MINT1175 175W Single Output Medical Series General Specifications (continued) Inrush Current Input Fuses Earth Leakage Current Efficiency Output Power Transient Response Signals 264Vac, cold start: will not exceed 50A F1, F2: 4A, 250VAC fuses provided on all models <200μA@264Vac, 60Hz, NC; <400µA SFC Overload Protection Short Circuit Protection Overvoltage Protection 90% typical at 115Vac 175W continuous, with 200 lfm airflow, 120W convection cooled – See chart for specific voltage model ratings. 50% load step. i/t: <0.2A/μS Max Volt Deviation = 3% Switching Frequency Isolation AC Power Fail, DC OK, Inhibit Hiccup Mode Provided - no damage will occur if the output is shorted. OVP firing reduces output voltage to <50% of nominal in <50ms. See chart for trip range. PFC: Fixed, 65kHz. Main Converter: Variable 35-200kHz, 65-70kHz at full load. Input-Output: 4000Vac, 2 x MOPP Input-Ground: 1800Vac, 1 x MOPP Output-Ground: 1500Vac Ripple and Noise See chart Operating Temperature -10°C to +70°C Start Up at -40°C, full load Output Voltage See chart Temperature Derating Derate output power linearly above 50°C to 50% at 70°C +/-5% from nominal Storage Temperature -40°C to +85°C Voltage Adjustability Minimum Load Total Regulation Vibration Dimensions Weight Altitude Not required Relative Humidity +/- 3%. See chart Operating: 0.003g²/Hz, 1.5grms overall, 3 axes, 10 min/axis Non-Operating: 0.026g²/Hz, 5.0grms overall, 3 axes, 1 hr/axis Shock Safety Standards W: 2.0” x L: 4.0” x H: 1.3” MTBF 210g Operating: -500 to 10,000 ft. Non-operating: -500 to 40,000 ft. 5% to 95%, non-condensing Operating: Half-sine, 20gpk, 10ms, 3 axes, 6 shocks total Non-Operating: Half-sine, 40 gpk, 10 ms, 3 axes, 6 shocks total EN/CSA/UL/IEC 60601-1, 3rd Edition 331,000 hours, 25°C Ambient, 110Vac input Auxiliary Signals AC Power Fail: Inhibit: www.slpower.com During normal operation, stays HIGH. Signal goes LOW with 5mS warning before loss of DC output from AC failure. DC OK: Open collector logic signal goes and stays HIGH 100mS to 500mS after main output reaches regulation. Connect to inhibit pin to output common to inhibit the DC output. MINT1175 REV 1.2, 17-DEC-12 2 Connector Information Input Connector J100 Ground J101 DC Output Connector J300 Fan Output Connector J500 Signal Connector J600 PIN 1) AC LINE PIN 2) EMPTY PIN 3) AC NEUTRAL 0.187” FASTON TAB PIN 1) +Vout PIN 4) -Vout PIN 2) +Vout PIN 5) -Vout PIN 3) +Vout PIN 6) -Vout PIN 1) +12Vfan PIN 2) RTN PIN 1) Inhibit PIN 2) PF/DC OK PIN 3, 4) Common Mating Connector: Molex 09-50-3031 Pins= 08-52-0072 Mating Connector: Molex 01-90020009 Mating Connector: AMP 640250-6 Pins = 640252-1 Mating Connector: Molex 1375820-2 Pins = 1375819 Mating Connector: Molex 1375820-4 Pins = 1375819 Fan Output – J500 J500 provides a [email protected] output to support a system cooling fan. The fan output is always available when AC input is present, so it also can be used for a 12V standby output is so desired. AC Power Failure/DC OK and Inhibit Signals – J600 J300–Pin 1 J300– Pin 1 Timing Sequence MINT1175 175W Single Output Medical Series Isolation Specifications Parameter Conditions/Description Insulation Safety Rating Input/Ground Input/Output Output/Ground Electric Strength Test Voltage Input/Ground Input/Output Output/Ground Min Nom Max Units - - Vac Vac Vac Min Nom Max Units 90 115/230 264 Vac 1 MOPP 2 MOPP Operational 1800 4000 1500 Input Specifications All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted. Parameter Conditions/Description Input Voltage Turn-On Input Voltage Ramping up 82.7 Turn-Off Input Voltage Ramping down 67.0 Input Frequency Vac Vac 47 50/60 63 Hz A Inrush Current Limitation 264Vac, cold start - - 50 Power Factor Vi nom, Io nom 0.9 - - Efficiency Vi nom, Io nom MINT1175A1206K01 MINT1175A1506K01 MINT1175A2406K01 MINT1175A4806K01 MINT1175A5606K01 - 90% - % Parameter Conditions/Description Min Nom Max Units Output Voltage Setpoint Accuracy MINT1175A1206K01 Vi nom,, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom Output Voltage Setpoint Accuracy MINT1175A1506K01 Vi nom,, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom Output Voltage Setpoint Accuracy MINT1175A2406K01 Vi nom,, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom Output Voltage Setpoint Accuracy MINT1175A4806K01 Vi nom,, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom Output Voltage Setpoint Accuracy MINT1175A5606K01 Vi nom,, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom Output Current V1 Output Current V2 MINT1175A1206K01 0 0 10.0 - 14.6 0.4 ADC ADC Output Current V1 Output Current V2 MINT1175A1506K01 0 0 8.0 - 11.7 0.4 ADC ADC Output Current V1 Output Current V2 MINT1175A2406K01 0 0 5.0 - 7.3 0.4 ADC ADC Output Current V1 Output Current V2 MINT1175A4806K01 0 0 2.5 - 3.6 0.4 ADC ADC Output Current V1 Output Current V2 MINT1175A5606K01 0 0 2.1 - 3.1 0.4 ADC ADC Static Line Regulation V1 Vi min-Vi max, Vi nom, 0-100% Io nom -1 - 1 % Vo nom Static Load Regulation V1 (Droop Characteristic) Vi min-Vi max, Vi nom, 0-100% Io nom -3 - 3 % Vo nom Output Specifications www.slpower.com MINT1175 REV 1.2, 17-DEC-12 5