TI ADS1259IPW

ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
Industrial, 14kSPS, 24-Bit Analog-to-Digital Converter
with Low-Drift Reference
Check for Samples: ADS1259
FEATURES
DESCRIPTION
•
•
•
The ADS1259 is a high-linearity, low-drift, 24-bit,
analog-to-digital converter (ADC) designed for the
needs of industrial process control, precision
instrumentation, and other exacting applications.
Combined with a signal amplifier (such as the
PGA280),
a
high-resolution,
high-accuracy
measurement system is formed that is capable of
digitizing a wide range of signals.
1
23
•
•
•
•
•
•
•
•
•
24 Bits, No Missing Codes
Output Data Rates From 10 To 14kSPS
High Performance:
– INL: 0.4ppm
– Reference Drift: 2ppm/°C
– Gain Drift: 0.5ppm/°C
– Offset Drift: 0.05μV/°C
– Noise: 0.7μVRMS at 60SPS
Simultaneous 50/60Hz Rejection at 10SPS
Single-Cycle Settling
Internal Oscillator
Out-of-Range Detection
Readback Data Integrity by Checksum and
Redundant Data Read Capability
SPI™-Compatible Interface
Analog Supply: +5V or ±2.5V
Digital Supply: +2.7V to +5V
Low Power: 13mW
APPLICATIONS
•
•
•
AINP
VREFP
VREFN REFOUT
DS
Modulator
AINN
SYNCOUT
2.5V
Reference
fCLK/8
Programmable
Digital Filter
Calibration
Engine
DVDD
Clock
Generator
XTAL2
Dissipating only 13mW in operation, the ADS1259
can be powered down, dissipating less than 25μW.
The ADS1259 is offered in a TSSOP-20 package and
is fully specified from –40°C to +105°C.
START
SCLK
DIN
CS
DGND
RELATED PRODUCTS
DRDY
DOUT
ADS1259
AVSS
XTAL1/CLKIN
RESET/PWDN
Control
and
Serial
Interface
Out-of-Range
Detection
The ADS1259 also provides an integrated low-noise,
very low drift 2.5V reference. The on-chip oscillator,
an external crystal, or an external clock can by used
as the ADC clock source.
Data and control communication are handled over a
4MHz, SPI-compatible interface capable of operating
with a minimum of three wires. Data integrity is
augmented by data bytes checksum and redundant
data read capability. Conversions are synchronized
either by command or by pin.
Industrial Process Control
Scientific Instrumentation
Test and Measurement
AVDD
The converter uses a fourth-order, inherently stable,
delta-sigma (ΔΣ) modulator that provides outstanding
noise and linearity performance. The data rates are
programmable up to 14kSPS including 10SPS,
50SPS, and 60SPS that provide excellent normal
mode line-cycle rejection. The digital filter can be
programmed for a fast settling mode where the
conversions settle in a single cycle, or programmed
for a high line-cycle rejection mode. A fast responding
input over-range detector flags the conversion data if
an input over-range should occur.
FEATURES
PRODUCT
24-bit ADC with integrated PGA
ADS1256
Wide-range PGA
PGA280
High-precision PGA; G = 1, 10, 100, 1000
PGA204
1
2
3
Please be aware that an important notice concerning availability, standard warranty, and use in critical applications of Texas
Instruments semiconductor products and disclaimers thereto appears at the end of this data sheet.
SPI is a trademark of Motorola.
All other trademarks are the property of their respective owners.
PRODUCTION DATA information is current as of publication date.
Products conform to specifications per the terms of the Texas
Instruments standard warranty. Production processing does not
necessarily include testing of all parameters.
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
This integrated circuit can be damaged by ESD. Texas Instruments recommends that all integrated circuits be handled with
appropriate precautions. Failure to observe proper handling and installation procedures can cause damage.
ESD damage can range from subtle performance degradation to complete device failure. Precision integrated circuits may be more
susceptible to damage because very small parametric changes could cause the device not to meet its published specifications.
ORDERING INFORMATION
For the most current package and ordering information see the Package Option Addendum at the end of this
document, or see the TI web site at www.ti.com.
ABSOLUTE MAXIMUM RATINGS (1)
Over operating free-air temperature range, unless otherwise noted.
ADS1259
MIN
MAX
UNIT
AVDD to AVSS
–0.3
+5.5
V
AVSS to DGND
–2.8
+0.3
V
DVDD to DGND
–0.3
+5.5
V
Input current, momentary
–100
+100
mA
Input current, continuous
–10
+10
mA
V
Analog input voltage to DGND
AVSS – 0.3
AVDD + 0.3
Digital input voltage to DGND
–0.3
DVDD + 0.3
V
+150
°C
Maximum junction temperature
Operating temperature range
–40
+125
°C
Storage temperature range
–60
+150
°C
(1)
2
Stresses above these ratings may cause permanent damage. Exposure to absolute maximum conditions for extended periods may
degrade device reliability. These are stress ratings only, and functional operation of the device at these or any other conditions beyond
those specified is not implied.
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
ELECTRICAL CHARACTERISTICS
Minimum/maximum specifications are at TA = –40°C to +105°C. Typical specifications are at TA = +25°C, AVDD = +2.5V, AVSS = –2.5V,
DVDD = +3.3V, fCLK = 7.3728MHz, VREF = 2.5V, and fDATA = 60SPS, unless otherwise noted.
ADS1259
PARAMETER
TEST CONDITIONS
MIN
TYP
ADS1259B
MAX
MIN
TYP
MAX
UNIT
ANALOG INPUTS
Full-scale input voltage range (FSR)
VIN = (AINP – AINN)
Absolute input voltage
(AINP, AINN to DGND)
±VREF
AVSS – 0.1
±VREF
AVDD + 0.1
AVSS – 0.1
V
AVDD + 0.1
V
Differential input impedance
120
120
kΩ
Common-mode input impedance
500
500
kΩ
SYSTEM PERFORMANCE
Resolution
No missing codes
Data rate (fDATA)
24
24
10
14400
Bits
10
14400
SPS (1)
±0.0003
±0.001
±0.00004 (2)
±0.0003
%FSR
Offset error
±40
±250
±40
±250
μV
Offset error after calibration (3)
±1
0.05
0.25
μV/°C
±0.05
±0.5
%
2.5
ppm/°C
Integral nonlinearity
Offset drift (4)
Best fit method
TA = –40°C to +105°C
Gain error (5)
Gain error after calibration (3)
Gain drift (4)
Common-mode rejection
Noise
0.05
0.25
±0.05
±0.5
±0.0002
TA = –40°C to +105°C
Normal mode rejection
±1
0.5
μV
±0.0002
2.5
0.5
%
See Figure 42
60Hz, ac (6)
100
See Table 1
120
100
0.7
120
dB
0.7
μV
AVDD, AVSS power-supply
rejection
60Hz, ac (6)
85
95
85
95
dB
DVDD power-supply rejection
60Hz, ac (6)
85
110
85
110
dB
OUT-OF-RANGE DETECTION
Threshold (7)
Level
±105
±105
%FSR
Accuracy
±0.5
±0.5
%FSR
VOLTAGE REFERENCE INPUTS
Reference input range (VREF)
VREF = (VREFP –
VREFN)
0.5
2.5
AVDD – AVSS +
200mV
0.5
2.5
AVDD – AVSS +
200mV
V
Negative reference absolute input
(VREFN to DGND)
AVSS –
100mV
VREFP – 0.5
AVSS –
100mV
VREFP – 0.5
V
Positive reference absolute input
(VREFP to DGND)
VREFN + 0.5
AVDD + 100mV
VREFN + 0.5
AVDD + 100mV
V
200nA +
60nA/V
200nA +
60nA/V
Internal or external clock
0.2
0.2
nA/°C
VREFOUT = (REFOUT –
AVSS)
2.5
2.5
V
Average reference input current (8)
Average reference input current drift
INTERNAL VOLTAGE REFERENCE
Reference output voltage
Accuracy
Temperature drift
±0.4
TA = +25°C
(4)
TA = –40°C to +105°C
10
40
4
TA = 0°C to +85°C
2
–10
Drive current sink and source
Load regulation
10
–10
±0.2
%
12
ppm/°C
5
ppm/°C
10
mA (9)
μV/mA
10
10
Turn-on settling time (10)
±0.001% settling
1
1
s
Long-term stability
0 to 1000 hours
70
70
ppm/1000hr
30
30
ppm
Thermal hysteresis (11)
(1)
(2)
(3)
(4)
(5)
(6)
SPS = samples per second.
Shaded cells indicate improved specifications of the ADS1259B.
Calibration accuracy is on the level of noise (signal and ADC), reduced by the effect of 16 reading averaging.
Reference drift specified by design and final production test. Drift calculated over the specified temperature range using box method.
Excludes internal reference error.
fDATA = 14.4kSPS. Placing a notch of the digital filter at 60Hz (setting fDATA = 10SPS or 60SPS) further improves the common-mode
rejection and power-supply rejection of this input frequency.
(7) Absolute input voltage range for out-of-range specification: AVSS + 150mV ≤ AINP or AINN ≤ AVDD – 150mV.
(8) Over the range: AVSS ≤ VREFP or VREFN ≤ AVDD. For reference voltage exceeding AVDD or AVSS, input current = 150nA/10mV.
(9) Limit the reference output current to ±10mA.
(10) CREFOUT = 1μF, CREFIN = 1μF.
(11) See the Thermal Hysteresis section.
Copyright © 2009–2011, Texas Instruments Incorporated
3
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
ELECTRICAL CHARACTERISTICS (continued)
Minimum/maximum specifications are at TA = –40°C to +105°C. Typical specifications are at TA = +25°C, AVDD = +2.5V,
AVSS = –2.5V, DVDD = +3.3V, fCLK = 7.3728MHz, VREF = 2.5V, and fDATA = 60SPS, unless otherwise noted.
ADS1259
PARAMETER
TEST CONDITIONS
MIN
TYP
ADS1259B
MAX
MIN
TYP
MAX
UNIT
CLOCK SOURCE (fCLK)
Nominal frequency
7.3728
Accuracy
±0.2
±2
2
7.3728
8
Frequency range
0.1
7.3728
Duty cycle
40
Internal oscillator
Frequency range
Crystal oscillator
Start-up time (12)
7.3728
MHz
±0.2
±2
%
2
7.3728
8
MHz
8
0.1
7.3728
8
MHz
60
40
60
%
20
20
ms
External clock
DIGITAL INPUT/OUTPUT (DVDD = 2.7V to 5.25V)
VIH
0.8 DVDD
DVDD
0.8 DVDD
DVDD
V
VIL
DGND
0.2 DVDD
DGND
0.2 DVDD
V
VOH
VOL
IOH = 1mA
0.8 DVDD
IOH = 8mA
0.75 DVDD
IOL = 1mA
V
0.75 DVDD
0.2 DVDD
IOL = 8mA
Input hysteresis
Input leakage
0.8 DVDD
0.2 DVDD
0.2 DVDD
0.1
0.1
0 < VDIGITAL INPUT < DVDD
V
0.2 DVDD
±10
V
V
V
±10
μA
POWER SUPPLY
AVSS
–2.6
0
–2.6
0
V
AVDD
AVSS + 4.75
AVSS + 5.25
AVSS + 4.75
AVSS + 5.25
V
DVDD
2.7
5.25
2.7
5.25
V
3.8
|mA|
AVDD, AVSS current
DVDD current (13)
Operating
(reference enabled)
2.3
Sleep mode
(reference enabled)
200
Sleep mode
(reference disabled)
1
3.8
2.3
200
40
1
|μA|
40
|μA|
Power-Down mode
1
40
1
40
|μA|
Operating
500
700
500
700
μA
Sleep mode
160
300
160
300
μA
Power-Down mode (14)
1
10
1
10
μA
Operating
13
22
13
22
mW
Sleep mode
(reference enabled)
1.5
Sleep mode
(reference disabled)
0.5
1.2
0.5
1.2
mW
Power-Down mode
10
240
10
240
μW
1.5
mW
Power dissipation
TEMPERATURE RANGE
Specified temperature range
–40
+105
–40
+105
°C
Operating temperature range
–40
+125
–40
+125
°C
Storage temperature range
–60
+150
–60
+150
°C
(12) Crystal operation using 18pF load capacitors.
(13) Specified with internal oscillator operating (internal oscillator current: 40µA, typ).
(14) External CLKIN, SCLK stopped. Digital inputs maintained at VIH or VIL voltage levels.
4
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
PIN CONFIGURATION
PW PACKAGE
TSSOP-20
(TOP VIEW)
AINP
1
20
AVDD
AINN
2
19
AVSS
RESET/PWDN
3
18
VREFN
START
4
17
VREFP
SYNCOUT
5
16
REFOUT
ADS1259
CS
6
15
DVDD
SCLK
7
14
DGND
DIN
8
13
BYPASS
DOUT
9
12
XTAL2
DRDY
10
11
XTAL1/CLKIN
ADS1259 Terminal Functions
PIN NAME
PIN #
FUNCTION
DESCRIPTION
AINP
1
Analog input
Positive analog input
AINN
2
Analog input
Negative analog input
RESET/PWDN
3
Digital input
Reset/Power-Down; reset is active low; hold low for power-down
START
4
Digital input
Start conversions, active high
SYNCOUT
5
Digital output
Sync clock output (fCLK/8)
CS
6
Digital input
SPI chip-select, active low
SCLK
7
Digital input
SPI clock input
DIN
8
Digital input
SPI data input
DOUT
9
Digital output
SPI data output
DRDY
10
Digital output
Data ready output, active low
XTAL1/CLKIN
11
Digital input
Internal oscillator: DGND
External clock: clock input
Crystal oscillator: external crystal1
XTAL2
12
Digital
External crystal2, otherwise no connection
BYPASS
13
Analog
Core voltage bypass
DGND
14
Digital
Digital ground
DVDD
15
Digital
Digital power supply
REFOUT
16
Analog output
VREFP
17
Analog input
Positive reference input
VREFN
18
Analog input
Negative reference input
AVSS
19
Analog
Negative analog power supply and negative reference output
AVDD
20
Analog
Positive analog power supply
Copyright © 2009–2011, Texas Instruments Incorporated
Positive reference output
5
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
SPI TIMING CHARACTERISTICS
tSPWH
tSCLK
tCSH
CS
tCSSC
tSPWL
SCLK
tDIST
DIN
B7
B6
B5
B4
tDIHD
DOUT
B7
B3
B2
B1
B0
B3
B2
B1
B0
tDOPD
B6
B5
B4
tCSDOD
tDOHD
tCSDOZ
Figure 1. Serial Interface Timing
TIMING REQUIREMENTS: SERIAL INTERFACE TIMING
At TA = –40°C to +105°C and DVDD = 2.7V to 5.25V, unless otherwise noted.
SYMBOL
6
MIN
CS low to first SCLK: setup time (1)
50
tSCLK
SCLK period
1.8
tSPWH
SCLK pulse width: high
90
tSPWL
SCLK pulse width: low (3)
tDIST
Valid DIN to SCLK falling edge: setup time
35
20
Valid DIN to SCLK falling edge: hold time
SCLK rising edge to valid new DOUT: propagation delay (4)
tDOHD
SCLK rising edge to DOUT invalid: hold time
0
(4)
0
CS low to DOUT driven: propagation delay
tCSDOZ
CS high to DOUT Hi-Z: propagation delay
ns
ns
tCLK
ns
ns
ns
ns
40
20
20
(2)
ns
60
CS high pulse
UNIT
tCLK
216
tDIHD
tCSDOD
MAX
90
tDOPD
tCSH
(1)
(2)
(3)
(4)
DESCRIPTION
tCSSC
ns
ns
tCLK
CS can be tied low.
tCLK = 1/fCLK.
Holding SCLK low longer than 216 × tCLK cycles resets the SPI interface (enabled by SPI register bit).
DOUT load = 20pF || 100kΩ to DGND.
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
TYPICAL CHARACTERISTICS
At TA = +25°C, AVDD = +2.5V, AVSS = –2.5V, DVDD = 3.3V, VREF = 2.5V, VREFN = AVSS, fCLK = 7.3728MHz, and
fDATA = 60SPS, unless otherwise noted.
NOISE DISTRIBUTION HISTOGRAM
NOISE DISTRIBUTION HISTOGRAM
200
180
160
160
Data Rate = 10SPS
Shorted Input
512 Samples
140
120
Occurrences
140
Occurrences
Data Rate = 60SPS
Shorted Input
512 Samples
120
100
80
60
100
80
60
40
40
20
20
0
-5.0
-4.5
-4.0
-3.5
-3.0
-2.5
-2.0
-1.5
-1.0
-0.5
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
-5.0
-4.5
-4.0
-3.5
-3.0
-2.5
-2.0
-1.5
-1.0
-0.5
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
0
Reading (mV)
Reading (mV)
Figure 2.
Figure 3.
NOISE DISTRIBUTION HISTOGRAM
NOISE DISTRIBUTION HISTOGRAM
1600
2800
2400
Data Rate = 400SPS
Shorted Input
4096 Samples
1400
1200
Occurrences
2000
Occurrences
Data Rate = 14.4kSPS
Shorted Input
4096 Samples
1600
1200
800
1000
800
600
400
200
0
0
25.0
22.5
20.0
17.5
15.0
12.5
10.0
7.5
5.0
2.5
0
-2.5
-5.0
-7.5
-10.0
-12.5
-15.0
-17.5
-20.0
-22.5
-25.0
-50
-45
-40
-35
-30
-25
-20
-15
-10
-5
0
5
10
15
20
25
30
35
40
45
50
400
Reading (mV)
Reading (mV)
Figure 4.
Figure 5.
EFFECTIVE NUMBER OF BITS vs TEMPERATURE
EFFECTIVE NUMBER OF BITS HISTOGRAM
25
16
Data Rate = 10SPS
14
24
Data Rate = 60SPS
30 Units
Occurrences
ENOB (rms)
12
23
Data Rate = 60SPS
22
21
10
8
6
4
20
2
Data Rate = 14.4kSPS
19
Figure 6.
Copyright © 2009–2011, Texas Instruments Incorporated
23.5
23.3
23.4
23.1
23.2
23.0
22.8
125
22.9
105
22.7
Temperature (°C)
85
22.5
65
22.6
45
22.3
25
22.4
5
22.2
-15
22.1
-35
22.0
0
-55
ENOB (rms)
Figure 7.
7
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
TYPICAL CHARACTERISTICS (continued)
At TA = +25°C, AVDD = +2.5V, AVSS = –2.5V, DVDD = 3.3V, VREF = 2.5V, VREFN = AVSS, fCLK = 7.3728MHz, and
fDATA = 60SPS, unless otherwise noted.
NOISE vs INPUT VOLTAGE
NOISE vs INPUT VOLTAGE
7
7
Data Rate = 60SPS
6
5
5
Noise (mVrms)
Noise (mVrms)
Data Rate = 10SPS
6
4
3
Ratiometric Configuration
Internal Reference
CREFIN = 1mF
2
Internal Reference
CREFIN = 1mF
4
2
REF5025
1
REF5025
1
0
-2.5 -2.0 -1.5 -1.0 -0.5
0
0.5
1.0
1.5
2.0
0
-2.5 -2.0 -1.5 -1.0 -0.5
2.5
Figure 8.
Figure 9.
8
REF5025
Noise (mVrms)
Noise (mVrms)
1.5
2.0
2.5
Shorted Input
9
8
1.0
NOISE vs REFERENCE VOLTAGE
10
Internal Reference
CREFIN = 1mF
6
Ratiometric Configuration
4
2
Data Rate = 14.4kSPS
7
6
5
Data Rate = 14.4kSPS
4
3
Data Rate = 10SPS
Data Rate = 400SPS
Data Rate = 16.6SPS
2
Data Rate = 14.4kSPS
0
-2.5 -2.0 -1.5 -1.0 -0.5
1
0
0
0.5
1.0
1.5
2.0
2.5
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
VIN (V)
Reference Voltage (V)
Figure 10.
Figure 11.
LINEARITY DEVIATION vs INPUT LEVEL
4.0
4.5
5.0
5.5
INTEGRAL NONLINEARITY vs TEMPERATURE
3
3.0
5 Units
T = +125°C
T = +85°C
T = +25°C
T = -40°C
2.5
1
2.0
INL (ppm)
Linearity Deviation (ppm)
0.5
VIN (V)
NOISE vs INPUT VOLTAGE
0
1.5
-1
1.0
-2
0.5
-3
-2.5 -2.0 -1.5 -1.0 -0.5
8
0
VIN (V)
10
2
Ratiometric Configuration
3
0
0
0.5
1.0
1.5
2.0
2.5
-55
-35
-15
5
25
45
Input Signal (V)
Temperature (°C)
Figure 12.
Figure 13.
65
85
105
125
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
TYPICAL CHARACTERISTICS (continued)
At TA = +25°C, AVDD = +2.5V, AVSS = –2.5V, DVDD = 3.3V, VREF = 2.5V, VREFN = AVSS, fCLK = 7.3728MHz, and
fDATA = 60SPS, unless otherwise noted.
OFFSET vs TEMPERATURE
GAIN vs TEMPERATURE
100
500
5 Units
400
75
5 Units
300
Gain Error (ppm)
25
0
-25
200
100
0
-100
-200
-300
-50
-400
-75
-500
-55
-35
-15
5
25
45
65
85
105
125
-55
-35
5
-15
Figure 14.
45
65
85
105
125
Figure 15.
OFFSET DRIFT DISTRIBUTION HISTOGRAM
GAIN DRIFT DISTRIBUTION HISTOGRAM
30
20
60 Units From Two Production Lots
60 Units From Two Production Lots
25
16
2.6
2.2
2.4
2.0
1.8
1.6
1.2
1.4
0
0.26
0.22
0.24
0.18
0.20
0.14
0.16
0.12
0.08
0.10
0
0.06
0
0.04
4
0
5
0.8
8
0.4
10
12
0.6
15
0.2
Occurrences
20
0.02
Occurrences
25
Temperature (°C)
Temperature (°C)
1.0
Offset (mV)
50
Gain Drift (ppm/°C)
Offset Drift (mV/°C)
Figure 16.
Figure 17.
INTEGRAL NONLINEARITY vs REFERENCE VOLTAGE
GAIN ERROR AND OFFSET vs REFERENCE VOLTAGE
50
3.0
150
40
2.5
1.5
1.0
20
50
10
0
0
-10
-50
-20
-30
0.5
Offset (mV)
Gain Error (ppm)
2.0
INL (ppm)
100
30
-100
-40
0
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
Reference Voltage (V)
Figure 18.
Copyright © 2009–2011, Texas Instruments Incorporated
4.5
5.0
5.5
-50
-150
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
Reference Voltage (V)
Figure 19.
9
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
TYPICAL CHARACTERISTICS (continued)
At TA = +25°C, AVDD = +2.5V, AVSS = –2.5V, DVDD = 3.3V, VREF = 2.5V, VREFN = AVSS, fCLK = 7.3728MHz, and
fDATA = 60SPS, unless otherwise noted.
POWER-SUPPLY AND COMMON-MODE REJECTION vs
FREQUENCY
140
CMR
3.5
DVDD
DVDD
100
Power-Supply Current (mA)
120
PSR and CMR (dB)
POWER-SUPPLY CURRENT vs TEMPERATURE
4.0
CMR
AVDD, AVSS
80
AVDD
60
AVSS
40
20
3.0
AVDD, AVSS (Internal Reference On)
2.5
2.0
1.5
1.0
0.5
0
DVDD
0
10
100
1k
10k
100k
1M
-55
-35
-15
5
Power-Supply and Common-Mode Frequency (Hz)
25
Figure 20.
INTERNAL REFERENCE VOLTAGE vs TEMPERATURE
105
125
INTERNAL REFERENCE VOLTAGE vs TEMPERATURE
ADS1259
Internal Reference Voltage (V)
Internal Reference Voltage (V)
85
2.504
ADS1259B
2.500
2.499
2.498
2.497
2.496
2.502
2.500
2.498
2.496
2.494
-50
-25
0
25
50
75
100
125
-55
-35
-15
5
Temperature (°C)
25
45
65
85
105
125
Temperature (°C)
Figure 22.
Figure 23.
OUT-OF-RANGE THRESHOLD DISTRIBUTION
HISTOGRAM
OUT-OF-RANGE THRESHOLD vs TEMPERATURE
2.0
18
Relative to ±105%
30 Units
1.6
5 Units
1.2
Threshold Error (%)
14
12
10
8
6
0.8
0.4
0
-0.4
-0.8
4
-1.2
2
-1.6
0
-2.0
-2.0
-1.8
-1.6
-1.4
-1.2
-1.0
-0.8
-0.6
-0.4
-0.2
0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
Occurrences
65
Figure 21.
2.501
16
45
Temperature (°C)
-55
-35
-15
5
25
45
65
85
105
125
Temperature (°C)
Threshold Error (%)
Figure 24.
10
Figure 25.
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
TYPICAL CHARACTERISTICS (continued)
At TA = +25°C, AVDD = +2.5V, AVSS = –2.5V, DVDD = 3.3V, VREF = 2.5V, VREFN = AVSS, fCLK = 7.3728MHz, and
fDATA = 60SPS, unless otherwise noted.
REFERENCE INPUT CURRENT vs TEMPERATURE
REFERENCE INPUT CURRENT vs REFERENCE VOLTAGE
250
VREFP, VREFN Input Current (nA)
VREFP, VREFN Input Current (nA)
0
-25
-50
VREFP Input Current
-75
-100
-125
VREFN Input Current
-150
-175
-200
-225
200
150
100
50
-50
-100
-150
-55
-35
5
-15
25
45
65
85
105
125
VREFP Input Current
(VREFN = AVSS)
-200
-250
-250
VREFN Input Current
(VREFP = AVDD)
0
0.5
1.0
1.5
2.0
Figure 26.
DIFFERENTIAL INPUT IMPEDANCE vs TEMPERATURE
3.5
4.0
4.5
5.0
INTERNAL REFERENCE SETTLING TIME
0.010
CREFIN = 1mF X7R
0.008
128
Settling (% of Final Value)
Differential Input Impedance (kW)
3.0
Figure 27.
130
Internal Oscillator
126
External Crystal
124
122
0.006
0.004
0.002
0
-0.002
-0.004
-0.006
-0.008
120
-55
-35
-15
5
25
45
65
85
105
-0.010
125
0
1
2
3
4
5
6
Time (s)
Temperature (°C)
Figure 28.
Figure 29.
INTERNAL OSCILLATOR vs TEMPERATURE
INTERNAL REFERENCE LONG-TERM STABILITY
0
Reference Voltage Stability (ppm)
7.50
7.45
Internal Oscillator (MHz)
2.5
VREF (V)
Temperature (°C)
7.40
7.35
7.30
7.25
−20
−30
−40
−50
−60
−70
−80
−90
−100
7.20
-55
-35
-15
5
25
45
65
Temperature (°C)
Figure 30.
Copyright © 2009–2011, Texas Instruments Incorporated
85
105
125
48 units
−10
0
100 200 300 400 500 600 700 800 900 1000
Hours
G001
Figure 31.
11
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
OVERVIEW
The ADS1259 is a high-linearity, low drift
analog-to-digital converter (ADC) designed for the
needs of industrial process control, precision
instrumentation, and similar applications. The
converter provides high-resolution, 24-bit output data
at sample rates ranging from 10SPS to 14.4kSPS.
Figure 32 shows a block diagram of the ADS1259.
The device allows unipolar or bipolar analog
power-supply configuration (AVDD – AVSS = 5V
total). The analog supplies may be set to single +5V
to accept unipolar (or offset-bipolar) signals or the
supplies can be set to ±2.5V to accept true bipolar
signals. The operating range of the digital power
supply (DVDD) is 2.7V to 5V.
An internal low dropout regulator (LDO) powers the
digital core from the DVDD supply while the device
I/O operates directly from DVDD. BYPASS is the
LDO output and requires a 0.1μF or larger capacitor
to ground.
The inherently stable, fourth-order, ΔΣ modulator
measures the differential input signal [VIN = (AINP –
AINN)] against the differential reference [VREF =
(VREFP – VREFN)]. A fast responding out-of-range
detector flags the output data if the input should
over-range while converting.
The digital filter receives the modulator signal and
provides the digital output. The filter consists of a
fifth-order sinc filter followed by a programmable
averager, selectable as either a sinc1 or sinc2. In
sinc1 mode, the filter settles in a single conversion.
The programmable averaging yields output data rates
from 10SPS to 14.4kSPS.
The ADS1259 integrates a low-drift, low-noise +2.5V
reference. The internal reference can drive loads up
to ±10mA. The ADS1259 also operates from an
external reference if desired. The reference input is
buffered to reduce loading of external circuits.
An onboard oscillator is provided as the clock source
for the device. Optionally, an external crystal can be
used. As a third clock option, the device can be
driven by an external clock source. SYNCOUT is an
output that provides a 1/8 rate clock intended to drive
the chopping clock input of the PGA280.
Gain and offset registers scale the digital filter output
to produce the final code value. On-command
calibration corrects for system offset and gain errors.
An SPI-compatible serial interface provides the
control and configuration as well as the data interface
to the ADS1259. Onboard registers combined with
commands are used to control and configure the
device.
The RESET/PWDN pin is dual function. A momentary
low resets the device and, if the pin is held low,
powers down the device. The START pin, as well as
commands, controls the conversions.
AVDD
VREFP
+1.8V
(Digital Core)
2.5V
Reference
REFOUT
AINP
DS
Modulator
AINN
BYPASS DVDD
VREFN
Out-of-Range
Detection
Programmable
Digital Filter
LDO
SYNCOUT
fCLK/8
Clock
Generator
XTAL2
RESET/PWDN
Calibration
Engine
FLAG
XTAL1/CLKIN
START
Control
and
Serial
Interface
DRDY
SCLK
DIN
DOUT
CS
ADS1259
AVSS
DGND
Figure 32. ADS1259 Block Diagram
12
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
NOISE PERFORMANCE
MODULATOR
The ADS1259 offers excellent noise performance that
can be optimized by adjusting the data rate and by
selection of the digital filter mode. As the averaging is
increased by reducing the data rate, the noise drops
correspondingly. Additionally, because the sinc2
digital filter provides more filtering than the sinc1
digital filter, sinc2 provides lower noise conversions.
Table 1 shows the noise as a function of data rate
and filter mode.
The
high-performance
modulator
is
an
inherently-stable, fourth-order, ΔΣ, 2 + 2 pipelined
structure, as shown in Figure 33. It shifts the
quantization noise to a higher frequency (out of the
passband) where digital filtering can easily remove it.
Table 1 expresses typical noise data in several ways:
RMS noise, effective number of bits (ENOB), and
noise-free bits. ENOB is calculated from Equation 1:
ln
fMOD = fCLK/8
2nd-Order
DS
1st-Stage
Analog Input (VIN)
To Digital Filter
2nd-Order
DS
2nd-Stage
FSR
RMS Noise
ENOB =
4th-Order Modulator
ln(2)
Figure 33. Fourth-Order Modulator
Where:
FSR = 2VREF
(1)
The calculation of noise-free bits uses the same
formula as Equation 1, except that the peak-to-peak
noise value is used instead of RMS noise.
ADC
The analog-to-digital converter (ADC) section of the
ADS1259 is composed of two blocks: a high accuracy
modulator and a programmable digital filter.
The modulator first stage converts the analog input
voltage into a pulse-code modulated (PCM) stream.
When the level of differential analog input (AINP –
AINN) is near the level of the reference voltage
(VREFP – VREFN), the 1s density of the PCM data
stream is at its highest. When the level of the
differential analog input is near zero, the PCM 0s and
1s densities are nearly equal. At the two extremes of
the analog input levels (+FS and –FS), the 1s density
of the PCM streams are approximately +90% and
+10%, respectively.
The modulator second stage produces a 1s density
data stream designed to cancel the quantization
noise of the first stage. The data streams of the two
stages are then combined in the digital filter stage.
Table 1. Typical Noise Data vs Data Rate and Digital Filter (1)
(1)
(2)
(3)
SINC1 DIGITAL FILTER
DATA
RATE
(SPS)
SAMPLE
SIZE (2)
SINC2 DIGITAL FILTER
NOISE
(μVRMS)
NOISE
(μVPP)
ENOB
(RMS)
NOISEFREE BITS
NOISE
(μVRMS)
NOISE
(μVPP)
ENOB
(RMS)
NOISEFREE BITS
10
128
0.5
1.8
23.3
21.4
0.45
1.6
23.4
21.6
16.6
256
0.55
2.4
23.1
21.0
0.5
2
23.3
21.3
50
512
0.65
3.5
22.9
20.4
0.6
3
23.0
20.7
60
512
0.7
4
22.8
20.3
0.65
3.5
22.9
20.4
400
4096
1.4
9.5
21.8
19.0
1.2
8.3
22.0
19.2
1200
8192
2.3
17
21.1
18.2
2
14
21.3
18.4
3600
8192
3.9
32
20.3
17.3
3.4
27
20.5
17.5
14400
8192
6.2
50
19.6
16.6
(3)
(3)
(3)
(3)
Noise data taken with shorted analog inputs and internal 2.5V reference using the circuit of Figure 64.
Data sample sizes used for analysis.
Same as sinc1 mode.
Copyright © 2009–2011, Texas Instruments Incorporated
13
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
The ADS1259 modulator is inherently stable and
therefore has predictable recovery behavior resulting
from an input overdrive condition. The modulator
does not exhibit the self-resetting behavior of other
modulator types, which often results in unstable
output conversion results when overdriven.
The ADS1259 modulator outputs a 1s density data
stream at 90% duty cycle with the positive full-scale
input signal applied (10% duty cycle with the negative
full-scale signal). If the input is overdriven past 90%
modulation, but below 100% modulation (10% and
0% for negative overdrive, respectively), the
modulator remains stable and continues to output the
1s density data stream. The digital filter may or may
not clip the output codes to +FS or –FS, depending
on the duration of the overdrive. When the input is
returned to the normal range from a long duration
overdrive (worst case), the modulator returns
immediately to the normal range, but the group delay
of the digital filter delays the return of the conversion
result to within the linear range (one reading for the
sinc1 filter and two readings for completely settled
data).
If the inputs are sufficiently overdriven to drive the
modulator to full duty cycle (that is, all 1s or all 0s),
the modulator enters a stable saturated state. The
digital output code may clip to +FS or –FS, again
depending on the duration. A small duration overdrive
may not always clip the output code. When the input
returns to the normal range, the modulator requires
up to 12 modulator clock cycles (fMOD) to exit
saturation and return to the linear region. The digital
filter requires two additional conversions (sinc1, more
for sinc2) for fully settled data.
In the extreme case of over-range, either input is
overdriven exceeding that either analog supply
voltage plus an internal ESD diode drop. The internal
ESD diodes begin to conduct and the signal on the
input is clipped. If the differential input signal range is
not exceeded, the modulator remains in linear
operation. If the differential input signal range is
exceeded, the modulator is saturated but stable, and
outputs all 1s or 0s. When the input overdrive is
removed, the diodes recovery quickly and the
14
ADS1259 recovers as normal. Note that the linear
input range is ±100mV beyond the analog supply
voltages; with input levels greater than this range,
use care to limit the input current to 100mA peak
transient (10mA continuous).
INPUT OUT-OF-RANGE DETECTION (FLAG)
The ADS1259 has a fast-responding out-of-range
circuit that triggers when the differential input exceeds
+105% or –105% of FSR (±1.05 VREF). The
out-of-range circuit latches the result of the
comparator output and appends the result as either
the LSB of conversion data or as bit 7 of the data
checksum byte. After the conversion data are read, or
after a new conversion is started, the comparator
latch is reset. Figure 34 and Figure 35 show the
detection block diagram and the detection operation,
respectively. See the Data Checksum Byte and FLAG
Bit section for more detail.
AINP
å
IABSI
1.05 VREF
AINN
J
Q
fMOD/2
Data Read
Reset
FLAG
K
Figure 34. Input Out-Of-Range Detect Block
Diagram
AINP - AINN (% VREF)
MODULATOR OVERLOAD BEHAVIOR
www.ti.com
+105
(Conversions)
0
-105
FLAG
Bit
1
0
Figure 35. Input Out-Of-Range Detect Operation
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
ANALOG INPUTS (AINP, AINN)
ESD diodes protect the analog inputs. To keep these
diodes from turning on, make sure the voltages on
the input pins do not go below AVSS by more than
300mV, and likewise do not exceed AVDD by more
than 300mV.
The ADS1259 measures the differential input signal
VIN = (AINP – AINN) against the differential reference
VREF = (VREFP – VREFN) using internal capacitors
that are continuously charged and discharged.
Figure 37 shows the simplified schematic of the ADC
input circuitry; the right side of the figure illustrates
the input circuitry with the capacitors and switches
replaced by an equivalent circuit. Figure 36
demonstrates the ON/OFF timings for the switches of
Figure 37.
AVSS – 300mV < (AINP or AINN) < AVDD + 300mV.
Note that the valid input range is:
AVSS – 100mV < (AINP or AINN) < AVDD + 100mV
tSAMPLE = 1/fMOD
ON
In Figure 37, S1 switches close during the input
sampling phase. With switch S1 closed, CA1 charges
to AINP, CA2 charges to AINN, and CB charges to
(AINP – AINN). For the discharge phase, S1 opens
first and then S2 closes. CA1 and CA2 discharge to
approximately to AVSS + 2.5V and CB discharges to
0V. This two-phase sample/discharge cycle repeats
with a period of tSAMPLE = 1/fMOD. fMOD is the operating
frequency of the modulator, where fMOD = fCLK/8.
S1
OFF
ON
S2
OFF
Figure 36. S1 and S2 Switch Timing for Figure 37
Although optimized for differential signals, the
ADS1259 inputs may be driven with a single-ended
signal by fixing one input to AVSS or mid-supply. Full
dynamic range is achieved when the inputs are
differentially driven ±VREF.
The charging of the input sampling capacitors draws
a transient current from the source driving the
ADS1259 ADC inputs. The average value of this
current can be used to calculate an effective
impedance (REFF) where REFF = VIN/IAVERAGE. These
impedances scale inversely with fMOD. For example, if
fMOD is reduced by a factor of two, the impedances
double. Note that the sampling capacitors can vary
±15% over production lots and typically vary 1% with
temperature. The variations of the sampling
capacitors have a corresponding effect on the analog
input impedance.
AVDD
As a result of the switched-capacitor input structure of
the ADS1259, a buffer is recommended to drive the
analog inputs. An input filter comprised of 20Ω to 50Ω
resistors and 10nF capacitors should be used
between the buffer and the ADS1259 inputs.
(fMOD = 0.9216MHz)
AVSS + 2.5V
AVSS + 2.5V
S2
AINP
REFF A = 500kW
CA1 = 2pF
ESD Diodes
Equivalent
Circuit
S1
AINP
REFF B = 130kW
CB = 8pF
S1
AINN
AINN
REFF A = 500kW
CA2 = 2pF
ESD Diodes
S2
AVSS
AVSS + 2.5V
REFF =
1
fMOD ´ CX
and fMOD = fCLK/8
AVSS + 2.5V
RDIFF = REFF B || 2REFF A = 120kW
RCOM = REFF A = 500kW
Figure 37. Simplified ADC Input Structure
Copyright © 2009–2011, Texas Instruments Incorporated
15
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
REFERENCE
the ADC reference input pins, VREFP and VREFN.
(Note that these device pins are not intended to drive
external circuits.) An external 1μF capacitor,
connected from VREFP to VREFN, is recommended
for noise reduction. The capacitor can be increased
for increased noise filtering, but the settling time of
the reference may also increase. The settling time
should be considered upon activating the internal
reference.
The ADS1259 includes an onboard voltage reference
with a low temperature coefficient. The reference
voltage is 2.5V with the capability of sinking and
sourcing 10mA via the REFOUT pin. The ADS1259
can also operate from an external reference. The
external reference is the default selection. Refer to
Figure 38 for a reference block diagram.
Internal Reference
See Figure 29 for typical reference settling CREFIN =
1µF. The capacitor dielectric absorption results in
increased settling time for RC filter circuits.
The reference output is provided between pins
REFOUT and AVSS. Because the reference output
return shares the same pin as AVSS, route the
reference return trace and the AVSS trace
independently as Kelvin-connected printed circuit
board (PCB) traces. For stability reasons, connect a
1μF capacitor between REFOUT and AVSS.
To activate the internal reference, set the register bit
RBIAS = 1. This enables the reference bias. Once
biased, the internal reference can then be selected as
the ADC reference by the register bit EXTREF.
EXTREF = 0 closes the internal switches.
An internal switch connects the internal reference to
AVDD
REFOUT
(+)
+
Reference Output
(-)
Reference Bias
RBIAS Register Bit
(1 = Bias On)
2.5V
Reference
1 mF
(CREFOUT)
AVSS
Reference Select
EXTREF Register Bit
(1 = Switch Open for External Reference)
2kW
(+)
Reference Input
(-)
VREFP
(+)
+ 1mF
(CREFIN)
ADC Reference Input
(-)
VREFN
Figure 38. Reference Block Diagram
Table 2. Reference Selection for Figure 38
ADS1259 REFERENCE
RBIAS REGISTER BIT
EXTREF REGISTER BIT
Internal
1
0
External
(1)
16
See
(1)
1
If the reference output is not required, set RBIAS = 0. If the reference output is enabled (RBIAS = 1), an external 1µF capacitor must be
used between REFOUT and AVSS.
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
Reference Drift
External Reference
The ADS1259 internal reference is designed for
minimal drift error, which is defined as the change in
reference voltage over temperature. The drift is
calculated using the box method, as described by
Equation 2.
VREFMAX - VREFMIN
Drift =
x 106 (ppm)
VREFNOM ´ Temp Range
To select the ADS1259 for external reference
operation, set the EXTREF register bit = 1 (default). If
desired, the internal reference can continue to
provide a +2.5V reference output via the REFOUT
and AVSS pins. In this case, set the RBIAS register
bit = 1 to power the internal reference. If the internal
reference is activated, an external 1μF capacitor from
REFOUT to AVSS is required.
Where:
VREFMAX, VREFMIN, and VREFNOM are the maximum,
minimum, and nominal reference output voltages,
respectively, over the specified temperature
range.
(2)
For external reference applications, place a 1μF
(minimum) capacitor close to the VREFP and VREFN
pins.
The ADS1259 internal reference features a maximum
drift coefficient of 5ppm/°C over 0°C to +85°C
operating range and 12ppm/°C over –40°C to +105°C
operating range.
Thermal Hysteresis
Thermal hysteresis of the internal reference is defined
as the change in voltage after operating the device at
+25°C, cycling the device through the specified
temperature range, and returning to +25°C. It can be
expressed as Equation 3.
VHYST =
|VPRE - VPOST|
VNOM
´ 106(ppm)
Because the ADS1259 measures the signal inputs
(AINP and AINN) against the reference inputs
(VREFP and VREFN), reference noise and drift may
degrade overall system performance. In ratiometric
measurement applications, reference noise and drift
have a cancelling effect. In absolute measurement
applications, reference noise and drift directly effect
the conversion results.
Voltage Reference Inputs (VREFP, VREFN)
ESD diodes protect the reference inputs. To keep
these diodes from turning on, make sure the voltages
on the reference pins do not go below AVSS by more
than 300mV, and likewise do not exceed AVDD by
more than 300mV.
The absolute maximum reference input range is:
AVSS – 300mV < (VREFP or VREFN) < AVDD + 300mV(4)
Where:
VHYST = thermal hysteresis (in units of ppm).
VNOM = nominal reference voltage (+2.5V).
VPRE = reference voltage measured at +25°C
pretemperature cycling.
VPOST = reference voltage measured after the
device has been cycled from +25°C through the
temperature range of 0°C and +85°C and
returned to +25°C.
(3)
Copyright © 2009–2011, Texas Instruments Incorporated
Note that the valid operating range of the reference
inputs are shown in the Electrical Characteristics
table.
17
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
DIGITAL FILTER
The programmable low-pass digital filter receives the
modulator output and produces a high-resolution
digital output. By adjusting the amount of filtering,
tradeoffs can be made between resolution and data
rate: filter more for higher resolution, filter less for
higher data rate.
The filter consists of two sections: a fixed decimation
sinc5 filter followed by a variable decimation filter,
configurable as sinc1 or sinc2, as illustrated in
Figure 39. The sinc5 filter has fixed decimation of 64
and reduces the data rate of the modulator from
fCLK/8 to fCLK/512. The second filter stage receives the
data from the sinc5 filter. The second filter stage has
programmable averaging (or decimation) and can be
configured in either sinc1 or sinc2 mode. The
decimation ratio of this stage sets the final output
data rate. As detailed in Table 3, the DR[2:0] register
bits program the decimation ratio and the final output
data rate. The output data rates are identical for both
sinc1 and sinc2 filters.
Table 3. Decimation Ratio of Final Filter Stage
DR[2:0] REGISTER
BITS
DECIMATION
RATIO (R)
DATA RATE (SPS)
111
1
14400
110
4
3600
101
12
1200
100
36
400
011
240
60
010
288
50
001
864
16.6
000
1440
10
Modulator Rate = fCLK/8
The SINC2 register bit selects either the sinc1 or sinc2
filter. The sinc1 filter settles in one conversion cycle
while the sinc2 filter settles in two conversion cycles.
However, the sinc2 filter has the benefit of wider
frequency notches which improve line cycle rejection.
FREQUENCY RESPONSE
The low-pass digital filter sets the overall frequency
response of the ADS1259. The filter response is the
product of the fixed and programmable filter sections,
and is given by Equation 5:
½H(f)½ = ½Hsinc5(f)½ ´ ½HsincN(f)½ =
5
512p ´ f
sin
fCLK
64 ´ sin
8p ´ f
fCLK
N
sin
´
512p ´ R ´ f
fCLK
R ´ sin
512p ´ f
fCLK
where:
N = 1 (sinc1)
N = 2 (sinc2)
R = Decimation ratio (refer to Table 3)
(5)
The digital filter attenuates noise on the modulator
output, including noise from within the ADS1259 and
external noise present within the ADS1259 input
signal. Adjusting the filtering by changing the
decimation ratio used in the programmable filter
changes the filter bandwidth. With a higher number of
decimation, the bandwidth is reduced and more noise
is attenuated.
fCLK/512
1
sinc Filter
Analog
Modulator
5
Output
Data Rate = fCLK/(R ´ 512)
sinc Filter
(decimate by 64)
2
sinc Filter
SINC2 Register Bit
DR[2:0] Register Bits
(Program Decimation)
1
(0 = sinc )
Figure 39. Block Diagram of Digital Filter
18
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
The sinc5 filter produces wide notches at fCLK/512 and
multiples thereof. At these frequencies the filter has
zero gain. Figure 40 shows the response data rate =
14.4kSPS.
-20
Magnitude (dB)
With decimation of the second stage, the wide
notches produced by the sinc5 filter remain, but a
number of narrow notches are superimposed in the
response. The first of the notches occur at the data
rate. The number of superimposed notches is
determined by the decimation ratio, minus 1.
0
sinc2
10
4.3
3.1
16.6(3)
16.6
7.3
5.2
50
50
22
16
60
60
27
19
400
400
177
127
1200
1200
525
380
3600
3600
1440
1100
14400
14400
2930
See (4)
10
fCLK = 7.3728MHz.
Notch at 50Hz and 60Hz.
Notch at 50Hz.
Same as sinc1.
10
20
30
40
50
60
Frequency (kHz)
Figure 40. Frequency Response
for Data Rate = 14.4kSPS
0
sinc
2
-20
sinc
1
-40
-60
-80
-100
-120
–3dB BANDWIDTH (Hz)
sinc1
(2)
(1)
(2)
(3)
(4)
FIRST NOTCH
(Hz)
-100
0
-140
0
10
20
30
40
50
60
Frequency (kHz)
Figure 41. Frequency Response
(Data Rate = 3600SPS, R = 4)
0
-10
-20
Magnitude (dB)
DATA RATE
(SPS)
-80
-140
Magnitude (dB)
Table 4. First Notch Frequency and –3dB Filter
Bandwidth(1)
-60
-120
The second stage filter has notches (or zeroes) at the
data rate and multiples thereof. Figure 41 shows the
response of the second stage filter combined with the
sinc5 stage. Decimation of 4 produces three
equally-spaced notches between each main notch of
the sinc5 filter. The frequency response of the other
data rates (higher decimation ratios) produces a
similar pattern, but with more equally-spaced notches
between the main sinc5 notches. Table 4 lists the first
notch frequency and the –3dB bandwidth.
Figure 42 illustrates the detail of the magnitude
response with data rate = 60SPS. Note that input
frequencies within the ±1% 60Hz bandwidth are
attenuated 40dB by the sinc1 filter and 80dB by the
sinc2 filter.
-40
-30
sinc
-40
-50
sinc
-60
1
2
-70
-80
-90
-100
55
56
57
58
59
60
61
62
63
64
65
Frequency (Hz)
Figure 42. Magnitude Response
for Data Rate = 60SPS
Copyright © 2009–2011, Texas Instruments Incorporated
19
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
ALIASING
CLOCK SOURCE
The low-pass characteristic of the digital filter repeats
at multiples of the modulator rate (fMOD = fCLK/8).
Figure 43 shows the responses plotted out to
7.3728MHz at the data rate of 14.4kSPS. Notice how
the responses near dc, 0.9216MHz, 1.8432MHz,
2.7698MHz, etc, are the same as given by f =
NfMOD ± fDATA where N = 0, 1, 2, etc. The digital filter
attenuates high-frequency noise on the ADS1259
inputs up to the frequency where the response
repeats. However, noise or frequency components
existing in the signal where the response repeats
alias into the passband. Often, a simple RC antialias
filter is sufficient to reject these input frequencies.
There are three ways to provide the ADS1259 clock:
the internal oscillator, an external clock, or an
external crystal/ceramic resonator. The ADS1259
selects the clock source automatically. Figure 44
shows the clock select block. If either external clock
sources are present, the internal oscillator is disabled
and the external clock source is selected. If no
external clock is present, the internal oscillator is
selected. The ADS1259 continuously monitors the
clock source. The clock source can be polled by the
EXTCLK bit (bit 6 of register CONFIG2), 0 = internal
oscillator, 1 = external clock.
0
Magnitude (dB)
-20
The data rate and corresponding filter notches scale
by the accuracy of clock frequency. Consideration
should be given to the clock accuracy and the
corresponding effect to the notch frequency locations.
-40
-60
Clock
Detect
XTAL1/CLKIN
-80
-100
Internal
Oscillator
XTAL2
S0
-120
ENB
S1
SEL
MUX
-140
0
0.9
1.8
2.8
3.7
4.6
5.5
6.5
System Clock
7.4
Frequency (MHz)
Figure 43. Frequency Response to 7.3728MHz
(Data Rate = 14400SPS)
Figure 44. Equivalent Circuitry of the Clock
Source
Internal Oscillator
Figure 45 shows the internal oscillator connection.
XTAL1/CLKIN is grounded and XTAL2 is not
connected (floating). The internal oscillator draws
approximately 40μA from the DVDD supply. Note that
the internal oscillator has ±2% accuracy over
temperature. The oscillator accuracy has a
corresponding effect on line-cycle notch frequency
locations.
XTAL1/CLKIN
XTAL2
Figure 45. Internal Oscillator Connection
20
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
External Clock
SYNCOUT
Figure 46 shows the external clock connection. The
clock is applied to XTAL1/CLKIN and XTAL2 floats.
Make sure a clean clock input is applied to the
ADS1259, free of overshoot and glitches. A series
resistor often helps to reduce overshoot and should
be placed close to the driving end of the clock
source.
SYNCOUT is a digital output pin intended to
synchronize the chopping frequency of the PGA280
to the sampling frequency of the ADS1259.
Synchronizing the PGA280 to the ADS1259 places
the PGA280 chopped 1/f noise at an exact null in the
ADS1259 frequency response, where the PGA280 1/f
noise is rejected.
External
Clock
50W
XTAL1/CLKIN
XTAL2
Figure 46. External Clock Connection
Crystal Oscillator
Figure 47 shows the crystal oscillator connection. The
crystal connects to XTAL1/CLKIN and XTAL2 and the
capacitors connect to ground. The crystal and
capacitors should be placed close to the device pins
with short, direct traces. Neither the XTAL1/CLKIN
nor the XTAL2 pins can be used to drive any other
logic. Table 5 lists the recommended crystal for the
ADS1259. If using other crystals, verify the oscillator
start-up behavior.
XTAL1/CLKIN
Crystal
(7.3728MHz)
C1
XTAL2
C2
C1, C2: 5pF to 20pF
Figure 47. Crystal Connection
Table 5. Recommended Crystal
MANUFACTURER
FREQUENCY
PART NUMBER
ECS
7.3728MHz
ECS-73-18-10
SYNCOUT frequency is equal to the ADS1259 clock
rate divided by 8 (fSYNCOUT = fCLK/8). The output clock
is enabled by the register bit SYNCOUT. Disabling
the output stops the clock but the output remains
actively driven low. In power-down mode, the
SYNCOUT output becomes an input. As with all
digital inputs, the pin must not be allowed to float. An
external 1MΩ pull-down resistor is recommended to
ground the input in power-down mode.
The SYNCOUT clock is reset when START is
received and whenever registers CONFIG[2:0] are
changed. Connect SYNCOUT to the PGA280
SYNCIN pin through a 4.7kΩ series resistor. Place
the resistor as close as possible to the ADS1259
SYNCOUT pin.
SLEEP MODE
SLEEP mode is started by sending the SLEEP
command. In SLEEP mode, the device enters a
reduced power state and only a minimum of circuitry
is kept active. The WAKEUP command exits the
SLEEP mode and after which 512 fCLK cycles are
counted before the ADS1259 is ready for
communication. The register settings are unaffected
in SLEEP.
SLEEP does not change the RBIAS register bit. For
quick conversions after WAKEUP, keep the internal
reference bias on before entering SLEEP. Otherwise,
after exiting SLEEP mode, allow time for the
reference to settle. Alternatively, to minimize power
consumption during SLEEP, set the internal reference
bias off prior to engaging SLEEP. Note that in SLEEP
mode the SPI timeout function is disabled.
BYPASS
The digital core of the ADS1259 is powered by an
internal low dropout regulator (LDO). The DVDD
supply is the LDO input and the BYPASS pin is the
LDO output. A 1μF capacitor must be connected from
the LDO output to DGND. No other load current
should be drawn from the BYPASS pin.
Copyright © 2009–2011, Texas Instruments Incorporated
21
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
RESET/PWDN
RESET
The RESET/PWDN pin has two functions: device
power-down and device reset. Momentarily holding
the pin low resets the device and holding the pin low
for 216 fCLK cycles activates the Power-Down mode.
There are three methods to reset the ADS1259: cycle
the power supplies, take RESET/PWDN low, or send
the RESET opcode command.
When using the RESET/PWDN pin, take it low to
force a reset. Make sure to follow the minimum pulse
width timing specifications before taking the RESET
pin back high.
POWER-DOWN MODE
In power-down mode, internal circuit blocks are
disabled (including the oscillator, reference, and SPI)
and the device enters a micro-power state. To
engage power-down mode, hold the RESET/PWDN
pin low for 216 fCLK cycles. Note that the register
contents are not saved because they are reset when
RESET/PWDN goes high.
The RESET command takes effect on the eighth
falling SCLK edge of the opcode command. On reset,
the configuration registers are initialized to the default
states and the conversion cycle restarts. After reset,
allow eight fCLK cycles before communicating to the
ADS1259. Note that when using the reset command,
the SPI interface itself may require reset before
accepting the command. See the SPI Timing
Characteristics section for details.
Keep the digital inputs at defined VIH or VINL logic
levels (do not 3-state). To minimize power-supply
leakage current, disable the external clock. Note that
the ADS1259 digital outputs remain active in
power-down. The analog signal inputs may float.
POWER-ON SEQUENCE
To exit power-down, take RESET/PWDN high. Wait
216 fCLK cycles before communicating to the
ADS1259, as shown in Figure 48.
The ADS1259 has three power supplies: AVDD,
AVSS, and DVDD. The supplies can be sequenced in
any order but be sure that at any time the analog
inputs do not exceed AVDD or AVSS and the digital
inputs do not exceed DVDD. After the last power
supply has crossed the respective power-on
threshold, 216 fCLK cycles are counted before
releasing the internal reset. After the internal reset is
released, the ADS1259 is ready for operation.
Figure 49 shows the power-on sequence of the
ADS1259.
tLOW
RESET/PWDN
tRHSC
SCLK
Figure 48. RESET/PWDN Timing
Table 6. Timing Characteristics for Figure 48
SYMBOL
DESCRIPTION
MIN
UNIT
tLOW
Pulse width low for reset
4
tCLK
16
tCLK
tLOW
Pulse width low for power-down
tRHSC
Reset high to SPI communication start
8
tCLK
tRHSC
Exit power-down to SPI communication start
216
tCLK
AVDD - AVSS
DVDD
2
3.5V nom
1V nom
CLK
2
Internal Reset
16
´ tCLK
ADS1259 Operational
Figure 49. Power-On Sequence
22
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
START
START is a digital input that controls the ADS1259
conversions. Conversions are started when START is
taken high and are stopped when START is taken
low. If START is toggled during a conversion, the
conversion is restarted. DRDY goes high when
START is taken high. Figure 50 andTable 7 show the
START timing.
Note that reasserting START within 22 tCLK cycles of
the DRDY falling edge causes DRDY to fall soon
after. This conversion result should be discarded. The
next DRDY falling edge, as given in Table 9, is the
valid conversion data.
tSDSU
DRDY
tPWH
tSTDR
Gate Control Mode (PULSE Bit = 0, Default)
Conversions begin when either the START pin is
taken high or when the START command is sent.
Conversions continue indefinitely until the START pin
is taken low or the STOP command is transmitted. As
seen in Figure 51, DRDY is forced high when the
conversion starts and falls low when data are ready.
When stopped, the conversion in process completes
and further conversions are halted. Figure 50 and
Table 7 show the timing of DRDY and START.
tDSHD
START
START Pin
tPWL
or
Command
When using commands to control conversions, hold
the START pin low. The ADS1259 features two
modes to control conversions: Gate Control mode
and Pulse Control mode. The mode is selected by the
PULSE register bit.
(1)
START
STOP
STOP
or
START
Command(1)
(1) START and STOP commands take effect on the seventh SCLK
falling edge.
or
STOP
START
Halted
Converting
Halted
Figure 50. START to DRDY Timing
DRDY
CONVERSION CONTROL
The conversions of the ADS1259 are controlled by
either the START pin or by the START command.
(1) START and STOP opcode commands take effect on the
seventh SCLK falling edge.
Figure 51. Gate Control Mode
Table 7. START Timing (See Figure 50)
SYMBOL
DESCRIPTION
MIN
tSDSU
START pin low or STOP opcode to DRDY setup time to halt further
conversions
16
tCLK
tDSHD
START pin low or STOP opcode hold time to complete current
conversion (gate mode)
16
tCLK
tPWH,
START pin pulse width high, low
4
L
tSTDR
START pin rising edge to DRDY rising edge
Copyright © 2009–2011, Texas Instruments Incorporated
MAX
UNIT
tCLK
4
tCLK
23
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
Pulse Control Mode (PULSE Bit = 1)
Settling Time Using START
In the Pulse Control mode, the ADS1259 performs a
single conversion when either the START pin is taken
high or when the START command is sent. As seen
in Figure 52, DRDY goes high when the conversion is
started. When the conversion is complete, DRDY
goes low and further conversions are halted. To start
a new conversion, transition the START pin back to
high, or transmit the START opcode again.
When START goes high (via pin or command) a
delay may be programmed before the conversion
filter cycle begins. The programmable delay may be
useful to provide time for external circuits (such as
after an external signal mux change), before the
reading is started. Register bits DELAY[2:0] set the
initial delay time as shown in Table 8.
Table 8. Initial START Delay
START Pin
OR
OR
START
Halted
START
Single
Conversion
Halted
Single
Conversion
DRDY
(1) START opcode command takes effect on the seventh SCLK
falling edge.
Figure 52. Pulse Control Mode
CONVERSION SETTLING TIME
The ADS1259 features a digital filter architecture in
which settling time can be traded for wide filter
notches, resulting in improved line-cycle rejection.
This trade-off is determined by the selection of the
sinc1 or sinc2 filter. The sinc1 filter settles in a single
cycle while the sinc2 filter provides wide-width filter
notches. The settling time of the ADS1259 is different
if START is used to begin conversions or if the
ADS1259 is free-running the conversions. These
modes are explained in the Settling Time Using
START and Settling Time While Continuously
Converting sections.
VIN = AINP - AINN
Settled VIN
START
Pin
or
7th Falling SCLK Edge of Opcode
START
Command
DELAY[2:0]
tDELAY (tCLK)
tDELAY (µs)(1)
000
0
0
001
64
8.68
010
128
17.4
011
256
34.7
100
512
69.4
101
1024
139
110
2048
278
111
4096
556
(1) fCLK = 7.3728MHz.
After the programmable delay, the digital filter is reset
and a new conversion is started. DRDY goes low
when data are ready. There is no need to ignore or
discard data; the data are completely settled. The
total time to perform the first conversion is the sum of
the programmable delay time and the settling of the
digital filter. That is, the value of Table 8 and Table 9
combined. Figure 53 shows the timing and Table 9
shows the settling time with programmable delay
equal to '0'.
Table 9. Settling Time Using START
DATA RATE
(SPS)
SETTLING TIME (tSET) (ms)(1)
sinc1
sinc2
10
100
200
16.6
60.3
120
50
20.3
40.4
60
17.0
33.7
5.42
400
2.85
1200
1.18
2.10
3600
0.632
0.980
14,400
0.424
0.563
(1) fCLK = 7.3728MHz, DELAY[2:0] = 000.
DRDY
DOUT
tSET(1)
Settled
Data
(1) tSET = initial start delay plus the new conversion cycle time.
Figure 53. Data Retrieval Time After START
24
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
Settling Time While Continuously Converting
OFFSET AND GAIN
If there is a step change on the input signal while
continuously converting, the next data represent a
combination of the previous and current input signal
and should therefore be discarded; see Figure 54 for
this step change. Table 10 shows the number of
conversion cycles for completely settled data while
continuously converting.
The ADS1259 features low offset (40μV, typ) and low
gain errors (0.05%, typ). The offset and gain errors
can be corrected by sending calibration commands to
the ADS1259; see the Calibration section.
Table 10. Settling Time While Continuously
Converting DRDY Periods(1)
DATA RATE
(SPS)
SETTLING TIME (tSET) (Conversions)
sinc1
sinc2
10
2
3
16.6
2
3
50
2
3
60
2
3
400
2
3
1200
2
3
3600
3
4
14,400
6
7
The ADS1259 also features very low offset drift
(0.05μV/°C, typ) and very low gain drift (0.5ppm/°C,
typ). The offset and gain drift are calculated using the
box method, as described by Equation 6 and
Equation 7:
VOFFMAX - VOFFMIN
Offset Drift =
Temp Range
(6)
Gain Drift =
GainErrorMAX - GainErrorMIN
Temp Range
where:
VOFFMAX, VOFFMIN, GainErrorMAX, and GainErrorMIN
are the maximum and minimum offset and gain
error readings recorded over the Temp Range
(–40°C to +105°C)
(7)
(1) Settling time is defined as the number of DRDY periods after
the input signal has settled following an input step change.
For best data throughput in multiplexed applications, issue a
START condition (START pin or Start command) after the
input has settled following a multiplexer change; see the
Setling Time Using START section.
New VIN
VIN = AINP - AINN
Old VIN
DRDY
DOUT
Old VIN Data
tSET
Mix of
Old and New
VIN Data
Fully Settled
New VIN Data
Settled
Data
Figure 54. Step Change on VIN while Continuously Converting
Copyright © 2009–2011, Texas Instruments Incorporated
25
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
Table 11. Offset Calibration Values
OFFSET AND FULL-SCALE CALIBRATION
REGISTERS
The conversion data are scaled by offset and gain
registers before yielding the final output code. As
shown in Figure 55, the output of the digital filter is
first subtracted by the offset register (OFC) and then
multiplied by the full-scale register (FSC). Equation 8
shows the scaling:
FSC[2:0]
Final Output Data = (Input - OFC[2:0]) ´
400000h
(8)
The values of the offset and full-scale registers are
set by writing to them directly, or they are set by
calibration commands.
OFC[2:0] Registers
The offset calibration is a 24-bit word, composed of
three 8-bit registers, as shown in Table 13. The offset
is in twos complement format with a maximum
positive value of 7FFFFFh and a maximum negative
value of 800000h. This value is subtracted from the
conversion data. A register value of 00000h has no
offset correction (default value). Note that while the
offset calibration register value can correct offsets
ranging from –FS to +FS (as Table 11 shows), to
avoid input overload, the analog inputs cannot
exceed 105% full-scale.
AINP
Modulator
AINN
Digital
Filter
OFC REGISTER
FINAL OUTPUT CODE(1)
7FFFFFh
800001h
000001h
FFFFFFh
000000h
000000h
FFFFFFh
000001h
800001h
7FFFFFh
(1) Ideal output code excluding noise and inherent offset error.
FSC[2:0] Registers
The full-scale calibration is a 24-bit word, composed
of three 8-bit registers, as shown in Table 14. The
full-scale calibration value is 24-bit, straight binary,
normalized to 1.0 at code 400000h. Table 12
summarizes the scaling of the full-scale register. A
register value of 400000h (default value) has no gain
correction (gain = 1). Note that while the gain
calibration register value corrects gain errors above 1
(gain correction < 1), the full-scale range of the
analog inputs cannot exceed 105% full-scale to avoid
input overload.
Table 12. Full-Scale Calibration Register Values
FSC REGISTER
GAIN FACTOR
800000h
2.0
400000h
1.0
200000h
0.5
000000h
0
+
Output Data
Clipped to 24 Bits
S
´
OFC
Register
FSC Register
400000h
-
Final Output
Figure 55. Calibration Block Diagram
Table 13. Offset Calibration Word
REGISTER
BYTE
OFC0
LSB
B7
B6
B5
B4
BIT ORDER
B3
B2
B1
B0 (LSB)
OFC1
MID
B15
B14
B13
B12
B11
B10
B9
B8
OFC2
MSB
B23 (MSB)
B22
B21
B20
B19
B18
B17
B16
Table 14. Full-Scale Calibration Word
REGISTER
BYTE
FSC0
LSB
B7
B6
B5
B4
B3
B2
B1
B0 (LSB)
FSC1
MID
B15
B14
B13
B12
B11
B10
B9
B8
FSC2
MSB
B23 (MSB)
B22
B21
B20
B19
B18
B17
B16
26
BIT ORDER
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
CALIBRATION
The ADS1259 has commands to correct for system
offset and gain errors. Calibration can be performed
at any time the ADS1259 and associated circuitry
(such as the input amplifier, external reference, power
supplies, etc) have stabilized. Options include
calibrating after power-up, after temperature changes,
or calibration at regular intervals. To calibrate:
• Set the gate control mode (PULSE bit = 0)
• Start the ADS1259 conversions
• Apply the appropriate input to the ADS1259 (zero
or full-scale)
• Allow time for the input to completely settle
• Send the OFSCAL (offset calibration) or GANCAL
(full-scale calibration) command, as appropriate
• Wait for calibration to complete as given by the
time listed in Table 15. DRDY goes low when
calibration is complete. The conversion result at
this time uses the new offset or full-scale
calibration words.
Figure
56
shows the calibration
calibration, do not send commands.
timing.
During
The internal full-scale calibration word is bypassed
during offset calibration. Do not exceed +105% of
full-scale range for gain calibration. Note that the
out-of-range threshold is unaffected by gain
calibration.
Table 15. Calibration Timing
tCAL CALIBRATION TIME (ms)
DATA RATE
(SPS)
sinc1
sinc2
14400
1.89
2.19
3600
5.43
6.15
1200
14.9
16.7
400
43.2
48.4
60
284
318
50
341
380
16.6
1020
1140
10
1700
1900
1. fCLK = 7.3728MHz.
DRDY
(DOUT with CS = 0)
tCAL
Calibration complete
and first data ready.
Perform offset calibration prior to the gain calibration.
DIN
CAL Command
Figure 56. Calibration Timing
Copyright © 2009–2011, Texas Instruments Incorporated
27
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
SERIAL INTERFACE
DATA INPUT (DIN)
The SPI-compatible serial interface consists of four
signals: CS, SCLK, DIN, and DOUT or three signals,
in which case CS may be tied low. The interface is
used to read conversion data, configure registers,
and control the ADS1259 operation.
DIN is the input data pin and is used with SCLK to
send data to the ADS1259 (opcode commands and
register data). The device latches input data on the
falling edge of SCLK.
DATA OUTPUT (DOUT)
SERIAL COMMUNICATION
The ADS1259 communications occur by clocking
commands into the device (on DIN) and reading
register and conversion data (on DOUT). The SCLK
input is used to clock the data into and out of the
device. CS disables the ADS1259 serial port but
otherwise does not affect the ADC operation. The
communication protocol to the ADS1259 is
half-duplex. That is, data are transmitted to and from
the device one direction at a time.
Communications to and from the ADS1259 occurs on
8-bit boundaries. If an unintentional SCLK transition
should occur (such as from a possible noise spike),
the ADS1259 serial port may not respond properly.
The port can be reset by one of the following ways:
1. Take CS high and then low to reset the interface
2. Hold SCLK low for 216 fCLK cycles to reset the
interface
3. Take RESET/PWDN low and back high to overall
reset the device
4. Cycle the power supplies to overall reset the
device
CHIP SELECT (CS)
The chip select (CS) selects the ADS1259 for SPI
communication. To select the device, pull CS low. CS
must remain low for the duration of the serial
communication. When CS is taken high, the serial
interface is reset, input commands are ignored, and
DOUT enters a high-impedance state. If the
ADS1259 does not share the serial bus with another
device, CS may be tied low. Note that DRDY remains
active when CS is high.
DOUT is the output data pin and is used with SCLK
to read conversion and register data from the
ADS1259. In addition to providing data output, in
RDATAC mode DOUT indicates when data are
ready. Data are ready when DOUT transitions low. In
this manner, DOUT functions the same as DRDY
(with CS = 0), as shown in Figure 57. When reading
data, the data are shifted out on the rising edge of
SCLK. DOUT is in a 3-state condition when CS is
high.
DATA READY (DRDY)
DRDY is an output that indicates when conversion
data are available for reading (falling edge active).
DRDY is asserted on an output pin and also a
register bit. To poll the DRDY register bit, set the stop
read data continuous mode and then read the
CONFIG2 register. When the DRDY bit is low, data
can be read. The data read operation must complete
within 20 fCLK cycles of the next DRDY falling edge.
After power-on or after reset, DRDY defaults high.
When reading data in Gate Control mode, DRDY is
reset high on the first SCLK rising edge. If data are
not retrieved, DRDY pulses high during the new data
update time, as shown in Figure 57. Do not retrieve data
during this time as the data are invalid.
In Pulse Control mode, DRDY remains low until a
new conversion is started. The previous conversion
data may be read 20 tCLK prior to the DRDY falling
edge.
20 tCLK
DRDY Pin
Data Updating
(1)(2)
SERIAL CLOCK (SCLK)
The serial clock (SCLK) is a Schmitt-triggered input
used to clock data into and out of the ADS1259. Even
though the input is relatively noise immune, it is
recommended to keep SCLK as clean as possible to
prevent glitches from accidentally shifting the data. If
SCLK is held low for 216 fCLK periods, the serial
interface resets. After reset the next communication
cycle can be started. The timeout can be used to
recover communication when the serial interface is
interrupted. The SPI timeout is enabled by register bit
SPI. When the serial interface is idle, hold SCLK low.
28
(1) DOUT functions in the same manner as the DRDY pin if CS is
low and in the RDATAC mode.
(2) The DRDY bit functions in the same manner as the DRDY pin
(SDATAC mode only).
Figure 57. DRDY and DOUT With No Data
Retrieval
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
DATA FORMAT
DATA CHECKSUM BYTE AND FLAG BIT
The ADS1259 outputs 24 bits of conversion data in
binary twos complement format, MSB first. The data
LSB has a weight of VREF/(223 – 1). A positive
full-scale input produces an output code of 7FFFFFh
and the negative full-scale input produces an output
code of 800000h. The output clips at these codes for
signals that exceed full-scale. Table 16 summarizes the
ideal output codes for different input signals.
An optional checksum byte can be appended to the
conversion data bytes. The checksum makes the
data word length four bytes in length instead of three.
The checksum byte is enabled by the register bit
CHKSUM. The checksum itself is the least significant
byte sum of the three conversion data bytes, offset by
9Bh. Note that the checksum byte option only applies
to the readback conversion data, not to register data.
The checksum is either seven bits or eight bits,
depending if the FLAG register bit is enabled. If the
FLAG bit is enabled the checksum is seven bits, with
bit 7 replaced by the out-of-range flag. Figure 58 and
Table 17 describe the combinations of the FLAG and
CHKSUM register bits.
Table 16. Ideal Output Code versus Input Signal
DIFFERENTIAL INPUT SIGNAL VIN
(AINP – AINN)
IDEAL OUTPUT
CODE(1)
≥ VREF
+VREF
(223 - 1)
0
7FFFFFh
000001h
Checksum = MSB data byte + Mid data byte + LSB
data byte + 9Bh.
000000h
32-Bit Conversion Data (CHKSUM = 1)
-VREF
(223 - 1)
FFFFFFh
24-Bit Conversion Data (CHKSUM = 0)
MSB
£ -VREF
2
MID
LSB
CHECKSUM
23
23
800000h
Flag = 1; Bit 0 of LSB Conversion Data
2 -1
(1) Excludes effects of noise, linearity, offset, and gain errors.
DATA INTEGRITY
Data readback integrity is augmented by a checksum
byte and redundant data read capability. The
checksum byte is the sum of three data conversion
bytes, offset by 9Bh. Additionally, the data conversion
bytes may be read multiple times by continuing to
shift data past the initial read of 24 bits (32 bits if
checksum is enabled).
Copyright © 2009–2011, Texas Instruments Incorporated
or
Bit 7 of Checksum
Figure 58. Checksum Byte and Out-of-Range Flag
Table 17. Checksum Byte and Over-Range Flag
FLAG
REGISTER
BIT
CHKSUM
REGISTER
BIT
0
0
No checksum byte, no out-of-range
flag
0
1
8-bit checksum byte, no
out-of-range flag
1
0
No checksum byte, out-of-range
flag replaces LSB (bit 0) of
conversion data
1
1
7-bit checksum byte, out-of-range
replaces MSB (bit 7) of checksum
byte.
DESCRIPTION
29
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
DATA RETRIEVAL
New conversion data are available when DRDY goes
low. Read the data within 20 fCLK cycles of the next
DRDY falling edge or the data are incorrect. Do not
read data during this interval. The conversion data
may be read in two ways: Data Read in Continuous
mode and Data Read in Stop Continuous mode.
Data Read Operation in Continuous Mode
The Read Data Continuous mode is cancelled by
sending the Stop Read Data Continuous command
(SDATAC). This operation occurs simultaneously with
ADC conversion data on DOUT which can be
ignored. Once the SDATAC command is sent, other
commands may be sent to the ADS1259. Observe
the SCLK and DRDY timing requirements, when
reading data in this mode, as shown in Figure 59 and
Table 18.
In Read Data Continuous mode the conversion data
may be shifted out directly without the need of the
data read command. When DRDY (and DOUT, if CS
is low) assert low, the conversion data are ready. The
data are shifted out on DOUT on the rising edges of
SCLK, with the most significant bit (MSB) clocked out
first. In Gate Convert Mode, DRDY returns to high on
the first falling edge of SCLK. In Pulse Convert mode,
DRDY remains low until a new conversion starts.
As shown in Figure 60, the conversion data consist of
three or four bytes (data MSB first), depending on
whether the checksum byte is included. The data
may be read multiple times by continuing to shift the
data. The data read operation must be completed
with 20 fCLK cycles of next DRDY falling edge.
tDRSC
DRDY
SCLK
tSCDR
Figure 59. SCLK to DRDY Timing
Table 18. SCLK and DRDY Timing Characteristics
for Figure 57
SYMBOL
tSCDR
(1)
tDRSC
(1)
DESCRIPTION
MIN
UNIT
SCLK low before DRDY
low(1)
20
tCLK
DRDY falling edge to SCLK
rising edge(1)
40
ns
(1) These requirements apply only to reading conversion data in
RDATAC mode.
Data Ready
DRDY
Next Data Ready
(1)
CS
(2)
1
SCLK
DOUT
DIN
9
17
25
33
(4)
tUPDATE
(3)
Hi-Z
(5)
DATA MSB
DATA MID
DATA LSB
CHECKSUM
(6)
(7)
DATA MSB
(8)
(1) In Gate Convert Conversion mode, DRDY returns to high on the first falling edge of SCLK. In Pulse Convert mode, DRDY remains low
until the next conversion is started.
(2) CS may be held low. If CS is low, DOUT asserts low with DRDY.
(3) Data are updated on the rising edge of SCLK. DOUT is low until the first rising edge of SCLK.
(4) tUPDATE = 20/fCLK. Do not read data during this time.
(5) During this interval, DOUT follows DRDY.
(6) Optional data checksum byte.
(7) Optional repeat of previous conversion data.
(8) Hold DIN low, except for transmission of the SDATAC (STOP Read Data Continuous command).
Figure 60. Data Read Operation in Continuous Mode
30
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
Data Read Operation in Stop Continuous Mode
As shown in Figure 61, after sending the RDATA
command the data are shifted out on DOUT on the
rising edges of SCLK. The MSB is clocked out on the
first rising edge of SCLK. In Gate Control mode,
DRDY returns to high on the first falling edge of
SCLK. In Pulse Control mode, DRDY remains low
until a new conversion is started.
In Stop Read Data Continuous mode, a read data
command (RDATA) must be sent for each new data
read operation. New conversion data are ready when
DRDY falls low or the DRDY register bit transitions
low. The data read operation may then occur. The
read data command must be sent at least 20 fCLK
cycles before the DRDY falling edge or the data are
incorrect. Do not the read data command during this
time.
The conversion data consist of three or four bytes
(MSB first), depending on whether the checksum byte
is included. The data may be read multiple times by
continuing to shift the data.
Data Ready
DRDY
Next Data Ready
(1)
(3)
tUPDATE
CS
(2)
1
9
17
25
33
41
SCLK
DOUT
DIN
(7)
Hi-Z
(4)
012h
DATA MSB
DATA MID
DATA LSB
CHECKSUM
(5)
(6)
DATA MSB
(8)
(1) In Gate Control mode, DRDY returns to high on the first falling edge of SCLK. In Pulse Control mode, DRDY remains low until the next
conversion is started. The DRDY pin or DRDY register bit can also be polled to determine when data are ready.
(2) CS may be held low.
(3) tUPDATE = 20/fCLK. Do not issue the Read Data opcode during this time.
(4) During this interval, DOUT does not follow DRDY (stop continuous mode).
(5) Optional conversion data checksum.
(6) Optional repeat of previous conversion data.
(7) DIN data are latched on the falling edge of SCLK. Data are output on the rising edges of SCLK.
(8) Read Data command = 012h.
Figure 61. Data Read Operation in STOP Continuous Mode
Copyright © 2009–2011, Texas Instruments Incorporated
31
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
COMMAND DEFINITIONS
The commands summarized in Table 19 control and configure the operation of the ADS1259. The commands
are stand-alone, except for the register read and register write operations which require a second command byte
plus data. CS can be taken high or held low between opcode commands but must stay low for the entire
command operation. Note that the Read Data Continuous mode must be cancelled by the Stop Read Data
Continuous mode opcode (SDATAC) before sending further commands.
Table 19. Command Definitions (1)
COMMAND
TYPE
DESCRIPTION
FIRST OPCODE BYTE
WAKEUP
Control
Wake up from SLEEP mode
0000 001x (02h or 03h) (2)
SLEEP
Control
Begin SLEEP mode
0000 010x (04h or 05h) (2)
RESET
Control
Reset to power-up values
0000 011x (06h or 07h) (2)
START
Control
START conversion
0000 100x (08h or 09h) (2)
STOP
Control
STOP conversion
0000 101x (0Ah or 0Bh) (2)
RDATAC
Control
Set Read Data Continuous mode
0001 0000 (10h)
SDATAC
Control
Stop Read Data Continuous mode
0001 0001 (11h)
RDATA
Data
Read data by opcode
0001 001x (12h or 13h) (2)
(1)
(2)
SECOND OPCODE BYTE
RREG
Register
Read nnnn register at address rrrr
0010 rrrr (20h + 0000 rrrr)
0000 nnnn (00h + nnnn)
WREG
Register
Write nnnn register at address rrrr
0100 rrrr (40h + 0000 rrrr)
0000 nnnn (00h + nnnn)
OFSCAL
Calibration
Offset calibration
0001 1000 (18h)
GANCAL
Calibration
Gain calibration
0001 1001 (19h)
nnnn = number of registers to be read/written – 1. For example, to read/write 3 registers, set nnnn = 2 (0010).
rrrr = starting register address for read/write opcodes.
These commands are decoded on the seventh bit of the opcode. The eighth bit is a don't care bit. All other commands are decoded on
the eighth bit.
WAKEUP: Exit SLEEP Mode
Description: This command exits the low-power SLEEP mode; see the SLEEP Mode section.
SLEEP: Enter SLEEP Mode
Description: This command enters the low-power SLEEP mode. See the SLEEP Mode section.
RESET: Reset Registers to Default Values
Description: This command resets the digital filter cycle and returns all register settings to the default values.
START: Start Conversions
Description: This command starts data conversions. If PULSE bit = 1, then a single conversion is performed. If
PULSE bit = 0, then conversions continue until the STOP command is sent. Tie the START pin low to control
conversions by command.
STOP: Stop Conversions
Description: This command stops conversions. When the STOP command is sent, the conversion in progress
completes and further conversions are stopped. If conversions are already stopped, this command has no effect.
See the Conversion Control section. Tie the START pin low to control conversions by command.
RDATAC: Read Data Continuous
Description: This command enables the Read Data Continuous mode (default). See the Read Data Continuous
Mode section for details. Disable this mode with the SDATAC command before sending other commands.
SDATAC: Stop Read Data Continuous
Description: This command cancels the Read Data Continuous mode.
32
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
RDATA: Read Data
Description: Issue this command opcode after DRDY goes low to read the conversion result (in Stop Read Data
Continuous mode). See the Read Data Mode section for more details.
RREG: Read from Registers
Description: These opcode bytes read register data. The Register Read command is a two-byte opcode
followed by the output of the register data. The first byte contains the command opcode and the register address.
The second byte of the opcode specifies the number of registers to read – 1.
First opcode byte: 0010 rrrr, where rrrr is the starting register address.
Second opcode byte: 0000 nnnn, where nnnn is the number of registers to read.
The 17th SCLK rising edge of the operation clocks out the MSB of the first register.
(1)
CS
1
9
17
25
SCLK
DIN
OPCODE 1
DOUT
OPCODE 2
REG DATA
REG DATA + 1
(1) CS may be tied low.
Figure 62. RREG Command Example: Read Two Registers Starting from Register 00h (CONFIG0)
(OPCODE 1 = 0010 0000, OPCODE 2 = 0000 0001)
Copyright © 2009–2011, Texas Instruments Incorporated
33
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
WREG: Write to Register
Description: These two opcode bytes write register data. The Register Write command is a two-byte opcode
followed by the register data. The first byte contains the command opcode and the register address. The second
byte of the opcode specifies the number of registers to write – 1.
First opcode byte: 0100 rrrr, where rrrr is the starting register address.
Second opcode byte: 0000 nnnn, where nnnn is the number of registers to write
After the opcode bytes, the register data follows (in MSB-first format).
(1)
CS
1
9
17
25
SCLK
DIN
OPCODE 1
OPCODE 2
REG DATA 1
REG DATA 2
DOUT
(1) CS may be tied low.
Figure 63. WREG Command Example: Write Two Registers Starting from 00h (CONFIG0)
(OPCODE 1 = 0100 0000, OPCODE 2 = 0000 0001)
OFSCAL: Offset Calibration
Description: This command performs an offset calibration. Apply a zero signal and allow the input to stabilize
before sending the command; see the Calibration section for more details.
GANCAL: Gain Calibration
Description: This command performs a gain calibration. Apply a full-scale signal and allow the input to stabilize
before sending the command; see the Calibration section for more details.
34
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
REGISTER MAP
The operation of the ADS1259 is controlled through a set of registers. Collectively, the registers contain all the
information needed to configure the part, such as data rate, calibration, etc. Table 20 shows the register map.
Table 20. Register Map
ADDRESS
REGISTER
RESET
VALUE
BIT 7
BIT 6
BIT 5
BIT 4
BIT 3
BIT 2
BIT 1
0h
CONFIG0
10XX0101b
1
0
ID1
ID0
0
RBIAS
0
SPI
1h
CONFIG1
00001000b
FLAG
CHKSUM
0
SINC2
EXTREF
DELAY2
DELAY1
DELAY0
2h
CONFIG2
XX000000b
DRDY
EXTCLK
SYNCOUT
PULSE
0
DR2
DR1
DR0
3h
OFC0
00000000b
OFC07
OFC06
OFC05
OFC04
OFC03
OFC02
OFC01
OFC00
4h
OFC1
00000000b
OFC15
OFC14
OFC13
OFC12
OFC11
OFC10
OFC09
OFC08
5h
OFC2
00000000b
OFC23
OFC22
OFC21
OFC20
OFC19
OFC18
OFC17
OFC16
6h
FSC0
00000000b
FSC07
FSC06
FSC05
FSC04
FSC03
FSC02
FSC01
FSC00
7h
FSC1
00000000b
FSC15
FSC14
FSC13
FSC12
FSC11
FSC10
FSC09
FSC08
8h
FSC2
01000000b
FSC23
FSC22
FSC21
FSC20
FSC19
FSC18
FSC17
FSC16
BIT 0
CONFIG0: CONFIGURATION REGISTER 0 (Address = 0h)
7
6
5
4
3
2
1
0
1
0
ID1
ID0
0
RBIAS
0
SPI
Reset value = 10XX0101b.
Bit 7
Reserved (read-only)
Always returns '1'.
Bit 6
Reserved (read-only)
Always returns '0'.
Bits 5-4
ID[1:0]: Factory-programmed identification bits (read-only)
(Note that these bits may change without notification.)
Bit 3
Reserved
Always write '0'.
Bit 2
RBIAS: Internal reference bias
0 = Internal reference bias disabled
1 = Internal reference bias enabled (default)
Bit 1
Reserved
Always write '0'.
Bit 0
SPI: SCLK timeout of SPI interface
0 = SPI timeout disabled
1 = SPI timeout enabled (default), when SCLK is held low for 216 clock cycles
Copyright © 2009–2011, Texas Instruments Incorporated
35
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
CONFIG1: CONFIGURATION REGISTER 1 (Address = 1h)
7
6
5
4
3
2
1
0
FLAG
CHKSUM
0
SINC2
EXTREF
DELAY2
DELAY1
DELAY0
Reset value = 00001000b.
Bit 7
FLAG: Out-of-range flag
0 = Disabled (default)
1 = Enabled: replaces bit 24 (LSB) of the conversion data with the out-of-range bit; if the
CHKSUM byte is enabled, bit 7 of the checksum byte
Bit 6
CHKSUM: Checksum
0 = Disabled (default)
1 = Conversion data checksum byte included in readback
Bit 5
Reserved
Always write '0'.
Bit 4
SINC2: Digital filter mode
0 = sinc1 filter (default)
1 = sinc2 filter
Bit 3
EXTREF: Reference select
0 = Internal
1 = External (default)
Bits 2-0
DELAY[2:0]: START conversion delay
000
001
010
011
100
101
110
111
36
=
=
=
=
=
=
=
=
No delay (default)
64 tCLK
128 tCLK
256 tCLK
512 tCLK
1024 tCLK
2048 tCLK
4096 tCLK
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
CONFIG2: CONFIGURATION REGISTER 2 (Address = 2h)
7
6
5
4
3
2
1
0
DRDY
EXTCLK
SYNCOUT
PULSE
0
DR2
DR1
DR0
Reset value = XX000000b.
Bit 7
DRDY: Data ready (read-only)
This bit duplicates the state of the DRDY pin. Poll this bit to indicate that data are ready. When
DRDY is low, data are ready.
Bit 6
EXTCLK: Clock source (read-only)
0 = Device clock source is internal oscillator
1 = Device clock source is external clock
Note that the ADS1259 selects the clock source automatically.
Bit 5
SYNCOUT: SYNCOUT clock enable
0 = SYNCOUT disabled (default)
1 = SYNCOUT enabled
Note that if disabled, the output is driven low.
Bit 4
PULSE: Conversion Control mode select
0 = Gate Control mode (default)
1 = Pulse Control mode
Bit 3
Reserved
Always write '0'
Bits 2-0
DR[2:0] Data rate setting
000 = 10SPS (default)
001 = 16.6SPS
010 = 50SPS
011 = 60SPS
100 = 400SPS
101 = 1200SPS
110 = 3600SPS
111 = 14400SPS
NOTE: fCLK = 7.3728MHz
Copyright © 2009–2011, Texas Instruments Incorporated
37
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
OFC0: OFFSET CALIBRATION BYTE 0, LEAST SIGNIFICANT BYTE (Address = 3h)
7
6
5
4
3
2
1
0
OFC07
OFC06
OFC05
OFC04
OFC03
OFC02
OFC01
OFC00
Reset value = 00000000b.
OFC1: OFFSET CALIBRATION BYTE 1 (Address = 4h)
7
6
5
4
3
2
1
0
OFC15
OFC14
OFC13
OFC12
OFC11
OFC10
OFC09
OFC08
Reset value = 00000000b.
OFC2: OFFSET CALIBRATION BYTE 2, MOST SIGNIFICANT BYTE (Address = 5h)
7
6
5
4
3
2
1
0
OFC23
OFC22
OFC21
OFC20
OFC19
OFC18
OFC17
OFC16
Reset value = 00000000b.
FSC0: FULL-SCALE CALIBRATION BYTE 0, LEAST SIGNIFICANT BYTE (Address = 6h)
7
6
5
4
3
2
1
0
FSC07
FSC06
FSC05
FSC04
FSC03
FSC02
FSC01
FSC00
Reset value = 00000000b.
FSC1: FULL-SCALE CALIBRATION BYTE 1 (Address = 7h)
7
6
5
4
3
2
1
0
FSC15
FSC14
FSC13
FSC12
FSC11
FSC10
FSC09
FSC08
Reset value = 00000000b.
FSC2: FULL-SCALE CALIBRATION BYTE 2, MOST SIGNIFICANT BYTE (Address = 8h)
7
6
5
4
3
2
1
0
FSC23
FSC22
FSC21
FSC20
FSC19
FSC18
FSC17
FSC16
Reset value = 01000000b.
38
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
BASIC CONNECTION
The ADS1259 basic connections are shown in Figure 64. The diagram shows the ADS1259 operating with an
internal oscillator and with internal reference. Dual ±2.5V analog power supplies are also shown. Pins 6-9 are the
SPI port connection. The remaining digital I/O pins connect to the controller I/O. Note that the minimum
configuration of the digital I/O may include only SCLK, DIN, and DOUT.
+2.5V
(1)
(+)
20W to 50W
1
10nF
Signal Input
(-)
(2)
2
AINN
3
20W to 50W
Controller I/O
4
4.7kW 5
Drives the PGA280
SYNCIN Pin
ADS1259
AINP
VREFN
START
VREFP
SYNCOUT
CS
7
Controller SPI Port
8
9
Controller I/O
AVSS
RESET/PWDN
6
1MW
AVDD
SCLK
DIN
REFOUT
DVDD
DGND
BYPASS
DOUT
XTAL2
DRDY
XTAL1/CLKIN
10
1mF
20
-2.5V
19
1 mF
18
17
+
1mF
+
1mF
+
16
15
+3.3V
1mF
14
13
2.5V
Reference Output
+
1mF
12
11
1MW
(1) It is recommended to buffer the ADS1259 inputs. The output isolation resistors may be incorporated within the amplifier feedback loop.
(2) Low distortion C0G or film capacitor recommended.
Figure 64. ADS1259 Basic Connection Diagram
LAYOUT
Place the input buffer and input decoupling capacitors close to the ADS1259 inputs. The bypass capacitors for
power-supply and reference decoupling should also be placed close to the device. In some cases, it may be
necessary to use a split ground plane in which digital return currents of external components are routed away
from the ADS1259. In this case, connect the grounds at the power supply.
CONFIGURATION GUIDE
Configuration of the ADS1259 involves configuring the device hardware (power supply, I/O pins, etc) and device
register settings. The registers are configured by commands sent via the device SPI port.
Power Supplies
The ADS1259 analog section operates either with a single +5V or dual ±2.5V supplies. The digital section
operates from +2.7V to +5V. The digital and analog power supplies may be tied together (+5V only).
Reference
Select either the internal reference or an external reference for the ADS1259 (see the Reference section). The
default is external reference. Figure 64 depicts the internal reference connection.
Clock
Choose the desired clock source (see the Clock Source section). Figure 64 depicts the internal clock operation.
Copyright © 2009–2011, Texas Instruments Incorporated
39
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
SYNCOUT Pin
Connect the SYNCOUT pin to the SYNCIN pin of the PGA280, using a 4.7kΩ series resistor (placed close to the
ADS1259). The 1MΩ pull-down resistor is required when the ADS1259 is in power-down mode.
RESET/PWDN Pin
This pin must be high in normal operation. If it is desired to completely power down the device, or to have a
hardware reset control, then connect this pin to a controller. If these functions are not needed, tie the pin high.
(Note that the device can both be reset and SLEEP mode engaged by commands.)
START Pin
If it is desired to control conversions by pin, connect this line to the controller. Otherwise, this line can be tied
high to free-run conversions. The conversions can also be controlled by software commands. In this case, tie the
START pin low.
DRDY Pin
DRDY is an output that indicates when data are ready for readback. Note that the DOUT pin (and also the DRDY
register bit) indicates when data are ready as well, so DRDY connection to a controller is optional.
CS Pin
If the ADS1259 is a single device connected to the SPI bus, then CS can be tied low. Otherwise, for applications
where the ADS1259 shares the bus with another device, CS must be connected.
DOUT Pin
When the ADS1259 SPI is deselected (CS = 1), the DOUT pin is in 3-state mode. A pull-down resistor may be
necessary to prevent floating the controller input pin.
Miscellaneous Digital I/O
Avoid ringing on the digital inputs and outputs. Resistors in series with the trace driving end helps to reduce
ringing by controlling impedances.
SOFTWARE GUIDE
After the power supplies have fully established, allow a minimum of 216 system clocks before beginning
communication to the device. The registers can then be configured by commands via the SPI port. The following
steps detail a suggested procedure to initialize the ADS1259.
1. Send the SDATAC command <11h>. This command cancels the RDATAC mode. RDATAC mode must be
cancelled before the register write commands.
2. Send the register write command. The following example shows the register write as a block of nine bytes,
starting at register 0 (CONFIG0).
BYTES
DATA
OPERATION
1, 2
01000000, 00001000
Write register opcode bytes, starting at address 0, 9-byte block
3
00000101
CONFIG0; register data, bias the reference, SPI timeout
4
01010000
CONFIG1; checksum enabled, sinc2 filter selection, internal
reference
5
00000011
CONFIG2; Gate Convert mode, 60SPS
6, 7, 8
00000000, 00000000,
00000000
OFC[2:0]; 3 bytes for offset, no offset correction
9, 10, 11
00000000, 00000000,
01000000
FSC[2:0]; 3 bytes for gain, no full-scale correction
40
Copyright © 2009–2011, Texas Instruments Incorporated
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
3. Optional readback verification of the register data
READ register command: <20h>, <08>
The nine bytes of readback data that follow represent the nine register bytes.
4. Take the START pin high or send the START command to start conversions.
5. Optionally, send the RDATAC command <10h>. This permits reading of conversion data without the need of
the read data command. Otherwise, the read data opcode must be sent to read each conversion result.
6. When the DRDY pin or the DRDY bit goes low, or when DOUT transitions low, read the data.
PGA280 APPLICATION
Figure 65 shows the ADS1259 connected to the PGA280. The PGA280 is a programmable gain, fully-differential
instrumentation amplifier that is ideally suited to drive the ADS1259. The amplifier features ±5V to ±18V supply
input section that accepts wide ranging signal levels and features a +5V output section that matches the
ADS1259 low-voltage inputs. The ADS1259 +2.5V REFOUT drives the PGA280 VOCM pin to level shift the
signal.
The ADS1259 provides a clock output (SYNCOUT) that drives the PGA280 (GPIO6) chopping clock input. An
optional extended CS (ECS) function feature of the PGA280 (GPIO0) allows use of one CS to alternately select
each device for SPI communication. Additionally, the optional BUFA trigger output of the PGA280 (GPIO5) starts
the ADS1259 conversions. The trigger can be delayed to occur after an input multiplexer change. The delay
allows settling of the PGA280 before the ADC conversion begins.
VSN
(1)
1mF
VSOP
9
VOP
INP1
2
50W
10
INN1
7
MUX
PGA
VOCM
3
VON
1
INP2
8
1
10nF
(3)
1mF 50W
INN2
16
GPIO5
PGA280
GPIO0
17
14
16
24
13
18
AVDD
VREFP
VREFN
RESET/PWDN
5
DRDY
8
DIN
DOUT
SYNCOUT
START
SCLK
ADS1259
100kW
9
7
SPI
6
1MW
(1)
Controller
10
REFOUT
CS
+3.3V
5
3
AINP
AINN
4
19
VSON
SCLK
SDO
SDI
15
DVDD
DGND
CS
12
4.7kW
18
17
2
(2)
GPIO6
20
1mF
19
15
14
BYPASS
VSP
4
XTAL2
11
1 mF
+
XTAL1
6
+5V
(1)
DGND
-15V
DVDD
(1)
AVSS
+15V
11
12
13
1mF
(1) Refer to the PGA280 product data sheet for power-supply bypassing recommendations.
(2) Locate this resistor as close as possible to pin 5 of the ADS1259.
(3) C0G or film capacitor.
Figure 65. PGA280 Driving the ADS1259
Copyright © 2009–2011, Texas Instruments Incorporated
41
ADS1259
SBAS424D – JUNE 2009 – REVISED AUGUST 2011
www.ti.com
REVISION HISTORY
NOTE: Page numbers for previous revisions may differ from page numbers in the current version.
Changes from Revision C (March 2010) to Revision D
Page
•
Added Internal Voltage Reference, Long-term stability parameter to Electrical Characteristics table ................................. 3
•
Added Internal Voltage Reference, Thermal hysteresis parameter and footnote 11 to Electrical Characteristics table ...... 3
•
Added Figure 31 ................................................................................................................................................................. 11
•
Added Thermal Hysteresis subsection to Reference section ............................................................................................. 17
Changes from Revision B (January 2010) to Revision C
Page
•
Changed ADS1259B and ADS1259 Internal Voltage Reference, Accuracy parameter in the Electrical
Characteristics ...................................................................................................................................................................... 3
•
Changed ADS1259 Internal Voltage Reference, Temperature drift parameter in the Electrical Characteristics ................. 3
•
Added Figure 23, ADS1259 internal reference voltage versus temperature graph ............................................................ 10
•
Added PGA280 Application section .................................................................................................................................... 41
42
Copyright © 2009–2011, Texas Instruments Incorporated
PACKAGE OPTION ADDENDUM
www.ti.com
19-Aug-2011
PACKAGING INFORMATION
Orderable Device
Status
(1)
Package Type Package
Drawing
Pins
Package Qty
Eco Plan
(2)
Lead/
Ball Finish
MSL Peak Temp
ADS1259BIPW
ACTIVE
TSSOP
PW
20
70
Green (RoHS
& no Sb/Br)
CU NIPDAU Level-1-260C-UNLIM
ADS1259BIPWR
ACTIVE
TSSOP
PW
20
2000
Green (RoHS
& no Sb/Br)
CU NIPDAU Level-1-260C-UNLIM
ADS1259IPW
ACTIVE
TSSOP
PW
20
70
Green (RoHS
& no Sb/Br)
CU NIPDAU Level-1-260C-UNLIM
ADS1259IPWR
ACTIVE
TSSOP
PW
20
2000
Green (RoHS
& no Sb/Br)
CU NIPDAU Level-1-260C-UNLIM
(3)
Samples
(Requires Login)
(1)
The marketing status values are defined as follows:
ACTIVE: Product device recommended for new designs.
LIFEBUY: TI has announced that the device will be discontinued, and a lifetime-buy period is in effect.
NRND: Not recommended for new designs. Device is in production to support existing customers, but TI does not recommend using this part in a new design.
PREVIEW: Device has been announced but is not in production. Samples may or may not be available.
OBSOLETE: TI has discontinued the production of the device.
(2)
Eco Plan - The planned eco-friendly classification: Pb-Free (RoHS), Pb-Free (RoHS Exempt), or Green (RoHS & no Sb/Br) - please check http://www.ti.com/productcontent for the latest availability
information and additional product content details.
TBD: The Pb-Free/Green conversion plan has not been defined.
Pb-Free (RoHS): TI's terms "Lead-Free" or "Pb-Free" mean semiconductor products that are compatible with the current RoHS requirements for all 6 substances, including the requirement that
lead not exceed 0.1% by weight in homogeneous materials. Where designed to be soldered at high temperatures, TI Pb-Free products are suitable for use in specified lead-free processes.
Pb-Free (RoHS Exempt): This component has a RoHS exemption for either 1) lead-based flip-chip solder bumps used between the die and package, or 2) lead-based die adhesive used between
the die and leadframe. The component is otherwise considered Pb-Free (RoHS compatible) as defined above.
Green (RoHS & no Sb/Br): TI defines "Green" to mean Pb-Free (RoHS compatible), and free of Bromine (Br) and Antimony (Sb) based flame retardants (Br or Sb do not exceed 0.1% by weight
in homogeneous material)
(3)
MSL, Peak Temp. -- The Moisture Sensitivity Level rating according to the JEDEC industry standard classifications, and peak solder temperature.
Important Information and Disclaimer:The information provided on this page represents TI's knowledge and belief as of the date that it is provided. TI bases its knowledge and belief on information
provided by third parties, and makes no representation or warranty as to the accuracy of such information. Efforts are underway to better integrate information from third parties. TI has taken and
continues to take reasonable steps to provide representative and accurate information but may not have conducted destructive testing or chemical analysis on incoming materials and chemicals.
TI and TI suppliers consider certain information to be proprietary, and thus CAS numbers and other limited information may not be available for release.
In no event shall TI's liability arising out of such information exceed the total purchase price of the TI part(s) at issue in this document sold by TI to Customer on an annual basis.
Addendum-Page 1
PACKAGE MATERIALS INFORMATION
www.ti.com
14-Jul-2012
TAPE AND REEL INFORMATION
*All dimensions are nominal
Device
Package Package Pins
Type Drawing
SPQ
Reel
Reel
A0
Diameter Width (mm)
(mm) W1 (mm)
B0
(mm)
K0
(mm)
P1
(mm)
W
Pin1
(mm) Quadrant
ADS1259BIPWR
TSSOP
PW
20
2000
330.0
16.4
6.95
7.1
1.6
8.0
16.0
Q1
ADS1259IPWR
TSSOP
PW
20
2000
330.0
16.4
6.95
7.1
1.6
8.0
16.0
Q1
Pack Materials-Page 1
PACKAGE MATERIALS INFORMATION
www.ti.com
14-Jul-2012
*All dimensions are nominal
Device
Package Type
Package Drawing
Pins
SPQ
Length (mm)
Width (mm)
Height (mm)
ADS1259BIPWR
TSSOP
PW
20
2000
367.0
367.0
38.0
ADS1259IPWR
TSSOP
PW
20
2000
367.0
367.0
38.0
Pack Materials-Page 2
IMPORTANT NOTICE
Texas Instruments Incorporated and its subsidiaries (TI) reserve the right to make corrections, enhancements, improvements and other
changes to its semiconductor products and services per JESD46C and to discontinue any product or service per JESD48B. Buyers should
obtain the latest relevant information before placing orders and should verify that such information is current and complete. All
semiconductor products (also referred to herein as “components”) are sold subject to TI’s terms and conditions of sale supplied at the time
of order acknowledgment.
TI warrants performance of its components to the specifications applicable at the time of sale, in accordance with the warranty in TI’s terms
and conditions of sale of semiconductor products. Testing and other quality control techniques are used to the extent TI deems necessary
to support this warranty. Except where mandated by applicable law, testing of all parameters of each component is not necessarily
performed.
TI assumes no liability for applications assistance or the design of Buyers’ products. Buyers are responsible for their products and
applications using TI components. To minimize the risks associated with Buyers’ products and applications, Buyers should provide
adequate design and operating safeguards.
TI does not warrant or represent that any license, either express or implied, is granted under any patent right, copyright, mask work right, or
other intellectual property right relating to any combination, machine, or process in which TI components or services are used. Information
published by TI regarding third-party products or services does not constitute a license to use such products or services or a warranty or
endorsement thereof. Use of such information may require a license from a third party under the patents or other intellectual property of the
third party, or a license from TI under the patents or other intellectual property of TI.
Reproduction of significant portions of TI information in TI data books or data sheets is permissible only if reproduction is without alteration
and is accompanied by all associated warranties, conditions, limitations, and notices. TI is not responsible or liable for such altered
documentation. Information of third parties may be subject to additional restrictions.
Resale of TI components or services with statements different from or beyond the parameters stated by TI for that component or service
voids all express and any implied warranties for the associated TI component or service and is an unfair and deceptive business practice.
TI is not responsible or liable for any such statements.
Buyer acknowledges and agrees that it is solely responsible for compliance with all legal, regulatory and safety-related requirements
concerning its products, and any use of TI components in its applications, notwithstanding any applications-related information or support
that may be provided by TI. Buyer represents and agrees that it has all the necessary expertise to create and implement safeguards which
anticipate dangerous consequences of failures, monitor failures and their consequences, lessen the likelihood of failures that might cause
harm and take appropriate remedial actions. Buyer will fully indemnify TI and its representatives against any damages arising out of the use
of any TI components in safety-critical applications.
In some cases, TI components may be promoted specifically to facilitate safety-related applications. With such components, TI’s goal is to
help enable customers to design and create their own end-product solutions that meet applicable functional safety standards and
requirements. Nonetheless, such components are subject to these terms.
No TI components are authorized for use in FDA Class III (or similar life-critical medical equipment) unless authorized officers of the parties
have executed a special agreement specifically governing such use.
Only those TI components which TI has specifically designated as military grade or “enhanced plastic” are designed and intended for use in
military/aerospace applications or environments. Buyer acknowledges and agrees that any military or aerospace use of TI components
which have not been so designated is solely at the Buyer's risk, and that Buyer is solely responsible for compliance with all legal and
regulatory requirements in connection with such use.
TI has specifically designated certain components which meet ISO/TS16949 requirements, mainly for automotive use. Components which
have not been so designated are neither designed nor intended for automotive use; and TI will not be responsible for any failure of such
components to meet such requirements.
Products
Applications
Audio
www.ti.com/audio
Automotive and Transportation www.ti.com/automotive
Amplifiers
amplifier.ti.com
Communications and Telecom www.ti.com/communications
Data Converters
dataconverter.ti.com
Computers and Peripherals
www.ti.com/computers
DLP® Products
www.dlp.com
Consumer Electronics
www.ti.com/consumer-apps
DSP
dsp.ti.com
Energy and Lighting
www.ti.com/energy
Clocks and Timers
www.ti.com/clocks
Industrial
www.ti.com/industrial
Interface
interface.ti.com
Medical
www.ti.com/medical
Logic
logic.ti.com
Security
www.ti.com/security
Power Mgmt
power.ti.com
Space, Avionics and Defense
www.ti.com/space-avionics-defense
Microcontrollers
microcontroller.ti.com
Video and Imaging
www.ti.com/video
RFID
www.ti-rfid.com
OMAP Mobile Processors
www.ti.com/omap
TI E2E Community
e2e.ti.com
Wireless Connectivity
www.ti.com/wirelessconnectivity
Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265
Copyright © 2012, Texas Instruments Incorporated