TRUMPOWER TMP120-30

TRUMPower
120 WATT MEDICAL POWER ADAPTERS
D E S C R IP T IO N
TMP120 SERIES
The TMP120 series of AC/DC switching power
adapters are for 96 to 120 watts of continuous
output power. They are enclosed in a 94V-1 rated
case with an IEC 320/C14 or C18 inlet to mate with
interchangeable AC power cord for world-wide
use. All models meet CISPR 11 and FCC class B
emission limits, and are designed for medical
applications.
Class I, Class II
(LVD)
RoHS
F E AT U R E S
‧
‧
‧
‧
‧
‧
‧
Low safety ground leakage current
Optional output connectors
100% burn-in
9 0 t o 2 6 4 VAC w i d e i n p u t r a n g e
O v e r v o l ta g e a n d o v e r c u r r e n t p r o t e c t i o n s
Optional IEC 320/C18 inlet with class II insulation
C o m p l i a n t w i t h C E C a n d E n e r g y S ta r e f f i c i e n c y
l e v e l I V r e q u i r e m e n ts
‧ No load power consumption less than 0.5W
‧ Av e r a g e a c t i v e e f f i c i e n c y ≥ 8 5 %
‧ C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
S a f e t y Sta n d a r d A p p r o va l s :
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
I N P U T S P E C I F I C AT I O N S
GENERAL SPECIFICATIONS
I n p u t v o l ta g e :
Switching frequency:
Power factor:
Efficiency:
Hold-up time:
Line regulation:
Input frequency:
Input current:
90 to 264VAC
47 to 63Hz
1.60A (rms) for 115 VAC
0.80A (rms) for 230 VAC
90µA max. @ 115 VAC/60Hz
Earth leakage current:
150µA max. @ 230 VAC/50Hz
(Class I)
E n c l o s u r e l e a k a g e c u r r e n t : 90µA max. @ 264 VAC/50Hz
(Class II)
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
See rating chart
120 watts maximum
2% peak to peak maximum
Provided on output #1 only;
set at 112 -140% of its nominal
output voltage
All outputs protected to short
circuit conditions
All outputs ±0.04%/ºC maximum
Maximum excursion of 4% or
better on all models, recovering
to 1% of final value within 500µs
after a 25% step load change
E N V I R O N M E N TA L S P E C I F I C AT I O N S
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
50KHz ~ 110KHz
0.98 typical at 115VAC
85% min.
15 msec minimum at 115VAC
±0.5% maximum at full load
Inrush current:
60 amps @ 115VAC or 90 amps
@230VAC, at 25ºC cold start
Withstand voltage:
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
150,000 hours minimum at full load at
25ºC ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
E N 5 5 0 11 :
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
E N 6 1 0 0 0 - 4 - 11 :
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
0 ºC t o + 6 0 ºC
- 4 0 ºC t o + 8 5 ºC
5% to 95% non-condensing
Derate from 100% at +40ºC
linearly to 50% at +60ºC
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com
MEDICAL GRADE
TMP120 SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Output
Product No.
Vnom.
Imin.
Imax.
Tol.
Maximum
Output Power
TMP120-12
12V
0A
8.0A
5%
96W
TMP120-15
15V
0A
7.0A
5%
105W
TMP120-18
18V
0A
6.67A
5%
120W
TMP120-19
19V
0A
6.32A
5%
120W
TMP120-20
20V
0A
6.0A
5%
120W
TMP120-24
24V
0A
5.0A
5%
120W
TMP120-30
30V
0A
4.0A
5%
120W
TMP120-36
36V
0A
3.34A
5%
120W
TMP120-48
48V
0A
2.5A
5%
120W
NOTES:
1. Ripple and noise: Peak-to-peak with 20MHz bandwidth and 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor at
rated line voltage and load ranges.
2. To order a model with IEC320/C18 inlet and class II insulation, add “C2”, e.g. TMP120-12-C2.
MECHANICAL SPECIFICATIONS
NOTES:
1. Dimensions shown in inch [mm]
2. Tolerance 0.02 [0.5] maximum
3. Weight: 780 grams approx.
4. Standard output connector is a male 4-pin power DIN. Please consult TRUMPower for other connector options
5. TMP120-12 and TMP120-15 output connectors are without lock.
PIN CHART
PIN
Product No.
TMP120-12
TMP120-18
TMP120-20
TMP120-30
TMP120-48
TMP120-15
TMP120-19
TMP120-24
TMP120-36
1
2
3
4
Shell
Return
Output +V
Return
Output +V
AC Ground
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com