LEM LA25-SP11

Current Transducer LA 25-NP/SP11
IPN
=
1 A
For the electronic measurement of currents : DC, AC, pulsed,
mixed, with a galvanic isolation between the primary circuit
(high power) and the secondary circuit (electronic circuit).
16095
Electrical data
IPN
IP
RM
Primary nominal r.m.s. current
Primary current, measuring range
Measuring resistance
with ± 15 V
ISN
KN
VC
IC
Vd
Vb
@ ± 1.0 A max
@ ± 1.5 A max
Secondary nominal r.m.s. current
Conversion ratio
Supply voltage (± 5 %)
Current consumption
R.m.s. voltage for AC isolation test, 50 Hz, 1 mn
R.m.s. rated voltage 1), safe separation
basic isolation
1
0 .. ± 1.5
R M min
R M max
A
A
100
100
Ω
Ω
320
190
25
25 : 1000
± 15
10 + IS
2.5
600
1700
• Closed loop (compensated) multi-
mA
V
mA
kV
V
V
Typical accuracy @ IPN , TA = 25°C
Linearity
± 0.5
< 0.2
IO
IOM
IOT
Offset current 2) @ IP = 0, TA = 25°C
Residual current 3) @ IP = 0, after an overload of 3 x IPN
Thermal drift of IO
0°C .. + 25°C
+ 25°C .. + 70°C
Typ
± 0.05
± 0.05
± 0.06
± 0.10
tr
f
Response time 4) @ 90 % of IP max
Frequency bandwidth (- 1 dB)
<1
DC .. 150
ε
L
turns current transducer using the
Hall effect
• Insulated plastic case recognized
according to UL 94-V0.
Special features
• IPN = 1 A
• IP = 0 .. ± 1.5 A
• K N = 25 : 1000.
Advantages
Accuracy - Dynamic performance data
X
Features
%
%
Max
± 0.15
± 0.15
± 0.25
± 0.35
mA
mA
mA
mA
µs
kHz
•
•
•
•
•
•
•
Excellent accuracy
Very good linearity
Low temperature drift
Optimized response time
Wide frequency bandwidth
No insertion losses
High immunity to external
interference
• Current overload capability.
Applications
General data
TA
TS
RP
RS
LP
R IS
m
Notes :
• AC variable speed drives and servo
Ambient operating temperature
Ambient storage temperature
Primary coil resistance @ TA = 25°C
Secondary coil resistance @ TA = 70°C
Primary insertion inductance
Isolation resistance @ 500 V, TA = 25°C
Mass
Standards
1)
2)
3)
4)
0 .. + 70
°C
- 25 .. + 85
°C
< 51
mΩ
110
Ω
31
µH
> 1500
MΩ
22
g
EN 50178 : 1997
Pollution class 2
Measurement carried out after 15 mn functioning
The result of the coercive field of the magnetic circuit
With a di/dt of 100 A/µs.
motor drives
• Static converters for DC motor drives
• Battery supplied applications
• Uninterruptible Power Supplies
(UPS)
• Switched Mode Power Supplies
(SMPS)
• Power supplies for welding
applications.
060405/2
LEM reserves the right to carry out modifications on its transducers, in order to improve them, without previous notice.
LEM
Page 1/2
w w w .lem.com
Dimensions LA 25-NP/SP11
Bottom view
(in mm. 1 mm = 0.0394 inch)
Right view
Top view
swiss
made
Standard 00
or N° SP..
Year
Week
Secondary terminals
Terminal + : supply voltage + 15 V
Terminal M : measure
Terminal - : supply voltage - 15 V
Connection
in
out
Back view
Mechanical characteristics
Remark
• General tolerance
• Fastening & connection of primary
• IS is positive when IP flows from terminal + to terminal -.
± 0.2 mm
2 pins
0.635 x 0.635 mm
• Fastening & connection of secondary 3 pins ∅ 1 mm
• Recommended PCB hole
1.2 mm
060405/2
LEM reserves the right to carry out modifications on its transducers, in order to improve them, without previous notice.
LEM
Page 2/2
w w w .lem.com