PMP90 - PROTEK POWER

60-90 WATT MEDICAL & ITE POWER SUPPLIES
DESCRIPTION
PMP90 SERIES
(LVD)
The PMP90 series of AC/DC switching power supplies are for
RoHS
60-90 watts of continuous output power. They are enclosed in
a 94V-1 rated polyphenylene-oxide case with an IEC320/C14
or IEC320/C18 inlet to mate with interchangeable cord for
world-wide use. All models meet EN55011, EN55022 and FCC
class B emission limits, and are designed for medical and ITE
applications, not for life-supporting equipment.
FEATURES
z
z
z
z
z
z
z
z
z
Low safety ground leakage current
Both Class I and Class II models are certified
to medical and ITE safety standards.
W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
Optional output connectors
100% burn-in
Overvoltage protection
Overcurrent protection
C o m p l i a n t w i t h C E C a n d E n e r g y St a r E f f i c i e n c y
level IV requirements (except models
P M P 9 0 - 1 0 , P M P 9 0 - 1 0 - 1 a n d P M P 9 0 - 11 )
* No load power consumption less than 0.5 W
* Average active efficiency greater than 85%
Compliant with RoHS requirements
SAFETY STANDARD APPROVALS
UL 60601-1, CSA C22.2 No. 601.1
File No. E178020
TÜV EN 60601-1
UL 60950-1, CSA C22.2 No. 60950-1
File No. E137410
TÜV EN 60950-1
GENERAL SPECIFICATIONS
INPUT SPECIFICATIONS
Switching frequency:
50-110 KHz
Input voltage:
90-264 VAC
Power factor:
0.98 Typical
Input frequency:
47-63 Hz
Efficiency:
85% min. (except 77% min. for PMP90-10,
Input current:
1.20 A (rms) for 115 VAC
PMP90-10-1, PMP90-11 and PMP90-12)
0.60 A (rms) for 230 VAC
Hold-up time:
15 ms minimum at 110 VAC
Earth leakage current:
180 μA max. @ 264 VAC, 63 Hz
Line regulation:
±0.5% maximum at full load
Touch current:
100 μA max. @ 264 VAC, 63 Hz
Inrush current:
50 A @ 115 VAC or 100 A @ 230 VAC, at
Withstand voltage:
4000 VAC from input to output,
25℃ cold start
OUTPUT SPECIFICATIONS
Output voltage /current:
Maximum output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
1500 VAC from input to ground,
See rating chart.
See rating chart.
100 mVp-p maximum on 5 V, 6 V and
9 V outputs, 1% maximum on other
voltage outputs (12 V, 13.5 V …, 48 V
etc.)
Provided and set at 112-140% of its
nominal output voltage
Protected to short circuit conditions
500 VAC from output to ground
MTBF:
EMC Performance (IEC60601-1-2)
EN55011/EN55022:
±0.04% /℃ maximum
Maximum excursion of 4% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step
load change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
150,000 hours at full load at 25℃ ambient ,
calculated per MIL-HDBK-217F
0℃ to +60℃
-40℃ to +85℃
5% to 95% non-condensing
Derate from 100% at +40℃
linearly to 50% at +60℃
Class B conducted, class B radiated
FCC:
Class B conducted, class B radiated
VCCI:
Class B conducted, class B radiated
EN61000-3-2:
Harmonic distortion, class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8 KV air and ±6 KV contact
EN61000-4-3:
Radiated immunity, 3 V/m
EN61000-4-4:
Fast transient/burst, ±2 KV
EN61000-4-5:
Surge, ±1 KV diff., ±2 KV com.
EN61000-4-6:
Conducted immunity, 3 Vrms
EN61000-4-8:
Magnetic field immunity, 3 A/m
EN61000-4-11:
Voltage dip immunity, 30% reduction for 500
ms, 60% reduction for 100 ms, and >95%
reduction for 10 ms
49
UNIVERSAL INPUT
PMP90 MEDICAL & ITE SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
Model
Class I
PMP90-10
PMP90-10-1
PMP90-11
PMP90-12
PMP90-12-1
PMP90-13
PMP90-13-1
PMP90-13-2
PMP90-14
PMP90-16
PMP90-18
(1)
Class II
PMP90F-10
PMP90F-10-1
PMP90F-11
PMP90F-12
PMP90F-12-1
PMP90F-13
PMP90F-13-1
PMP90F-13-2
PMP90F-14
PMP90F-16
PMP90F-18
V1
5.0 V
6.0 V
9.0 V
12.0 V
13.5 V
15.0 V
18.0 V
19.0 V
24.0 V
30.0 V
48.0 V
Min.
Current
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
Output
Max.
Current
12.0 A
10.0 A
7.78 A
6.67 A
6.30 A
5.67 A
5.00 A
4.74 A
3.75 A
3.00 A
1.87 A
Tol.
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Average Active
Efficiency (typical)
@ 115 / 230 Vac
Max. Power
60 W
78 / 79 %
60 W
78 / 79 %
70 W
81 / 82 %
80 W
86 / 85%
85 W
86 / 85%
85 W
86 / 85 %
90 W
86 / 85 %
90 W
86 / 86 %
90 W
87 / 86 %
90 W
87 / 86 %
90 W
88 / 87 %
Ripple &
(2)
Noise
100 mV
100 mV
100 mV
120 mV
135 mV
150 mV
180 mV
190 mV
240 mV
300 mV
480 mV
NOTES:
1.
Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet.
2.
Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load
ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.
MECHANICAL SPECIFICATIONS
OUTPUT POWER DERATING CURVE
NOTES:
1.
Dimensions shown in inches [mm]
2.
Tolerance 0.02 [0.5] maximum
3.
Weight: 658 grams (1.45 lbs.) approx.
4.
Output connector is 4 pin plug without lock, mating with Kycon P/N KPJX-4S-S socket or equivalent.
5.
Refer to Section titled “OPTIONAL OUTPUT CONNECTORS”. Add the suffix assigned for a selected connector to a
wanted model number, e.g. PMP90-14-B1, for ordering.
PIN CHART
MODEL
PMP90-10
PMP90-10-1
PMP90-11
PMP90-12
PMP90-12-1
PMP90-13
PMP90-13-1
PMP90-13-2
PMP90-14
PMP90-16
PMP90-18
PIN
PMP90F-10
PMP90F-10-1
PMP90F-11
PMP90F-12
PMP90F-12-1
PMP90F-13
PMP90F-13-1
PMP90F-13-2
PMP90F-14
PMP90F-16
PMP90F-18
1
2
3
4
V1 Return
+V1
V1 Return
+V1
50