Medical Open Frame 65 Watts THE XPERTS IN POWER PBM65 Series 4000 VAC Isolation • • Low Leakage Current • Meets Class B, EN55011 Conducted • Medical Safety Agency Approvals • Both Line and Neutral Fused • Fixed Switching Frequency Specification Input General Input Voltage Input Frequency Input Current • • • Inrush Current • Earth Leakage Current • 85-264 VAC 47-63 Hz 1.6 A max at 115 VAC 1.0 A max at 230 VAC 17 A max at 115 VAC 40 A max at 230 VAC 90 µA max at 115 VAC, 60 Hz 150 µA max at 230 VAC, 50 Hz Output Output Voltage Output Voltage Adjustment Output Power Minimum Load Line Regulation Load Regulation Ripple and Noise Transient Response Temperature Coefficient Hold Up Time Overvoltage Protection Overcurrent Protection Short Circuit Protection Remote Sense • • 5.1 to 28 VDC ±5% • • • • • • • 65 Watts No minimum load required ±0.5% max at full load ±2% typical ±1% pk-pk max (See Note 2) 4% maximum deviation, 500 µs recovery time to within 1% for a 25% load change ±0.04%/°C • • 10 ms minimum at 110 VAC 112-132% with auto recovery • 125-200% with auto recovery • Trip & restart with auto recovery • Compensates for up to 0.5 V drop Efficiency • MTBF Switching Frequency Isolation Voltage • • • Size • Weight • 66% minimum for 5 V models 75% minimum for all others 400,000 hrs min MIL-HDBK-217F 42 kHz ±5 kHz 4000 VAC Input to Output 1500 VAC Input to Output 500 VAC Output to Ground 5.0 x 3.0 x 1.2” 127.00 x 76.20 x 30.48 mm 330 g approx (PCB format) Environmental Operating Temperature • Cooling Relative Humidity Storage Temperature • • • 0 °C to +70 °C Derate linearly from 100% load at +50 °C to 50% load at +70 °C Convection cooled 5-95% non-condensing -40 °C to +85 °C EMC & Safety Safety Approvals • EMI/EMC • ESD • Radiated Immunity EFT/Burst Surge • • • UL2601-1, CSA C22.2 No 601.1, EN60601-1 CE Mark (LVD) EN55011 (CISPR11) Class B and FCC Part 18 Level B radiated & conducted EN61000-4-2, Level 3 air, Level 2 contact, Perf Criteria A EN61000-4-3, 3 V/m, Perf Criteria A EN61000-4-4, Level 3, Perf Criteria A EN61000-4-5, Level 3, Perf Criteria A • • www.xpplc.com [email protected] Maximum Output Power 51 W 65 W 65 W 65 W 65 W 65 W Output Voltage 5.1 VDC 12.0 VDC 15.0 VDC 18.0 VDC 24.0 VDC 28.0 VDC PBM65 Output Current 10.0 A 5.5 A 4.4 A 3.7 A 2.8 A 2.4 A Total Regulation(1) 3% 2% 2% 2% 2% 2% Model Number PBM65US05 PBM65US12 PBM65US15 PBM65US18 PBM65US24 PBM65US28 Notes • 1. Total regulation includes line and load regulation and set point accuracy. 2. Ripple and noise measured with 20 MHz bandwidth with 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor. AC-DC Power Supplies OUTPUT VOLTAGE & CURRENT RATINGS Mechanical Details Medical Open Frame 65 Watts PBM65 Series P3 PIN CONNECTIONS Pin Function 1 Output 1 2 Output 1 3 Output 1 4 Return 5 Return 6 Return NOTES: 1. Dimensions shown in inches (mm). 2. Tolerance 0.02 (0.5) maximum. 3. Input connector mates with molex housing 09-50-3031 and Molex 2878 series crimp terminal. 4. Output connector mates with molex housing 09-50-3061 and Molex 2878 series crimp terminal. 5. P4 is for sense connector mates with molex housing 22-01-1023 and molex 40445 series crimp terminal. 6. Loom kits with 300 mm wire available, order part number PBM65 LOOM. 7. For mating connectors only, order part number PBM65 CONKT. 8. Covers available, order part number PBM65 COVER. 9. Weight: 330 g (PCB format). www.xpplc.com [email protected] Rev 3.0