SENSORTECHNICS 40PC500G1A

40PC Series
Fully signal conditioned pressure transducer
FEATURES
· Pressure ranges from 0 ... ± 50 mmHg
to 0 ... 500 psi gage
· High accuracy
· TTL power supply
· Ratiometric 0.5 to 4.5 V output
· Dual-Inline pinning for easy
PCB-mounting
· Excellent media compatibility
MEDIA COMPATIBILITY
P1: dry gases only
P2: all media compatible with glass, silicon, stainless steel, Invar, Sn/Ni plating or Sn/Ag solder
SPECIFICATIONS
ELECTRICAL CONNECTION
Maximum ratings
Supply voltage (VS)
Output current
Source
Sink
Output load capacitance
Temperature limits
Operating
Storage
Compensated
Humidity (non-condensing)
November 2007 / 573
4.75 ... 5.25 V
0.5 mA
1.0 mA
10 nF
NOTES:
-45 to +125°C
-55 to +125°C
-45 to +125°C
1. Square corner
marks pin 1 (Vs)
0 - 95 %RH
2. Output is short
circuit protected
1/8
www.sensortechnics.com
40PC Series
Fully signal conditioned pressure transducer
PRESSURE RANGE SPECIFICATIONS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Part number
Operating pressure range
Proof pressure1
Sensitivity (typ.)
40P C 001B 2A
0 ... ±50 mmHg
170 mmHg
40.0 mV/mmHg
40PC006G2A
0 ... 300 mmHg
650 mmHg
13.33 mV/mmHg
40PC015G2A
0 ... 15 psig
4 5 p si
266.7 mV/psi
40P C 015V 2A
0 ... -15 psig
3 0 p si
-266.7 mV/psi
40PC030G2A
0 ... 30 psig
6 0 p si
133.3 mV/psi
40PC100G2A
0 ... 100 psig
2 0 0 p si
40.0 mV/psi
40PC150G2A
0 ... 150 psig
3 0 0 p si
25.7 mV/psi
40PC250G2A
0 ... 250 psig
5 0 0 p si
16.0 mV/psi
40PC500G2A
0 ... 500 psig
1 0 0 0 p si
8.0 mV/psi
Specification notes:
1.
2.
3.
4.
5.
6.
2/8
Proof pressure is the maximum pressure which may be applied without causing damage to the sensing element.
Non-linearity - the maximum deviation of measured output at constant temperature, from "Best Straight Line" through
three points (offset pressure, full scale pressure and 1/2 full scale pressure).
Thermal effects tested and guaranteed in the specified temperature ranges relative to 25°C. All specifications shown
are relative to 25°C.
Span is the algebraic difference between lowest and highest specified pressure.
Total accuracy is the maximum deviation from the 25°C reference transfer function at any pressure or temperature over
the specified ranges. This calculation includes null, span, linearity, hysteresis, repeatability, null shift and span shift.
Response time for 0 to full-scale pressure step change, readings taken at 10 % and 90 % of full-scale pressure.
November 2007 / 573
www.sensortechnics.com
40PC Series
Fully signal conditioned pressure transducer
40PC001B2A
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Min.
Operating pressure
-50
Zero pressure offset
2.45
Typ.
2.50
4
Unit
+50
mmHg
2.55
4.0
Span
Output
Thermal effects (0 to 50°C)3
at -50 mmHg
0.5
at +50 mmHg
4.5
V
Offset
±1.50
Span
±1.50
2
Non-linearity (BSL)
Hysteresis and repeatability
±0.15
Ratiometricity error (VS = 4.75 ... 5.25 V)
±0.25
Current consumption (no load)
Response time
Max.
6
40PC006G2A
±0.8
%FSS
10.0
mA
1
ms
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Min.
Typ.
Max.
Unit
300
mmHg
Operating pressure
0
Zero pressure offset
0.45
0.50
0.55
3.95
4.00
4.05
4
Span
Full scale output
Thermal effects3
4.5
Offset
Span
-18 to +63°C
±1.25
-45 to +125°C
±2.75
-18 to +63°C
±0.75
-45 to +125°C
±2.0
Combined offset and span -18 to +63°C
Total accuracy5
±1.5
-45 to +125°C
±4.0
-45 to +125°C
±4.0
Non-linearity (BSL)2
±0.15
Ratiometricity error (VS = 4.75 ... 5.25 V)
±0.25
Current consumption (no load)
November 2007 / 573
%FSS
±0.2
Hysteresis and repeatability
Response time6
V
10.0
1
mA
ms
3/8
www.sensortechnics.com
40PC Series
Fully signal conditioned pressure transducer
40PC015G2A
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
Zero pressure offset
Span4
Full scale output
Thermal effects3
0
0.39
3.89
Offset
Span
Combined offset and span
Total accuracy5
Non-linearity (BSL)2
Hysteresis and repeatability
Ratiometricity error (VS = 4.75 ... 5.25 V)
Current consumption (no load)
Response time6
40PC015V2A
Min.
-18
-45
-18
-45
-18
-45
-45
to
to
to
to
to
to
to
Typ.
0.50
4.00
4.5
+63°C
+85°C
+63°C
+85°C
+63°C
+85°C
+85°C
Max.
Unit
15
0.61
4.11
p si g
±2.0
±2.75
±0.75
±1.0
±2.0
±3.0
±4.0
±0.2
V
%FSS
±0.15
±0.25
10.0
mA
ms
Max.
Unit
-15
0.54
4.1
p si g
1
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
Zero pressure offset
Span4
Full scale output
Thermal effects3
0
0.46
3.9
Offset
Span
Combined offset and span
Total accuracy5
Non-linearity (BSL)2
Hysteresis and repeatability
Ratiometricity error (VS = 4.75 ... 5.25 V)
Current consumption (no load)
Response time6
4/8
Min.
Typ.
0.50
4.0
4.5
-18 to +25°C
-45 to +85°C
-18 to +25°C
+25 to +85°C
-45 to +25°C
-18 to +25°C
+25 to +85°C
-40 to +25°C
+45 to +85°C
±0.75
±1.0
±2.0
±1.25
±2.25
±1.5
±1.0
±1.75
±2.25
±0.3
V
%FSS
±0.15
±0.25
10.0
1
mA
ms
November 2007 / 573
www.sensortechnics.com
40PC Series
Fully signal conditioned pressure transducer
40PC030G2A
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
Zero pressure offset
Span4
Full scale output
Thermal effects3
0
0.46
3.92
Offset
Span
Combined offset and span
Total accuracy5
Non-linearity (BSL)2
Hysteresis and repeatability
Ratiometricity error (VS = 4.75 ... 5.25 V)
Current consumption (no load)
Response time6
40PC100G2A
Min.
-18
-45
-45
-18
-45
-45
-18
-45
-45
to
to
to
to
to
to
to
to
to
Typ.
0.50
4.00
4.5
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+63°C
+125°C
+125°C
Max.
Unit
30
0.54
4.08
p si g
±1.5
±2.0
±2.25
±1.5
±2.0
±2.25
±2.0
±2.25
±2.25
±0.3
V
%FSS
±0.15
±0.25
10.0
mA
ms
Max.
Unit
100
0.54
4.09
p si g
1
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
Zero pressure offset
Span4
Full scale output
Thermal effects3
0
0.46
3.91
Offset
Span
Combined offset and span
Total accuracy5
Non-linearity (BSL)2
Hysteresis and repeatability
Ratiometricity error (VS = 4.75 ... 5.25 V)
Current consumption (no load)
Response time6
November 2007 / 573
Min.
-18
-45
-45
-18
-45
-45
-18
-45
-45
-45
to
to
to
to
to
to
to
to
to
to
Typ.
0.50
4.0
4.5
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+125°C
±1.25
±2.0
±3.0
±0.75
±1.0
±2.0
±1.5
±2.5
±3.0
±3.0
±0.1
V
%FSS
±0.15
±0.25
10.0
1
mA
ms
5/8
www.sensortechnics.com
40PC Series
Fully signal conditioned pressure transducer
40PC150G2A
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
Zero pressure offset
Span4
Full scale output
Thermal effects3
0
0.46
3.93
Offset
Span
Combined offset and span
Total accuracy5
Non-linearity (BSL)2
Hysteresis and repeatability
Ratiometricity error (VS = 4.75 ... 5.25 V)
Current consumption (no load)
Response time6
40PC250G2A
Min.
-18
-45
-45
-18
-45
-45
-18
-45
-45
-45
to
to
to
to
to
to
to
to
to
to
Typ.
0.50
4.0
4.5
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+125°C
Max.
Unit
150
0.54
4.07
p si g
±0.75
±1.0
±1.5
±0.75
±1.0
±1.5
±0.75
±1.0
±1.5
±2.5
±0.1
V
%FSS
±0.15
±0.25
10.0
mA
ms
Max.
Unit
250
0.54
4.07
p si g
1
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
Zero pressure offset
Span4
Full scale output
Thermal effects3
0
0.46
3.93
Offset
Span
Combined offset and span
Total accuracy5
Non-linearity (BSL)2
Hysteresis and repeatability
Ratiometricity error (VS = 4.75 ... 5.25 V)
Current consumption (no load)
Response time6
6/8
Min.
-18
-45
-45
-18
-45
-45
-18
-45
-45
-45
to
to
to
to
to
to
to
to
to
to
Typ.
0.50
4.0
4.5
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+125°C
±0.75
±1.0
±2.0
±0.75
±1.0
±2.0
±0.75
±1.0
±3.0
±3.0
±0.1
V
%FSS
±0.15
±0.25
10.0
1
mA
ms
November 2007 / 573
www.sensortechnics.com
40PC Series
Fully signal conditioned pressure transducer
40PC500G2A
PERFORMANCE CHARACTERISTICS
(unless otherwise noted, VS = 5.00 V, RL > 100 kΩ, tamb = 25°C)
Characteristics
Operating pressure
Zero pressure offset
Span4
Full scale output
Thermal effects3
Min.
0
0.46
3.91
Offset
Span
Combined offset and span
Total accuracy5
Non-linearity (BSL)2
Hysteresis and repeatability
Ratiometricity error (VS = 4.75 ... 5.25 V)
Current consumption (no load)
Response time6
-18
-45
-45
-18
-45
-45
-18
-45
-45
-45
to
to
to
to
to
to
to
to
to
to
Typ.
0.50
4.0
4.5
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+63°C
+85°C
+125°C
+125°C
Max.
Unit
500
0.54
4.09
p si g
±1.0
±1.5
±2.5
±0.75
±1.0
±1.5
±1.0
±1.75
±3.0
±3.0
±0.2
V
%FSS
±0.15
±0.25
10.0
1
mA
ms
OUTLINE DRAWING
mass: 4 g
November 2007 / 573
dimensions in mm (inches)
7/8
www.sensortechnics.com
40PC Series
Fully signal conditioned pressure transducer
ORDERING INFORMATION
40PC
40PC series
Pressure range
001: 0 ... ±50 mmHg
006: 0 ... 300 mmHg
015:
0 ... 15 psi
015:
0 ... -15 psi
030:
0 ... 30 psi
100:
0 ... 100 psi
150:
0 ... 150 psi
250:
0 ... 250 psi
500:
0 ... 500 psi
xxx
x
xA
Lead forming
1: unformed (straight)
2: formed away from port
3: formed towards port
2: stocking devices
1, 3: call Sensortechnics for availability
Pressure mode
B: bidirectional gage pressure: available for ±50 mmHg only
G: gage pressure
V: vacuum gage: available for -15 psi only
Sensortechnics reserves the right to make changes to any products herein. Sensortechnics does not assume any liability arising out of the
application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
8/8
November 2007 / 573