TRUMPOWER TMP60-T30

TRUMPower
30-60 WATT MEDICAL POWER ADAPTERS
TMP60 SERIES
DESCRIPTION
(LVD)
The PMP60 is a series of AC/DC switching power adapters
consists of 30 to 60 watts single, dual and triple outputs.
They are enclosed in a 94V-1 rated polyphenylene-oxide
case with an IEC 320/C14 inlet to mate with interchangeable
cord for world-wide use. All models meet CISPR 11 and
FCC class B emission limits, and are designed for medical
applications where low earth leakage current is required.
RoHS
F E AT U R E S
‧
‧
‧
‧
‧
‧
‧
‧
‧
Desktop form factor
Single, dual, or triple outputs
Optional output connectors
100% burn-in
W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
Input surge current protection
Overvoltage protection
Overcurrent protection
C o m p l i a n t w i t h C E C a n d E n e r g y St a r E f f i c i e n c y
level IV requirements
* No load power consumption less than 0.5W
* Av e r a g e a c t i v e e f f i c i e n c y ≧ 8 5 %
‧ Compliant with RoHS requirement
S a f e t y Sta n d a r d A p p r o va l s :
UL 60601-1
CSA C22.2 No. 601.1
I N P U T S P E C I F I C AT I O N S
Input voltage:
Input frequency:
Input current:
Leakage current:
8 5 t o 2 6 4 VA C
47 to 63 Hz
1 . 3 0 A ( r m s ) f o r 11 5 VA C
0 . 7 0 A ( r m s ) f o r 2 3 0 VA C
9 0 u A m a x . @ 11 5 VA C , 6 0 H z
1 5 0 u A m a x . @ 2 3 0 VA C , 5 0 H z
TÜV EN60601-1
GENERAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Switching frequency: 40KHz - 130KHz
Output voltage/current:
See rating chart
Efficiency:
Total output power:
See rating chart
Ripple and noise:
2﹪ peak to peak maximum
Hold-up time:
Line regulation:
Overvoltage protection:
Provided on output #1 only; set
at 112-140﹪ of its nominal
output voltage
Overcurrent protection:
All outputs protected to short
circuit conditions
Temperature coefficient:
All outputs ±0.04﹪ /℃
maximum
Transient response:
Maximum excursion of 4﹪ or
better on all models, recovering
to 1﹪ of final value within
500us after a 25﹪ step load
change
85﹪ minimum on single output models,
65﹪ minimum on the others
10 msec minimum at 110VAC
±0.5﹪ maximum at full load
Inrush current:
50 amps @ 115VAC, or 100 amps
@230VAC, at 25℃ cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
150,000 hours minimum at full load at
25℃ ambient , calculated per
MIL-HDBK-217F
EMC performance (EN60601-1-2)
E N 5 5 0 11 :
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion , Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD,± 8KV air and ± 6KV contact
EN61000-4-3:
Radiated immunity, 3V/m for 80~2500MHz
EN61000-4-4:
Fast transient/burst, ± 2KV
EN61000-4-5:
Surge, ± 1KV diff., ± 2KV com.
E N V I R O N M E N TA L S P E C I F I C AT I O N S
EN61000-4-6:
Conducted immunity, 3Vrms
Operating temperature: 0℃ to +60℃
St o r a g e t e m p e r a t u r e :
-40℃ to+85℃
5﹪ to 95﹪ non-condensing
Relative humidity:
Derate from 100﹪ at +40℃
Derating:
linearly to 50﹪at +60℃
EN61000-4-8:
Magnetic field immunity, 3A/m
E N 6 1 0 0 0 - 4 - 11 :
Voltage dips, 30﹪ reduction for 500ms,
60% reduction for 100ms and >95﹪
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com
TMP60 Series
Medical Grade
OUTPUT VOLTAGE/CURRENT RATING CHART
*Product No.
TMP60-Sxx
TMP60-Sxx
TMP60-Sxx
TMP60-Sxx
TMP60-Sxx
TMP60-Sxx
TMP60-Sxx
TMP60-D23
TMP60-T30
TMP60-T31
TMP60-T32
TMP60-T36
Notes:
Vnom.
11-13V
13-17V
17-21V
21-27V
27-33V
33-39V
46-50V
+5V
+3.3V
+5V
+5V
+5V
Output #1
Imin.
Imax.
0A
5.46A
0A
4.62A
0A
3.53A
0A
2.86A
0A
2.23A
0A
1.82A
0A
1.31A
1A
5.0A
1A
6.0A
1A
5.0A
1A
5.0A
1A
5.0A
Tol.
5%
5%
5%
5%
3%
3%
3%
5%
5%
5%
5%
5%
Vnom.
+12V
+5V
+12V
+15V
+24V
Output #2
Imin. Imax.
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
0.5A
3.0A
0.5A
3.0A
0.5A
3.0A
0.4A
2.3A
0.3A
1.5A
Tol.
Vnom.
5%
5%
5%
5%
5%
+12V
-12V
-15V
+12V
Output #3
Imin.
Imax.
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
0.1A
0.7A
0.1A
0.7A
0.1A
0.7A
0.1A
0.7A
Tol.
10%
10%
10%
10%
Maximum
Output
Power
60W
60W
60W
60W
60W
60W
60W
40W
30W
40W
40W
40W
(1) For single output models: Specify the output voltage xx when ordering, e.g. TMP60-S12 for 12V output, TMP60-S19.5 for 19.5V output.
(2) Ripple and noise: Peak-to-peak with 20MHz bandwidth and 10uF tantalum capacitor in parallel with a 0.1uF ceramic capacitor at rated line
voltage and load ranges
(3) The output voltages of a multiple output model may go outside of the stated tolerance when an output load current is out of stated limits. All
models may be operated at no-load without damage.
MECHANICAL SPECIFICATIONS
NOTES:
1. Dimensions show in inch
[mm]
2. Tolerance 0.02 [0.5]
maximum
3. Weight: 600 grams approx.
4. For other output connector
options, please contact
TRUMPower.
5. To order a model with on/off
switch, add suffix ”S” to the
model number, e.g.
TMP60-S30-S.
PIN CHART
1
2
3
4
5
TMP60-Sxx
RETURN
RETURN
OUTPUT
#1
RETURN
OUTPUT
#1
TMP60-D23
COMMOM
RETURN
COMMOM
RETURN
OUTPUT
#1
N.C.
OUTPUT
#2
COMMOM
RETURN
COMMOM
RETURN
OUTPUT
#1
OUTPUT
#3
OUTPUT
#2
P/N
PIN
TMP60-T30
TMP60-T36
TMP60-T31
TMP60-T33
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: [email protected] ● Website: www.TRUMPower.com