ETC AF033A

METRAport3A
Folding Analog Multimeter
3-349-302-03
1/4.04
•
9 voltage measuring ranges, V :
100 mV, 300 mV, 1 V, 3 V, 10 V, 30 V, 100 V, 300 V and 600 V
•
7 current measuring ranges, A :
10 µA, 100 µA, 1mA, 10 mA, 100 mA, 1A and 10 A
•
5 resistance measuring ranges: 1 Ω ... 2 kΩ/ 10 Ω ... 20 kΩ/
100 Ω ... 200 kΩ/ 1 kΩ ... 2 MΩ/ 10 kΩ ... 20 MΩ
•
9 level measuring ranges: -40 dB ... +62 dB
•
High input resistance of 10 MΩ
for no-load voltage measurement
•
Reflective scale in folding lid which can be tilted for easy
reading and provides transport protection as well,
accuracy class 1.5 =
•
Overload protection: 600 V
in all functions and range settings
•
Automatic battery shutdown when instrument is folded closed
Applications
The multimeter is equipped with an indicator display and an
electronic amplifier. It can be used in a wide variety of practical
electronics applications, for example in R&D, manufacturing,
equipment operation, in the test lab and for service calls, as well
as for training and vocational education.
Description
The multimeter has 46 measuring ranges for direct and
alternating voltages of up to 600 V, alternating voltage levels from
-40 to +62 dB, direct and alternating current of up to 10 A and
resistances of up to 20 MΩ. It has a constant input resistance of
10 MΩ in all voltage measuring ranges.
The multimeter is calibrated in RMS values for periodic sinusoidal
quantities. It makes use of full-wave rectification for the evaluation
of the arithmetic mean value.
All measuring ranges can be selected with the central measuring
range selector switch. The ranges are laid out in a clear-cut
fashion around the selector switch.
The measuring device and the display unit are enclosed in two
separate housing components which are connected to each
other by means of a hinge which can be snapped into various
positions. In this way, the ideal angle for ease of reading can be
selected with the instrument in any position.
GOSSEN METRAWATT GMBH
300 V CAT III
600 V CAT II
IEC/EN 61010-1, second edition
Several well coordinated safety devices protect the instrument
against damage caused by incorrect operation and overloading
within the specified overload limit values:
• Oversized precision resistors
•
Fuse link in combination with protective rectifier diodes
•
Overvoltage arrester / positive temperature coefficient
resistors (PTCs)
The instrument can be operated independent of mains current
with a commercially available 9 V flat cell battery. Long battery
service life is assured thanks to the minimal current consumption
of the instrument’s electronic components.
The multimeter’s rugged design provides for outstanding
protection where harsh mechanical stressing prevails. When
folded together, the instrument is provided with extra mechanical
protection for the measuring device, as well as for the display unit.
The connector jacks are protected against accidental contact.
Suitable measurement cables with 4 mm test probes are included
as standard equipment.
Applicable Regulations and Standards
IEC/EN 61010-1:2001
VDE 0411-1:2002
Safety requirements for electrical equipment for
measurement, control and laboratory use
DIN EN 61326
VDE 0843, part 20
Electrical equipment for control technology and
laboratory use – EMC requirements
DIN EN 60529
DIN VDE 0470, part 1
Test instruments and test procedures
– Degrees of protection provided by enclosures (IP code)
METRAport3A
Folding Analog Multimeter
Characteristic Values
Reference Conditions
Level 2)
Voltage 1)
100 mV
300 mV
1 V
3 V
10 V
30 V
100 V
300 V
600 V
Constant
Measuring Span
-20
-10
0
+10
+20
+30
+40
+50
+60
-40 ... ÷ 1 8
-30 ... ÷ 8
-20 ... + 2
-10 ... + 12
0 ... + 22
+10 ... + 32
+20 ... + 42
+30 ... + 52
+40 ... + 62
dB
dB
dB
dB
dB
dB
dB
dB
dB
Input
Resistance
Ri
dB
dB
dB
dB
dB
dB
dB
dB
dB
10 MΩ / /
50 pF
Overload
Protection
to 3)
600 V
600 V
600 V
600 V
600 V
600 V
600 V
600 V
600 V
Ambient temperature
Relative humidity
Normal position
of use
Measured quantity
frequency
Measured quantity
waveshape
Battery voltage
Other influencing
quantities
+23° C ±2 K
40 ... 60%
Instrument and scale horizontal ±1°
45 ... 65 Hz
Sine
7.5 V ±0.1 V
per IEC 60051/EN 60051
1)
The 100 mV to 10 V measuring ranges can also be used for current measurement
in accordance with the following table (e.g. reverse current or insulation current).
These measuring ranges are rated at accuracy class 5. Frequency range
at : 15 ... 100 Hz.
Same overload protection as for voltage measuring ranges.
2) Measured value = displayed value + constant, 0 dB
0.775 V, i.e. 1 mW at
600 Ω
3)
Protected with PTC resistor
~
Range
100 mV
300 mV
1 V
3 V
10 V
30 V
Current at
Upper Range Limit
10 nA
30 nA
100 nA
300 nA
1 µA
3 µA
Input
Resistance Ri
10.0 kΩ
1.0 kΩ
100.0 Ω
10.0 Ω
1.4 Ω
480 mΩ
26 mΩ
Current
10
100
1
10
100
1
10
µA
µA
mA
mA
mA
A
A
Resistance
Range
Ω
Ω
Ω
kΩ
kΩ
x 1
x 10
x 100
x 1
x 10
Reading Range
1
10
100
1
10
Ω ... 2 kΩ
Ω ... 20 kΩ
Ω ... 200 kΩ
kΩ ... 2 MΩ
kΩ ... 20 MΩ
Ri
10 MΩ
Temperature
+5 ... +23 ... +35° C
Limiting temperatures For accuracy +5 ... +35° C
For operation 0 ... +40° C
For storage -25 ... +65° C
(without batteries)
Position
Additional influence error of max. ±1% of
the scale length if the scale is tilted
between 0 and ±120° relative to horizontal
Frequency
Additional influence error of max. ±5% of
full scale value in the following ranges:
100 mV, 3 V ...
10 µA
...
300 mV,
Voltage Drop
∆U
100 mV
100 mV
100 mV
100 mV
140 mV
0.480 V
260 mV
Value at
Scale
Middle
(Ri)
45.6 Ω
456.0Ω
4.56 kΩ
45.6 kΩ
456.0kΩ
OpenCircuit
Voltage
U0
100mV
100mV
100mV
1 V
1 V
Overload Protection
to
600 V 3)
600 V 3)
600 V 3)
600 V 3)
600 V 3)
600 V 3)
600 V 4)
ShortCircuit
Current
IK
2.2 mA
0.22mA
22 µA
22 µA
2.2 µA
Overload
Protection
to
600 V 3)
600 V 3)
600 V 3)
600 V 3)
600 V 3)
3)
Protection by means of FF 1.6A/600V AC fuse link in combination with rectifier
diodes.
4)
FF 16A/600V fuse link, 10 A: max. 10 minutes
Accuracy under Reference Conditions per IEC 60051/EN 60051
Class 1.5 for zero-frequency quantities, class 2.5 for periodic
sinusoidal quantities and class 1.5 for resistance (intrinsic error
relative to 69 mm scale length) corresponding to max. 10%
intrinsic error of the measured value within the range of the
exaggerated scale arc.
2
Influencing Quantities and Nominal Ranges of Use
600 V:
10 A:
1 V:
15 Hz ...
1 kHz
15 Hz ...
1 kHz
15 Hz ... 200 Hz
Auxiliary voltage
No additional influence error. The upper
limit must be set with the potentiometer for
all measuring ranges in the case of
resistance measurement.
Series-mode interference
voltage damping
For V : > 60 dB, AC 50 Hz
For V~: > 120 dB, DC
Common-mode interference
voltage damping
> 120 dB, DC and 50 Hz AC
Other influencing
quantities
per IEC 60051/EN 60051
Power Supply
Battery operation
Battery service life
Battery test
Nominal voltage: 9 V ,
9 V flat cell battery per IEC 6LR61 (6F22),
alkaline manganese, zinc-carbon or NiCd
rechargeable battery.
With zinc-carbon: approx. 500 hours, with
alkaline manganese: approx. 1000 hours,
with NiCd: approx. 200 hours, in the Ω x 1
range: ¼ of specified service life
Pointer must be within the battery test
field.
GOSSEN METRAWATT GMBH
METRAport3A
Folding Analog Multimeter
Fuses
Ranges to 1 A and Ω
10 A range
Order Information
FF 1.6A/600V, 6 mm x 32 mm in
combination with rectifier diodes for the
10 µA ... 1 A and Ω x1 ... kΩ x10 ranges,
switching capacity: 50 kA at 600 V~
FF 16A/600V AC, 6 mm x 32 mm,
switching capacity: 50 kA at 600 V~
Measuring category
Operating voltage
Fouling factor
Test voltage
Type
Article Number
Folding analog multimeter for
demanding requirements
METRAport3A
M113A
Consumable Materials
Fuse link (shipped in package of 10) FF(UR)1.6A/700V AC
Z109E
Fuse link (shipped in package of 10) FF(UR)16A/600V AC
Z109A
Accessories
Electrical Safety
Safety class
Description
Carrying pouch
II per IEC/EN 61010-1:2001/VDE 04111:2002
II
III
600 V
300 V
2
3.5 kV~ per IEC/EN 61010-1:2001/
VDE 0411-1:2002
Electromagnetic Compatibility (EMC)
Interference emission EN 61326:2002 class B
Interference immunity EN 61326:2002
IEC 61000-4-2:1995/A1:1998,
Feature B
8 kV atmospheric discharge
4 kV contact discharge
IEC 61000-4-3:1995/A1:1998,
Feature B
3 V/m
F822
GTY 3172 095 P01
Current measuring adapter for safe,
trouble-free measurement of current
consumption via the mains plug of
connected power consumers
SM16
GTM 9070190E0002
High-voltage probe, 3 kV/3 V, ,
also suitable as low pass filter for
frequency converter signals
GTZ 3431 001R0011
HV3
High-voltage probe,
30 kV / 30 V DC, For safe measurement of direct voltage of up to 30 kV HV30
GTZ 3431 001R0001
Temperature probes with integrated
or plug-on sensors (see Measuring
Instruments and Testers catalog)
Z3431...
GTZ 3431...
For additional information regarding accessories please see:
• Measuring Instruments and Testers catalog
• www.gossenmetrawatt.com
Accessories
SM16 Current Measuring Adapter
Mechanical Design
Protection
Dimensions
Weight
Housing: IP 40, connector jacks: IP 20
146 x 118 x 44 mm (folded closed)
approx. 0.45 kg without battery
HV3 High-Voltage Probe
Standard Equipment
1
1
1
1
multimeter
9 V battery
KS17 measurement cable
set operating instructions
HV30 High-Voltage Probe
GOSSEN METRAWATT GMBH
3
METRAport3A
Folding Analog Multimeter
Current Measuring Accessories
Suitable for
METRAport
All current sensors and transformers are equipped with a connector cable (1.2 to 1.5 m long) with 4 mm safety banana plugs
Type
Designation
Measuring Range
Max.
Transformation
Meas.
Category Wire Dia. Ratio
Frequency
Range
Intrinsic Error
±(% rdg. + ...)
Article
Number
3A
AC/DC Current Sensors with Voltage Output
Z201A
Clip-on current sensor
with battery mode (30 h)
0.01 ... 20 A~/30 A–
300 V /
CAT III
19 mm
100 mV/A
DC ... 400 Hz 1% + 0.002 A
... 20 kHz
Z201A
●
Z202A
Clip-on current sensor
with 2 measuring ranges,
battery mode (50 h)
0.1 ... 20 A~/30 A–,
1 ... 200 A~/300 A–
300 V /
CAT III
19 mm
10 mV/A,
1 mV/A
DC ... 2 kHz
... 10 kHz
Z202A
●
Z203A
Clip-on current sensor
with 2 measuring ranges,
battery mode (50 h)
1 ... 200 A~/300 A–,
1 ... 1000 A~/A–
300 V /
CAT III
31 mm
1 mV/A
DC ...10 kHz 1% +0.5 A
Z203A
●
Z13B
Clip-on current sensor
with 2 measuring ranges,
battery mode (50 h)
0,2 ... 40 A~/60 A–, 300 V /
0.5 ... 400 A~/600 A–, CAT IV
50 mm
10 mV/A,
1 mV/A
DC ... 65 Hz 1.5% + 0.5 A
... 10 kHz
2.5%
Z13B
●
1% + 0.03 A,
1% + 0.3 A
AC Current Sensors with Voltage Output
WZ12B
Clip-on current sensor
10 mA~ ... 100 A~
300 V /
CAT III
15 mm
0.1 mV/mA
45 ... 65
... 500 Hz
1.5% +0.1 mA
Z219B
●
WZ12C
Clip-on current sensor
with 2 measuring ranges
1 mA~ ... 15 A~,
1 ... 150 A~
300 V /
CAT III
15 mm
1 mV/mA,
1 mV/A
45 ... 65
... 400 Hz
3% + 0.15 mA,
2% + 0.1 A
Z219C
●
WZ11B
Clip-on current sensor
with 2 measuring ranges
0.5 ... 20 A~,
5 ... 200 A~
600 V /
CAT III
20 mm
100 mV/A,
10 mV/A
30 .. .48 ... 1 ... 3%
65 ... 500 Hz
Z208B
●
Z3512A
Clip-on current sensor
with 4 measuring ranges
1 mA ... 1/10 A~
100/1000 A~
600 V /
CAT III
52 mm
1 V/A, 100 mV/A, 10 ... 48 ...
10 mV/A, 1 mV/A 65 ... 3 kHz
Z225A
●
AF033A
AmpFLEX flexible current
sensor with 2 measuring
ranges, battery (150 h)
5 ... 30 A~,
5 ... 300 A~
1000 V / Length
100 mV/A,
CAT III
600 mm 10 mV/A
10 ...100 Hz 1% + 0.5 A,
... 20 kHz
1% +0.5 A
Z207A
●
AF11A
AmpFLEX flexible current
sensor, battery (150 h)
5 ... 1000 A~
1000 V / Length
1 mV/A
CAT III
450 mm
10 ...100 Hz 1% + 2 A
... 20 kHz
Z207D
●
AF33A
AmpFLEX flexible current
sensor with 2 measuring
ranges, battery (150 h)
5 ... 300 A~,
5 ... 3000 A~
1000 V / Length
10 mV/A,
CAT III
900 mm 1 mV/A
10 ...100 Hz 1% + 0.5 A,
... 20 kHz
1% + 2 A
Z207B
●
AF101A
AmpFLEX flexible current
sensor with 2 measuring
ranges, battery (150 h)
5 A~... 1 k A~,
50 A~... 10 k A~
1000 V / Length
1 mV/A,
CAT III
1200 mm 0.1 mV/A
10 ...100 Hz 1% + 2 A,
... 20 kHz
1% + 10 A
Z207C
●
0.5 ... 3%,
0.2 ... 1%
AC Current Transformer with Current Output
WZ12A
Clip-on current transformer
15 ... 180 A~
300 V /
CAT III
15 mm
1 mA/A
45 ... 65
... 400 Hz
3%
Z219A
●
WZ12D
Clip-on current transformer
30 mA ... 150 A~
300 V /
CAT III
15 mm
1 mA/A
45 ... 65
... 500 Hz
2.5% +0.1 mA
Z219D
●
WZ11A
Clip-on current transformer
1 ... 200 A~
600 V /
CAT III
20 mm
1 mA/A
48 ... 65
... 400 Hz
1 ... 3%
Z208A
●
Z3511
Clip-on current transformer
4 ... 500 A~
600 V /
CAT III
30 x 63
mm
1 mA/A
48 ... 65
... 1 kHz
3% +0.4 A
GTZ 3511
000 R0001
●
Z3512
Clip-on current transformer
0.5 ... 1000 A~
600 V /
CAT III
52 mm
1 mA/A
30 .. .48 ...
65 ... 5 kHz
0.5% ... 0.7%
GTZ 3512
000 R0001
■
Z3514
Clip-on current transformer
1 ... 2000 A ~
600 V /
CAT III
64 x 150 1 mA/A
mm
30 .. .48 ...
65 ... 5 kHz
0.5% +0.1 A
GTZ 3514
000 R0001
■
● Without restriction
■ As of 50 A
WZ11/12 ...
Z201 ... Z203
Edited in Germany • Subject to change without notice • PDF version available on the Internet
GOSSEN METRAWATT GMBH
Thomas-Mann-Str. 16 - 20
90471 Nürnberg, Germany
Phone: +49-(0)-911-8602-0
Fax: +49-(0)-911-8602-669
E-mail: [email protected]
www.gossenmetrawatt.com
Z13B