TRUMPOWER TMC40

TRUMPower
30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
DESCRIPTION
The TMC40 series of compact, open PCB constructed, AC-DC
TMC40 SERIES
switching power supplies are capable of delivering 30-48
watts of continuous output power at convection cooling. They
operate at 90-264 VAC input voltage without the need of
voltage selection, and are suited for medical, information
technology and industrial applications. Approval to both
EN60601-1
and
EN60950-1
safety
standards
improves
design-in time and reduces end equipment compliance costs.
FEATURES
z
Medical and ITE approvals
z
Compact size 2” x4” x1.18”
z
Single, dual and triple outputs
z
Full-range input 90-264 VAC
z
Low earth leakage current
z
z
Level B emissions
RoHS compliant
SAFETY STANDARD APPROVALS
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
90-264 VAC
47-63 Hz
0.9 A (rms) for 100 VAC
0.5 A (rms) for 240 VAC
Earth Leakage current:
150 µA max. @ 264 VAC, 63 Hz
GENERAL SPECIFICATIONS
Switching frequency:
Efficiency:
62±5 KHz
80-88% typical except TMC40-T31-3 and
TMC40-T31-5 at 75% typical
Hold-up time:
12 msec minimum at 110 VAC
Line regulation:
±0.5% maximum at full
loadKHz
62±5
25 A @ 115 VAC, or 50
A @ 230
VAC, except
80-88%
typical
OUTPUT SPECIFICATIONS
Inrush current:
Output voltage/current:
Maximum output power:
Withstand voltage:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
See rating chart
See rating chart
2% peak to peak on 5.0 V model,
1% peak to peak on other models.
Provided on output #1 only; set at
112–132% of its nominal output voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% /°C maximum
Maximum excursion of 4% or better on all
models, recovering to 1% of final value
within 500 µs after a 25% step load
change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
o
o
-10 C to +70 C
o
o
-40 C to +85 C
5% to 95% non-condensing
o
Derate from 100% to +50 C linearly to
o
50% at +70 C
o
at 25 C cold start
TMC40-T31-3
and
4000 VAC from input to
output
TMC40-T31-5
at 75%
1500 VAC from input to
ground
typical
500 VAC from output to ground
MTBF:
o
400,000 hours at full load at 25 C ambient,
calculated per MIL-HDBK-217F
EMC Performance
EN55011 / EN55022:
Class B conducted, class B radiated
FCC:
Class B conducted, class B radiated
VCCI:
Class B conducted, class B radiated
EN61000-3-2:
Harmonic distortion, class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8 KV air and ±6 KV contact
EN61000-4-3:
Radiated immunity, 3 V/m
EN61000-4-4:
Fast transient/burst, ±2 KV
EN61000-4-5:
Surge, ±1 KV diff., ±2 KV com.
EN61000-4-6:
Conducted immunity, 3 Vrms
EN61000-4-8:
Magnetic field immunity, 3 A/m
EN61000-4-11:
Voltage dips,
30% reduction for 500 ms
60% reduction for 100 ms
>95% reduction for 10 ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com
UNIVERSAL INPUT
TMC40 MEDICAL & ITE SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Output #1
P/N
Output #2
Vnom.
Imin.
Imax.
Tol.
TMC40-S05
5 V
0A
8.0 A
2%
(N/A)
(N/A)
40 W
TMC40-S12
12 V
0A
3.5 A
2%
(N/A)
(N/A)
42 W
TMC40-S15
15 V
0A
3.0 A
2%
(N/A)
(N/A)
45 W
TMC40-S24
24 V
0A
2.0 A
2%
(N/A)
(N/A)
48 W
TMC40-S48
48 V
0A
1.0 A
2%
(N/A)
(N/A)
48 W
TMC40-D23
+5 V
0.5 A
6A
3%
+12 V
0.1 A
2A
5%
(N/A)
40 W
TMC40-D25
+5 V
0.5 A
6A
3%
+24 V
0.1 A
1A
5%
TMC40-T31
+5 V
0.5 A
6A
3%
+12 V
0.1 A
2A
5%
-12 V
0A
0.3 A
4%
40 W
+3.3 V
0.8 A
6A
3%
+5 V
0.1 A
2A
5%
+12 V
0A
0.3 A
4%
30 W
TMC40-T31-3
Vnom.
Imin.
Maximum
Output
Output #3
Imax.
Tol.
Vnom.
Imin.
Imax.
Tol.
(N/A)
Power
40 W
TMC40-T31-5
+5 V
0.5 A
6A
3%
+3.3 V
0A
1.5 A
5%
+12 V
0A
0.3 A
4%
30 W
TMC40-T32
+5 V
0.5 A
6A
3%
+15 V
0.1 A
1.5 A
5%
-15 V
0A
0.3 A
4%
40 W
TMC40-T39
+5 V
0.5 A
6A
3%
+24 V
0.1 A
1A
5%
-12 V
0A
0.3 A
4%
40 W
NOTE: 1. Safety approvals are for PCB form only. To order unit with cover fitted, add suffix “C”.
MECHANICAL SPECIFICATIONS
NOTES:
1.
2.
3.
4.
5.
6.
7.
Dimensions shown in inches [mm]
Tolerance 0.02 [0.5] maximum
Input connector mates with Molex housing
09-50-3031 and Molex 2878 series crimp
terminal.
Output connector mates with Moles housing
09-50-3061 and Moles 2878 series crimp
terminal.
Ground tab is 0.25 [6.35] x 0.032 [0.8]
To ensure compliance with level B
emissions, connect the two “*” marked
mounting holes with metallic standoffs to
chassis.
Weight: 205 grams (0.45 lbs.) approx.
PIN CHART
PIN
1
2
3
4
5
6
+V1
+V1
RTN
RTN
N.C
N.C
+V1
+V1
RTN
RTN
N.C
+V2
+V1
+V1
RTN
RTN
-V3
+V2
+V1
+V1
RTN
RTN
+V3
+V2
MODEL
TMC40-S05
TMC40-S24
TMC40-S12
TMC40-S48
TMC40-D23
TMC40-D25
TMC40-T31
TMC40-T32
TMC40-T31-3 ,TMC40-T31-5
TMC40-S15
TMC40-T39
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com