POWERBOX PM200-45-1A

MEDICAL & MILITARY
PM200 Series
2 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Features
• Low safety ground leakage current.
• Autoranging input.
• Overvoltage protection.
• Overcurrent protection.
• Power fail detection (PFD) signal.
• Optional cover and fan assembly.
• 100% burn-in at full rated load.
• Short curcuit protected with auto-recvery
Specifications
OPERATING
INPUT
Input Voltage
Frequency
47Hz–67Hz
Input Current
4.5A (rms) for 115Vac.
2.6A (rms) for 230Vac.
450,000 hrs @ 25°C
Efficiency
70% minimum.
ENVIRONMENTAL
Operating Temperature
0°C to 70°C
Leakage current
90µA max. @ 115Vac. 60Hz.
150µA max. @ 230Vac. 50Hz.
Derating
Derate to 50% at 70°C
Isolation
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Storage Temperature
-40°C to 85°C
Relative Humidity
5%–95% non-condensing
Cooling
25 CFM forced air, provided on “C” version, to be
provided for “B”version by user.
Inrush Current
20A @ 115V, 40A @ 230V
STANDARDS AND APPROVALS
OUTPUT
342
MTBF
90-130/180-264Vac (Autoranging)
Output Voltage
See table
Safety standards
UL2601-1, CSA C22.2, TUV EN60601-1
Output Voltage Adjustment
V1 ±10%
C-Tick
AS/NZS CISPR11 Group 1, Class A
Output Current
See table
EMC standards
EN60601-1-2: 2001,EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
Ripple & Noise
2% peak to peak
EMI standards
EN55011 Class B, FCC Class B, VCCI Class B
Line Regulation
±0.5% maximum at full load.
Load Regulation
See table
Transient Response
Max. deviation 4% recovering to 1% within 500µs
after 25% load charge
Overvoltage Protection
Provided on V1 only. Set at 110%–132% of
nominal
Overcurrent Protection
All outputs protected for short circuit conditions
Temperature Coefficient
All outputs ±0.04% per °C
Holdup Time
Typically 12mS at 110VAC
PFD Signal
Power fail detection signal. TTL logic low upon
loss of input power.
MECHANICAL
Dimensions
229x127x65mm
Weight
1.4Kgs for “B” version.
1.6Kgs for “C” version.
MEDICAL & MILITARY
PM200 Series
2 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Selection Table
MAX.
MODEL
OUTPUT #1
OUTPUT #2 (2)
IMAX TOL VNOM
OUTPUT #4 (3)
IMIN IMAX TOL VNOM
IMIN
IMAX
OUTPUT
VNOM
IMIN
PM200-10
5V
3.0A
40A
2%
200W
PM200-12
12V
1.2A
16.7A
2%
200W
PM200-13
15V
1.0A
13.4A
2%
200W
PM200-14
24V
0.6A
8.4A
2%
200W
PM200-16
30V
0.5A
6.7A
2%
200W
PM200-18
48V
0.5A
4.2A
2%
PM200-23
+5V
3.0A
30A
2%
+12V
0A
8A
4%
200W
PM200-24
+5V
3.0A
30A
2%
+15V
0A
6A
4%
200W
PM200-25
+5V
3.0A
30A
2%
+24V
0A
4A
4%
200W
PM200-30
+5V
3.0A
30A
2%
+12V
0A
8A
4%
-5V
0A
6A
4%
200W
PM200-31
+5V
3.0A
30A
2%
+12V
0A
8A
4%
-12V
0A
4A
4%
200W
PM200-32
+5V
3.0A
30A
2%
+15V
0A
6A
4%
-15V
0A
4A
4%
200W
PM200-33
+5V
3.0A
30A
2%
+15V
0A
6A
4%
-12V
0A
4A
4%
PM200-40
+5V
3.0A
30A
2%
+12V
0A
8A
4%
-12V
0A
4A
4%
5V
0A
6A
4%
200W
PM200-41
+5V
3.0A
30A
2%
+15V
0A
6A
4%
-15V
0A
4A
4%
24V
0A
4A
4%
200W
PM200-42
+5V
3.0A
30A
2%
+12V
0A
8A
4%
-12V
0A
4A
4%
12V
0A
4A
4%
200W
PM200-44
+5V
3.0A
30A
2%
+12V
0A
8A
4%
-15V
0A
4A
4%
15V
0A
4A
4%
200W
PM200-45
+5V
3.0A
30A
2%
+12V
0A
8A
4%
-12V
0A
4A
4%
24V
0A
4A
4%
200W
NUMBER(1)(5)(6)
IMAX TOL VNOM IMIN
OUTPUT #3
TOL POWER (4)
200W
200W
NOTES: (1) Add suffix “B” for U-bracket format and “C” for enclosed format.
(2) Peak current is 12A on +12V, 9A on +15V and 6A on +24V.
(3) Output #4 is floating. It can be connected externally for positive or negative output.
(4) 200 watts for “C” version with a cover and fan assembly 150 watts for “B” version without moving air (maximum current of ouput #1 & #2 derated to 70%),
or 200 watts with 25 CFM forced air provided by user.
(5) When the remote Sense facility is not used, +Sense must be connected to +V, and -Sense to Return, on P2 connector.
(6) All models may be operated at no-load. At no-load, output voltage tolerance increases to 10%.
Pin Chart
P1
MODEL/PIN
PM200-10
PM200-13
PM200-16
PM200-23
PM200-25
PM200-30
PM200-32
PM200-40A
PM200-42A
PM200-45-1A
PM200-12
PM200-14
PM200-18
PM200-24
PM200-31
PM200-33
PM200-41
PM200-44
P2
6
P3
1
2
3
1,2
3,4,5
7
AC
Live
AC
Neutral
AC
Ground
-Sense
RET.
Output 1 Output 1 Output 1 +Sense
AC
Live
AC
Live
AC
Live
AC
Neutral
AC
Neutral
AC
Neutral
AC
Output 1
Ground
AC
Output 1
Ground
AC
Output 1
Ground
COM
RET.
COM
RET.
COM
RET.
Output 2
N.C.
Output 2 Output 3
Output 2 Output 3
8
9
1
2
3
4
Fan
COM
RET.
COM
RET.
P.F.D.
COM
RET.
COM
RET.
COM
RET.
COM
RET.
P.F.D.
COM
RET.
COM
RET.
P.F.D.
N.C.
N.C.
Fan
N.C.
N.C.
Fan
Output 4
Output 4 Fan
Return
P.F.D.
Technical Illustration
Your dependable power partner – www.powerbox.com.au