SLPOWER GSM15

GSM15 Medical
15 Watt Global Performance Medical Switcher
GLOBAL PERFORMANCE SWITCHERS
FEATURES:
DC Power Supplies
• Industry’s smallest 15 W medically approved switcher
• Compact size (3.00” x 2.10” x 0.92”)
• Wide-range ac input: 90-264 Vac
• Less than 75 μA leakage current @ 120 Vac
• Approved to UL2601-1, EN60601-1
• EMI to FCC, CISPR 11 Class B
• Overvoltage protection standard
• RoHS compliant models available (G suffix)
•
marked to LVD
SPECIFICATIONS
Ac Input
90-264 Vac, 47-63 Hz single phase. Class I or class II grounding.
Temperature Coefficient
0.03% / °C typical.
Input Current
Maximum input current at 90 Vac, 60 Hz with full rated output load not to
exceed 0.6 A.
EMI/EM Compliance
All models include built-in EMI filtering to meet the following EMC requirements of
IEC601-1-2.
Input Protection
Internal ac fuse provided on all units. Designed to blow only if a catastrophic
failure occurs in the unit -- Fuse does not blow on unsustained overload or
short circuit.
Performance Requirement
EMC Standard
Typical Margin
Conducted Emissions
EN55011, Class B; FCC Class B
2 dB Class II Gnd
6 dB Class I Gnd
Surge Discharge
EN61000-4-2, Level 3
2 kV
RF Field Susceptibility
EN61000-4-3, Level 3
2V
Fast Transients/Bursts
EN61000-4-4, Level 3
500 V
Efficiency
69-85% depending on model.
Surge Susceptibility
EN61000-4-5, Level 3
500 V
Overload Protection
Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit. Factory set to begin power limiting at 23 W.
Conducted RF Susceptibility
EN61000-4-6
25%
Voltage Sags & Surges
EN61000-4-11
5%
Inrush Current
Inrush is limited by internal thermistors. The inrush at 240 Vac, averaged over
the first ac half-cycle under cold start conditions will not exceed 37 A.
Overvoltage Protection
Built in OVP on all models. Approximately 120-140% of output voltage.
Output Noise
0.5% rms, 1% Pk-Pk, 20 MHz Bandwidth, differential mode. Measured with
noise probe directly across output terminals of the power supply.
Medical Safety Approvals
All models are Certified to be in compliance with the applicable requirements of
UL2601-1, IEC60601-1, CSA-C22.2 No. 601-1, EN60601-1.
Leakage Current
The maximum leakage current for GSM15 series will be as follows;
132Vac/60Hz UL2601-1 test method
Transient Response
Main Output - 500 µs max. response time for return to within 0.5% of final
value for a 50% load step change, ∆i/ ∆t< 0.2 A/μs. Maximum voltage deviation is 3.5%.
Hold-up Time
10 ms minimum from loss of ac input voltage at full load, nominal line (120
Vac).
GND
Connection
Normal
Single Fault
Class I
75 μA
105 μA
Class II
39 μA
54 μA
GND
Connection
Normal
Single Fault
Class I
128 μA
180 μA
Class II
66 μA
94 μA
264Vac/50Hz IEC60601-1 test method
All specifications are typical at nominal input, full load at 25OC unless otherwise stated
Medical
Model
Voltage
Output
Min.
Normal (A)
Peak (B)
Initial Set
Point
OVP Setpoint
Total
Regulation
Ripple and
Noise
GSM15-5
5.1 V
0A
2.35 A
3A
2.5%
7.2 V
2%
1%
GSM15-12
12 V
0A
1.25 A
1.5 A
2.5%
16 V
2%
1%
GSM15-15
15 V
0A
1.0 A
1.2 A
2.5%
21 V
2%
1%
GSM15-24
24 V
0A
0.625 A
0.75 A
2.5%
32 V
2%
1%
GSM15-28
28 V
0A
0.54 A
0.64 A
2.5%
280 V
2%
1%
Notes:
A. Rating with unrestricted convection cooling.
B. Peak Power for 60 sec. 10% duty cycle or continuous rating with 150 LFM of airflow.
C. Output voltages preset at factory, not user adjustable.
D. Add “G” suffix to model nuber for RoHS compliant model.
GSM15 MECHANICAL SPECIFICATIONS
INPUT: J1 AMP P/N 640456-4
PIN 1) AC LINE
PIN 2) N/C
PIN 3) N/C
PIN 4) AC NEUTRAL
GND: 0.098 DIA. THRU HOLE
OUTPUT: J2 AMP P/N 640456-4
PIN 1) COMMON Return
PIN 2) COMMON Return
PIN 3) OUTPUT #1 + Vout
PIN 4) OUTPUT#1 +Vout
MATING CONNECTOR AMP P/N
MTA – 100 Recepticle
NOTE: 3A MAXIMUM RECOMMENDED
CURRENT PER CONNECTOR PIN
Overall Dimensions:
3.00 x 2.10 x .92 inches
76.20mm x 53.34mm x 23.37mm
Weight: 0.25 LBS. [.113 kg]
MAX.
ENVIRONMENTAL SPECIFICATIONS
OPERATING
NON-OPERATING
Temperature (A)
0 to 50° C
-40 to +85°C
Humidity (A)
0 to 95% RH
0 to 95% RH
Shock (B)
20 gpk
40 gpk
Altitude
-500 to 10,000 ft
-500 to 40,000 ft
Vibration (C)
1.5 g rms’ 0.003 g2/Hz
5 g rms’ 0.026 g2/Hz
A. Units should be allowed to warm up/operate under non-condensing
conditions before application of power.
B. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3
orthogonal axes, total 6 shocks.
C. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1
hr./axis non-operating.
SL Power Electronics Corp., 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com Rev. 1/07.
Data Sheet © 2007 SL Power Electronics Corp. The information and specifications contained in this data sheet are believed to be correct at time of publication.
However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.