MINT1180 Series Datasheet

MINT1180
180W Single Output
Medical Series
Features
3” x 5” x 1.3” Package
180W Convection Cooled Power
Ideal for 1U Applications
Class I and Class II versions
Universal Input 90-264 VAC
Approved to CSA/EN/IEC/UL60601-1, 3rd
Edition, 2 x MOPP Isolation
Approved to CSA/EN/IEC/UL60950-1 2nd
Edition
Efficiency 90% typical
3 Year Warranty
RoHS Compliant
Description
The MINT1180 product family is the most cost-effective medically approved supply in the market with 180 watts
of convection cooled output power in 3”X5” footprint. The series comes in eight different models ranging from
12 to 48V and has been approved to IEC60601-1 3rd Edition, rated for 2 MOPP (Means of Patient Protection)
isolation. This new single output power supply family has both Class I and II input versions and accepts wide
input range of 90 to 264Vac. The typical 90% efficiency and convection cooled make this supply an excellent
solution for HomeCare applications where the use of fans is not desired.
Model Selection
Model Number
Volts
Output
Current*
Minimum Load
Ripple &
Noise**
Total
Regulation
OVP Threshold
MINT1180A1275K01
12V
15.0A
0A
120mV pk-pk
±3%
14.0 ± 1.1V
MINT1180A1575K01
15V
12.0A
0A
150mV pk-pk
±3%
18.5 ± 1.2V
MINT1180A1875K01
18V
10.0A
0A
180mV pk-pk
±3%
21.5 ± 2.0V
MINT1180A2475K01
24V
7.50A
0A
240mV pk-pk
±3%
29.0 ± 2.5V
MINT1180A2875K01
28V
6.40A
0A
280mV pk-pk
±3%
33.5 ± 2.5V
MINT1180A3275K01
32V
5.62A
0A
320mV pk-pk
±3%
36.0 ± 3.0V
MINT1180A4875K01
48V
3.75A
0A
480mV pk-pk
±3%
56.0 ± 3.0V
Notes:
* Total convection power is 180 Watts.
** Measured with noise probe directly across output terminals, and load terminated with 0.1µF ceramic and 10µF low ESR capacitors.
General Specifications
AC Input
Input Current
www.slpower.com
100-240Vac, ±10%, 47-63Hz, 1∅
120–370Vdc
Turn On Time
Less than 3 sec. @115Vac & Full Load
115Vac: 1.8A, 230Vac: 0.9A
Hold-up Time
>16mS at 180W, 120Vac
MINT1180 REV 1.8, 15-OCT-13
1
MINT1180
180W Single Output
Medical Series
General Specifications (continued)
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
Output Power
Transient
Response
Signals
264Vac, cold start: will not exceed 55A
Overload
Protection
F1, F2: 4.00A, 250VAC fuses provided on all
models
Short Circuit
Protection
<275µA@264Vac, 60Hz, NC; <400µA SFC
Overvoltage
Protection
88% typical
Switching
Frequency
180W convection cooled
500µS typ. for return to within 0.5% of
nominal, 50% load step. ∆i/∆t <0.2A/µS.
Max Volt Deviation = 3%
Isolation
N/A
120%-150% of rating, Hiccup Mode
Provided - no damage will occur if the output is
shorted, auto recovery
OVP latch at 110%-130% of rated output
voltage.
PFC: Fixed, 65kHz
Main Converter: Variable 35-200kHz, 65-70kHz
at full load.
Input-Output: 4000Vac, 2 x MOPP
Input-Ground: 1800Vac, 1 x MOPP
Output-Ground: 1500Vac
Ripple and Noise
See chart
Operating
Temperature
-10°C to +70°C
Start Up at -40°C, full load
Output Voltage
See chart
Temperature
Derating
Derate output power linearly above 50°C to 50%
at 70°C
Voltage
Adjustability
Fixed output
Storage
Temperature
-40°C to +85°C
Minimum Load
Not required
Altitude
Total Regulation
+/- 3% combined line, load and initial setting.
Vibration
Operating: 0.003g²/Hz, 1.5grms overall, 3
axes, 10 min/axis
Non-Operating: 0.026g²/Hz, 5.0grms overall,
3 axes, 1 hr/axis
Dimensions
Weight
Relative Humidity
Shock
Safety Standards
W: 3.0” x L: 5.0” x H: 1.3”
MTBF
325g
Operating: -500 to 10,000 ft.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6
shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3
axes, 6 shocks total
EN/CSA/UL/IEC 60601-1, 3rd Edition
EN/CSA/UL/IEC 60950-1, 2nd Edition
214,194 hours, 25°C, 110Vac
Output vs. Temperature Derating Curve
180W convection cooled, derate output power to 50% at 70°C.
Output Power (W)
200
150
100
50
0
-10
0
Convection
www.slpower.com
10
20
30
40
50
60
70
Temperature (°C)
MINT1180 REV 1.8, 15-OCT-13
2
MINT1180
180W Single Output
Medical Series
EMI/EMC Compliance
Conducted Emissions
EN55011/22 Class B, FCC Part 15, Subpart B, Class B
Radiated Emissions
EN55011/22; FCC Part 15, Class A with 6dB Margin
Static Discharge Immunity
EN61000-4-2, Criteria A, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity
EN61000-4-3, 3V/m. Criteria A.
EFT/Burst Immunity
EN61000-4-4, 2kV/5kHz, Criteria A
Line Surge Immunity
EN61000-4-5, 1kV differential, 2kV common-mode, Criteria A
Conducted RF Immunity
EN61000-4-6, 3Vrms, Criteria A
Power Frequency Magnetic Field Immunity
EN61000-4-8, 3A/m, Criteria A
Voltage Dip Immunity
EN61000-4-11, 5% Vnom: 0.5cycle; 40% Vnom: 5 cycles, 70% Vnom:
25 cycles, Criteria A
Line Harmonic Emissions
EN61000-3-2, Class A, B, C, & D
Flicker Test
EN61000-3-3, Complies (dmax<6%)
Mechanical Drawing
Notes:
1. All dimensions in inches (mm), tolerance
is +/-0.000”.
2. Mounting holes should be grounded for
EMI purposes.
3. FG is safety ground connection.
4. The power supply requires mounting on
metal standoffs 0.20” (5mm) in height, min.
[4] x 0.156
4 PLCS
MTG HOLES
Connector Information
Input Connector
J100
Ground
(FG)
DC Output Connector
J300
PIN 1) AC LINE
PIN 2) EMPTY
PIN 3) AC NEUTRAL
0.25” FASTON TAB
Term. 1,2,3: RTN
Term. 4,5,6: +Vout
Mating Connector:
AMP 640250-3
Pins: 640252-2
Mating Connector:
Molex 190020001
Mating Connector:
AMP 640250-6
Pins: 640252-2
Data Sheet © 2013 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
MINT1180 REV 1.8, 15-OCT-13
3