GNT30 MEDICAL SERIES SINGLE OUTPUT POWER SUPPLY SUMMARY Low Power Medical Power Supplies 30 Watt AC/DC Universal Input Power Supply • 90-264 Vac input range • Overvoltage and short circuit protection • Approved to UL, CSA and EN Standards • Small 50 mm x 80 mm footprint • Class II input available (G2T30) The GNT30 Series is a 30 W universal input AC/DC Power supply in a very small footprint. With the medical approvals the GNT30 is ideal for a variety of medical device applications including, small single board computers, battery charging, and running small motors, pumps, and solenoids. The series with full approval to EN60601-1 Standard improves design-in time, and reduces end system compliance costs. SPECIFICATIONS All specifications are typical at nominal input, full load at 25 OC unless otherwise stated OUPUT SPECIFICATIONS Output Power INPUT SPECIFICATIONS 150 l.f.m. airflow 30 Watts max.(See note 4 Input Voltage range Universal input 100-240 nom (90 - 264 max) Vac Total Regulation See table Turn-on time 1.5 s max. Input frequency 55 Hz +/- 10 Hz 3.5% max. dev. 1 ms recovery to 1% Input surge current 32 A max. Ground Leakage current 132 Vac 60 Hz 100 mA (180 mA single fault) Temperature Coefficient +/-0.03%/oC Input current 120 Vac 0.6 A Overvoltage protection 125% +/-10% 230 Vac 0.37 A F1,F2 2.5 A Transient response Short Circuit protection 0.25 A/ms 50% load deviation Power cycling Yes EMC INFORMATION Input fuse ENVIRONMENTAL SPECIFICATIONS Conducted emissions EN55011 Level B Radiated emissions EN55011 Level A Line freq. harmonics EN61000-3-2 Complies Voltage fluctuations EN61000-3-3 Complies ESD Air EN61000-4-2 8 kV contact ESD Contact EN61000-4-2 6 kV air Radiated immunity EN61000-4-3 3 V/m Fast Transients (EFT) EN61000-4-4 Thermal Performance operating ambient (see chart) 0-50 oC non-operating -40 to +85 oC 0 - 50 oC 150 l.f.m (See note 4.) 30 Watts 50 -70 oC ambient 150 l.f.m derate 2% /oC Relative Humidity non-condensing 0% - 95% RH 2 kV Maximum Altitude operating / non-operating 10,000 ft. / 40,000 ft. max. Operating Non-operating 1.5 grms 20-2000 Hz 5 grms 20-2000 Hz Operating Non-operating 20 gpk 10 ms, half-sine 40 gpk 10 ms half-sine Line surge immunity EN61000-4-5 1 kV diff /2 kV cm Vibration Conducted immunity EN61000-4-6 3 Vrms Power freq. mag. field EN61000-4-8 3 A/m Shock Voltage dip immunity EN61000-4-11 5 cycles 40 % vnom 0.5 Watts. 25 cycles 70 % vnom 20 W GENERAL SPECIFICATIONS Hold-up time Efficiency isolation voltage 120 Vac, 60 Hz 120 Vac 30 W output input to ouput input to ground fixed Switching frequency Safety Approvals Maximum weight 16 ms at 30 Watts output >78 % 4000 Vac 1500 Vac 120 KHz, +/- 5 KHz UL/EN/IEC 60601-1 CSA22.2 No.601 UL/EN/IEC 60950-1 CSA22.2 No.60950-1 100 g (0.22 lbs) MODEL NUMBER 40oC Convection 50oC Convection 50oC Airflow RIPPLE REGULATION GNT30-5 GNT30-12 GNT30-15 GNT30-24 GNT30-28 GNT30-48 5 V 6.0 A 12 V 2.5 A 15 V 2.0 A 24 V 1.25 A 28 V 1.07 A 48 V 0.625 A 50 mV 120 mV 150 mV 240 mV 280 mV 480 mV 2% 2% 2% 2% 2% 2% 5 V 4.0 A 12 V 2.5 A 15 V 2.0 A 24 V 1.25 A 28 V 1.07 A 48 V 0.625 A 5 V 3.2 A 12 V 1.66 A 15 V 1.06 A 24 V 0.83 A 28 V 0.8 A 48 V 0.416 A NOTES 1. When the input voltage is less than 90 Vac the operating temperature range is 0oC to 40oC. The ripple and regulation specs. may not be met. 2. Peak output current rating is for 1 minute and 10 % duty cycle. 3. Heatsink temperatures should not be allowed to exceed 90 oC 4. Output Voltage is adjustable +/- 5% of nominal. See installation instructions (www.condorpower.com) MECHANICAL NOTES INPUT J1 AMP P/N 640456-3 TB1 PIN 1 AC LINE PIN 1 AC LINE PIN 2 N/C PIN 2 AC NEUTRAL PIN 3 AC NEUTRAL 0.187 Faston GROUND GROUND OUTPUT J2 AMP P/N 640456-4 TB2 PIN 1 COMMON PIN 1 COMMON PIN 2 COMMON PIN 2 OUTPUT 1 PIN 3 OUTPUT 1 PIN 4 OUTPUT 1 MATING CONNECTORS CONTACT INPUT HOUSING 1375820-3 1375819-1 OUTPUT HOUSING 1375820-4 1375819-1 Condor DC Power Supplies Inc. 2311 Statham Parkway, Oxnard CA, 93033, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 Email: [email protected] GNT30 Rev H 11/11/05. Data Sheet © 2005 Condor DC Power Supplies Inc. The information and specifications contained in this data sheet are believed to be correct at time of publication. However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.