SLPOWER GNT30

GNT30 MEDICAL SERIES
SINGLE OUTPUT POWER SUPPLY
SUMMARY
Low Power Medical Power Supplies 30 Watt AC/DC Universal Input Power Supply
• 90-264 Vac input range
• Overvoltage and short circuit protection
• Approved to UL, CSA and EN Standards
• Small 50 mm x 80 mm footprint
• Class II input available (G2T30)
The GNT30 Series is a 30 W universal input AC/DC Power
supply in a very small footprint. With the medical approvals
the GNT30 is ideal for a variety of medical device applications
including, small single board computers, battery charging, and
running small motors, pumps, and solenoids. The series with full
approval to EN60601-1 Standard improves design-in time, and
reduces end system compliance costs. SPECIFICATIONS
All specifications are typical at nominal input, full load at 25 OC unless otherwise stated
OUPUT SPECIFICATIONS
Output Power
INPUT SPECIFICATIONS
150 l.f.m. airflow
30 Watts max.(See note 4
Input Voltage range
Universal input
100-240 nom (90 - 264 max)
Vac
Total Regulation
See table
Turn-on time
1.5 s max.
Input frequency
55 Hz +/- 10 Hz
3.5% max. dev. 1 ms recovery
to 1%
Input surge current
32 A max.
Ground Leakage current
132 Vac 60 Hz
100 mA (180 mA single fault)
Temperature Coefficient
+/-0.03%/oC
Input current
120 Vac
0.6 A
Overvoltage protection
125% +/-10%
230 Vac
0.37 A
F1,F2
2.5 A
Transient response
Short Circuit protection
0.25 A/ms 50% load
deviation
Power cycling
Yes
EMC INFORMATION
Input fuse
ENVIRONMENTAL SPECIFICATIONS
Conducted emissions
EN55011
Level B
Radiated emissions
EN55011
Level A
Line freq. harmonics
EN61000-3-2
Complies
Voltage fluctuations
EN61000-3-3
Complies
ESD Air
EN61000-4-2
8 kV contact
ESD Contact
EN61000-4-2
6 kV air
Radiated immunity
EN61000-4-3
3 V/m
Fast Transients (EFT)
EN61000-4-4
Thermal Performance
operating ambient
(see chart)
0-50 oC
non-operating
-40 to +85 oC
0 - 50 oC
150 l.f.m (See note 4.)
30 Watts
50 -70 oC ambient
150 l.f.m
derate 2% /oC
Relative Humidity
non-condensing
0% - 95% RH
2 kV
Maximum Altitude
operating / non-operating
10,000 ft. / 40,000 ft. max.
Operating
Non-operating
1.5 grms 20-2000 Hz
5 grms 20-2000 Hz
Operating
Non-operating
20 gpk 10 ms, half-sine
40 gpk 10 ms half-sine
Line surge immunity
EN61000-4-5
1 kV diff /2 kV cm
Vibration
Conducted immunity
EN61000-4-6
3 Vrms
Power freq. mag. field
EN61000-4-8
3 A/m
Shock
Voltage dip immunity
EN61000-4-11
5 cycles 40 % vnom 0.5 Watts.
25 cycles 70 % vnom 20 W
GENERAL SPECIFICATIONS
Hold-up time
Efficiency
isolation voltage
120 Vac, 60 Hz
120 Vac 30 W output
input to ouput
input to ground
fixed
Switching frequency
Safety Approvals
Maximum weight
16 ms at 30 Watts output
>78 % 4000 Vac
1500 Vac
120 KHz, +/- 5 KHz
UL/EN/IEC 60601-1 CSA22.2 No.601
UL/EN/IEC 60950-1 CSA22.2 No.60950-1
100 g (0.22 lbs)
MODEL NUMBER 40oC Convection 50oC Convection
50oC Airflow
RIPPLE
REGULATION
GNT30-5
GNT30-12
GNT30-15
GNT30-24
GNT30-28
GNT30-48
5 V 6.0 A
12 V 2.5 A
15 V 2.0 A
24 V 1.25 A
28 V 1.07 A
48 V 0.625 A
50 mV
120 mV
150 mV
240 mV
280 mV
480 mV
2%
2%
2%
2%
2%
2%
5 V 4.0 A
12 V 2.5 A
15 V 2.0 A
24 V 1.25 A
28 V 1.07 A
48 V 0.625 A
5 V 3.2 A
12 V 1.66 A
15 V 1.06 A
24 V 0.83 A
28 V 0.8 A
48 V 0.416 A
NOTES
1. When the input voltage is less than 90 Vac the operating temperature range is 0oC to 40oC. The ripple and regulation
specs. may not be met.
2. Peak output current rating is for 1 minute and 10 % duty cycle.
3. Heatsink temperatures should not be allowed to exceed 90 oC
4. Output Voltage is adjustable +/- 5% of nominal. See installation instructions (www.condorpower.com)
MECHANICAL NOTES
INPUT
J1 AMP P/N 640456-3 TB1
PIN 1 AC LINE
PIN 1 AC LINE
PIN 2 N/C
PIN 2 AC NEUTRAL
PIN 3 AC NEUTRAL
0.187 Faston
GROUND
GROUND
OUTPUT J2 AMP P/N 640456-4 TB2
PIN 1 COMMON
PIN 1 COMMON
PIN 2 COMMON
PIN 2 OUTPUT 1
PIN 3 OUTPUT 1
PIN 4 OUTPUT 1
MATING CONNECTORS
CONTACT
INPUT
HOUSING 1375820-3
1375819-1
OUTPUT
HOUSING 1375820-4
1375819-1
Condor DC Power Supplies Inc. 2311 Statham Parkway, Oxnard CA, 93033, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 Email: [email protected]
GNT30 Rev H 11/11/05.
Data Sheet © 2005 Condor DC Power Supplies Inc. The information and specifications contained in this data sheet are believed to be correct at time of publication. However, Condor accepts no
responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.