Emissions
Leakage Current
100uA max. @ 264VAC, 50hz.
FCC Class B
EN55022 Class B / CISPR22 Class B
Inrush Current (cold start @ ambient 25°C)
Immunity
30A for 115VAC at max. load
60A for 230VAC at max. load
> 80.6% average efficiency (Level IV)
EN 50082-1: (EN55024)
EN 61000-4-2, Level 3
EN 61000-4-3, Level 2
EN 61000-4-4, Level 2
EN 61000-4-5, Level 3
EN 61000-4-6, Level 2
EN 61000-4-8
EN 61000-4-11
EN 61000-3-2: Harmonics
EN 61000-3-3: Flicker
Input Power Saving
0.50W max. for 115VAC, 60Hz at no load (Level IV)
0.75W max. for 230VAC, 50Hz at no load (Level III)
OUTPUT:
Efficiency
Hold-up Time
Dielectric Withstand (Hi-pot) Test
10mS Min. @ 115VAC, 60Hz
Primary to Secondary: 4000VAC for 1 minute, 10mA
Over-Voltage Protection
DC Output Connector
Included, all models
2.1mm x 5.5mm center positive standard
Over-Current Protection
AC Input Clips (sold separately)
Output equipped with short circuit protection – auto
restart
US: RPA
Europe: RPE
UK: RPK
Australia: RPS
Dimension Diagram Unit: mm