FUJIKURA FCX-MQ-F-AC

Oxygen sensor module
FCX-M Data sheet
■Features
・High accuracy
・Long life (More than 3 years in normal air) *1
・Continues calibration and maintenance not required *2
・Calibration gas not required
・From ppm to 95%O2, wide measurement range
・No interfarance to media
・No pressure dependence
■Applications
・Oxygen concentrator
・Nitrogen purifier
・Alarm for oxygen shortage
・Culture oven ro Incubator
・Combustion control
・Soldering system
FCX-M series
Model
RoHS compliance
■Specifications
Item
FCX-MP-F-AC
FCX-MQ-F-AC
Measurement range
Accuracy
50 to 1000ppmO2
+/-5%FS
50 to 1000 mV
0.05 to 1 %O2
+/-5%FS
50 to 1000 mV
Output (Logarithmic)
3
Response time *
4
Power supply *
Power consumption
Operating temperature
Operating humiditiy
*5
Dimensions
Weight
Item
Measurement range
Accuracy
Output (Linear)
3
Response time *
4
Power supply *
Power consumption
Operating temperature
Operating humiditiy
Dimensions
Weight
*
1
2
*
3
*
4
*
5
*
*5
Vout=-999.5x10
2
x LN 1- ppmO
106
3
FCX-MV-F
FCX-MV-F-AC
0.1 to 25 %O2
+/-1%FS
1 to 250 mV
FCX-MW-F
FCX-MW-F-AC
0.1 to 95 %O2
+/-0.5%FS
0 to 2669 mV
3
Vout= -99.5x10
x LN 1- %O2
100
90sec. (for 90% value)
30sec. (for 90% value)
Depends on operating environment.
Annual calibration or maintenance still recommended.
From 10 to 90% step change.
Regurated DC power supply.The current capacity must be 1A.
Without vapour condensation.
Note ; Please read above instructions before using the sensor.
Fujikura reserves right to change specifications without notice.
If you have any questions regarding technical issues or specifications,
please contact us.
Fujikura Ltd. Sensor Department
5-1 Kiba 1-chome, Koto-ku, Tokyo 135-8512, Japan
Phone +81-(0)3-5606-1072 E-mail : [email protected]
75 to 95 %O2
+/-2%O2
1235 to 2669 mV
FCX-MEP2-F
FCX-MEP2-F-AC
0.1 to 25 %O2
+/-1%FS
4 to 20 mA
Output= -57x
Vout= -891x LN 1- %O2
100
DC5+/-0.2V
5W max.
-10 to 50 deg.C
0 to 85%RH
50x16x104 mm
Approx. 65grams
FCX-MVL-F
FCX-MWL-F
FCX-MVL-F-AC
FCX-MWL-F-AC
0.1 to 25 %O2
0.1 to 95 %O2
+/-1%FS
+/-1%FS
1 to 250 mV
1 to 950 mV
Vout= 10xO2%
30sec. (for 90% value)
DC5+/-0.2V
5W max.
-10 to 50 deg.C
0 to 85%RH
50x38x104 mm
Approx. 120grams
FCX-MWGP-2A-F
LN 1- %O2
100
+4
30sec. (for 90% value)
DC5+0.1/-0.25V
DC12+/-3V
4W max.
3W
0 to 60 deg.C
-10 to 50 deg.C
40 to 85%RH
0 to 85%RH
40x23x80 mm
45x20x52 mm
Approx. 40grams
Approx. 40grams
Instruction for use ;
1. Sensor modules are calibrated with N2-O2 balance gases. Other gases may affect
the sensor performance by increase errors.
2. Do NOT separate the sensing element from the circuit board and lead wires length
must be remained unchanged.
3. Do NOT use in harsh environmental conditions containing halogen atoms(F,Cl,Br) ,
SOx ,Nox or H2S which make the sensor inferior in a short time.
4. Flammable gases ,such as H2 ,CO ,methane or Alcohol make a big error. And please
note that the temperature at the sensing part goes uo to 450 degrees C during the
operation,which may exceed the flash point and cause explosion.
5. Silicone gas or vapour including Siloxane also make the sensor inferior in a short time.
6. Dust or oil mist cause failures of the products,errors or a slow response. They must be
eliminated with a filter.
7. Water drop's contact may give a damage to sensor. It must be eliminated with a filter.
8. Temperature at the sensor mesh surface reach 50~80 degrees C during operation.
Please take precautions against burning yourseif.
9. A strong shock such as drop may cut internal wires or break the sensor pellet which is
made of ceramic. Do NOT apply a shock of 10G or more to the sensor.
10. The warranty period is one year from the ex-factory date. This warranty does NOT
apply to the sensors as follows ;
1) There are any defects to faults caused by an improper dealing during the
transpotation after Fujikura has delivered the sensor to place where the buyer had
instructed.
2) There are any defects or faults caused by the buyer's misuses ,abuse or neglect.
3) The buyer fixed or remake the sensors.
4) There are any indentical or consequential damages which are given in the usage.
5) There are any defects or faults caused by natural disasters such as life ,earthquake,
flood or thunder.
11. Fujikura's oxygen sensors are NOT developed ,designed ,manufactured ,sold ,
intended or authorized for use as components in systems intended for the surgical
implant into the body ,other applications intended to support or sustain life ,fail-safe
applications in which the failures ,breakages or where misused of the sensors could
create a situation where personal injury or death ,explosion or fire ,or serious social
damage may occur. Fujikura and its subcontractors and distributors accept NO
responsibility the buyer's selection and use of Fujikura's oxygen sensors without
Fujikura's written approval in any such unintended or unauthorized and situation where
personal injury or death ,explosion or fire ,or serious social damage may occur.
12. A fail-safe design is strongly required when customers use the sensor in
medecal applications or alarm systems for oxygen shortage even except above.
Caution!
Please carefully read the followings before using the oxygen sensor products.
1.Applications for medical appliances, life-support equipment and low oxygen detectors
(1)
Fujikura products are not designed, intended or approved for use as components of surgical or
life support systems, or other applications that may cause injury or death as a result of failure.
In unapproved applications or uses where the customer implies, directly or indirectly, resultant
injuries or deaths are due to Fujikura, Fujikura affiliates and agencies (citing for example, a
design or manufacture fault), Fujikura, Fujikura affiliates and agencies shall be free from
responsibility relating to any claims, costs, losses, and compensation.
(2)
When a Fujikura product is to be used in medical appliances and oxygen detectors other than
those mentioned above, it is strongly advised that fail-safe designs are established.
Fujikura
should be consulted for the necessary information.
2.Service life and guarantee period
(1)
The end of service life shall be defined as the time when the output no longer meets
the specified precision.
(2)
The guarantee period is for one year from the date of shipment.
During the
guarantee period, should defects occur under normal conditions of use as specified in
the manual and within the service life, the product will be repaired or replaced
without charge.
However, a repair or replacement fee will be charged in the
following cases.
Defect or damage due to inappropriate transportation or handling after delivery.
Defect or damage caused by misuse, abuse or careless handling.
Defect or damage due to unauthorized repairs or changes in configuration
Damage to the cosmetic appearance caused during use
Damage from fire, earthquake, flood or other natural disasters and abnormal
voltage.
3.Operational precautions
3.1 Measurement of atmospheric gases
(1) Calibration gas
The sensor should be adjusted with a calibration gas that is a mixture of nitrogen, N2
(or Argon, Ar) and oxygen, O2.
Other balance gases may result in incorrect
measurements.
(2) Combustible gases
Fujikura Ltd. Sensor Department
1-5-1, Kiba, Koto-ku, Tokyo 135-8512, Japan
phone +81-3-5606-1072
E-mail : [email protected]
1
An atmosphere continuing combustible gases such as methane, alcohol, hydrogen and
carbon monoxide may cause errors in measurement.
Since the sensor element
functions at 450°C, gases that ignite below that temperature must not be used.
(3) Silicon gases
Never use silicon gases containing siloxane, as these gases react with the sensor and
produce oxides, destroying the performance of the sensor over a very short period.
(4) Fluorocarbons
Do not use freons and others that contain halogens (F, Cl and Br), as these gasses
react with materials inside the sensor and damage the performance.
(5) Sulfur oxides and hydrogen
Never use sulfur oxides (SOx) and hydrogen (H2), as they react with the sensor and
destroy the performance of the sensor over a very short period.
3.2 Operating conditions
(1) Dust and oil mist
Employ a filter system to eliminate dust and oil mists that clog the sensor and
analyzer filter, resulting in problems, measurement errors and incorrect responses.
(2) Water and condensation moisture
Contact of the sensor with water may destroy the sensor. Exclude water from the
system.
3.3 Others
(1)
Do not touch the sensor mesh while in operation, since the sensor mesh is heated to 50
- 80°C.
(2)
Do not subject the sensor to a shock of 10G or greater which may cause breaks in the
wiring and cracks in the sensor chip.
(3)
The sensor element is made of a ceramic material. Never expose it to heat suddenly
as this could destroy the element.
4. Others
Any product described in the catalogues may be altered without prior notice to improve
reliability, function or design. Fujikura is not responsible for any incidents due to
application of products and circuits described in the catalogues. No Fujikura patents or
rights are licensed to a third party.
1st October , 2011
Fujikura Ltd. Sensor Department
1-5-1, Kiba, Koto-ku, Tokyo 135-8512, Japan
phone +81-3-5606-1072
E-mail : [email protected]