HP U1240

Agilent U1240 Series Handheld
Digital Multimeters
Helping You Check and Fix More Installation
and Maintenance Bugs
Data Sheet
Key features
Check more, fix more
• 10,000-count display
• 0.09% basic DCV accuracy
• True RMS AC measurement
• Basic functions — ACV, DCV,
ACI, DCI, resistance, frequency,
diode, continuity tests
• Advanced functions —
Capacitance, temperature,
MINMAX recording
Ease of use
The Agilent U1240 Series handheld digital multimeters enable you to check
more with wider measurement ranges. They feature true RMS readings on their
10,000-count displays. The adjustable backlighting allows you to complete your
jobs even in subdued lighting conditions, or to simply prolong battery life. Your
maintenance tasks are greatly simplified due to the built-in switch counter,
harmonic ratio, dual and differential temperature capabilities, with just a press
of the button. The meters have a high safety rating with CAT III 1000 V and
CAT IV 600 V protection and are certified to CE and CSA standards. On top of
that, the U1240 Series comes with a certificate of calibration and test report –
at no extra cost.
The latest multimeters in this series, the U1241B and U1242B, now come in
vivid orange cases, offering the capabilities and functions you need.
Functions and ranges at a glance
• Adjustable backlighting —
two intensity levels
Ambient temperature
• Manual data logging (U1242B
only)
MINMAX Recording
Capacitance:
0.1 nF to 10 mF
• Built-in switch counter,
harmonic ratio (U1242B only),
dual/differential temperature
capabilities (U1242B only)
Resistance:
0.1 Ω to 100 MΩ
ACµA: 0.1 µA to 10 mA
Built to last
Diode Test
Switch Counter
Backlight:
Two intensity levels
• Overmold body casing
• CAT III 1000 V and
CAT IV 600 V safety protection
• Certified to CE and CSA
standards
• Operating temperature:
–10 to 55 °C
DCV/ACV: 0.1 mV to 1000 V
Harmonic Ratio
(U1242B only)
T1–T2 Differential
Temperature
(U1242B only)
Frequency: 1 Hz to 200 kHz
Manual Data Logging
(U1242B only)
ACmA: 0.01 to 1000 mA
ACA: 0.01 to 10 A
Temperature:
–40 to 1000 °C
Electrical Specifications
DC specifications
Function
Voltage1
Current
Range
Resolution
1000.0 mV
0.1 mV
–
10.000 V
0.001 V
–
100.00 V
0.01 V
–
1000.0 V
0.1 V
–
0.15% + 5
1000.0 µA
0.1 µA
< 0.06 V (50 Ω)
0.1% + 3
10000 µA
1 µA
< 0.55 V (50 Ω)
0.1% + 3
100.00 mA
0.01 mA
< 0.18 V (0.5 Ω)
0.2% + 3
440.0 mA
0.1 mA
< 0.8 V (0.5 Ω)
0.5% + 3
10.000 A3
0.001 A
< 0.4 V (0.01 Ω)
0.6% + 5
0.1 Ω
0.5 mA
10.000 kΩ
0.001 kΩ
50 µA
100.00 kΩ
0.01 kΩ
4.91 µA
1000.0 kΩ
0.1 kΩ
447 nA
10.000 MΩ
0.001 MΩ
112 nA
0.8% + 3
2
5
1000.0 Ω
5
4
Resistance
6
Diode test7
Test current/
burden voltage
Accuracy ± (% of reading + counts
of least significant digit)
U1241B
U1242B
0.09% + 5
0.09% + 2
0.3% + 3
100.00 MΩ
0.01 MΩ
112 nA
1.5% + 3
1V
0.001 V
approximately 0.5 mA
0.3% + 2
1. Input impedance: 10 MΩ (nominal).
2. Current can be measured up to 440 mA continuously. An additional 0.2% needs to be added to the specified accuracy if the signal measured is in the
range of 440 mA to 1100 mA for 30 seconds maximum. After measuring a current of > 440 mA, leave the meter to cool down for twice the measuring time used before applying a low current measurement.
3. Current can be measured up to 10 A continuously with a maximum operating temperature of 50 °C. An additional 0.3% needs to be added to the
specified accuracy if the signal measured is in the range of 10 A to 19.999 A for 15 seconds maximum. After measuring a current of > 10 A, leave the
meter to cool down for 60 seconds before applying a low current measurement.
4. The maximum open voltage is < 2.8 V. For instant continuity, the built-in buzzer sounds when resistance is < 10.0 Ω.
5. The accuracy of 1 kΩ and 10 kΩ is specified after Null function, which is used to substrate the test lead resistance and the thermal effect.
6. For the range of 100 MΩ, the R.H. is specified for < 60%. The temperature coefficient will be 0.15 times of specified accuracy as > 50 MΩ.
7. Overload protection: 1000 V RMS for circuits < 0.3 A short circuit current. The built-in buzzer sounds when reading is approximately below 50 mV and
audible single tone for normal forward biased diode or semiconductor junction as 0.3 V ≤ Reading ≤ 0.8 V.
2
Electrical Specifications
AC specifications
Function
8, 12
AC voltage
True RMS
AC current9, 12
True RMS
Range
Resolution
Test current/
burden voltage
1000.0 mV
0.1 mV
–
10.000 V
0.001 V
–
100.00 V
0.01 V
–
1000.0 V
0.1 V
–
1000.0 µA
0.1 µA
< 0.06 V (50 Ω)
10000 µA
1 µA
< 0.55 V (50 Ω)
100.00 mA
0.01 mA
< 0.18 V (0.5 Ω)
10
Accuracy ± (% of reading + counts of
least significant digit)
40 to 500 Hz 500 Hz to 1 kHz 1 to 2 kHz
2% + 5
1% + 5
1% + 5
–
2% + 5
–
440.0 mA
0.1 mA
< 0.8 V (0.5 Ω)
10.000 A11
0.001 A
< 0.4 V (0.01 Ω)
1% + 5
1.5% + 5
–
8. Input impedance: 10 MΩ (nominal) in parallel with < 100 pF, with overload protection of 1000 V RMS
9. Crest factor ≤ 3. For non-sinusoidal waveforms with crest factor up to 3, add 2% reading + 2% full scale typical.
10. Current can be measured from 50 mA to 440 mA continuously. An additional 0.2% needs to be added to the specified accuracy if the signal
measured is in the range of 440 mA to 1100 mA for 30 seconds maximum. After measuring a current of > 440 mA, leave the meter to cool down for
twice the measuring time used before application of low current measurement.
11. Current can be measured from 0.5 A up to 10 A continuously with a maximum operating temperature of 50 °C. An additional 0.3% needs to be added
to the specified accuracy if the signal measured is in the range of 10 A to 19.999 A for 15 seconds maximum. After measuring a current of >10 A,
leave the meter to cool down for 60 seconds before applying a low current measurement.
12. AC voltage and AC current specifications are AC coupled. True RMS measurement is valid from 5 % of range to 100 % of range.
Temperature specifications
Thermocouple type
Range
Resolution
Accuracy ± (% of reading +
offset error)
K (for U1241B and U1242B)
–40 to 1000 ºC/–48 to 1832 ºF
0.1 °C /0.1 °F
1% + 1 ºC/1% + 1.8 °F
J (for U1242B only)
–40 to 1000 ºC/–48 to 1832 ºF
0.1 ºC/0.1 ºF
1% + 1 ºC/1% + 1.8 °F
Capacitance specifications
Range
Resolution
1000.0 nF
0.1 nF
10.000 µF
0.001 µF
100.00 µF
0.01 µF
1000.0 µF
0.1 µF
10.000 mF
0.001 mF
Accuracy ± (% of reading + counts of least significant digit)
1.2% + 4
2% + 4
Harmonic ratio specifications
Range
Frequency
Voltage
0.0% to 99.9%
40 to 500 Hz
100 mVAC to 1000 VAC
3
Electrical Specifications
Switch counter definition
Switch condition1, 2
Circuit switch
Display3
Switch threshold
Low level
Normally close
Lo
< 370 ohms
4
Close to open
Number of switch count
Low to high transition
High level
Normally open
Hi
> 430 ohms
Intermittent5
Open to close
Number of switch count
High to low transition
Intermittent
1.
2.
3.
4.
5.
Detects intermittent Opens or Closes lasting for at least 250 μsec.
Test current of 0.5 mA with maximum open circuit voltage of 2.8 V is used.
Maximum count reading: 199.99 M (display “OL” when achieving 2 x 108 and thereafter).
Count only low to high transition for initial switch condition of Low.
Count only high to low transition for initial switch condition of High.
Frequency specifications
Range
Resolution
100.00 Hz
0.01 Hz
1000.0 Hz
0.1 Hz
10.000 kHz
0.001 kHz
100.00 kHz
0.01 kHz
1000.0 kHz1
0.1 kHz
Accuracy
Minimum input frequency
0.03%+3
1 Hz
1. Effective frequency measurement of up to 200 kHz; refer to frequency sensitivity table below for details.
Frequency sensitivity during voltage measurement
Input range (Maximum input for specified accuracy =
10 x Range or 1000 V)
Minimum sensitivity (RMS sine wave)
20 Hz to 50 kHz
50 to 200 kHz
1000.0 mV
0.3 V
0.6 V
10.000 V
0.5 V
1.8 V
100.00 V
5V
10 V (< 100 kHz)
1000.0 V
50 V
100 V (< 100 kHz)
Frequency sensitivity during current measurement
Input range
Minimum sensitivity (RMS sine wave)
20 Hz to 20 kHz
1000.0 µA
100 µA
10000 µA
500 µA
100.00 mA
10 mA
440.00 mA
50 mA
10.000 A
1A
4
Electrical Specifications
Measuring rate
Function
Times/second
ACV
7
DCV (V or mV)
7
Ω
14
Diode
14
Capacitance
4 (< 100 µF)
DCA (μA, mA, A)
7
ACA (μA, mA, A)
7
Temperature
7 (single)
Frequency
1 (> 10 Hz)
General Specifications
Display
Dual display (secondary display is intended for temperature function display only)
consists of 4-digit liquid crystal display (LCD) with maximum reading of 11,000 counts.
Automatic polarity indication.
Power consumption
0.22 VA maximum
Battery type and life
Four single standard 1.5 V AAA batteries (Alkaline or Zinc Chloride type); 300 hours
typical
Operating environment
• Full accuracy at –10 to 55 °C; and to 80% RH for temperatures up to 30 °C, decreasing
linearly to 50% RH at 55 °C
• 0 – 2000 meters per IEC 61010-1 2nd Edition CAT III, 1000 V / CAT IV, 600 V IEC
61010-1 2nd Edition
Storage compliance
–20 to 70 °C
Safety compliance
• IEC 61010-1:2001 / EN61010-1:2001
• Canada: CSA C22.2 No. 61010-1:2004
Measurement category
CAT III 1000 V / CAT IV 600 V Overvoltage Protection, Pollution Degree 2
EMC compliance
• CISPR 11:1990/EN55011:1990
• Canada: ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
Common Mode Rejection Ratio (CMRR)
> 90 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)
Normal Mode Rejection Ration (NMRR)
> 60 dB at 50/60 Hz ± 0.1%
Crest factor
< 3.0
Temperature coefficient
0.1 × (specified accuracy) / °C (from –10 to 18 °C or 28 to 55 °C)
Shock and vibration
Tested to IEC/EN 60068-2
Dimensions (H x W x D)
193.8 mm x 92.2 mm x 58.0 mm
Weight
• 450 g with batteries
• 400 g without batteries
Warranty
• Three years for main unit
• Three months for standard shipped accessories
5
Ordering Information
U1241B
U1242B
Standard shipped accessories
•
•
•
•
•
Four 1.5 V AAA alkaline batteries
Certificate of Calibration (CoC)
Test probe leads (4-mm tips)
Quick Start Guide
Free test data (Option UK6)
Optional accessories
Measuring accessories (non-temperature)
U1161A
Extended test
lead kit
Includes two test leads (red and black), two test probes, mediumsized alligator clips and 4-mm banana plugs.
• Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
• Test probes: CAT lll 1000 V, CAT lV 600 V, 15 A
• Medium-sized alligator clips: CAT III 1000 V, CAT IV 600 V, 15 A
• 4-mm banana plugs: CAT II 600 V, 10 A
U1162A
Alligator clips
• One pair of insulated alligator clips (red and black).
Recommended for use with Agilent standard test leads.
• Rated CAT III 1000 V, CAT IV 600 V, 15 A
U1163A
SMT grabbers
• One pair of SMT grabbers (red and black). Recommended for use
with Agilent standard test leads.
• Rated CAT II 300 V, 3 A
U1164A
Fine-tip test
probes
• One pair of fine-tip test probes (red and black). Recommended for
use with Agilent standard test leads.
• Rated CAT II 300 V, 3 A
U1168A
Standard test
lead kit
Includes two test leads (red and black), 19-mm and 4-mm test
probes, alligator clips, fine-tip test probes, SMT grabbers and mini
grabber (black).
• Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
• Test probes (19-mm tip): CAT II 1000 V, 15 A
• Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A
(highly recommended for CAT IV environment)
• Alligator clips: CAT III 1000 V, CAT IV 600 V, 15 A
• Fine-tip test probes: CAT II 300 V, 3 A
• SMT grabber: CAT II 300 V, 3 A
• Mini grabber: CAT II 300 V, 3 A
U1169A
Test probe leads
Includes two test leads (red and black), and a pair each of 19-mm
and 4-mm test probes.
• Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
• Test probes (19-mm tip): CAT II 1000 V, 15 A
• Test probes (4-mm tip): CAT III 1000 V, CAT IV 600 V, 15 A
(highly recommended for CAT IV environment)
U1583B
AC current clamp
• Dual range: 40 A and 400 A
• Rated CAT III 600 V
• BNC-to-banana-plug adapter provided for use with DMMs
6
Ordering Information
Optional accessories
Measuring accessories (temperature)
U1180A
Thermocouple
adapter+lead kit,
J and K types
Includes thermocouple adapter, thermocouple bead J-type and
thermocouple bead K-type.
• T/C adapter J/K-type
• T/C bead J-type: –20 °C to 200 °C
• T/C bead K-type: –20 °C to 200 °C
U1181A
Immersion
temperature probe
• Type-K T/C for use in oil and other liquids
• Measurement range: –50 °C to 700 ºC
• Includes adapter U1184A for connection to DMM
U1182A
Industrial surface
temperature probe
• Type-K T/C for use on still surfaces
• Measurement range: –50 ºC to 400 ºC
• Includes adapter U1184A for connection to DMM
U1183A
Air temperature
probe
• Type-K T/C for use in air and non-caustic gas
• Measurement range: –50 ºC to 800 ºC
• Includes adapter U1184A for connection to DMM
U1184A
Temperature probe
adapter
• Mini-connector-to-banana-plug adapter for use with DMM
U1185A
J-type thermocouple
and adapter
• T/C adapter J/K-type
• T/C bead J-type: –20 °C to 200 °C
U1186A
K-type
thermocouple and
adapter
• T/C adapter J/K-type
• T/C bead J-type: –20 °C to 200 °C
Carrying case
U1172A
Transit case
(aluminium-clad)
The robust casing to transport your DMM and accessories
• Aluminum-clad, black panel construction
• Dimension: 18 inches (H) x 13 inches (W) x 6 inches (D)
• Weight: 4 kg
U1174A
Soft carrying case
The convenient way to carry your DMM and essential accessories
• Dimension: 9 inches (H) x 5 inches (W) x 3 inches (D)
Hanging kit
U1171A
Magnetic hanging
kit
For fastening the DMM to a steel surface so both hands are free
U1179A IR
Connectivity Bracke
Connect with U1177A IR-to-Bluetooth adapter or U1173A IR-USB
cable
7
www.agilent.com
www.agilent.com/find/handhelddmm
For more information on Agilent
Technologies’ products, applications or
services, please contact your local Agilent
office. The complete list is available at:
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
www.axiestandard.org
AdvancedTCA® Extensions for
Instrumentation and Test (AXIe) is
an open standard that extends the
AdvancedTCA for general purpose
and semiconductor test. Agilent
is a founding member of the AXIe
consortium.
Agilent Advantage Services is committed
to your success throughout your equipment’s lifetime. To keep you competitive,
we continually invest in tools and
processes that speed up calibration and
repair and reduce your cost of ownership.
You can also use Infoline Web Services
to manage equipment and services more
effectively. By sharing our measurement
and service expertise, we help you create
the products that change our world.
www.agilent.com/find/advantageservices
www.lxistandard.org
LAN eXtensions for Instruments puts
the power of Ethernet and the Web
inside your test systems. Agilent
is a founding member of the LXI
consortium.
www.pxisa.org
PCI eXtensions for Instrumentation
(PXI) modular instrumentation
delivers a rugged, PC-based highperformance measurement and
automation system.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilent’s
measurement expertise and product
breadth, combined with channel
partner convenience.
Agilent Electronic Measurement Group
DEKRA Certified
ISO 9001:2008
Quality Management Sys
System
www.agilent.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
(11) 4197 3600
01800 5064 800
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 375 8100
www.agilent.com/quality
Europe & Middle East
Belgium
32 (0) 2 404 93 40
Denmark
45 45 80 12 15
Finland
358 (0) 10 855 2100
France
0825 010 700*
*0.125 €/minute
Germany
Ireland
Israel
Italy
Netherlands
Spain
Sweden
United Kingdom
49 (0) 7031 464 6333
1890 924 204
972-3-9288-504/544
39 02 92 60 8484
31 (0) 20 547 2111
34 (91) 631 3300
0200-88 22 55
44 (0) 118 927 6201
For other unlisted countries:
www.agilent.com/find/contactus
Revised: January 6, 2012
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2012
Published in USA, March 1, 2012