DATA SHEET ABC40 SERIES 40W AC/DC FEATURES 40 W convection cooled -20 to 50 deg C full load operation 90-264 VAC input 2” x 4 “ x 1.2” (101.6 x 50.8 x 30.48 mm) No minimum load required - single output models Conducted EMI EN 55022-B, FCC Part 15 Level B ITE Safety Agency Approvals RoHS Compliant No load power < 0.3 W APPLICATIONS Lighting Automation Control Industrial Processing Robotics Applied Computing Wireless Communications Instrumentation Test and Measurement BCD.00281_AA_September-18-2013 1 www.power-one.com DATA SHEET TECHNICAL DATA: Input PARAMETER DESCRIPTION/CONDITION Input voltage range Input frequency range Input surge current Safety ground leakage current Universal Input 47 to 400 Hz 230 Vac (cold start) 230 Vac 120 Vac @ 40 W 230 Vac @ 40 W Input current: 90 - 264 Vac Note 3 60 A max. 1000 μA max 0.85 A rms 0.45 A rms Output PARAMETER DESCRIPTION/CONDITION Voltage Adjustment V1 Main output 50 to 100% load change, 50/60 Hz, 50% duty cycle, 0.1A / uSec 130% Typical 130% Typical V1 >85% V1: ± 3%, V2 & V3: ± 5% < 100 mSec Transient Response Over Voltage Protection Short Circuit Protection Remote Sense Efficiency Set Point Tolerance Rise Time ± 10% < 10%, recovery time < 5 mS V1 only 0.5 Vdc compensation Ordering Information PRODUCT FAMILY VOLTS (VDC) OUTPUT CURRENT MAX (A) MINIMUM LOAD (A) (4) RIPPLE & NOISE TOTAL REGULATION OVP THRESHOLD ABC40-1005G 5.1 8.0 0 1% ± 0.8% 130% Typical ABC40-1012G 12 3.5 0 1% ± 0.8% 130% Typical ABC40-1015G 15 2.7 0 1% ± 0.8% 130% Typical ABC40-1024G 24 1.7 0 1% ± 0.8% 130% Typical ABC40-1048G 48 0.83 0 1% ± 0.8% 130% Typical 5.2 6.0 0.5 1% ± 0.8% 130% Typical 12.5 2.0 0.1 1% ± 5.3% -12.8 0.5 0.0 1% ± 5.3% 5.2 6.0 0.5 1% ± 0.8% 23.8 1.0 0.1 1% ± 5.3% -12.8 0.5 0.0 1% ± 5.3% ABC40-3000G ABC40-3001G ABC40-3002G ABC40-3003G ABC40 Series 5.2 6.0 0.5 1% ± 0.8% 14.6 1.5 0.1 1% ± 5.3% -14.8 0.5 0.0 1% ± 5.3% 3.3 6.0 1.0 1.5% ± 0.8% 5.2 3.0 0.1 1% ± 5.3% -12.8 0.5 0.0 1% ± 5.3% 2 130% Typical 130% Typical 130% Typical www.power-one.com DATA SHEET Notes: 1. 2. 3. 4. 5. 6. 7. 8. 9. Maximum outputs for each output. Max power rating should not be exceeded. Output noise measurement is made with a 20 MHz bandwidth using a 6” twisted pair, terminated with a 10 uF tantalum capacitor in parallel with a 0.1 uF ceramic capacitor. Safety Approved: 47 to 63 Hz. Minimum load specified to meet cross regulation. Specifications are for nominal input voltage, 25° C and max load unless otherwise stated. Class 1 models have Earthing tab J4. Class 2 models (-2 suffix) have no Earthing tab. Ripple and Noise may be 3% at initial start-up at -20 C. Specifications subject to change without notice. Warranty 2 years. General Specifications PARAMETER DESCRIPTION/CONDITION Hold Up Time MTBF Switching Frequency Isolation Voltage Weight 6 mSec >100 khrs 67 kHz Typical Input to Output: Min 4242 Vdc 150 g (0.33 lbs) 120 / 230 Vac input MIL-HDBK-217F Input to Ground: Min 2120 Vdc Environmental PARAMETER Operating Temperature Humidity Altitude Conducted Emissions: Harmonic current correction: Radiated Emissions Static Discharge Conducted immunity: DESCRIPTION/CONDITION Operating Storage 95% Operating 10,000 ft. EN55022, FCC part 15- B EN61000-3-2, Class A EN55011-B, FCC part 15-B EN61000-4-2 Level 3 EN61000-4-6 -20 to 70° C. See derating chart below. -40 to +85° C. Non Condensing Non-operating 40,000 ft To be controlled in end system 2, 3, 4, 5 Level 3 Safety PARAMETER DESCRIPTION/CONDITION EN / UL / CSA EN60950-1+A12:2011, IEC60950-1 2nd+A1 2009, CSA-22.2 No 60950-01-07+ A1, UL60950-1-2011 BCD.00281_AA_September-18-2013 3 www.power-one.com DATA SHEET Figure 1 Output Power Vs. Temperature Figure 2 Dimension Drawing (Top and Side View) Mechanical INPUT = J1 EARTHING TAB = J4 DC OUTPUT = J2 Pin 1: AC Neutral Pin 2: AC Line Molex: 19705-4301 Pin 1 = V1 Pin 2 = V1 Pin 3 = RTN Pin 4 = RTN Pin 5 = V3 Pin 6 = V2 Mating Connector: Molex: 09-50-3031 Pins: 08-50-0106 Mating Connector: Molex: 190030001 Mating Connector: Tyco: 647402-6; Pins: 3-647409 REMOTE SENSE= J3 Pin 1 = +V1 Sense Pin 2 = -V1 Sense Mating Connector: Molex: 22-01-2021 Copyright © 2010 Power-One Inc. All rights reserved. Words and logos that are identified as trademarks and/or service marks are, unless noted otherwise, the trademarks and service marks of Power-One Inc. in the U.S. and other countries. All other product or service names are the property of their respective holders. Power-One products are protected under numerous U.S. and foreign patents and pending applications, maskwork rights, and copyrights. Power-One reserves the right to make changes to any products and services at any time without notice. Power-One assumes no responsibility or liability arising out of the application or use of any information, product, or service described herein except as expressly agreed to in writing by Power-One Inc. NUCLEAR AND MEDICAL APPLICATIONS - Power-One products are not designed, intended for use in, or authorized for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems without the express written consent of the respective divisional president of Power-One, Inc. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice. ABC40 Series 4 www.power-one.com