SCAN BLOCK - Hamamatsu Photonics

NEW SCAN BLOCK
C10516
Compact scan head for easy assembly into a system
OVERVIEW
The C10516 scan block is an optical block that combines galvano
scanners with a telecentric fθ (F-theta) lens designed for laser
beam scanning in the visible range. Coupling the C10516 to other
optical blocks allows forming a measurement system for laser
scanning fluorescence microscopes, reflecting microscopes, and
even confocal microscopes. The 21.5 mm diameter observation
area allows making wide-range sample observations. High magnification images can be observed with a photomultiplier tube by attaching the scan block to the C-mount port of a commercially
available microscope.
APPLICATIONS
Left: Scan block main unit, Right: Control unit
FEATURES
●Laser scanning microscopes
●Confocal laser scanning microscopes
●2D scan by two galvano scanners
●Contains a telecentric fθ (F-theta) lens
●Wide observation area of 21.5 mm diameter
●High-magnification observations in combination with a microscope objective lens
●Covers a wide spectral range
(Example) Biological microscopes
Industrial microscopes
DNA chip and protein chip readers
CONFIGURATION
The C10516 consists of a scan block and a control unit. Two galvano scanners and a telecentric fθ lens are built into
the scan block main unit. The control unit contains a driver board for driving and controlling the galvano scanners.
Laser
Scan Block
C10516
(Commercially Available Product)
Control
Signal
Control Unit
X
Command Input
Signal (X)
PC
(Sold Separately)
USB,
PCI, etc
Scan Block
Main Unit
Galvano
Scanner
PMT Module
(Sold Separately)
Optical Block
(Sold Separately)
(Note 2)
Y
Command Input
Signal (Y)
Commercially
Available Data
Acquisition Unit
(DAQ)
(Sold Separately)(Note 1)
Object
PMT Signal
Note 1: This system requires a data acquisition
unit (DAQ) that outputs analog signals
for controlling the galvano scanners and
acquires PMT signals.
Note 2: The optical blocks connected to the scan
block can be changed according to the
measurement purpose.
SPECIFICATIONS
●Telecentric fθ lens transmittance
Units
V
VA
nm
nm
mm
mm
mm
deg.
mm
mm
Lp/mm
V
mm/V
Hz
%
—
W/cm2
°C
°C
%
%
100
TPMOB0248EA
90
TRANSMITTANCE (%)
80
70
60
50
40
30
20
10
0
300
400
500
600
800
700
900
1000
WAVELENGTH (nm)
●Command input voltage and
observation field of view
TPMOC0248EA
-3 V
15 mm
15 mm
0V
MAXIMUM
MEASUREMENT
AREA
(inscribed quadrangle)
+3 V
+6 V
AThe telecentric fθ lens is designed with three wavelengths (488nm, 532nm, 632nm).
BVoltage supplied to the X and Y axis scanner signal input on the control unit.
CScan length at the field-of-view position for command input signal voltage that is supplied to the galvano scanners.
DPercentage of constant velocity movement zone where the galvano scanner enters the measurement area.
ENo condensation
MAXIMUM
MEASUREMENT
AREA
21.5 mm
-6 V
COMMAND INPUT SIGNAL (Y-Axis)
Specifications
Items
Power Input
AC100 to AC240
Power Consumption
85
Max.
488 to 632
Design Wavelength A
Transmittance (≥90 %)
440 to 720
Focal Length
50
Maximum Measurement Area
21.5
Optical
15 × 15
Maximum Measurement Area (inscribed quadrangle)
Characteristics
±12.4
Scan Angle of Maximum Measurement Area
4
Input Beam Aperture
20
Working Distance
80
Resolution
±6
Maximum Command Input Signal B
Galvano
2.5
Scan Length per Command Input Signal C
Scanner
400
Maximum Scan Speed
Characteristics
≥ 70
Fill Fraction D
Al
Mirror
Coating
100 to 150
Characteristics Damage Threshold
0 to +50
Operating Ambient Temperature
-10 to +50
Storage Temperature
Below 80
Operating Ambient Humidity E
Below 85
Storage Humidity E
-6 V
-3 V
0V
+3 V
+6 V
COMMAND INPUT SIGNAL (X-Axis)
CONNECTION EXAMPLES
The optical system can be assembled by connecting the necessary optical blocks to the light input-output port on the
scan block. The optical system connected to the scan block differs according to the object to be measured. Use the following examples as a general reference to connect optical blocks (sold separately) and commercially available optical
components to the scan block.
1Reflected light detection
Attaching other optical blocks to the light input port of the scan block allows observations in an area of 15 mm × 15 mm.
Laser
Control Unit
PMT module
4
A
Scan block
E
XY
Optical block
F
2
Scan Block
Main Unit
DAQ
PMT Module
1
* Hamamatsu provides various types of optical
blocks other than those listed in the connection
examples.
Refer to our “Optical Blocks” catalog for more
information.
B
Other manufacturers’ products
Product name
1 Data acquisition unit (DAQ)
2 Quarter-wave plate
3 Polarization beam splitter
4 Fiber output laser
C
D
G
3
H
Optical blocks*
Product name
A Scan block
B C-mount adapter block
C Joint block
D Polarizer holder block
E Fiber adapter block
F Joint block
G Beam splitter holder block
H Adapter block
Type No.
C10516
A10039
A10038-01
A11026
A10037
A10038-02
A10035-90
A10030-01
Qty
1
1
1
1
1
1
1
1
TPMOC0249EA
2Fluorescence detection using scan block attached to microscope port
The scan block can be connected to the C-mount camera port of a commercially available microscope via the
microscope port adapter that comes with the scan block. This system allows fluorescence image acquisition by laser
scanning.
Control Unit
Scan block
Laser
D
1
4
XY
2
DAQ
A
3
PMT module
B
Scan Block
Main Unit
Commercially
available
microscope
PMT Module
* Microscope port adapter
attached to scan block
B
C
E
F
G
Optical blocks*
Product name
A Scan block
B C-mount adapter block
C Joint block
D Beam aligner block
E Dichroic block
F Filter block
G Adapter block
Other manufacturers’ products
Product name
1 Data acquisition unit (DAQ)
2 Infinity-corrected objective lens
3 Iris diaphragm barrel
4 Fiber output laser
Type No.
C10516
A10039
A10038-01
A10760
A10034-xx
A10033-xx
A10030-01
Qty
1
2
1
1
1
1
1
TPMOC0250EA
3Fluorescence detection using confocal optical system
Using the scan block with a stage block and other optical blocks allows building a microscope system.
Scan block
Control Unit
Laser
D
5
1
Stage block
XY
PMT module
2
DAQ
3
A
2
G
B
H
Scan Block
Main Unit
J
K
* Microscope port adapter
attached to scan block
Other manufacturers’ products
Product name
1 Data acquisition unit (DAQ)
2 Infinity-corrected objective lens
3 Iris diaphragm barrel
4 Pinhole
5 Fiber output laser
I
PMT Module
B
C
E
B
F
Optical blocks*
Product name
A Scan block
B C-mount adapter block
C Joint block
D Beam aligner block
E Dichroic block
F Tube lens block
G Pinhole holder block
H Filter block
I Adapter block
J Stage block
K Tube lens block
4
Type No.
C10516
A10039
A10038-01
A10760
A10034-xx
A10859
A11027
A10033-xx
A10030-01
A10858
A10859-01
Qty
1
3
1
1
1
1
1
1
1
1
1
TPMOC0251EA
DIMENSIONAL OUTLINES (Unit: mm)
●Main Unit
●Control Unit
20
(WORKING DISTANCE)
(13.4)
150
270
(15)
80
1/4"-20UNC
DEPTH 8
(FOR SECURING SCAN BLOCK)
85
200
fθ LENS
POSITIONING PIN
20.00 ± 0.05
X
CONTROL
Y
(8)
(187.526)
43
23
TPMOA0080EA
MICROSCOPE
PORT ADAPTER
32.2
23
62
17.526
INTERNAL
PART OF
fθ LENZ
POWER
100
36
27
+0
50 - 0.04
fθ LENZ
FF PLANE
6
C-MOUNT (FEMALE THREAD), DEPTH 5
SET SCREW FOR HOLDING MICROSCOPE PORT ADAPTER
70
50
20
6
4-M4, DEPTH 5
(FOR SECURING SCAN BLOCK)
TPMOA0079EA
<RELATED PRODUCTS>
Sample Software
Hamamatsu provides sample software for acquiring images using the scan block. This software
was developed with National Instruments LabVIEW 2011 and so the software program was
created assuming that a data acquisition unit (DAQ) of National Instruments is used. We also offer
the program source free of charge to allow you to change the program or add functions as needed.
[Operating conditions] The following items must be installed:
LabVIEW 2011 or later
NI-DAQmx
NI-IMAQ 4.6.4 or later
Please contact us when you need this sample software.
Data Acquisition Unit (DAQ) (commercially available product)
A data acquisition unit (DAQ) is required to operate the scan block. Using a commercially available data acquisition unit (DAQ) helps you
operate the scan block. Select the DAQ with specifications that match your application.
[Recommended specifications]
Number of Analog Input Resolution Sampling Rate Number of Analog Output Resolution
Input Channels
Output Channels
(bits)
(S/second)
(bits)
1 or more a
12 or more
1 M or more
2
12 or more
Output Rate
(S/second)
1 M or more
Simultaneous
Sampling
Digital
I/O
a
b
aWhen acquiring images from multiple channels, we recommend using a simultaneous sampling DAQ. If a multiplexer type is used, the acquired images will shift in time.
bCounter input will be needed when using photon counting to acquire images.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2013 Hamamatsu Photonics K.K.
HAMAMATSU PHOTONICS K.K.
www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Division
314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [email protected]
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [email protected]
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [email protected]
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: [email protected]
North Europe: Hamamatsu Photonics Norden AB: Thorshamnsgatan 35 SE-164 40 Kista, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [email protected]
TPMO1057E01
Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: [email protected]
AUG. 2013 IP
China: Hamamatsu Photonics (China) Co., Ltd.: 1201 Tower B, Jiaming Center, 27 Dongsanhuan Road North, Chaoyang District, Beijing 100020, China, Telephone: (86)10-6586-6006, Fax: (86)10-6586-2866 E-mail: [email protected]