SLPOWER GPFC125H

GPFC125 Commercial/GPFM125 Medical
125 Watt Global Performance Switchers
FEATURES:
•
•
•
•
•
Compact 125 watt multiple output
Power density of 4 watts per cubic inch
Power Factor Correction to meet EN61000-3-2
Small package 6.00 x 3.5 x 1.5 inches
Conducted EMI exceeds FCC Class B and CISPR 22
Class B (Commercial models) and CISPR 11 Class B
(Medical models)
• Commercial approved to UL1950, IEC950, CSA22.2
No. 950 and EN60950.
• Medical approved to UL2601-1, IEC601-1, CSA22.2
No. 601.1, EN60601-1
•
marked to LVD
SPECIFICATIONS:
Ac Input
Remote Sense
90-264 Vac, 47-63 Hz single phase.
Provided as a standard feature on V1 of all models.
Input Current
Temperature Range
Maximum input current 2A at 90 Vac, 60 Hz with full rated load.
Input current harmonic content meets the requirements of
IEC1000-3-2. Active circuitry provides high power factor as
listed at .96 min., .99 typical. All data taken with full output
loading.
0 to 50 °C at full rated output power. For operation above 50 °C,
derate output power and current by 2.5 % per °C.
DC Output Power
125 W with 150 LFM of air, 50 W convection cooled, 80 W
convection cooled with unit mounted to min. 12 x 12 x .125 in.
thick aluminim plate. Peak ratings are for 60 s maximum
duration, 10% duty cycle. With optional cover installed,125 W,
200 LFM airflow is required(-C suffix).
Inrush Current
Inrush 240 Vac is less than 37 A, averaged over the first ac
half-cycle under cold start conditions. Limiting provided by
internal thermistors.
Input Protection
Internal ac fuse provided on all models. Fuse does not blow on
overload or short circuit — fuse blows only if a catastrophic
failure occurs in the unit.
Efficiency
72 to 85% at full rated load. Depending upon model and load
distribution.
Minimum Load
Supply will function with no load on any output. To maintain
regulation with full load on V2, 3 and 4 output V1 requires a
minimum load of (2A + V4 current). Lower minimum load may be
obtained if less then full load is drawn from auxiliary outputs.
Overload Protection
Fully protected against short circuit and output overload. Total
output power limited to approximately 150 Watts. Cycling type
limits on output 1, 2 & 3, linear fold back on output 4. Recovery
after fault is automatic.
Overvoltage Protection
Built in on all models - See Output table for individual model
output limits.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential mode.
Transient Response
500 ms typical response time for return to within 0.5% of final
value for a 50% load step change, ∆i/∆t< 0.2 A ms. Maximum
voltage deviation is 3.5%.
Temperature Coefficient
0.03% / °C typical on all outputs.
52
Altitude
Operating:
Non-Operating
-500 to 10,000 ft. MSL
-500 to 40,000 ft MSL
Shock and Vibration
All models are designed to meet the following specifications;
Random Vibration Operating: 0.003 g2/Hz, 1.5 grms overall, 3 axes, 10 min. / axis
Non-Operating: 0.026 g2/Hz, 5.0 grms overall, 3 axes, 1 hr. / axis
Shock Operating: Half-sine, 20 gpk, 10 ms, 3 axes, 6 shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3 axes, 6 shocks total
EMI/EMC Compliance
All models include built-in EMI filtering to meet the EMC
requirements below.
EMI SPECIFICATIONS
COMPLIANCE LEVEL
Conducted Emissions-GPFC125
Conducted Emissions-GPFM125
Static Discharge
RF Field Susceptibility
Fast Transients / Bursts
Surge Susceptibility
Conducted RF Susceptibility
Voltage Sags & Surges
Line Frequency Harmonics
EN55022, Class B; FCC Class B
EN55011,Class B; FCC Class B
EN61000-4-2, 6 kV contact 8 kV air
EN61000-4-3, 3V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
EN61000-4-6, 3V
EN61000-4-11
EN61000-3-2 Class A
Commercial Safety Approvals
All models are approved to UL1950, CSA22.2 No. 950-95,
IEC950, EN60950. CB certificate available. Exceeds FCC and
CISPR22 Class B conducted emissions requirement
Medical Safety Approvals
All models are Certified to be in compliance with the applicable
requirements of UL2601, CSA 22.2 No. 601.1-M90, IEC 601-1
(1988), EN 60601-1: 1990. CB certificate available.
Leakage Current
The maximum leakage current is as follows:
Test Condition
Normal
Single Fault
132 Vac @ 60 Hz input
70 µA
120 µA
264 Vac @ 50 Hz input
130 µA
240 µA
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com
GPFC125 Commercial/GPFM125 Medical 125 Watt Multiple Output
Commercial
Model
GPFC125A
Medical
Model
GPFM125A
GPFC125B
GPFM125B
GPFC125C
GPFM125C
GPFC125D
GPFM125D
GPFC125E
GPFM125E
GPFC125F
GPFM125F
GPFC125G
GPFM125G
GPFC125H
GPFM125H
Voltage
Output No.
1
2
3
4
Output Voltage
+5 V
+12 V
12 V (C)
-12 V
Output
Current (A)
9A
4A
2.5 A
0.3 A
Qutput
Current (B)
16 A
7A
4A
1A
Voltage
Adjustment
±5%
±5%
---
OVP
6.2 ±0.6 V
----
Total
Regulation
1%
1%
1%
1%
Ripple/
Noise
1%
1%
1%
1%
1
+5 V
9A
16 A
±5%
6.2 ±0.6 V
1%
1%
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
+12 V
12 V (C)
-5 V
+5 V
+12 V
15 V (C)
-15 V
+5 V
+24 V
12 V (C)
-12 V
+5 V
+24 V
15 V (C)
-15 V
+5 V
+15 V
15 V (C)
-5V
+5 V
+3.3 V
12 V (C)
-12 V
+3.3 V
+5 V
12 V (C)
-12 V
4A
2.5 A
0.3 A
9A
4A
2.5 A
0.3 A
9A
3A
2.5 A
0.3 A
9A
3A
2.5 A
0.3 A
9A
3.5 A
2.5 A
0.3 A
9A
7A
2.5 A
0.3 A
9A
4A
2.5 A
0.3 A
7A
4A
1A
16 A
7A
4A
1A
16 A
4.5 A
4A
1A
16 A
4.5 A
4A
1A
16 A
6A
4A
1A
16 A
10 A
4A
1A
16 A
7A
4A
1A
±5%
--±5%
±5%
--±5%
±5%
--±5%
±5%
--±5%
±5%
--±5%
±5%
--±5%
±5%
---
---6.2 ±0.6 V
---6.2 ±0.6 V
---6.2 ±0.6 V
---6.2 ±0.6 V
---6.2 ±0.6 V
4.2 ±0.6 V
--4.2 ±0.6 V
6.2 ±0.6 V
---
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
2%
1%
1%
2%
1%
1%
1%
Notes:
A. Continuous rating for unrestricted convection cooling.
B. Peak rating or continuous rating with 150 LFM air cooling.
C. Output 3 is isolated. Can be connected as + or - output.
GPFC125/GPFM125 MECHANICAL SPECIFICATIONS
INPUT: J1 AMP P.C.B. HEADER P/N 640445-5
PIN 1) AC LINE
PIN 2) N/C
PIN 3) AC NEUTRAL
PIN 4) N/C
PIN 5) AC GROUND
MATING CONNECTOR AMP P/N
HOUSING 640250-5
CONTACT 770476-1
SIGNALS: J2
AMP P.C.B. HEADER P/N 640456-4
MATING CONNECTOR P/N 640440-4
PIN 1) POWER FAIL
PIN 2) -SENSE
PIN 3) +SENSE
PIN 4) COMMON
OUTPUT: J3 AMP P.C.B HEADER P/N 1-640445-3
PIN 1) -V4OUT
PIN 2-4) +V1OUT
PINS 5-9) COMMON
PIN 10,11) +V2 OUT
PIN 12) +V3OUT
PIN 13) -V3RTN
MATING CONNECTOR AMP P/N
HOUSING 1-640250-3
CONTACT 770476-1
OPTIONAL COVER P/N 08-30466-2125
WEIGHT: 1.25 LBS [ .56 KG]
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com
53