H HiTRON UNIVERSAL INPUT AC-DC MEDICAL APPLICATION OPEN FRAME INTERNAL SWITCHING POWER SUPPLIES 65WATTS GREEN POWER TRIPLE OUTPUT HICM65 SERIES FEATURES: COMPACT 4”x 2” PACKAGE FREE AIR CONVECTION UP TO 83% HIGH EFFICIENCY -20°C TO +70°C WIDE OPERATING TEMP. MEET MEDICAL STANDARDS IEC60601-1 PCB & ENCLOSURE FORMAT OPTIONAL CE MARKING COMPLIANCE SPECIFICATION INPUT SPECIFICATION OUTPUT SPECIFICATION Input Voltage: Typical 90-264Vac. Input Connector: V-H connector. Input Frequency: 47-63Hz. Inrush Current: Typical 16A rms at 230Vac. Input Current: Typ. 1.1A at 115Vac, 0.55A at 230Vac. Dielectric Withstand: Meet IEC60601-1. EMI: Meet CISPR EN55022 Class B. Hold-up Time: Typ. 16mS at 115Vac,76mS at 230Vac. Earth Leakage Current: Less than 0.24mA at 230Vac. Enclosure Leakage Current: 0.01mA at 230Vac. No Load Power: Less than 1.5W at 230Vac. Output Voltage: See Ratings Chart. Output Current: See Ratings Chart. Output Connector: V-H connector. Output Wattage: Typical 65 watts. Line Regulation: Typical 0.5-2%. Load Regulation: VO1/2 typ. ±2.0%, VO3 typ. ±3.0% Total Regulation: VO1 typ. ±4.0%, VO2/3 typ. ±8.0%. Noise & Ripple: Typical 1.0% peak to peak. OVP: Built-in VO1 (Auto-recovery). Adjustability: Not available. Overload Protection (OLP): Fully protected against outputs overload and short circuit. Typ. OLP set at 130-150% of rating output power. Consult the factory for OLP setting. GENERAL SPECIFICATION Efficiency: Typ. 83% at 230Vac(various with the o/p volt.) Switching Frequency: 65K Hz. Circuit Topology: Fix-flyback Circuit. Transient Response: Output voltage returns in less than 0.01mS following a 25% load change. Safety Standard: IEC60601-1 Class I. Operating Temperature: -20°C to +70 °C with derating. (please refer to the derating chart). Storage Temperature: -25 to +85°C. Cooling: Free air convection. Construction: PCB and Enclosure optional. Power Density: 6.5Watts / Cubic Inch. NOTE: (1) All measurements are at nominal input, full load, and +25°C unless otherwise specified. (2) Load regulation is measured at 115Vac or 230Vac in percentage to indicate the change in output voltage as the load varied from half load to full load (%). (3) Due to requests in market and advances in technology, specifications subject to change without notice. -1- Edited on May 27, 2015 Version 0 OUTPUT VOLTAGE / CURRENT RATINGS CHART SINGLE OUTPUT MAIN VO1 ★ MODEL NO. (note 1) AUX. VO2 Min. Typ. AUX. VO3 Min. Typ. Volt. Max. Max. Min. Typ. Volt. Max. HICM65-T050II 1.2A 6.0A +5V 8.0A 0.50A 2.50A +12V 3.0A 0.1A 0.5A -12V 0.5A HICM65-T050KK 1.2A 6.0A +5V 8.0A 0.40A 2.00A +15V 2.5A 0.1A 0.5A -15V 0.5A HICM65-T050MI 1.2A 6.0A +5V 8.0A 0.25A 1.25A +24V 1.5A 0.1A 0.5A -12V 0.5A Volt. Remark: “★” OVP built-in Note: 1.Please add suffix to model number to define type: add “-C” for enclosure (metal box) version, otherwise will be PCB version. For example: HICM65-T050II is for PCB version, HICM65-T050II-C is for enclosure (metal box) version. 2. Derate output power by 20% for enclosure version. MECHANICAL DIMENSIONS: MM [INCHES] PCB Version DERATING CHART Enclosure (Metal Box) Version INPUT & OUTPUT CONNECTORS PIN ASSIGNMENT -2- ASSIGNMENT CNTR/PIN# AC-LINE CNT1-L AC-NEUTRAL CNT1-N AC-GROUND GND VO1 CNT2-5,6 VO2 CNT2-2 VO3 CNT2-1 DC COM CNT2-3,4 Edited on May 27, 2015 Version 0