AGILENT E1476A

Agilent E1476A 64-Channel
3-Wire T/C Relay Multiplexer
Data Sheet
Features
• 1-Slot, C-size, register based
• Low-thermal offset relay <2 µV
• 64 channels of temperature
with compensation
• 64 channels of voltage - 3-wire
high, low, and guard
• 64 channels 2-wire and
32 channels 4-wire resistances
• Includes QUIC easy-to-use
terminal block
Description
The Agilent E1476A High-Density
Reed Relay Multiplexer is a
C-size, 1-slot, register-based
VXI module. This low-offset,
thermocouple compensated
multiplexer is dynamically
configurable providing 64 channels of two-, three- or four-wire
(32 channels) of switching. This
multiplexer module consists of
a component card with switches
and a QUIC spring clamp terminal block that plugs onto the
component card. The E1476A is
ideal for applications needing a
relay multiplexer that is dynamically configurable, and makes
maximum high-quality, high-point
count measurements.
High-integrity voltage measurements are possible with
three-wire high, low, and guard
switching. In addition to making
two-wire resistance and precision
four-wire resistance measurements, you can make up to
64 channels of thermocouple
temperature measurements
with automatic cold junction
compensation.
Refer to the Agilent Technologies
Website for instrument driver
availability and downloading
instructions, as well as for
recent product updates, if
applicable.
Temperature Measurements
The reference thermistor is also
accessible by both banks, each
bank having a control switch
allowing for either a two-wire or
four-wire resistance measurement of the 5000 Ω reference
thermistor mounted on the
isothermal plane located in the
terminal block. Using a scanning
multimeter configuration, the
channel relays and five control
relays are programmed by SCPI
commands or by register read/
writes. SCPI command syntax
to make a temperature scan
of K type thermocouples is:
MEAS:TEMP? TC, K, (@100:163)
Configuration
Product Specifications
Each of the 64 channels provides
separate high, low, and guard
connections, all easily accessible
via the quick connect screwless
terminals on the companion
terminal block. The multiplexer
is organized in two banks of
32 with each bank having its own
voltage sense control switch and
one bank having a current source
control switch. This dual bank
configuration makes it possible
to use half the channels as sense
channels, while the other half are
used as current source channels,
thus obtaining 32 four-wire measurement channels, each with
high, low, and guard connections.
Input
One 6 cm (2.5-in) analog bus
cable (E1400-61605) is shipped
with each module to allow you to
interconnect the E1411B 5.5-digit
multimeter to one or more E1476A
multiplexers via its front panel
analog bus connector. For connection to an external voltmeter
or other VXI multimeter with
conventional front panel connectors, access to the analog bus
lines is available in the terminal
block. This allows you to connect
the analog bus signal lines to the
multimeter inputs using ordinary
hookup wire.
AC
DC:
Maximum voltage
120 Vdc
(any terminal to any other terminal or chassis):
AC rms: Maximum voltage
120 V rms
(any terminal to any other terminal or chassis):
Maximum current
(per channel common, non-inductive):
35 mA
Maximum power per channel:
4 VA
Maximum thermal
offset per channel, differential Hi-Lo:
<4 µV, <2 µV
(10 samples averaged)
Closed channel resistance:
100 Ω ±5 Ω
DC
Insulation resistance (between any two points): 10E9 Ω, 40ºC, 95% RH
Insulation resistance (Hi to Lo, power off):
n/a
Maximum bandwidth
(-3 dB, 50 Ω source/load):
100 kHz
Crosstalk (channel-to-channel):
100 kHz:
-70 dB
10 MHz:
-45 dB
Both:
n/a
Closed channel capacitance:
<175 pf H-L, <300 pf L-G,
<1500 pf G-C
General Characteristics
Relays:
Power down state:
Power up state:
Minimum relay life:
No load:
Rated load:
Reed relays
Break-before-make
Relays open on power down
Relays open on power up
5 x 10E9 operations
10E7 operations
Reference junction measurement
accuracy (18 to 28 ºC operating):
0.38 °C
Strain gage excitation:
n/a
Screw terminal wire size:
22 to 26 AWG (0.5, 0.75, 0.9 mm)
Scanning rate:
333 channels/s typ.
2
General Specifications
Ordering Information
VXI Characteristics
Description
Product No.
64-channel 3-wire T/C Relay Multiplexer
E1476A
Pre-QUIC-type terminal block
E1476A 106
Crimp-and-insert terminal block**
E1476A A3E**
Service Manual
E1476A OB3
Extra pre-QUIC-type terminal block (if ordered seperately)
E1476-80000
Extra QUIC-type terminal block (if ordered seperately)
E1476-80010
Extra crimp-and-insert terminal block (if ordered seperately)**
E1476-80011**
VXI device type:
Register-based,
A16, slave only
Size:
C
Slots:
1
Connectors:
P1/2
Shared memory:
None
VXI busses:
None
C-size capability:
n/a
Instrument Drivers
See the Agilent Technologies Website
http://www.agilent.com/find/inst_drivers
for availability and downloading.
Command module
firmware:
Downloadable
Command module firmware rev.:
Notes:
** Crimp-and-insert contacts are not included.
Crimp-and-insert contacts order 1252-6533 or ERNI 014728
Crimp-and-insert tools
Hand crimp tool 8710-2306 or ERNI 014374
Pin extractor tool 8710-2307 or ERNI 471555
Extra crimp and insert connectors
96 pin connector body 1252-6532 or ERNI 024069
160 pin connector body 1252-6531 or ERNI 024070
A.06
I-SCPI Win 3.1:
Yes
I-SCPI Series 700:
Yes
C-SCPI LynxOS:
Yes
C-SCPI Series 700:
Yes
Panel Drivers:
Yes
VXIplug&play Win Framework:
Yes
Agilent E1476A terminal block
VXIplug&play Win 95/NT Framework: Yes
VXIplug&play HP-UX Framework:
No
Module Current
+5 V:
+12 V:
-12 V:
+24 V:
-24 V:
-5.2 V:
-2 V:
IPM
IDM
0.1
0
0
0
0
0
0
0.1
0
0
0
0
0
0
Cooling Slot
Watts/slot:
4.00
∆P mm H2O:
0.10
Air Flow liter/s:
0.30
3
Agilent E1476A Circuit Diagram
Remove all doubt
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
www.agilent.com/find/open
Agilent Open simplifies the process
of connecting and programming
test systems to help engineers
design, validate and manufacture
electronic products. Agilent offers
open connectivity for a broad range
of system-ready instruments, open
industry software, PC-standard I/O
and global support, which are
combined to more easily integrate
test system development.
www.lxistandard.org
LXI is the LAN-based successor to
GPIB, providing faster, more efficient
connectivity. Agilent is a founding
member of the LXI consortium.
Our repair and calibration services
will get your equipment back to you,
performing like new, when promised. You will get full value out of
your Agilent equipment throughout its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confidence in your measurements.
Agilent offers a wide range of additional expert test and measurement services for your equipment,
including initial start-up assistance
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/find/removealldoubt
www.agilent.com
For more information on Agilent Technologies’
products, applications or services, please
contact your local Agilent office. The complete
list is available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
Europe & Middle East
Austria
Belgium
Denmark
Finland
France
0820 87 44 11
32 (0) 2 404 93 40
45 70 13 15 15
358 (0) 10 855 2100
0825 010 700*
Germany
01805 24 6333**
*0.125 € fixed network rates
**0.14 €/minute
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0) 20 547 2111
Spain
34 (91) 631 3300
Sweden
0200-88 22 55
Switzerland (French)
41 (21) 8113811(Opt 2)
Switzerland (German) 0800 80 53 53 (Opt 1)
United Kingdom
44 (0) 118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: October 24, 2007
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2008
Printed in USA, January 11 2008
5965-5607E